{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Miami",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99832",
      "recalling_firm": "Beckman Coulter, Inc.",
      "address_1": "11800 Sw 147th Ave",
      "address_2": "",
      "postal_code": "33196-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.",
      "recall_number": "Z-0160-2027",
      "product_description": "DxU 840m Iris, REF: C76947;  DxU 850m Iris, REF: C49513",
      "product_quantity": "2,088",
      "reason_for_recall": "The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement  to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual  preventive maintenance (PM) requirements without adequate reliability evaluation. When the check  valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor,  potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure),  biohazardous exposure, property damage, and delayed patient results.",
      "recall_initiation_date": "20260812",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "Product/UDI-DI:  DxU 840m/15099590748654, DxU 850m/15099590748647.  All Serial Numbers",
      "more_code_info": ""
    }
  ]
}