  "results": [
    {
      "status": "Ongoing",
      "city": "Round Lake",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99826",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "25212 W Il Route 120",
      "address_2": "",
      "postal_code": "60073-9799",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of IL, MO, TX.",
      "recall_number": "Z-0332-2027",
      "product_description": "TruSystem 7000 U Operating Table (dV), Product Code 1723633",
      "product_quantity": "62 units",
      "reason_for_recall": "The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.",
      "recall_initiation_date": "20260914",
      "center_classification_date": "20261103",
      "report_date": "20261111",
      "code_info": "UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662",
      "more_code_info": ""
    }
  ]
}