{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-07",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99779",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "International distribution to the country of Japan.",
      "recall_number": "Z-3299-2026",
      "product_description": "Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.",
      "product_quantity": "72 units",
      "reason_for_recall": "There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.",
      "recall_initiation_date": "20260824",
      "center_classification_date": "20260925",
      "report_date": "20261007",
      "code_info": "Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.",
      "more_code_info": ""
    }
  ]
}