{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-07",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tiffin",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99759",
      "recalling_firm": "American Contract Systems Inc",
      "address_1": "85 Shaffer Park Dr",
      "address_2": "",
      "postal_code": "44883-9290",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of Iowa, Minnesota, Wisconsin, and South Dakota.",
      "recall_number": "Z-3294-2026",
      "product_description": "Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays    ABDOMINAL HYSTERECTOMY - 205989\t  \tModel No. MHAH55AM  \t  LAVH-LAP COLON\t  \tModel No. SPLL54U  \t  VAG HYST PACK\t  \tModel No. SPVH62P",
      "product_quantity": "1,804 kits total",
      "reason_for_recall": "Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).",
      "recall_initiation_date": "20260810",
      "center_classification_date": "20260925",
      "report_date": "20261007",
      "code_info": "MHAH55AM    UDI-DI 191072236487    Lots 42-8529911   42-8649211   42-8687711   42-8797811   42-8857011   42-8869711    SPLL54U    UDI-DI 191072239181    Lots 07-2516111   07-2528611   07-2545611   07-2547911   07-2561411    SPVH62P    UDI-DI 191072238771    Lots 07-2544611   12-7436711",
      "more_code_info": ""
    }
  ]
}