  "results": [
    {
      "status": "Ongoing",
      "city": "La Ciotat Cedex",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99745",
      "recalling_firm": "INTERVASCULAR, SAS",
      "address_1": "1 Zone Industrielle Athelia",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Jordan, Kenya, Kosovo, Kuwait, Latvia, Luxembourg, Malaysia, Mauritius, Monaco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Yemen.",
      "recall_number": "Z-3258-2026",
      "product_description": "Cardioroot Woven Collagen Coated Vascular Graft;  Model Numbers: 1. HEWROOT0024, 2. HEWROOT0026, 3. HEWROOT0028, 4. HEWROOT0030, 5. HEWROOT0032, 6. HEWROOT0034;",
      "product_quantity": "16329 (3880 US, 12449 OUS)",
      "reason_for_recall": "Product failed to meet the specified acceptance criteria for Integral Water Permeability (IWP).",
      "recall_initiation_date": "20260812",
      "center_classification_date": "20260923",
      "report_date": "20260930",
      "code_info": "1. Model Number: HEWROOT0024; UDI-DI: 00384401013884; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;  2. Model Number: HEWROOT0026;  UDI-DI: 00384401013891; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;  3. Model Number: HEWROOT0028;  UDI-DI: 00384401013907; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;  4. Model Number: HEWROOT0030;  UDI-DI: 00384401013914; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;  5. Model Number: HEWROOT0032; UDI-DI: 00384401013921; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;  6. Model Number: HEWROOT0034; UDI-DI: 00384401013938; Serial/Lot Numbers: All CARDIOROOT Woven products within labeled shelf life;",
      "more_code_info": ""
    }
  ]
}