      "status": "Ongoing",
      "city": "Heidelberg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99595",
      "recalling_firm": "Heidelberg Engineering GmbH",
      "address_1": "Max-Jarecki-Str. 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.",
      "recall_number": "Z-3215-2026",
      "product_description": "Brand Name: Heidelberg Engineering GmbH  Product Name: SPECTRALIS with Flex Module  Model/Catalog Number: S3610, S2610, S2000, S1610    multimodal imaging platform optimized for the posterior segment of the eye",
      "product_quantity": "87",
      "reason_for_recall": "Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.",
      "recall_initiation_date": "20260806",
      "center_classification_date": "20260918",
      "report_date": "20260930",
      "code_info": "UDI-DI 04059237000235   Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611",
      "more_code_info": ""
    }
  ]
}