{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-07",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99423",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2932-2026",
      "product_description": "Philips Azurion, Model Numbers:  722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers:  R3.0 and R3.1; Interventional fluoroscopic x-ray system",
      "product_quantity": "2016 units",
      "reason_for_recall": "Two software related issues:   1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment  and  2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications",
      "recall_initiation_date": "20260716",
      "center_classification_date": "20260810",
      "report_date": "20260819",
      "code_info": "UDI:  (01)00884838116757(21)5, (01)00884838116771(21)686, (01)00884838116771(21)521, (01)00884838116771(21)812, (01)00884838116771(21)163, (01)00884838116801(21)127, (01)00884838116757(21)34, (01)00884838116733(21)61, (01)00884838116764(21)226, (01)00884838116771(21)327, (01)00884838116764(21)277, (01)00884838116771(21)780, (01)00884838116771(21)429, (01)00884838116764(21)276, (01)00884838116788(21)87, (01)00884838116801(21)37, (01)00884838116771(21)362, (01)00884838116771(21)502, (01)00884838116740(21)77, (01)00884838116757(21)106, (01)00884838116771(21)222, (01)00884838116771(21)545, (01)00884838116801(21)159, (01)00884838116801(21)345, (01)00884838116771(21)987, (01)00884838116771(21)457, (01)00884838116801(21)136, (01)00884838116757(21)49, (01)00884838116771(21)803, (01)00884838116771(21)901, (01)00884838116771(21)347, (01)00884838116764(21)335, (01)00884838116801(21)199, (01)00884838116801(21)24, (01)00884838116771(21)902, (01)00884838116771(21)529, (01)00884838116764(21)99, (01)00884838116740(21)37, (01)00884838116801(21)421, (01)00884838116771(21)718, (01)00884838116801(21)212, (01)00884838116764(21)122, (01)00884838116801(21)17, (01)00884838116764(21)344, (01)00884838116771(21)287, (01)00884838116757(21)140, (01)00884838116764(21)169, (01)00884838116771(21)822, (01)00884838116801(21)25, (01)00884838116801(21)180, (01)00884838116771(21)88, (01)00884838116788(21)127, (01)00884838116771(21)363, (01)00884838116801(21)227, (01)00884838116771(21)716, (01)00884838116771(21)185, (01)00884838116771(21)1145, (01)00884838116771(21)426, (01)00884838116801(21)235, (01)00884838116771(21)292, (01)00884838116771(21)252, (01)00884838116801(21)131, (01)00884838116771(21)332, (01)00884838116771(21)993, (01)00884838116771(21)97, (01)00884838116771(21)359, (01)00884838116771(21)1088, (01)00884838116764(21)350, (01)00884838116771(21)893, (01)00884838116801(21)288, (01)00884838116771(21)346, (01)00884838116771(21)695, (01)00884838116771(21)383, (01)00884838116771(21)300, (01)00884838116771(21)804, (01)00884838116801(21)504, (01)00884838116771(21)186, (01)00884838116764(21)267, (01)00884838116771(21)1354, (01)00884838116771(21)1246, (01)00884838116771(21)99, (01)00884838116771(21)413, (01)00884838116801(21)231, (01)00884838116771(21)547, (01)00884838116771(21)449, (01)00884838116757(21)99, (01)00884838116801(21)203, (01)00884838116771(21)75, (01)00884838116771(21)813, (01)00884838116801(21)138, (01)00884838116771(21)447, (01)00884838116771(21)915, (01)00884838116771(21)511, (01)00884838116788(21)117, (01)00884838116801(21)426, (01)00884838116801(21)20, (01)00884838116771(21)484, (01)00884838116801(21)249, (01)00884838116757(21)392, (01)00884838116771(21)626, (01)00884838116771(21)215, (01)00884838116764(21)96, (01)00884838116771(21)704, (01)00884838116764(21)108, (01)00884838116771(21)248, (01)00884838116771(21)594, (01)00884838116771(21)87, (01)00884838116771(21)333, (01)00884838116788(21)76, (01)00884838116771(21)1226, (01)00884838116771(21)927, (01)00884838116771(21)1037, (01)00884838116771(21)1140, (01)00884838116771(21)286, (01)00884838116771(21)173, (01)00884838116771(21)395, (01)00884838116801(21)200, (01)00884838116757(21)12, (01)00884838116771(21)68, (01)00884838116801(21)468, (01)00884838116771(21)656, (01)00884838116801(21)208, (01)00884838116771(21)459, (01)00884838116771(21)848, (01)00884838116757(21)138, (01)00884838116801(21)42, (01)00884838116771(21)71, (01)00884838116771(21)465, (01)00884838116801(21)205, (01)00884838116771(21)855, (01)00884838116801(21)183, (01)00884838116757(21)101, (01)00884838116771(21)377, (01)00884838116771(21)854, (01)00884838116771(21)517, (01)00884838116771(21)330, (01)00884838116801(21)363, (01)00884838116771(21)370, (01)00884838116757(21)91, (01)00884838116771(21)660, (01)00884838116771(21)501, (01)00884838116771(21)535, (01)00884838116771(21)463, (01)00884838116771(21)320, (01)00884838116771(21)1248, (01)00884838116801(21)81, (01)00884838116771(21)817, (01)00884838116771(21)516, (01)00884838116771(21)616, (01)00884838116764(21)271, (01)00884838116757(21)127, (01)00884838116771(21)475, (01)00884838116757(21)2, (01)00884838116771(21)270, (01)00884838116757(21)96, (01)00884838116757(21)316, (01)00884838116771(21)583, (01)00884838116771(21)887, (01)00884838116771(21)549, (01)00884838116801(21)148, (01)00884838116801(21)133, (01)00884838116764(21)322, (01)00884838116757(21)124, (01)00884838116757(21)317, (01)00884838116771(21)762, (01)00884838116771(21)617, (01)00884838116771(21)703, (01)00884838116771(21)178, (01)00884838116757(21)36, (01)00884838116771(21)171, (01)00884838116788(21)188, (01)00884838116757(21)136, (01)00884838116771(21)1076, (01)00884838116771(21)847, (01)00884838116771(21)1342, (01)00884838116771(21)592, (01)00884838116771(21)211, (01)00884838116771(21)331, (01)00884838116771(21)1203, (01)00884838116788(21)88, (01)00884838116771(21)726, (01)00884838116771(21)467, (01)00884838116771(21)228, (01)00884838116801(21)31, (01)00884838116771(21)1277, (01)00884838116771(21)162, (01)00884838116771(21)774, (01)00884838116771(21)85, (01)00884838116771(21)890, (01)00884838116801(21)342, (01)00884838116771(21)284, (01)00884838116764(21)212, (01)00884838116771(21)1023, (01)00884838116771(21)985, (01)00884838116771(21)824, (01)00884838116771(21)1130, (01)00884838116801(21)54, (01)00884838116771(21)1131, (01)00884838116771(21)67, (01)00884838116801(21)218, (01)00884838116764(21)153, (01)00884838116801(21)194, (01)00884838116771(21)469, (01)00884838116771(21)696, (01)00884838116771(21)472, (01)00884838116771(21)326, (01)00884838116771(21)648, (01)00884838116771(21)914, (01)00884838116771(21)758, (01)00884838116801(21)43, (01)00884838116771(21)1142, (01)00884838116801(21)158, (01)00884838116771(21)548, (01)00884838116764(21)259, (01)00884838116771(21)260, (01)00884838116771(21)619, (01)00884838116757(21)108, (01)00884838116757(21)141, (01)00884838116801(21)416, (01)00884838116771(21)269, (01)00884838116771(21)1309, (01)00884838116771(21)489, (01)00884838116771(21)1134, (01)00884838116764(21)269, (01)00884838116771(21)255, (01)00884838116771(21)629, (01)00884838116757(21)48, (01)00884838116771(21)1079, (01)00884838116771(21)730, (01)00884838116771(21)637, (01)00884838116771(21)778, (01)00884838116771(21)633, (01)00884838116771(21)92, (01)00884838116757(21)46, (01)00884838116771(21)345, (01)00884838116771(21)874, (01)00884838116771(21)257, (01)00884838116733(21)40, (01)00884838116771(21)101, (01)00884838116771(21)972, (01)00884838116771(21)493, (01)00884838116771(21)664, (01)00884838116757(21)350, (01)00884838116771(21)772, (01)00884838116771(21)560, (01)00884838116788(21)43, (01)00884838116771(21)79, (01)00884838116764(21)199, (01)00884838116771(21)924, (01)00884838116771(21)372, (01)00884838116771(21)714, (01)00884838116771(21)797, (01)00884838116771(21)794, (01)00884838116771(21)565, (01)00884838116801(21)21, (01)00884838116771(21)225, (01)00884838116771(21)657, (01)00884838116771(21)585, (01)00884838116771(21)638, (01)00884838116771(21)1211, (01)00884838116771(21)233, (01)00884838116771(21)620, (01)00884838116771(21)906, (01)00884838116771(21)433, (01)00884838116757(21)76, (01)00884838116771(21)1005, (01)00884838116801(21)40, (01)00884838116757(21)45, (01)00884838116771(21)208, (01)00884838116771(21)661, (01)00884838116771(21)435, (01)00884838116771(21)309, (01)00884838116771(21)418, (01)00884838116771(21)689, (01)00884838116771(21)867, (01)00884838116764(21)118, (01)00884838116771(21)862, (01)00884838116771(21)524, (01)00884838116788(21)102, (01)00884838116771(21)256, (01)00884838116801(21)60, (01)00884838116764(21)214, (01)00884838116771(21)159, (01)00884838116771(21)929, (01)00884838116764(21)283, (01)00884838116771(21)1254, (01)00884838116771(21)1209, (01)00884838116801(21)50, (01)00884838116757(21)6, (01)00884838116771(21)1126, (01)00884838116801(21)378, (01)00884838116771(21)768, (01)00884838116801(21)335, (01)00884838116771(21)581, (01)00884838116771(21)580, (01)00884838116801(21)506, (01)00884838116801(21)39, (01)00884838116771(21)753, (01)00884838116771(21)833, (01)00884838116801(21)191, (01)00884838116771(21)223, (01)00884838116740(21)38, (01)00884838116771(21)618, (01)00884838116801(21)280, (01)00884838116801(21)233, (01)00884838116801(21)48, (01)00884838116771(21)107, (01)00884838116771(21)849, (01)00884838116771(21)409, (01)00884838116771(21)1004, (01)00884838116771(21)209, (01)00884838116801(21)197, (01)00884838116771(21)442, (01)00884838116771(21)386, (01)00884838116771(21)194, (01)00884838116801(21)47, (01)00884838116801(21)221, (01)00884838116771(21)411, (01)00884838116771(21)698, (01)00884838116771(21)1069, (01)00884838116771(21)54, (01)00884838116771(21)904, (01)00884838116801(21)68, (01)00884838116771(21)76, (01)00884838116771(21)1294, (01)00884838116771(21)108, (01)00884838116771(21)1378, (01)00884838116771(21)70, (01)00884838116771(21)165, (01)00884838116771(21)329, (01)00884838116771(21)122, (01)00884838116801(21)204, (01)00884838116771(21)403, (01)00884838116801(21)444, (01)00884838116771(21)676, (01)00884838116788(21)71, (01)00884838116771(21)1066, (01)00884838116771(21)392, (01)00884838116764(21)41, (01)00884838116771(21)288, (01)00884838116801(21)230, (01)00884838116771(21)348, (01)00884838116771(21)263, (01)00884838116771(21)174, (01)00884838116771(21)321, (01)00884838116771(21)770, (01)00884838116771(21)262, (01)00884838116771(21)877, (01)00884838116764(21)182, (01)00884838116764(21)124, (01)00884838116771(21)519, (01)00884838116764(21)147, (01)00884838116771(21)520, (01)00884838116801(21)225, (01)00884838116801(21)427, (01)00884838116788(21)11, (01)00884838116771(21)509, (01)00884838116771(21)1063, (01)00884838116757(21)61, (01)00884838116771(21)1047, (01)00884838116771(21)864, (01)00884838116757(21)10, (01)00884838116771(21)1299, (01)00884838116740(21)30, (01)00884838116771(21)515, (01)00884838116757(21)43, (01)00884838116771(21)1111, (01)00884838116733(21)2, (01)00884838116771(21)485, (01)00884838116771(21)405, (01)00884838116771(21)376, (01)00884838116771(21)460, (01)00884838116764(21)347, (01)00884838116757(21)298, (01)00884838116771(21)853, (01)00884838116771(21)221, (01)00884838116764(21)278, (01)00884838116771(21)526, (01)00884838116801(21)220, (01)00884838116801(21)265, (01)00884838116764(21)111, (01)00884838116771(21)496, (01)00884838116771(21)886, (01)00884838116771(21)796, (01)00884838116771(21)151, (01)00884838116801(21)38, (01)00884838116801(21)463, (01)00884838116771(21)428, (01)00884838116771(21)922, (01)00884838116771(21)500, (01)00884838116771(21)586, (01)00884838116771(21)217, (01)00884838116771(21)492, (01)00884838116764(21)289, (01)00884838116771(21)232, (01)00884838116771(21)140, (01)00884838116771(21)371, (01)00884838116771(21)344, (01)00884838116801(21)32, (01)00884838116771(21)800, (01)00884838116771(21)646, (01)00884838116771(21)667, (01)00884838116771(21)1045, (01)00884838116771(21)827, (01)00884838116801(21)276, (01)00884838116801(21)116, (01)00884838116771(21)674, (01)00884838116764(21)172, (01)00884838116771(21)220, (01)00884838116757(21)40, (01)00884838116771(21)538, (01)00884838116788(21)57, (01)00884838116771(21)1104, (01)00884838116771(21)917, (01)00884838116771(21)675, (01)00884838116801(21)134, (01)00884838116771(21)1214, (01)00884838116801(21)260, (01)00884838116771(21)102, (01)00884838116771(21)342, (01)00884838116771(21)891, (01)00884838116771(21)462, (01)00884838116771(21)787, (01)00884838116764(21)192, (01)00884838116771(21)1080, (01)00884838116764(21)302, (01)00884838116764(21)218, (01)00884838116764(21)219, (01)00884838116771(21)430, (01)00884838116771(21)514, (01)00884838116771(21)319, (01)00884838116757(21)114, (01)00884838116788(21)89, (01)00884838116771(21)334, (01)00884838116771(21)416, (01)00884838116771(21)513, (01)00884838116771(21)518, (01)00884838116801(21)272, (01)00884838116801(21)273, (01)00884838116771(21)723, (01)00884838116771(21)641, (01)00884838116771(21)1000, (01)00884838116771(21)324, (01)00884838116801(21)339, (01)00884838116740(21)74, (01)00884838116740(21)52, (01)00884838116757(21)97, (01)00884838116801(21)10, (01)00884838116801(21)327, (01)00884838116771(21)289, (01)00884838116788(21)179, (01)00884838116771(21)187, (01)00884838116764(21)250, (01)00884838116726(21)9, (01)00884838116740(21)70, (01)00884838116801(21)289, (01)00884838116771(21)809, (01)00884838116771(21)111, (01)00884838116788(21)13, (01)00884838116801(21)67, (01)00884838116764(21)377, (01)00884838116771(21)1206, (01)00884838116764(21)123, (01)00884838116764(21)307, (01)00884838116764(21)308, (01)00884838116801(21)449, (01)00884838116740(21)138, (01)00884838116801(21)242, (01)00884838116771(21)1300, (01)00884838116764(21)37, (01)00884838116764(21)38, (01)00884838116771(21)719, (01)00884838116740(21)96, (01)00884838116801(21)518, (01)00884838116771(21)839, (01)00884838116733(21)22, (01)00884838116740(21)49, (01)00884838116764(21)183, (01)00884838116771(21)275, (01)00884838116771(21)750, (01)00884838116771(21)966, (01)00884838116771(21)1120, (01)00884838116757(21)333, (01)00884838116788(21)180, (01)00884838116801(21)486, (01)00884838116801(21)413, (01)00884838116757(21)8, (01)00884838116801(21)361, (01)00884838116801(21)29, (01)00884838116771(21)706, (01)00884838116764(21)273, (01)00884838116771(21)482, (01)00884838116771(21)534, (01)00884838116771(21)821, (01)00884838116788(21)74, (01)00884838116801(21)192, (01)00884838116764(21)304, (01)00884838116771(21)738, (01)00884838116764(21)354, (01)00884838116771(21)748, (01)00884838116771(21)786, (01)00884838116771(21)963, (01)00884838116771(21)1348, (01)00884838116771(21)759, (01)00884838116771(21)438, (01)00884838116801(21)223, (01)00884838116771(21)691, (01)00884838116801(21)13, (01)00884838116788(21)201, (01)00884838116771(21)732, (01)00884838116771(21)412, (01)00884838116764(21)17, (01)00884838116771(21)42, (01)00884838116771(21)873, (01)00884838116801(21)164, (01)00884838116771(21)1346, (01)00884838116771(21)682, (01)00884838116801(21)368, (01)00884838116771(21)763, (01)00884838116733(21)10, (01)00884838116764(21)76, (01)00884838116764(21)137, (01)00884838116771(21)385, (01)00884838116771(21)598, (01)00884838116771(21)679, (01)00884838116733(21)50, (01)00884838116757(21)113, (01)00884838116726(21)40, (01)00884838116771(21)939, (01)00884838116771(21)941, (01)00884838116771(21)783, (01)00884838116764(21)460, (01)00884838116771(21)45, (01)00884838116771(21)625, (01)00884838116764(21)353, (01)00884838116801(21)215, (01)00884838116740(21)50, (01)00884838116788(21)204, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116771(21)146, (01)00884838116771(21)1110, (01)00884838116757(21)111, (01)00884838116771(21)582, (01)00884838116757(21)118, (01)00884838116757(21)119, (01)00884838116757(21)139, (01)00884838116740(21)106, (01)00884838116788(21)105, (01)00884838116733(21)19, (01)00884838116733(21)20, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116757(21)50, (01)00884838116733(21)25, (01)00884838116764(21)300, (01)00884838116733(21)30, (01)00884838116771(21)964, (01)00884838116733(21)67, (01)00884838116757(21)22, (01)00884838116733(21)43, (01)00884838116764(21)236, (01)00884838116764(21)246, (01)00884838116764(21)247, (01)00884838116764(21)251, (01)00884838116771(21)1292, (01)00884838116801(21)348, (01)00884838116740(21)41, (01)00884838116726(21)7, (01)00884838116771(21)37, (01)00884838116771(21)160, (01)00884838116771(21)795, (01)00884838116764(21)12, (01)00884838116801(21)244, (01)00884838116801(21)322, (01)00884838116801(21)352, (01)00884838116771(21)78, (01)00884838116801(21)300, (01)00884838116788(21)118, (01)00884838116771(21)401, (01)00884838116764(21)201, (01)00884838116740(21)68, (01)00884838116801(21)383, (01)00884838116771(21)767, (01)00884838116801(21)330, (01)00884838116764(21)238, (01)00884838116801(21)173, (01)00884838116801(21)351, (01)00884838116771(21)838, (01)00884838116757(21)386, (01)00884838116771(21)295, (01)00884838116771(21)486, (01)00884838116764(21)82, (01)00884838116801(21)117, (01)00884838116788(21)158, (01)00884838116764(21)245, (01)00884838116788(21)135, (01)00884838116801(21)376, (01)00884838116771(21)642, (01)00884838116801(21)354, (01)00884838116733(21)59, (01)00884838116733(21)94, (01)00884838116788(21)136, (01)00884838116764(21)416, (01)00884838116801(21)433, (01)00884838116726(21)47, (01)00884838116764(21)8, (01)00884838116788(21)164, (01)00884838116788(21)168, (01)00884838116801(21)536, (01)00884838116801(21)537, (01)00884838116771(21)1116, (01)00884838116771(21)408, (01)00884838116733(21)45, (01)00884838116764(21)141, (01)00884838116801(21)201, (01)00884838116764(21)297, (01)00884838116764(21)298, (01)00884838116764(21)299, (01)00884838116771(21)205, (01)00884838116740(21)83, (01)00884838116764(21)406, (01)00884838116733(21)7, (01)00884838116771(21)737, (01)00884838116801(21)309, (01)00884838116771(21)114, (01)00884838116771(21)261, (01)00884838116801(21)278, (01)00884838116801(21)414, (01)00884838116764(21)26, (01)00884838116764(21)27, (01)00884838116788(21)94, (01)00884838116771(21)808, (01)00884838116788(21)104, (01)00884838116771(21)1331, (01)00884838116771(21)850, (01)00884838116771(21)397, (01)00884838116771(21)952, (01)00884838116801(21)271, (01)00884838116771(21)1052, (01)00884838116801(21)34, (01)00884838116764(21)235, (01)00884838116801(21)222, (01)00884838116801(21)294, (01)00884838116764(21)68, (01)00884838116771(21)683, (01)00884838116764(21)328, (01)00884838116801(21)446, (01)00884838116740(21)21, (01)00884838116771(21)503, (01)00884838116726(21)58, (01)00884838116801(21)435, (01)00884838116771(21)555, (01)00884838116771(21)556, (01)00884838116764(21)14, (01)00884838116771(21)36, (01)00884838116771(21)46, (01)00884838116764(21)98, (01)00884838116764(21)139, (01)00884838116788(21)70, (01)00884838116801(21)122, (01)00884838116726(21)24, (01)00884838116788(21)198, (01)00884838116801(21)441, (01)00884838116788(21)148, (01)00884838116771(21)678, (01)00884838116757(21)63, (01)00884838116771(21)1374, (01)00884838116801(21)374, (01)00884838116757(21)130, (01)00884838116757(21)132, (01)00884838116757(21)277, (01)00884838116757(21)261, (01)00884838116757(21)262, (01)00884838116764(21)117, (01)00884838116771(21)895, (01)00884838116771(21)974, (01)00884838116788(21)54, (01)00884838116771(21)709, (01)00884838116771(21)1281, (01)00884838116801(21)326, (01)00884838116733(21)15, (01)00884838116771(21)494, (01)00884838116771(21)757, (01)00884838116771(21)1072, (01)00884838116771(21)1247, (01)00884838116801(21)428, (01)00884838116764(21)279, (01)00884838116771(21)578, (01)00884838116801(21)438, (01)00884838116788(21)194, (01)00884838116771(21)755, (01)00884838116788(21)147, (01)00884838116764(21)355, (01)00884838116740(21)134, (01)00884838116740(21)78, (01)00884838116740(21)53, (01)00884838116726(21)35, (01)00884838116771(21)880, (01)00884838116757(21)89, (01)00884838116764(21)210, (01)00884838116740(21)47, (01)00884838116801(21)162, (01)00884838116771(21)1232, (01)00884838116801(21)93, (01)00884838116764(21)28, (01)00884838116801(21)245, (01)00884838116771(21)740, (01)00884838116801(21)241, (01)00884838116764(21)431, (01)00884838116764(21)453, (01)00884838116771(21)1335, (01)00884838116771(21)1396, (01)00884838116764(21)392, (01)00884838116764(21)391, (01)00884838116740(21)3, (01)00884838116788(21)47, (01)00884838116764(21)7, (01)00884838116771(21)27, (01)00884838116788(21)4, (01)00884838116801(21)207, (01)00884838116801(21)232, (01)00884838116771(21)981, (01)00884838116801(21)380, (01)00884838116771(21)947, (01)00884838116764(21)311, (01)00884838116764(21)138, (01)00884838116764(21)326, (01)00884838116726(21)16, (01)00884838116733(21)28, (01)00884838116771(21)316, (01)00884838116740(21)17, (01)00884838116740(21)18, (01)00884838116771(21)204, (01)00884838116764(21)31, (01)00884838116764(21)280, (01)00884838116764(21)281, (01)00884838116801(21)312, (01)00884838116726(21)50, (01)00884838116726(21)15, (01)00884838116764(21)165, (01)00884838116764(21)166, (01)00884838116740(21)25, (01)00884838116740(21)24, (01)00884838116757(21)86, (01)00884838116726(21)41, (01)00884838116771(21)745, (01)00884838116771(21)421, (01)00884838116771(21)907, (01)00884838116740(21)72, (01)00884838116764(21)62, (01)00884838116771(21)1423, (01)00884838116764(21)95, (01)00884838116771(21)632, (01)00884838116764(21)100, (01)00884838116764(21)29, (01)00884838116801(21)385, (01)00884838116801(21)248, (01)00884838116771(21)982, (01)00884838116757(21)69, (01)00884838116740(21)87, (01)00884838116740(21)93, (01)00884838116740(21)94, (01)00884838116771(21)945, (01)00884838116801(21)377, (01)00884838116740(21)112, (01)00884838116801(21)389, (01)00884838116771(21)259, (01)00884838116771(21)264, (01)00884838116771(21)265, (01)00884838116757(21)3, (01)00884838116771(21)710, (01)00884838116771(21)82, (01)00884838116733(21)90, (01)00884838116771(21)751, (01)00884838116801(21)216, (01)00884838116801(21)184, (01)00884838116771(21)353, (01)00884838116757(21)67, (01)00884838116801(21)219, (01)00884838116757(21)349, (01)00884838116771(21)701, (01)00884838116771(21)868, (01)00884838116764(21)203, (01)00884838116764(21)204, (01)00884838116764(21)205, (01)00884838116771(21)566, (01)00884838116771(21)567, (01)00884838116733(21)60, (01)00884838116771(21)1217, (01)00884838116771(21)1218, (01)00884838116771(21)1216, (01)00884838116771(21)720, (01)00884838116771(21)733, (01)00884838116771(21)739, (01)00884838116801(21)311, (01)00884838116771(21)91, (01)00884838116801(21)177, (01)00884838116771(21)134, (01)00884838116757(21)300, (01)00884838116757(21)39, (01)00884838116771(21)296, (01)00884838116771(21)190, (01)00884838116771(21)1100, (01)00884838116757(21)330, (01)00884838116771(21)304, (01)00884838116771(21)106, (01)00884838116771(21)307, (01)00884838116771(21)1357, (01)00884838116757(21)4, (01)00884838116757(21)44, (01)00884838116771(21)30, (01)00884838116801(21)56, (01)00884838116764(21)331, (01)00884838116771(21)700, (01)00884838116771(21)1121, (01)00884838116771(21)570, (01)00884838116771(21)192, (01)00884838116788(21)92, (01)00884838116764(21)389, (01)00884838116757(21)37, (01)00884838116771(21)658, (01)00884838116771(21)722, (01)00884838116801(21)57, (01)00884838116726(21)51, (01)00884838116764(21)120, (01)00884838116764(21)310, (01)00884838116740(21)34, (01)00884838116740(21)88, (01)00884838116801(21)452, (01)00884838116801(21)169, (01)00884838116801(21)91, (01)00884838116764(21)423, (01)00884838116771(21)764, (01)00884838116764(21)285, (01)00884838116764(21)286, (01)00884838116771(21)687, (01)00884838116771(21)93, (01)00884838116764(21)292, (01)00884838116801(21)167, (01)00884838116764(21)161, (01)00884838116764(21)252, (01)00884838116771(21)86, (01)00884838116771(21)158, (01)00884838116771(21)715, (01)00884838116764(21)294, (01)00884838116771(21)283, (01)00884838116740(21)114, (01)00884838116801(21)371, (01)00884838116764(21)175, (01)00884838116764(21)110, (01)00884838116764(21)112, (01)00884838116764(21)173, (01)00884838116771(21)103, (01)00884838116740(21)40, (01)00884838116801(21)44, (01)00884838116726(21)5, (01)00884838116764(21)398, (01)00884838116788(21)139, (01)00884838116788(21)137, (01)00884838116771(21)120, (01)00884838116757(21)93, (01)00884838116757(21)110, (01)00884838116771(21)415, (01)00884838116771(21)928, (01)00884838116733(21)17, (01)00884838116757(21)327, (01)00884838116757(21)328, (01)00884838116771(21)1158, (01)00884838116771(21)1033, (01)00884838116801(21)475, (01)00884838116771(21)869, (01)00884838116771(21)846, (01)00884838116771(21)143, (01)00884838116771(21)712, (01)00884838116771(21)224, (01)00884838116771(21)180, (01)00884838116771(21)238, (01)00884838116771(21)908, (01)00884838116771(21)688, (01)00884838116771(21)1164, (01)00884838116771(21)1089, (01)00884838116801(21)156, (01)00884838116771(21)282, (01)00884838116733(21)38, (01)00884838116801(21)324, (01)00884838116771(21)464, (01)00884838116771(21)756, (01)00884838116757(21)366, (01)00884838116764(21)336, (01)00884838116788(21)79, (01)00884838116740(21)137, (01)00884838116771(21)64, (01)00884838116771(21)66, (01)00884838116771(21)83, (01)00884838116771(21)90, (01)00884838116788(21)93, (01)00884838116771(21)587, (01)00884838116771(21)845, (01)00884838116771(21)53, (01)00884838116771(21)131, (01)00884838116771(21)179, (01)00884838116801(21)538, (01)00884838116764(21)485, (01)00884838116757(21)1, (01)00884838116726(21)3, (01)00884838116771(21)1144, (01)00884838116740(21)69, (01)00884838116771(21)588, (01)00884838116757(21)68, (01)00884838116801(21)375, (01)00884838116801(21)274, (01)00884838116740(21)123, (01)00884838116771(21)200, (01)00884838116771(21)243, (01)00884838116764(21)160, (01)00884838116788(21)172, (01)00884838116764(21)359, (01)00884838116771(21)1161, (01)00884838116757(21)11, (01)00884838116771(21)1173, (01)00884838116771(21)302, (01)00884838116764(21)408, (01)00884838116771(21)278, (01)00884838116771(21)277, (01)00884838116771(21)56, (01)00884838116771(21)499, (01)00884838116771(21)910, (01)00884838116740(21)61, (01)00884838116757(21)77, (01)00884838116801(21)275, (01)00884838116771(21)1184, (01)00884838116771(21)761, (01)00884838116740(21)142, (01)00884838116764(21)92, (01)00884838116764(21)93, (01)00884838116771(21)16, (01)00884838116740(21)26, (01)00884838116764(21)51, (01)00884838116771(21)690, (01)00884838116801(21)344, (01)00884838116771(21)920, (01)00884838116801(21)510, (01)00884838116764(21)75, (01)00884838116764(21)202, (01)00884838116801(21)237, (01)00884838116757(21)20, (01)00884838116771(21)455, (01)00884838116801(21)113, (01)00884838116733(21)56, (01)00884838116771(21)888, (01)00884838116771(21)562, (01)00884838116801(21)46, (01)00884838116771(21)851, (01)00884838116764(21)86, (01)00884838116771(21)961, (01)00884838116764(21)179, (01)00884838116771(21)702, (01)00884838116733(21)73, (01)00884838116733(21)74, (01)00884838116733(21)75, (01)00884838116733(21)76, (01)00884838116733(21)77, (01)00884838116733(21)78, (01)00884838116733(21)79, (01)00884838116733(21)80, (01)00884838116733(21)81, (01)00884838116733(21)82, (01)00884838116733(21)83, (01)00884838116733(21)84, (01)00884838116733(21)85, (01)00884838116733(21)86, (01)00884838116740(21)80, (01)00884838116764(21)181, (01)00884838116801(21)290, (01)00884838116764(21)241, (01)00884838116733(21)9, (01)00884838116740(21)115, (01)00884838116764(21)197, (01)00884838116771(21)552, (01)00884838116764(21)429, (01)00884838116733(21)88, (01)00884838116740(21)136, (01)00884838116764(21)58, (01)00884838116788(21)72, (01)00884838116740(21)6, (01)00884838116740(21)7, (01)00884838116740(21)8, (01)00884838116757(21)7, (01)00884838116788(21)27, (01)00884838116788(21)78, (01)00884838116771(21)424, (01)00884838116801(21)384, (01)00884838116801(21)386, (01)00884838116801(21)33, (01)00884838116764(21)176, (01)00884838116801(21)292, (01)00884838116771(21)558, (01)00884838116733(21)62, (01)00884838116733(21)64, (01)00884838116740(21)4, (01)00884838116740(21)5, (01)00884838116764(21)223, (01)00884838116757(21)260, (01)00884838116757(21)57, (01)00884838116757(21)58, (01)00884838116771(21)40, (01)00884838116801(21)66, (01)00884838116788(21)181, (01)00884838116771(21)1231, (01)00884838116764(21)211, (01)00884838116788(21)183, (01)00884838116740(21)132, (01)00884838116771(21)15, (01)00884838116740(21)99, (01)00884838116764(21)447, (01)00884838116764(21)136, (01)00884838116764(21)227, (01)00884838116788(21)170, (01)00884838116771(21)473, (01)00884838116801(21)541, (01)00884838116801(21)165, (01)00884838116757(21)71, (01)00884838116757(21)72, (01)00884838116801(21)234, (01)00884838116801(21)163, (01)00884838116771(21)544, (01)00884838116764(21)464, (01)00884838116771(21)1392, (01)00884838116764(21)452, (01)00884838116740(21)139, (01)00884838116801(21)176, (01)00884838116764(21)463, (01)00884838116771(21)859, (01)00884838116801(21)283, (01)00884838116801(21)284, (01)00884838116801(21)285, (01)00884838116764(21)81, (01)00884838116801(21)182, (01)00884838116771(21)504, (01)00884838116801(21)193, (01)00884838116757(21)332, (01)00884838116764(21)330, (01)00884838116764(21)16, (01)00884838116733(21)68, (01)00884838116771(21)468, (01)00884838116764(21)24, (01)00884838116771(21)1008, (01)00884838116740(21)12, (01)00884838116764(21)184, (01)00884838116757(21)14, (01)00884838116757(21)82, (01)00884838116771(21)645, (01)00884838116757(21)115, (01)00884838116733(21)63, (01)00884838116757(21)334, (01)00884838116801(21)360, (01)00884838116771(21)448, (01)00884838116764(21)129, (01)00884838116764(21)222, (01)00884838116764(21)158, (01)00884838116764(21)156, (01)00884838116771(21)1125, (01)00884838116771(21)1224, (01)00884838116771(21)399, (01)00884838116801(21)406, (01)00884838116764(21)486, (01)00884838116771(21)991, (01)00884838116771(21)992, (01)00884838116771(21)1001, (01)00884838116740(21)63, (01)00884838116726(21)31, (01)00884838116788(21)14, (01)00884838116788(21)20, (01)00884838116771(21)523, (01)00884838116764(21)254, (01)00884838116771(21)576, (01)00884838116764(21)256, (01)00884838116771(21)575, (01)00884838116771(21)51, (01)00884838116757(21)301, (01)00884838116764(21)253, (01)00884838116771(21)612, (01)00884838116757(21)70, (01)00884838116764(21)30, (01)00884838116771(21)181, (01)00884838116771(21)182, (01)00884838116771(21)579, (01)00884838116801(21)287, (01)00884838116771(21)1137, (01)00884838116801(21)347, (01)00884838116801(21)321, (01)00884838116764(21)284, (01)00884838116764(21)134, (01)00884838116764(21)143, (01)00884838116771(21)1127, (01)00884838116771(21)1270, (01)00884838116771(21)1367, (01)00884838116771(21)96, (01)00884838116740(21)73, (01)00884838116764(21)233, (01)00884838116764(21)162, (01)00884838116764(21)71, (01)00884838116771(21)235, (01)00884838116771(21)236, (01)00884838116771(21)237, (01)00884838116771(21)1167, (01)00884838116757(21)122, (01)00884838116757(21)289, (01)00884838116757(21)341, (01)00884838116757(21)370, (01)00884838116788(21)109, (01)00884838116757(21)323, (01)00884838116788(21)108, (01)00884838116757(21)145, (01)00884838116764(21)40, (01)00884838116771(21)1364, (01)00884838116801(21)425, (01)00884838116801(21)394, (01)00884838116757(21)112, (01)00884838116788(21)56, (01)00884838116771(21)177, (01)00884838116757(21)123, (01)00884838116757(21)239, (01)00884838116757(21)254, (01)00884838116757(21)336, (01)00884838116757(21)359, (01)00884838116757(21)372, (01)00884838116757(21)154, (01)00884838116757(21)368, (01)00884838116757(21)347, (01)00884838116757(21)258, (01)00884838116733(21)87, (01)00884838116801(21)397, (01)00884838116757(21)319, (01)00884838116757(21)305, (01)00884838116757(21)321, (01)00884838116757(21)371, (01)00884838116757(21)151, (01)00884838116757(21)340, (01)00884838116801(21)498, (01)00884838116801(21)398, (01)00884838116757(21)337, (01)00884838116757(21)322, (01)00884838116757(21)339, (01)00884838116757(21)311, (01)00884838116757(21)335, (01)00884838116771(21)953, (01)00884838116757(21)228, (01)00884838116801(21)443, (01)00884838116757(21)147, (01)00884838116757(21)292, (01)00884838116757(21)221, (01)00884838116757(21)248, (01)00884838116757(21)362, (01)00884838116757(21)320, (01)00884838116757(21)338, (01)00884838116757(21)313, (01)00884838116757(21)291, (01)00884838116757(21)148, (01)00884838116801(21)445, (01)00884838116757(21)229, (01)00884838116757(21)342, (01)00884838116757(21)346, (01)00884838116757(21)343, (01)00884838116757(21)348, (01)00884838116757(21)374, (01)00884838116757(21)358, (01)00884838116757(21)144, (01)00884838116733(21)57, (01)00884838116757(21)230, (01)00884838116757(21)307, (01)00884838116757(21)143, (01)00884838116757(21)167, (01)00884838116757(21)150, (01)00884838116757(21)344, (01)00884838116757(21)224, (01)00884838116757(21)235, (01)00884838116757(21)354, (01)00884838116757(21)173, (01)00884838116757(21)312, (01)00884838116801(21)419, (01)00884838116757(21)213, (01)00884838116757(21)324, (01)00884838116757(21)212, (01)00884838116757(21)152, (01)00884838116757(21)222, (01)00884838116757(21)146, (01)00884838116757(21)373, (01)00884838116801(21)396, (01)00884838116757(21)237, (01)00884838116757(21)149, (01)00884838116757(21)325, (01)00884838116757(21)345, (01)00884838116757(21)283, (01)00884838116757(21)219, (01)00884838116733(21)58, (01)00884838116757(21)157, (01)00884838116801(21)399, (01)00884838116771(21)883, (01)00884838116801(21)500, (01)00884838116764(21)48, (01)00884838116771(21)1022, (01)00884838116740(21)56, (01)00884838116771(21)1150, (01)00884838116764(21)376, (01)00884838116801(21)484, (01)00884838116771(21)506, (01)00884838116801(21)8, (01)00884838116726(21)29, (01)00884838116771(21)933, (01)00884838116740(21)14, (01)00884838116801(21)472, (01)00884838116771(21)246, (01)00884838116771(21)310, (01)0088483811680",
      "more_code_info": "1(21)269, (01)00884838116771(21)634, (01)00884838116757(21)294, (01)00884838116801(21)411, (01)00884838116801(21)531, (01)00884838116726(21)46, (01)00884838116771(21)1192, (01)00884838116771(21)414, (01)00884838116801(21)102, (01)00884838116733(21)54, (01)00884838116733(21)71, (01)00884838116771(21)495, (01)00884838116771(21)487, (01)00884838116771(21)600, (01)00884838116788(21)106, (01)00884838116771(21)1389, (01)00884838116801(21)297, (01)00884838116771(21)400, (01)00884838116801(21)316, (01)00884838116771(21)749, (01)00884838116726(21)4, (01)00884838116771(21)1106, (01)00884838116771(21)1018, (01)00884838116771(21)1390, (01)00884838116757(21)81, (01)00884838116771(21)876, (01)00884838116764(21)101, (01)00884838116726(21)38, (01)00884838116771(21)95, (01)00884838116764(21)180, (01)00884838116771(21)1410, (01)00884838116771(21)754, (01)00884838116771(21)930, (01)00884838116726(21)33, (01)00884838116726(21)27, (01)00884838116801(21)543, (01)00884838116771(21)104, (01)00884838116764(21)105, (01)00884838116733(21)55, (01)00884838116764(21)324, (01)00884838116771(21)481, (01)00884838116764(21)106, (01)00884838116764(21)133, (01)00884838116771(21)313, (01)00884838116801(21)132, (01)00884838116771(21)1153, (01)00884838116740(21)48, (01)00884838116788(21)30, (01)00884838116788(21)31, (01)00884838116757(21)131, (01)00884838116771(21)983, (01)00884838116771(21)1006, (01)00884838116801(21)157, (01)00884838116726(21)37, (01)00884838116764(21)57, (01)00884838116740(21)67, (01)00884838116764(21)61, (01)00884838116788(21)142, (01)00884838116764(21)186, (01)00884838116764(21)189, (01)00884838116771(21)498, (01)00884838116771(21)782, (01)00884838116771(21)792, (01)00884838116771(21)826, (01)00884838116771(21)497, (01)00884838116771(21)508, (01)00884838116801(21)224, (01)00884838116801(21)487, (01)00884838116801(21)432, (01)00884838116757(21)356, (01)00884838116757(21)309, (01)00884838116788(21)107, (01)00884838116764(21)370, (01)00884838116771(21)1257, (01)00884838116801(21)331, (01)00884838116764(21)365, (01)00884838116801(21)98, (01)00884838116771(21)39, (01)00884838116764(21)89, (01)00884838116764(21)90, (01)00884838116764(21)91, (01)00884838116801(21)168, (01)00884838116771(21)337, (01)00884838116771(21)949, (01)00884838116771(21)1114, (01)00884838116740(21)124, (01)00884838116801(21)454, (01)00884838116801(21)189, (01)00884838116764(21)145, (01)00884838116771(21)250, (01)00884838116764(21)439, (01)00884838116764(21)385, (01)00884838116801(21)30, (01)00884838116771(21)1091, (01)00884838116764(21)44, (01)00884838116788(21)187, (01)00884838116740(21)71, (01)00884838116771(21)281, (01)00884838116764(21)231, (01)00884838116764(21)338, (01)00884838116771(21)680, (01)00884838116764(21)288, (01)00884838116788(21)125, (01)00884838116801(21)188, (01)00884838116757(21)128, (01)00884838116740(21)79, (01)00884838116788(21)155, (01)00884838116801(21)488, (01)00884838116801(21)370, (01)00884838116801(21)146, (01)00884838116726(21)32, (01)00884838116771(21)195, (01)00884838116801(21)125, (01)00884838116771(21)491, (01)00884838116726(21)52, (01)00884838116771(21)21, (01)00884838116764(21)268, (01)00884838116771(21)776, (01)00884838116801(21)333, (01)00884838116801(21)334, (01)00884838116764(21)312, (01)00884838116801(21)434, (01)00884838116771(21)341, (01)00884838116771(21)1057, (01)00884838116740(21)60, (01)00884838116771(21)614, (01)00884838116764(21)85, (01)00884838116764(21)221, (01)00884838116740(21)28, (01)00884838116733(21)39, (01)00884838116771(21)791, (01)00884838116726(21)53, (01)00884838116757(21)31, (01)00884838116771(21)1092, (01)00884838116788(21)160, (01)00884838116801(21)437, (01)00884838116788(21)10, (01)00884838116771(21)631, (01)00884838116771(21)589, (01)00884838116740(21)66, (01)00884838116764(21)193, (01)00884838116764(21)194, (01)00884838116764(21)349, (01)00884838116764(21)83, (01)00884838116764(21)84, (01)00884838116764(21)87, (01)00884838116764(21)88, (01)00884838116771(21)692, (01)00884838116726(21)6, (01)00884838116771(21)1086, (01)00884838116771(21)775, (01)00884838116771(21)1229, (01)00884838116740(21)75, (01)00884838116764(21)19, (01)00884838116771(21)242, (01)00884838116740(21)81, (01)00884838116788(21)3, (01)00884838116771(21)820, (01)00884838116764(21)5, (01)00884838116764(21)6, (01)00884838116771(21)627, (01)00884838116771(21)942, (01)00884838116726(21)2, (01)00884838116771(21)1383, (01)00884838116801(21)405, (01)00884838116757(21)375, (01)00884838116764(21)18, (01)00884838116771(21)490, (01)00884838116740(21)133, (01)00884838116788(21)192, (01)00884838116788(21)99, (01)00884838116771(21)1147, (01)00884838116757(21)120, (01)00884838116757(21)121, (01)00884838116771(21)137, (01)00884838116771(21)138, (01)00884838116771(21)1267, (01)00884838116771(21)417, (01)00884838116771(21)314, (01)00884838116771(21)635, (01)00884838116771(21)308, (01)00884838116764(21)178, (01)00884838116764(21)343, (01)00884838116764(21)314, (01)00884838116726(21)39, (01)00884838116771(21)41, (01)00884838116771(21)245, (01)00884838116801(21)27, (01)00884838116764(21)332, (01)00884838116757(21)35, (01)00884838116801(21)16, (01)00884838116801(21)145, (01)00884838116771(21)1397, (01)00884838116771(21)149, (01)00884838116771(21)65, (01)00884838116726(21)20, (01)00884838116726(21)26, (01)00884838116740(21)45, (01)00884838116771(21)665, (01)00884838116757(21)297, (01)00884838116771(21)717, (01)00884838116726(21)43, (01)00884838116771(21)1085, (01)00884838116771(21)721, (01)00884838116788(21)41, (01)00884838116740(21)20, (01)00884838116764(21)257, (01)00884838116764(21)55, (01)00884838116771(21)38, (01)00884838116771(21)1195, (01)00884838116771(21)145, (01)00884838116764(21)373, (01)00884838116771(21)1237, (01)00884838116771(21)1326, (01)00884838116757(21)293, (01)00884838116757(21)360, (01)00884838116801(21)329, (01)00884838116801(21)415, (01)00884838116733(21)27, (01)00884838116801(21)407, (01)00884838116757(21)28, (01)00884838116757(21)30, (01)00884838116757(21)54, (01)00884838116771(21)1017, (01)00884838116764(21)321, (01)00884838116771(21)651, (01)00884838116764(21)239, (01)00884838116764(21)240, (01)00884838116764(21)243, (01)00884838116771(21)445, (01)00884838116771(21)1384, (01)00884838116771(21)650, (01)00884838116771(21)17, (01)00884838116764(21)295, (01)00884838116771(21)743, (01)00884838116764(21)130, (01)00884838116771(21)355, (01)00884838116757(21)19, (01)00884838116757(21)102, (01)00884838116757(21)104, (01)00884838116764(21)432, (01)00884838116764(21)234, (01)00884838116757(21)107, (01)00884838116733(21)52, (01)00884838116740(21)19, (01)00884838116771(21)881, (01)00884838116771(21)1199, (01)00884838116764(21)9, (01)00884838116801(21)100, (01)00884838116757(21)296, (01)00884838116740(21)51, (01)00884838116801(21)123, (01)00884838116771(21)305, (01)00884838116764(21)390, (01)00884838116771(21)1058, (01)00884838116771(21)148, (01)00884838116771(21)1009, (01)00884838116757(21)137, (01)00884838116771(21)572, (01)00884838116757(21)29, (01)00884838116757(21)56, (01)00884838116771(21)727, (01)00884838116740(21)29, (01)00884838116757(21)51, (01)00884838116771(21)443, (01)00884838116771(21)12, (01)00884838116771(21)13, (01)00884838116764(21)372, (01)00884838116764(21)371, (01)00884838116771(21)1190, (01)00884838116771(21)1191, (01)00884838116771(21)819, (01)00884838116764(21)200, (01)00884838116771(21)452, (01)00884838116801(21)23, (01)00884838116788(21)91, (01)00884838116764(21)32, (01)00884838116801(21)338, (01)00884838116771(21)254, (01)00884838116801(21)96, (01)00884838116757(21)32, (01)00884838116801(21)181, (01)00884838116740(21)27, (01)00884838116764(21)159, (01)00884838116764(21)217, (01)00884838116740(21)100, (01)00884838116764(21)152, (01)00884838116801(21)45, (01)00884838116764(21)340, (01)00884838116757(21)299, (01)00884838116740(21)58, (01)00884838116740(21)11, (01)00884838116757(21)134, (01)00884838116740(21)85, (01)00884838116764(21)260, (01)00884838116764(21)337, (01)00884838116788(21)153, (01)00884838116764(21)56, (01)00884838116740(21)57, (01)00884838116740(21)35, (01)00884838116801(21)357, (01)00884838116764(21)20, (01)00884838116771(21)49, (01)00884838116801(21)258, (01)00884838116771(21)1030, (01)00884838116740(21)62, (01)00884838116771(21)328, (01)00884838116771(21)169, (01)00884838116726(21)10, (01)00884838116771(21)1095, (01)00884838116801(21)130, (01)00884838116757(21)105, (01)00884838116764(21)225, (01)00884838116771(21)649, (01)00884838116757(21)326, (01)00884838116757(21)75, (01)00884838116764(21)291, (01)00884838116740(21)54, (01)00884838116771(21)105, (01)00884838116740(21)33, (01)00884838116801(21)358, (01)00884838116771(21)892, (01)00884838116733(21)13, (01)00884838116740(21)92, (01)00884838116771(21)157, (01)00884838116771(21)73, (01)00884838116733(21)12, (01)00884838116801(21)114, (01)00884838116764(21)50, (01)00884838116764(21)393, (01)00884838116771(21)1235, (01)00884838116764(21)11, (01)00884838116771(21)23, (01)00884838116764(21)434, (01)00884838116757(21)16, (01)00884838116740(21)1, (01)00884838116740(21)2, (01)00884838116801(21)154, (01)00884838116740(21)97, (01)00884838116764(21)72, (01)00884838116788(21)202, (01)00884838116801(21)279, (01)00884838116771(21)611, (01)00884838116801(21)395, (01)00884838116771(21)1302, (01)00884838116771(21)958, (01)00884838116771(21)959, (01)00884838116771(21)446, (01)00884838116771(21)731, (01)00884838116771(21)677, (01)00884838116771(21)741, (01)00884838116771(21)837, (01)00884838116801(21)263, (01)00884838116788(21)133, (01)00884838116757(21)117, (01)00884838116764(21)313, (01)00884838116801(21)264, (01)00884838116764(21)395, (01)00884838116771(21)1029, (01)00884838116764(21)107, (01)00884838116771(21)115, (01)00884838116771(21)338, (01)00884838116771(21)571, (01)00884838116764(21)35, (01)00884838116764(21)394, (01)00884838116771(21)1236, (01)00884838116764(21)420, (01)00884838116726(21)11, (01)00884838116771(21)439, (01)00884838116726(21)21, (01)00884838116757(21)308, (01)00884838116757(21)315, (01)00884838116771(21)219, (01)00884838116771(21)384, (01)00884838116801(21)353, (01)00884838116771(21)599, (01)00884838116764(21)46, (01)00884838116764(21)47, (01)00884838116801(21)533, (01)00884838116757(21)295, (01)00884838116788(21)131, (01)00884838116764(21)451, (01)00884838116771(21)44, (01)00884838116771(21)1271, (01)00884838116764(21)74, (01)00884838116771(21)1176, (01)00884838116764(21)43, (01)00884838116764(21)60, (01)00884838116771(21)1365, (01)00884838116801(21)418, (01)00884838116771(21)132, (01)00884838116771(21)1376, (01)00884838116771(21)1238, (01)00884838116771(21)1050, (01)00884838116740(21)32, (01)00884838116801(21)126, (01)00884838116771(21)713, (01)00884838116764(21)208, (01)00884838116764(21)209, (01)00884838116771(21)478, (01)00884838116771(21)479, (01)00884838116771(21)480, (01)00884838116788(21)82, (01)00884838116788(21)83, (01)00884838116788(21)84, (01)00884838116771(21)378, (01)00884838116771(21)406, (01)00884838116764(21)104, (01)00884838116771(21)19, (01)00884838116771(21)975, (01)00884838116771(21)724, (01)00884838116771(21)604, (01)00884838116771(21)241, (01)00884838116771(21)57, (01)00884838116771(21)742, (01)00884838116771(21)835, (01)00884838116764(21)78, (01)00884838116771(21)135, (01)00884838116771(21)374, (01)00884838116771(21)505, (01)00884838116801(21)128, (01)00884838116771(21)303, (01)00884838116764(21)290, (01)00884838116771(21)387, (01)00884838116771(21)388, (01)00884838116771(21)1090, (01)00884838116801(21)86, (01)00884838116764(21)272, (01)00884838116801(21)94, (01)00884838116764(21)249, (01)00884838116764(21)77, (01)00884838116764(21)80, (01)00884838116771(21)142, (01)00884838116764(21)174, (01)00884838116771(21)802, (01)00884838116771(21)58, (01)00884838116771(21)474, (01)00884838116771(21)390, (01)00884838116801(21)379, (01)00884838116764(21)303, (01)00884838116764(21)163, (01)00884838116771(21)339, (01)00884838116771(21)239, (01)00884838116740(21)59, (01)00884838116771(21)247, (01)00884838116726(21)25, (01)00884838116726(21)13, (01)00884838116733(21)1, (01)00884838116764(21)97, (01)00884838116726(21)8, (01)00884838116771(21)531, (01)00884838116764(21)198, (01)00884838116771(21)249, (01)00884838116771(21)364, (01)00884838116771(21)365, (01)00884838116801(21)160, (01)00884838116764(21)102, (01)00884838116801(21)310, (01)00884838116771(21)127, (01)00884838116771(21)569, (01)00884838116771(21)47, (01)00884838116801(21)41, (01)00884838116771(21)685, (01)00884838116771(21)229, (01)00884838116771(21)360, (01)00884838116771(21)765, (01)00884838116771(21)784, (01)00884838116801(21)410, (01)00884838116771(21)773, (01)00884838116788(21)100, (01)00884838116726(21)23, (01)00884838116733(21)65, (01)00884838116801(21)170, (01)00884838116801(21)82, (01)00884838116771(21)596, (01)00884838116801(21)63, (01)00884838116757(21)303, (01)00884838116764(21)146, (01)00884838116764(21)148, (01)00884838116764(21)149, (01)00884838116771(21)361, (01)00884838116771(21)425, (01)00884838116771(21)1388, (01)00884838116801(21)268, (01)00884838116764(21)450, (01)00884838116788(21)126, (01)00884838116771(21)950, (01)00884838116771(21)52, (01)00884838116801(21)328, (01)00884838116771(21)294, (01)00884838116771(21)840, (01)00884838116801(21)277, (01)00884838116726(21)49, (01)00884838116771(21)1124, (01)00884838116771(21)766, (01)00884838116757(21)73, (01)00884838116771(21)844, (01)00884838116771(21)189, (01)00884838116771(21)1194, (01)00884838116764(21)362, (01)00884838116764(21)470, (01)00884838116771(21)1401, (01)00884838116764(21)357, (01)00884838116771(21)1155, (01)00884838116771(21)18, (01)00884838116764(21)109, (01)00884838116771(21)312, (01)00884838116801(21)499, (01)00884838116771(21)944, (01)00884838116801(21)450, (01)00884838116801(21)529, (01)00884838116771(21)1337, (01)00884838116801(21)303, (01)00884838116788(21)66, (01)00884838116788(21)67, (01)00884838116801(21)111, (01)00884838116801(21)172, (01)00884838116801(21)112, (01)00884838116788(21)140, (01)00884838116801(21)259, (01)00884838116788(21)96, (01)00884838116771(21)1099, (01)00884838116801(21)318, (01)00884838116801(21)103, (01)00884838116771(21)978, (01)00884838116788(21)63, (01)00884838116801(21)75, (01)00884838116788(21)120, (01)00884838116788(21)34, (01)00884838116801(21)84, (01)00884838116757(21)100, (01)00884838116801(21)52, (01)00884838116771(21)156, (01)00884838116801(21)388, (01)00884838116788(21)203, (01)00884838116801(21)356, (01)00884838116764(21)15, (01)00884838116788(21)28, (01)00884838116788(21)17, (01)00884838116788(21)18, (01)00884838116801(21)73, (01)00884838116788(21)85, (01)00884838116771(21)1049, (01)00884838116788(21)112, (01)00884838116801(21)80, (01)00884838116771(21)301, (01)00884838116788(21)65, (01)00884838116801(21)175, (01)00884838116801(21)306, (01)00884838116801(21)305, (01)00884838116788(21)144, (01)00884838116771(21)943, (01)00884838116801(21)92, (01)00884838116771(21)152, (01)00884838116771(21)279, (01)00884838116788(21)40, (01)00884838116801(21)155, (01)00884838116788(21)143, (01)00884838116771(21)197, (01)00884838116788(21)45, (01)00884838116771(21)33, (01)00884838116801(21)55, (01)00884838116771(21)871, (01)00884838116788(21)44, (01)00884838116771(21)184, (01)00884838116788(21)42, (01)00884838116801(21)461, (01)00884838116764(21)351, (01)00884838116801(21)85, (01)00884838116801(21)366, (01)00884838116801(21)367, (01)00884838116771(21)32, (01)00884838116764(21)13, (01)00884838116801(21)535, (01)00884838116788(21)52, (01)00884838116801(21)142, (01)00884838116788(21)113, (01)00884838116788(21)49, (01)00884838116801(21)105, (01)00884838116788(21)101, (01)00884838116771(21)1162, (01)00884838116771(21)20, (01)00884838116801(21)108, (01)00884838116801(21)53, (01)00884838116788(21)53, (01)00884838116801(21)144, (01)00884838116801(21)198, (01)00884838116801(21)141, (01)00884838116801(21)299, (01)00884838116801(21)298, (01)00884838116801(21)124, (01)00884838116788(21)48, (01)00884838116801(21)101, (01)00884838116788(21)81, (01)00884838116801(21)320, (01)00884838116788(21)103, (01)00884838116764(21)39, (01)00884838116771(21)31, (01)00884838116788(21)152, (01)00884838116771(21)909, (01)00884838116801(21)340, (01)00884838116788(21)25, (01)00884838116801(21)390, (01)00884838116801(21)365, (01)00884838116801(21)62, (01)00884838116801(21)261, (01)00884838116801(21)88, (01)00884838116801(21)89, (01)00884838116801(21)109, (01)00884838116788(21)24, (01)00884838116801(21)83, (01)00884838116764(21)206, (01)00884838116801(21)171, (01)00884838116771(21)26, (01)00884838116764(21)67, (01)00884838116788(21)6, (01)00884838116801(21)317, (01)00884838116801(21)240, (01)00884838116801(21)76, (01)00884838116788(21)38, (01)00884838116788(21)171, (01)00884838116788(21)39, (01)00884838116801(21)106, (01)00884838116801(21)90, (01)00884838116788(21)7, (01)00884838116801(21)251, (01)00884838116801(21)229, (01)00884838116771(21)563, (01)00884838116771(21)29, (01)00884838116788(21)5, (01)00884838116801(21)49, (01)00884838116771(21)1323, (01)00884838116801(21)314, (01)00884838116788(21)22, (01)00884838116801(21)77, (01)00884838116801(21)409, (01)00884838116788(21)141, (01)00884838116788(21)8, (01)00884838116771(21)870, (01)00884838116801(21)364, (01)00884838116801(21)74, (01)00884838116801(21)387, (01)00884838116788(21)110, (01)00884838116771(21)1166, (01)00884838116771(21)971, (01)00884838116764(21)248, (01)00884838116788(21)58, (01)00884838116771(21)272, (01)00884838116801(21)382, (01)00884838116771(21)553, (01)00884838116801(21)519, (01)00884838116801(21)78, (01)00884838116788(21)111, (01)00884838116788(21)98, (01)00884838116757(21)13, (01)00884838116788(21)138, (01)00884838116788(21)146, (01)00884838116788(21)209, (01)00884838116788(21)64, (01)00884838116771(21)746, (01)00884838116788(21)62, (01)00884838116801(21)161, (01)00884838116788(21)9, (01)00884838116801(21)59, (01)00884838116801(21)206, (01)00884838116771(21)736, (01)00884838116771(21)559, (01)00884838116788(21)35, (01)00884838116801(21)104, (01)00884838116801(21)70, (01)00884838116801(21)107, (01)00884838116771(21)856, (01)00884838116801(21)250, (01)00884838116801(21)64, (01)00884838116788(21)145, (01)00884838116801(21)469, (01)00884838116801(21)151, (01)00884838116801(21)152, (01)00884838116788(21)19, (01)00884838116801(21)293, (01)00884838116801(21)69, (01)00884838116788(21)23, (01)00884838116788(21)21, (01)00884838116801(21)471, (01)00884838116801(21)153, (01)00884838116801(21)301, (01)00884838116788(21)162, (01)00884838116788(21)33."
    }
  ]
}