{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-07",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99422",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-2931-2026",
      "product_description": "Philips Azurion  Rystem, Model Numbers:  722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 722225, 722226, 722227, 722228, 722280*, 722281*, and 722282*  (*These models are distributed only in China), with Software Version Number R2.2.10 and R2.1.10* (*This software version is distributed only in China); Interventional Fluoroscopic X-Ray System",
      "product_quantity": "10872 units",
      "reason_for_recall": "Two software related issues affecting Azurion R2.2.10 Systems:    1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and   2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications",
      "recall_initiation_date": "20260716",
      "center_classification_date": "20260810",
      "report_date": "20260819",
      "code_info": "UDI:  (01)00884838099258(21)1229, (01)00884838099258(21)450, (01)00884838085251(21)302, (01)00884838099258(21)1576, (01)00884838099258(21)1577, (01)00884838099258(21)1578, (01)00884838099241(21)324, (01)00884838099241(21)585, (01)00884838099241(21)332, (01)00884838099258(21)750, (01)00884838099241(21)312, (01)00884838099241(21)85, (01)00884838099258(21)409, (01)00884838085268(21)1023, (01)00884838099272(21)125, (01)00884838099272(21)127, (01)00884838099272(21)126, (01)00884838099272(21)122, (01)00884838099272(21)123, (01)00884838099258(21)295, (01)00884838099272(21)124, (01)00884838099272(21)120, (01)00884838099258(21)277, (01)00884838099258(21)262, (01)00884838099258(21)261, (01)00884838099258(21)272, (01)00884838099258(21)275, (01)00884838085367(21)547, (01)00884838099258(21)630, (01)00884838099258(21)1003, (01)00884838085251(21)585, (01)00884838099258(21)1315, (01)00884838085268(21)645, (01)00884838085251(21)410, (01)00884838085268(21)646, (01)00884838099258(21)3526, (01)00884838099258(21)2744, (01)00884838099258(21)3698, (01)00884838099258(21)959, (01)00884838099258(21)381, (01)00884838085268(21)1013, (01)00884838085268(21)1412, (01)00884838099258(21)232, (01)00884838099258(21)1371, (01)00884838099241(21)533, (01)00884838085268(21)1387, (01)00884838099272(21)153, (01)00884838099258(21)1019, (01)00884838085268(21)460, (01)00884838099258(21)260, (01)00884838085251(21)753, (01)00884838085268(21)35, (01)00884838085268(21)40, (01)00884838085268(21)43, (01)00884838099272(21)67, (01)00884838099258(21)1049, (01)00884838099258(21)3045, (01)00884838099258(21)3044, (01)00884838099258(21)3107, (01)00884838099258(21)2258, (01)00884838099258(21)1484, (01)00884838099272(21)209, (01)00884838099258(21)2573, (01)00884838099258(21)3514, (01)00884838099258(21)1083, (01)00884838099258(21)3314, (01)00884838085268(21)703307, (01)00884838099258(21)2436, (01)00884838099258(21)2184, (01)00884838099241(21)479, (01)00884838099272(21)271, (01)00884838099241(21)189, (01)00884838099258(21)246, (01)00884838085251(21)523, (01)00884838099241(21)1038, (01)00884838099241(21)47, (01)00884838099258(21)475, (01)00884838099241(21)1146, (01)00884838085268(21)703904, (01)00884838099258(21)1810, (01)00884838085268(21)890, (01)00884838099265(21)104, (01)00884838099258(21)762, (01)00884838099258(21)3153, (01)00884838085251(21)840, (01)00884838085251(21)293, (01)00884838099258(21)526, (01)00884838099258(21)2241, (01)00884838099258(21)588, (01)00884838099258(21)469, (01)00884838099272(21)675, (01)00884838099258(21)962, (01)00884838085268(21)1137, (01)00884838085268(21)1048, (01)00884838099241(21)352, (01)00884838099258(21)539, (01)00884838085350(21)221, (01)00884838099258(21)421, (01)00884838085350(21)42, (01)00884838085350(21)172, (01)00884838085350(21)173, (01)00884838099258(21)288, (01)00884838085268(21)401, (01)00884838099265(21)153, (01)00884838099241(21)391, (01)00884838085367(21)467, (01)00884838099272(21)112, (01)00884838099258(21)90, (01)00884838085268(21)1054, (01)00884838085268(21)1010, (01)00884838085367(21)178, (01)00884838085268(21)536, (01)00884838085268(21)534, (01)00884838085268(21)535, (01)00884838085268(21)266, (01)00884838099258(21)2784, (01)00884838099258(21)1635, (01)00884838085367(21)335, (01)00884838099258(21)50, (01)00884838085282(21)209, (01)00884838085251(21)291, (01)00884838099258(21)239, (01)00884838099258(21)2520, (01)00884838099258(21)589, (01)00884838085268(21)818, (01)00884838099258(21)91, (01)00884838099210(21)332, (01)00884838085268(21)412, (01)00884838099258(21)2945, (01)00884838099258(21)1556, (01)00884838099258(21)840, (01)00884838085268(21)617, (01)00884838099241(21)328, (01)00884838099258(21)362, (01)00884838099258(21)43, (01)00884838099258(21)1869, (01)00884838099241(21)681, (01)00884838099272(21)1091, (01)00884838099272(21)1052, (01)00884838099258(21)1564, (01)00884838085268(21)246, (01)00884838085268(21)488, (01)00884838085268(21)53, (01)00884838085268(21)57, (01)00884838099258(21)1692, (01)00884838085268(21)905, (01)00884838085367(21)147, (01)00884838099258(21)2778, (01)00884838085268(21)55, (01)00884838099258(21)44, (01)00884838099258(21)210, (01)00884838085268(21)992, (01)00884838085268(21)691, (01)00884838099258(21)2421, (01)00884838099258(21)2432, (01)00884838099258(21)287, (01)00884838099272(21)113, (01)00884838085268(21)848, (01)00884838085367(21)205, (01)00884838085268(21)809, (01)00884838099272(21)121, (01)00884838099241(21)1130, (01)00884838099258(21)3291, (01)00884838099241(21)1052, (01)00884838099258(21)2882, (01)00884838085268(21)1033, (01)00884838099258(21)3155, (01)00884838085268(21)891, (01)00884838085251(21)536, (01)00884838085251(21)537, (01)00884838085268(21)836, (01)00884838099258(21)3211, (01)00884838099258(21)904, (01)00884838085268(21)1155, (01)00884838085268(21)466, (01)00884838085268(21)1157, (01)00884838099272(21)158, (01)00884838099258(21)403, (01)00884838099272(21)139, (01)00884838085268(21)1395, (01)00884838099272(21)1145, (01)00884838099272(21)1078, (01)00884838085268(21)703559, (01)00884838099265(21)46, (01)00884838099241(21)855, (01)00884838099258(21)1356, (01)00884838099241(21)581, (01)00884838099241(21)388, (01)00884838085251(21)870, (01)00884838099258(21)1955, (01)00884838099258(21)2800, (01)00884838099258(21)2801, (01)00884838099258(21)1808, (01)00884838085268(21)263, (01)00884838099241(21)1138, (01)00884838099258(21)1838, (01)00884838099258(21)2368, (01)00884838099258(21)951, (01)00884838099258(21)723, (01)00884838085268(21)703391, (01)00884838099265(21)28, (01)00884838085251(21)613, (01)00884838085367(21)170, (01)00884838085268(21)420, (01)00884838099258(21)3697, (01)00884838099241(21)905, (01)00884838099258(21)216, (01)00884838099258(21)1874, (01)00884838099258(21)3062, (01)00884838099258(21)3063, (01)00884838099258(21)1150, (01)00884838099258(21)790, (01)00884838099241(21)570, (01)00884838099258(21)2881, (01)00884838099258(21)226, (01)00884838085350(21)109, (01)00884838085268(21)673, (01)00884838099234(21)67, (01)00884838085268(21)298, (01)00884838085268(21)297, (01)00884838085268(21)299, (01)00884838099234(21)524, (01)00884838099258(21)3042, (01)00884838099258(21)2434, (01)00884838099210(21)111, (01)00884838099258(21)3374, (01)00884838085282(21)238, (01)00884838085268(21)564, (01)00884838085268(21)838, (01)00884838099258(21)1266, (01)00884838099258(21)365, (01)00884838099258(21)512, (01)00884838099258(21)266, (01)00884838099258(21)954, (01)00884838099241(21)77, (01)00884838085268(21)1117, (01)00884838085268(21)1032, (01)00884838099234(21)49, (01)00884838099258(21)301, (01)00884838099258(21)1046, (01)00884838085367(21)455, (01)00884838085268(21)820, (01)00884838099258(21)1081, (01)00884838099258(21)993, (01)00884838099258(21)994, (01)00884838099258(21)1103, (01)00884838099241(21)495, (01)00884838099241(21)448, (01)00884838099241(21)571, (01)00884838085268(21)987, (01)00884838085251(21)598, (01)00884838099258(21)958, (01)00884838085251(21)679, (01)00884838085268(21)1310, (01)00884838099258(21)901, (01)00884838085268(21)1211, (01)00884838085268(21)1213, (01)00884838085268(21)47, (01)00884838099258(21)3067, (01)00884838099258(21)514, (01)00884838099258(21)553, (01)00884838085350(21)250, (01)00884838099234(21)679, (01)00884838099234(21)678, (01)00884838099258(21)2390, (01)00884838099258(21)832, (01)00884838085268(21)1365, (01)00884838085268(21)1408, (01)00884838085268(21)881, (01)00884838099241(21)171, (01)00884838099258(21)1569, (01)00884838099258(21)618, (01)00884838099241(21)1051, (01)00884838099258(21)1636, (01)00884838099241(21)439, (01)00884838099258(21)82, (01)00884838099258(21)617, (01)00884838085268(21)774, (01)00884838085251(21)224, (01)00884838099258(21)753, (01)00884838099258(21)754, (01)00884838099258(21)40, (01)00884838099258(21)2227, (01)00884838099258(21)2280, (01)00884838099265(21)72, (01)00884838099234(21)36, (01)00884838099234(21)10, (01)00884838085268(21)856, (01)00884838099234(21)64, (01)00884838099258(21)2879, (01)00884838085367(21)494, (01)00884838099258(21)369, (01)00884838099258(21)1929, (01)00884838099258(21)1781, (01)00884838085268(21)922, (01)00884838099258(21)3642, (01)00884838099258(21)2376, (01)00884838099258(21)527, (01)00884838085268(21)58, (01)00884838099258(21)2972, (01)00884838099258(21)3244, (01)00884838099258(21)1938, (01)00884838099258(21)1025, (01)00884838099258(21)297, (01)00884838099258(21)110, (01)00884838099258(21)3393, (01)00884838099258(21)2786, (01)00884838099258(21)2765, (01)00884838099258(21)1300, (01)00884838099265(21)208, (01)00884838085282(21)201, (01)00884838085268(21)471, (01)00884838099258(21)3336, (01)00884838099258(21)584, (01)00884838099272(21)239, (01)00884838099258(21)1362, (01)00884838099241(21)126, (01)00884838085251(21)829, (01)00884838099258(21)1255, (01)00884838099258(21)1259, (01)00884838099258(21)859, (01)00884838099258(21)2971, (01)00884838099258(21)957, (01)00884838099234(21)110, (01)00884838099258(21)425, (01)00884838085251(21)249, (01)00884838085268(21)1012, (01)00884838085268(21)755, (01)00884838099272(21)105, (01)00884838099241(21)464, (01)00884838099258(21)837, (01)00884838099258(21)1934, (01)00884838099241(21)628, (01)00884838099258(21)1224, (01)00884838085268(21)633, (01)00884838085268(21)703540, (01)00884838099258(21)611, (01)00884838099241(21)36, (01)00884838099258(21)2929, (01)00884838085268(21)296, (01)00884838099258(21)1917, (01)00884838099241(21)1088, (01)00884838099258(21)2759, (01)00884838099272(21)678, (01)00884838085251(21)546, (01)00884838099241(21)580, (01)00884838099258(21)3582, (01)00884838099258(21)3490, (01)00884838099203(21)181, (01)00884838099241(21)1241, (01)00884838099241(21)1179, (01)00884838085268(21)703558, (01)00884838099258(21)2856, (01)00884838099272(21)362, (01)00884838099258(21)831, (01)00884838099272(21)114, (01)00884838099241(21)1065, (01)00884838099258(21)2996, (01)00884838099258(21)2906, (01)00884838099258(21)2918, (01)00884838099258(21)3282, (01)00884838085268(21)703336, (01)00884838085268(21)884, (01)00884838085268(21)991, (01)00884838085268(21)842, (01)00884838085268(21)574, (01)00884838085268(21)753, (01)00884838099258(21)826, (01)00884838099258(21)1706, (01)00884838099258(21)566, (01)00884838099258(21)565, (01)00884838085268(21)703493, (01)00884838085350(21)192, (01)00884838085268(21)1086, (01)00884838085268(21)840, (01)00884838085350(21)92, (01)00884838085350(21)103, (01)00884838085268(21)790, (01)00884838085251(21)478, (01)00884838085251(21)479, (01)00884838085268(21)407, (01)00884838085268(21)481, (01)00884838085268(21)408, (01)00884838099241(21)799, (01)00884838099258(21)2495, (01)00884838099258(21)742, (01)00884838099258(21)2201, (01)00884838099241(21)577, (01)00884838085367(21)564, (01)00884838085268(21)703350, (01)00884838099258(21)434, (01)00884838099258(21)2420, (01)00884838099258(21)2419, (01)00884838099258(21)164, (01)00884838085251(21)513, (01)00884838099258(21)1481, (01)00884838099258(21)912, (01)00884838099258(21)2559, (01)00884838099258(21)2993, (01)00884838085268(21)877, (01)00884838099241(21)1107, (01)00884838099241(21)1060, (01)00884838099258(21)3020, (01)00884838085251(21)278, (01)00884838099258(21)2511, (01)00884838085268(21)703395, (01)00884838099258(21)3043, (01)00884838099272(21)758, (01)00884838099258(21)2438, (01)00884838085268(21)703323, (01)00884838085268(21)703775, (01)00884838099258(21)224, (01)00884838099258(21)2166, (01)00884838099258(21)190, (01)00884838099272(21)403, (01)00884838099258(21)153, (01)00884838085350(21)114, (01)00884838085367(21)418, (01)00884838099258(21)1931, (01)00884838085367(21)95, (01)00884838085268(21)590, (01)00884838085268(21)713, (01)00884838085268(21)711, (01)00884838085268(21)710, (01)00884838085268(21)709, (01)00884838085251(21)438, (01)00884838085251(21)437, (01)00884838085268(21)716, (01)00884838099258(21)1106, (01)00884838085268(21)1216, (01)00884838085367(21)440, (01)00884838099272(21)702, (01)00884838099258(21)1032, (01)00884838099272(21)108, (01)00884838085268(21)483, (01)00884838099258(21)789, (01)00884838099258(21)788, (01)00884838099258(21)3302, (01)00884838099258(21)79, (01)00884838085251(21)456, (01)00884838099241(21)1114, (01)00884838085251(21)833, (01)00884838085251(21)566, (01)00884838085268(21)1098, (01)00884838099258(21)1271, (01)00884838099258(21)739, (01)00884838099241(21)323, (01)00884838099258(21)1605, (01)00884838099241(21)626, (01)00884838099272(21)545, (01)00884838085268(21)1392, (01)00884838099241(21)252, (01)00884838099258(21)656, (01)00884838085268(21)241, (01)00884838099258(21)667, (01)00884838099241(21)261, (01)00884838099258(21)650, (01)00884838099258(21)609, (01)00884838099258(21)395, (01)00884838099272(21)380, (01)00884838099272(21)197, (01)00884838085268(21)703408, (01)00884838099258(21)2723, (01)00884838099258(21)2039, (01)00884838085251(21)488, (01)00884838085268(21)1363, (01)00884838085268(21)1108, (01)00884838085367(21)247, (01)00884838099258(21)182, (01)00884838099258(21)2504, (01)00884838085282(21)258, (01)00884838085367(21)225, (01)00884838085251(21)327, (01)00884838085251(21)550, (01)00884838099272(21)281, (01)00884838085268(21)703404, (01)00884838085268(21)1355, (01)00884838085268(21)703387, (01)00884838085268(21)1414, (01)00884838099234(21)403, (01)00884838099258(21)75, (01)00884838085275(21)122, (01)00884838085268(21)863, (01)00884838099258(21)3305, (01)00884838099258(21)380, (01)00884838099265(21)41, (01)00884838099258(21)1995, (01)00884838099258(21)2911, (01)00884838099258(21)1318, (01)00884838099258(21)1337, (01)00884838099258(21)1249, (01)00884838099258(21)1063, (01)00884838099265(21)122, (01)00884838085268(21)350, (01)00884838099258(21)1039, (01)00884838099272(21)688, (01)00884838099258(21)2176, (01)00884838099258(21)2177, (01)00884838085268(21)528, (01)00884838085268(21)703465, (01)00884838099234(21)328, (01)00884838099272(21)593, (01)00884838099258(21)1939, (01)00884838099258(21)1375, (01)00884838099258(21)1374, (01)00884838099272(21)592, (01)00884838099272(21)495, (01)00884838099258(21)1524, (01)00884838099272(21)339, (01)00884838099258(21)814, (01)00884838099272(21)319, (01)00884838099258(21)813, (01)00884838099265(21)147, (01)00884838099241(21)125, (01)00884838085268(21)1380, (01)00884838099272(21)687, (01)00884838099258(21)1382, (01)00884838099258(21)998, (01)00884838099241(21)207, (01)00884838085251(21)571, (01)00884838085268(21)908, (01)00884838085268(21)1238, (01)00884838099258(21)2649, (01)00884838085268(21)1378, (01)00884838099258(21)3273, (01)00884838099258(21)3152, (01)00884838099258(21)1220, (01)00884838099258(21)3296, (01)00884838099272(21)921, (01)00884838099241(21)329, (01)00884838099258(21)286, (01)00884838099258(21)289, (01)00884838099258(21)1679, (01)00884838085268(21)262, (01)00884838099258(21)613, (01)00884838099272(21)182, (01)00884838099234(21)481, (01)00884838099258(21)3510, (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085350(21)225, (01)00884838085268(21)598, (01)00884838099234(21)373, (01)00884838099258(21)1234, (01)00884838085350(21)69, (01)00884838099258(21)1064, (01)00884838099258(21)3048, (01)00884838099258(21)3046, (01)00884838099265(21)376, (01)00884838099272(21)183, (01)00884838085268(21)859, (01)00884838085268(21)858, (01)00884838085367(21)64, (01)00884838085268(21)275, (01)00884838085268(21)703861, (01)00884838099258(21)1693, (01)00884838099258(21)363, (01)00884838099234(21)149, (01)00884838099241(21)1001, (01)00884838099258(21)2181, (01)00884838099258(21)2981, (01)00884838099210(21)257, (01)00884838099272(21)690, (01)00884838099210(21)256, (01)00884838099210(21)254, (01)00884838099272(21)325, (01)00884838099258(21)580, (01)00884838099258(21)3354, (01)00884838099258(21)1604, (01)00884838085268(21)520, (01)00884838085268(21)335, (01)00884838099234(21)213, (01)00884838099258(21)850, (01)00884838085367(21)149, (01)00884838099258(21)1258, (01)00884838099258(21)1260, (01)00884838099258(21)1036, (01)00884838085268(21)1018, (01)00884838085268(21)593, (01)00884838099272(21)100, (01)00884838099258(21)1160, (01)00884838099272(21)272, (01)00884838099258(21)1346, (01)00884838099258(21)1347, (01)00884838099258(21)1117, (01)00884838085251(21)545, (01)00884838085251(21)544, (01)00884838085268(21)911, (01)00884838099272(21)680, (01)00884838099258(21)1082, (01)00884838099258(21)333, (01)00884838099258(21)306, (01)00884838099258(21)273, (01)00884838099258(21)2448, (01)00884838099258(21)218, (01)00884838085268(21)480, (01)00884838085268(21)703329, (01)00884838085268(21)1410, (01)00884838085350(21)214, (01)00884838085367(21)42, (01)00884838085350(21)41, (01)00884838099258(21)2668, (01)00884838099258(21)1614, (01)00884838099258(21)547, (01)00884838099272(21)724, (01)00884838085268(21)213, (01)00884838099241(21)801, (01)00884838085268(21)325, (01)00884838099258(21)3453, (01)00884838099258(21)1625, (01)00884838099258(21)228, (01)00884838099258(21)1223, (01)00884838099241(21)560, (01)00884838099241(21)561, (01)00884838099234(21)45, (01)00884838099241(21)205, (01)00884838085268(21)496, (01)00884838085268(21)887, (01)00884838085268(21)903, (01)00884838099258(21)25, (01)00884838085268(21)703413, (01)00884838099258(21)1059, (01)00884838099258(21)967, (01)00884838085268(21)1432, (01)00884838099258(21)391, (01)00884838099258(21)2952, (01)00884838099234(21)537, (01)00884838085251(21)465, (01)00884838085251(21)376, (01)00884838099258(21)2121, (01)00884838099258(21)612, (01)00884838099258(21)2912, (01)00884838099258(21)950, (01)00884838099258(21)946, (01)00884838099258(21)2873, (01)00884838099258(21)2687, (01)00884838099210(21)265, (01)00884838085367(21)153, (01)00884838085268(21)594, (01)00884838099272(21)350, (01)00884838085268(21)1309, (01)00884838099258(21)191, (01)00884838085268(21)1312, (01)00884838099258(21)3450, (01)00884838099258(21)3158, (01)00884838099258(21)2433, (01)00884838085268(21)999, (01)00884838099203(21)3, (01)00884838099258(21)1009, (01)00884838099258(21)995, (01)00884838099258(21)454, (01)00884838099272(21)891, (01)00884838099258(21)2830, (01)00884838099258(21)2829, (01)00884838085268(21)1424, (01)00884838099272(21)280, (01)00884838099234(21)615, (01)00884838099258(21)3262, (01)00884838099258(21)1253, (01)00884838099258(21)3283, (01)00884838099258(21)1488, (01)00884838085268(21)674, (01)00884838085268(21)465, (01)00884838099258(21)966, (01)00884838099258(21)3345, (01)00884838099265(21)262, (01)00884838085268(21)1084, (01)00884838099258(21)2594, (01)00884838099258(21)1940, (01)00884838099234(21)159, (01)00884838099258(21)3578, (01)00884838085268(21)353, (01)00884838085268(21)1230, (01)00884838085268(21)499, (01)00884838099258(21)2870, (01)00884838099241(21)379, (01)00884838099258(21)886, (01)00884838099241(21)440, (01)00884838085268(21)813, (01)00884838085268(21)703508, (01)00884838099234(21)18, (01)00884838099258(21)2656, (01)00884838099258(21)867, (01)00884838085367(21)324, (01)00884838085268(21)33, (01)00884838099258(21)738, (01)00884838099258(21)2883, (01)00884838099272(21)270, (01)00884838085367(21)323, (01)00884838099258(21)2908, (01)00884838099234(21)231, (01)00884838085268(21)1346, (01)00884838099272(21)254, (01)00884838099258(21)3503, (01)00884838099258(21)3504, (01)00884838099234(21)180, (01)00884838099258(21)244, (01)00884838099227(21)265, (01)00884838085268(21)1401, (01)00884838085268(21)1189, (01)00884838085268(21)361, (01)00884838099272(21)107, (01)00884838099258(21)1270, (01)00884838099234(21)61, (01)00884838099234(21)273, (01)00884838099234(21)204, (01)00884838085268(21)267, (01)00884838085268(21)403, (01)00884838099258(21)1518, (01)00884838099258(21)360, (01)00884838085367(21)320, (01)00884838085367(21)89, (01)00884838085268(21)526, (01)00884838085268(21)439, (01)00884838085367(21)85, (01)00884838085268(21)1021, (01)00884838085268(21)736, (01)00884838099258(21)229, (01)00884838099258(21)2793, (01)00884838099272(21)144, (01)00884838099272(21)761, (01)00884838099258(21)1975, (01)00884838099258(21)428, (01)00884838099258(21)3564, (01)00884838099265(21)331, (01)00884838099241(21)204, (01)00884838099258(21)2722, (01)00884838099258(21)2311, (01)00884838099258(21)476, (01)00884838099241(21)192, (01)00884838099241(21)197, (01)00884838085268(21)1370, (01)00884838085268(21)608, (01)00884838099258(21)1559, (01)00884838099258(21)238, (01)00884838085251(21)31, (01)00884838085268(21)68, (01)00884838099258(21)968, (01)00884838099258(21)42, (01)00884838085268(21)703324, (01)00884838085268(21)611, (01)00884838085268(21)414, (01)00884838085268(21)415, (01)00884838085268(21)413, (01)00884838085268(21)703306, (01)00884838085268(21)453, (01)00884838085367(21)109, (01)00884838099258(21)1334, (01)00884838099258(21)1924, (01)00884838099258(21)1926, (01)00884838085268(21)895, (01)00884838099234(21)538, (01)00884838099258(21)2145, (01)00884838099272(21)681, (01)00884838099272(21)691, (01)00884838099210(21)255, (01)00884838099258(21)2831, (01)00884838085251(21)921, (01)00884838099258(21)166, (01)00884838085268(21)703615, (01)00884838085268(21)587, (01)00884838085268(21)1399, (01)00884838085268(21)1297, (01)00884838085367(21)407, (01)00884838085268(21)428, (01)00884838099258(21)2331, (01)00884838099258(21)56, (01)00884838099258(21)3362, (01)00884838099241(21)177, (01)00884838085268(21)883, (01)00884838099272(21)779, (01)00884838099258(21)2901, (01)00884838099258(21)105, (01)00884838085251(21)700, (01)00884838099258(21)291, (01)00884838099258(21)817, (01)00884838099258(21)444, (01)00884838085268(21)1267, (01)00884838085268(21)758, (01)00884838085268(21)422, (01)00884838085268(21)1040, (01)00884838099258(21)2058, (01)00884838099258(21)640, (01)00884838099258(21)298, (01)00884838085268(21)634, (01)00884838085268(21)714, (01)00884838085268(21)573, (01)00884838099241(21)1342, (01)00884838099272(21)294, (01)00884838099241(21)37, (01)00884838085367(21)239, (01)00884838085251(21)165, (01)00884838085268(21)274, (01)00884838099210(21)101, (01)00884838099210(21)100, (01)00884838099210(21)99, (01)00884838099258(21)2980, (01)00884838085367(21)505, (01)00884838085268(21)525, (01)00884838099258(21)240, (01)00884838085251(21)445, (01)00884838099258(21)1516, (01)00884838099258(21)876, (01)00884838099272(21)884, (01)00884838085268(21)885, (01)00884838085367(21)271, (01)00884838085268(21)843, (01)00884838085268(21)1276, (01)00884838099258(21)94, (01)00884838099234(21)76, (01)00884838099258(21)564, (01)00884838085268(21)703478, (01)00884838085268(21)1102, (01)00884838099272(21)845, (01)00884838085268(21)1142, (01)00884838099258(21)33, (01)00884838085268(21)919, (01)00884838085268(21)792, (01)00884838099258(21)856, (01)00884838085268(21)703528, (01)00884838085268(21)754, (01)00884838085268(21)599, (01)00884838085251(21)539, (01)00884838099258(21)1245, (01)00884838085282(21)261, (01)00884838085282(21)260, (01)00884838099241(21)340, (01)00884838099258(21)385, (01)00884838099258(21)2532, (01)00884838099258(21)2524, (01)00884838099241(21)1097, (01)00884838099241(21)357, (01)00884838099241(21)282, (01)00884838099272(21)559, (01)00884838099258(21)3376, (01)00884838099258(21)2503, (01)00884838099258(21)715, (01)00884838099241(21)1015, (01)00884838099272(21)116, (01)00884838099265(21)112, (01)00884838085268(21)1171, (01)00884838099272(21)343, (01)00884838099272(21)98, (01)00884838085367(21)443, (01)00884838085268(21)1096, (01)00884838085268(21)1097, (01)00884838085251(21)658, (01)00884838085251(21)656, (01)00884838085251(21)657, (01)00884838085251(21)654, (01)00884838099258(21)1047, (01)00884838099258(21)461, (01)00884838099241(21)214, (01)00884838085268(21)817, (01)00884838085268(21)766, (01)00884838085251(21)436, (01)00884838085251(21)862, (01)00884838099258(21)2135, (01)00884838099258(21)577, (01)00884838085251(21)1007, (01)00884838099241(21)44, (01)00884838099258(21)41, (01)00884838099258(21)2969, (01)00884838099258(21)129, (01)00884838085268(21)78, (01)00884838099241(21)336, (01)00884838099265(21)110, (01)00884838099258(21)439, (01)00884838099241(21)206, (01)00884838099258(21)432, (01)00884838099241(21)210, (01)00884838099241(21)1231, (01)00884838099241(21)343, (01)00884838099258(21)603, (01)00884838099203(21)7, (01)00884838099234(21)502, (01)00884838099258(21)2449, (01)00884838085251(21)786, (01)00884838099258(21)1038, (01)00884838099258(21)792, (01)00884838099258(21)595, (01)00884838099234(21)150, (01)00884838085367(21)104, (01)00884838099258(21)2979, (01)00884838099258(21)3007, (01)00884838099258(21)713, (01)00884838099258(21)741, (01)00884838099258(21)401, (01)00884838099241(21)390, (01)00884838099241(21)365, (01)00884838085268(21)703419, (01)00884838099258(21)2595, (01)00884838099258(21)400, (01)00884838099258(21)3373, (01)00884838099258(21)415, (01)00884838099258(21)1317, (01)00884838099258(21)3398, (01)00884838099210(21)310, (01)00884838099258(21)963, (01)00884838099258(21)965, (01)00884838099258(21)1228, (01)00884838099258(21)2308, (01)00884838099272(21)301, (01)00884838099258(21)1265, (01)00884838099265(21)207, (01)00884838099241(21)1067, (01)00884838099241(21)1106, (01)00884838099258(21)3179, (01)00884838099265(21)405, (01)00884838099265(21)404, (01)00884838099265(21)375, (01)00884838099241(21)1144, (01)00884838099258(21)3123, (01)00884838099258(21)3120, (01)00884838099258(21)2995, (01)00884838099258(21)1510, (01)00884838099241(21)631, (01)00884838099241(21)187, (01)00884838099241(21)186, (01)00884838085268(21)1322, (01)00884838085251(21)308, (01)00884838099234(21)132, (01)00884838099272(21)117, (01)00884838085367(21)370, (01)00884838099272(21)876, (01)00884838099258(21)178, (01)00884838099258(21)494, (01)00884838099241(21)87, (01)00884838099258(21)1662, (01)00884838099258(21)2316, (01)00884838099241(21)373, (01)00884838099258(21)571, (01)00884838085268(21)878, (01)00884838085268(21)1170, (01)00884838099234(21)399, (01)00884838085268(21)703435, (01)00884838085268(21)1107, (01)00884838099258(21)2619, (01)00884838099258(21)2620, (01)00884838099258(21)151, (01)00884838099258(21)554, (01)00884838085367(21)322, (01)00884838099258(21)1574, (01)00884838099258(21)892, (01)00884838085268(21)703310, (01)00884838099258(21)1174, (01)00884838085268(21)944, (01)00884838099258(21)657, (01)00884838099258(21)274, (01)00884838099265(21)68, (01)00884838099258(21)115, (01)00884838085268(21)703521, (01)00884838085268(21)654, (01)00884838085367(21)136, (01)00884838085251(21)396, (01)00884838099258(21)3146, (01)00884838099258(21)3145, (01)00884838085268(21)273, (01)00884838099258(21)1357, (01)00884838099258(21)215, (01)00884838085268(21)703305, (01)00884838099258(21)674, (01)00884838099258(21)1086, (01)00884838099258(21)1093, (01)00884838099258(21)1279, (01)00884838099258(21)3038, (01)00884838099258(21)1329, (01)00884838099258(21)342, (01)00884838085268(21)715, (01)00884838099258(21)282, (01)00884838099241(21)514, (01)00884838099203(21)178, (01)00884838099258(21)1102, (01)00884838099241(21)599, (01)00884838099258(21)1554, (01)00884838099241(21)601, (01)00884838099241(21)602, (01)00884838099258(21)865, (01)00884838085251(21)576, (01)00884838099258(21)138, (01)00884838099241(21)69, (01)00884838099272(21)1021, (01)00884838099258(21)770, (01)00884838099258(21)516, (01)00884838099258(21)478, (01)00884838085268(21)703495, (01)00884838085268(21)1285, (01)00884838099241(21)638, (01)00884838099258(21)187, (01)00884838085367(21)373, (01)00884838085251(21)121, (01)00884838099258(21)871, (01)00884838099258(21)873, (01)00884838099265(21)177, (01)00884838085367(21)442, (01)00884838085268(21)854, (01)00884838099258(21)49, (01)00884838085251(21)383, (01)00884838085251(21)239, (01)00884838085268(21)19, (01)00884838085268(21)904, (01)00884838099258(21)154, (01)00884838099258(21)1783, (01)00884838099258(21)1482, (01)00884838099258(21)2246, (01)00884838099241(21)728, (01)00884838099265(21)291, (01)00884838099241(21)937, (01)00884838099241(21)944, (01)00884838099265(21)299, (01)00884838099241(21)951, (01)00884838099241(21)945, (01)00884838099258(21)2247, (01)00884838099258(21)2248, (01)00884838085268(21)1005, (01)00884838099258(21)2229, (01)00884838099258(21)278, (01)00884838099241(21)852, (01)00884838099272(21)244, (01)00884838099258(21)2764, (01)00884838099258(21)433, (01)00884838099258(21)1168, (01)00884838099258(21)2727, (01)00884838099241(21)256, (01)00884838085268(21)804, (01)00884838085251(21)491, (01)00884838085251(21)125, (01)00884838099258(21)3525, (01)00884838085268(21)703497, (01)00884838085268(21)730, (01)00884838099258(21)2522, (01)00884838099258(21)2523, (01)00884838099272(21)781, (01)00884838099272(21)782, (01)00884838099241(21)679, (01)00884838099272(21)332, (01)00884838099258(21)1395, (01)00884838099258(21)1175, (01)00884838099258(21)964, (01)00884838099258(21)552, (01)00884838099258(21)1571, (01)00884838099258(21)624, (01)00884838085367(21)524, (01)00884838099258(21)885, (01)00884838099272(21)534, (01)00884838099258(21)1994, (01)00884838099258(21)189, (01)00884838085367(21)308, (01)00884838099258(21)1100, (01)00884838099272(21)341, (01)00884838099258(21)2839, (01)00884838099272(21)374, (01)00884838099258(21)217, (01)00884838099272(21)78, (01)00884838085268(21)1400, (01)00884838099241(21)596, (01)00884838099234(21)533, (01)00884838099203(21)54, (01)00884838099258(21)192, (01)00884838085275(21)131, (01)00884838085268(21)954, (01)00884838085275(21)84, (01)00884838099258(21)3130, (01)00884838085367(21)354, (01)00884838099272(21)11, (01)00884838085367(21)75, (01)00884838085268(21)375, (01)00884838099234(21)211, (01)00884838099258(21)2502, (01)00884838099241(21)932, (01)00884838099258(21)755, (01)00884838099241(21)67, (01)00884838099258(21)3210, (01)00884838099241(21)503, (01)00884838085268(21)957, (01)00884838099272(21)441, (01)00884838085268(21)703332, (01)00884838085251(21)500, (01)00884838085268(21)697, (01)00884838099258(21)3603, (01)00884838099258(21)3486, (01)00884838085251(21)645, (01)00884838085268(21)658, (01)00884838099258(21)108, (01)00884838099258(21)1278, (01)00884838085251(21)379, (01)00884838099258(21)373, (01)00884838099258(21)2154, (01)00884838099258(21)339, (01)00884838099258(21)471, (01)00884838099241(21)432, (01)00884838099258(21)947, (01)00884838099258(21)103, (01)00884838085268(21)423, (01)00884838085268(21)1022, (01)00884838099241(21)131, (01)00884838099258(21)1140, (01)00884838099272(21)129, (01)00884838099258(21)825, (01)00884838099258(21)1519, (01)00884838099227(21)135, (01)00884838099227(21)88, (01)00884838099258(21)234, (01)00884838099258(21)47, (01)00884838085251(21)399, (01)00884838099258(21)3794, (01)00884838099272(21)180, (01)00884838099272(21)918, (01)00884838099258(21)1651, (01)00884838099258(21)1615, (01)00884838099258(21)1637, (01)00884838099241(21)301, (01)00884838099258(21)1925, (01)00884838099241(21)704, (01)00884838099258(21)1657, (01)00884838099241(21)456, (01)00884838099258(21)665, (01)00884838099210(21)69, (01)00884838099258(21)637, (01)00884838099258(21)2846, (01)00884838085367(21)279, (01)00884838099258(21)1264, (01)00884838085268(21)897, (01)00884838085268(21)896, (01)00884838099258(21)3341, (01)00884838085268(21)1411, (01)00884838085251(21)803, (01)00884838099241(21)659, (01)00884838099258(21)1606, (01)00884838099258(21)1612, (01)00884838099258(21)1613, (01)00884838099258(21)3219, (01)00884838085251(21)415, (01)00884838085268(21)1430, (01)00884838099258(21)3337, (01)00884838099272(21)442, (01)00884838099258(21)338, (01)00884838099241(21)242, (01)00884838099241(21)746, (01)00884838085268(21)621, (01)00884838099258(21)3541, (01)00884838099258(21)2549, (01)00884838099258(21)2569, (01)00884838099258(21)1490, (01)00884838099265(21)400, (01)00884838099241(21)1161, (01)00884838099258(21)128, (01)00884838099258(21)1101, (01)00884838099258(21)2974, (01)00884838099258(21)3588, (01)00884838099241(21)1265, (01)00884838099241(21)1090, (01)00884838099258(21)1118, (01)00884838085268(21)703426, (01)00884838099272(21)220, (01)00884838099265(21)29, (01)00884838099203(21)6, (01)00884838085367(21)400, (01)00884838099258(21)1467, (01)00884838099241(21)457, (01)00884838099241(21)455, (01)00884838099258(21)1012, (01)00884838099203(21)93, (01)00884838099258(21)2913, (01)00884838099272(21)555, (01)00884838099265(21)141, (01)00884838099258(21)265, (01)00884838099234(21)17, (01)00884838085275(21)132, (01)00884838085268(21)628, (01)00884838085251(21)296, (01)00884838099258(21)1158, (01)00884838099258(21)2435, (01)00884838085268(21)795, (01)00884838085350(21)98, (01)00884838099210(21)12, (01)00884838085268(21)272, (01)00884838085268(21)1007, (01)00884838099258(21)2045, (01)00884838099258(21)1960, (01)00884838085268(21)940, (01)00884838099258(21)446, (01)00884838099258(21)119, (01)00884838085268(21)250, (01)00884838085268(21)46, (01)00884838099272(21)189, (01)00884838099241(21)221, (01)00884838099258(21)484, (01)00884838099258(21)483, (01)00884838099272(21)190, (01)00884838085282(21)314, (01)00884838099258(21)",
      "more_code_info": "696, (01)00884838099258(21)2860, (01)00884838099272(21)871, (01)00884838099258(21)1252, (01)00884838099258(21)869, (01)00884838085268(21)1289, (01)00884838099258(21)1376, (01)00884838099258(21)1377, (01)00884838099258(21)1514, (01)00884838085268(21)605, (01)00884838099258(21)1550, (01)00884838099258(21)2185, (01)00884838099241(21)454, (01)00884838085268(21)65, (01)00884838099258(21)3023, (01)00884838099241(21)829, (01)00884838099258(21)2900, (01)00884838085268(21)360, (01)00884838085268(21)472, (01)00884838085268(21)468, (01)00884838085268(21)469, (01)00884838085268(21)828, (01)00884838099258(21)587, (01)00884838099258(21)1985, (01)00884838099258(21)3178, (01)00884838099241(21)816, (01)00884838085268(21)737, (01)00884838085251(21)136, (01)00884838099258(21)2740, (01)00884838099258(21)1839, (01)00884838099241(21)318, (01)00884838099258(21)608, (01)00884838099258(21)438, (01)00884838099258(21)436, (01)00884838099234(21)47, (01)00884838099258(21)2141, (01)00884838099258(21)2140, (01)00884838099210(21)226, (01)00884838099258(21)112, (01)00884838099258(21)2548, (01)00884838099234(21)87, (01)00884838099234(21)38, (01)00884838099258(21)1257, (01)00884838099227(21)276, (01)00884838099258(21)1739, (01)00884838099258(21)1037, (01)00884838099258(21)422, (01)00884838085268(21)925, (01)00884838099265(21)216, (01)00884838099265(21)217, (01)00884838099258(21)464, (01)00884838099241(21)262, (01)00884838099258(21)841, (01)00884838099241(21)229, (01)00884838099241(21)230, (01)00884838099258(21)335, (01)00884838099258(21)326, (01)00884838085350(21)81, (01)00884838099234(21)619, (01)00884838099234(21)617, (01)00884838099234(21)618, (01)00884838099258(21)3515, (01)00884838085268(21)857, (01)00884838099258(21)134, (01)00884838085268(21)32, (01)00884838099210(21)19, (01)00884838099210(21)15, (01)00884838085251(21)890, (01)00884838085268(21)703416, (01)00884838099258(21)453, (01)00884838099258(21)893, (01)00884838099241(21)308, (01)00884838085268(21)703327, (01)00884838099258(21)1540, (01)00884838099258(21)507, (01)00884838085268(21)34, (01)00884838099234(21)128, (01)00884838099272(21)446, (01)00884838099203(21)69, (01)00884838099241(21)637, (01)00884838085251(21)697, (01)00884838085251(21)626, (01)00884838099258(21)2675, (01)00884838085251(21)559, (01)00884838099272(21)842, (01)00884838099258(21)147, (01)00884838099258(21)148, (01)00884838099258(21)1565, (01)00884838099258(21)447, (01)00884838085268(21)1222, (01)00884838085268(21)1268, (01)00884838099258(21)107, (01)00884838099258(21)2754, (01)00884838085268(21)1063, (01)00884838099272(21)1162, (01)00884838099258(21)1018, (01)00884838085268(21)703335, (01)00884838099258(21)2377, (01)00884838085268(21)938, (01)00884838085268(21)769, (01)00884838099258(21)1879, (01)00884838099234(21)522, (01)00884838099258(21)1004, (01)00884838085268(21)655, (01)00884838099258(21)1214, (01)00884838099258(21)417, (01)00884838085268(21)229, (01)00884838099258(21)374, (01)00884838099258(21)396, (01)00884838099258(21)482, (01)00884838085251(21)532, (01)00884838085251(21)538, (01)00884838099258(21)730, (01)00884838099234(21)32, (01)00884838085268(21)738, (01)00884838085268(21)1204, (01)00884838099272(21)255, (01)00884838099258(21)3016, (01)00884838099258(21)1486, (01)00884838099258(21)2572, (01)00884838099234(21)600, (01)00884838099210(21)23, (01)00884838099203(21)9, (01)00884838099258(21)1405, (01)00884838099258(21)163, (01)00884838099210(21)17, (01)00884838099234(21)25, (01)00884838099234(21)23, (01)00884838099258(21)1230, (01)00884838099258(21)492, (01)00884838085268(21)1311, (01)00884838085367(21)196, (01)00884838085367(21)510, (01)00884838099258(21)233, (01)00884838085268(21)572, (01)00884838099258(21)2226, (01)00884838099258(21)1935, (01)00884838085268(21)776, (01)00884838085282(21)229, (01)00884838085268(21)538, (01)00884838085251(21)185, (01)00884838099258(21)3492, (01)00884838099234(21)62, (01)00884838099258(21)2567, (01)00884838099258(21)2332, (01)00884838099258(21)2533, (01)00884838099258(21)3147, (01)00884838099258(21)3132, (01)00884838099241(21)124, (01)00884838085367(21)110, (01)00884838099234(21)181, (01)00884838099234(21)177, (01)00884838099258(21)583, (01)00884838099241(21)198, (01)00884838099258(21)3269, (01)00884838099258(21)662, (01)00884838099234(21)48, (01)00884838099272(21)848, (01)00884838085268(21)797, (01)00884838085268(21)989, (01)00884838099258(21)836, (01)00884838099258(21)78, (01)00884838085268(21)946, (01)00884838099258(21)1663, (01)00884838085367(21)446, (01)00884838085268(21)888, (01)00884838099258(21)267, (01)00884838099258(21)268, (01)00884838099210(21)55, (01)00884838085367(21)78, (01)00884838085268(21)393, (01)00884838085268(21)394, (01)00884838085251(21)166, (01)00884838085268(21)1062, (01)00884838099272(21)208, (01)00884838099258(21)346, (01)00884838099210(21)45, (01)00884838099258(21)2239, (01)00884838099258(21)1555, (01)00884838099272(21)554, (01)00884838099258(21)1276, (01)00884838085268(21)703396, (01)00884838085268(21)52, (01)00884838099241(21)433, (01)00884838085367(21)330, (01)00884838099258(21)3384, (01)00884838085350(21)123, (01)00884838085367(21)73, (01)00884838099234(21)189, (01)00884838099265(21)71, (01)00884838099258(21)443, (01)00884838085268(21)703420, (01)00884838099258(21)1782, (01)00884838099258(21)2439, (01)00884838099258(21)2452, (01)00884838099241(21)889, (01)00884838099241(21)883, (01)00884838099258(21)252, (01)00884838085251(21)257, (01)00884838085251(21)258, (01)00884838099258(21)3224, (01)00884838099258(21)3090, (01)00884838099272(21)955, (01)00884838099258(21)2281, (01)00884838085251(21)684, (01)00884838085251(21)359, (01)00884838099258(21)2609, (01)00884838099258(21)1436, (01)00884838099258(21)1653, (01)00884838085268(21)281, (01)00884838085367(21)288, (01)00884838099258(21)1525, (01)00884838099265(21)37, (01)00884838099272(21)360, (01)00884838085268(21)939, (01)00884838099258(21)2718, (01)00884838085282(21)291, (01)00884838085268(21)703407, (01)00884838099258(21)1246, (01)00884838099258(21)1256, (01)00884838085367(21)285, (01)00884838099258(21)414, (01)00884838085282(21)270, (01)00884838099241(21)566, (01)00884838099258(21)797, (01)00884838099258(21)1014, (01)00884838099258(21)2370, (01)00884838099258(21)1866, (01)00884838099234(21)153, (01)00884838085268(21)1219, (01)00884838099258(21)953, (01)00884838085268(21)955, (01)00884838099241(21)1084, (01)00884838099258(21)763, (01)00884838099272(21)337, (01)00884838085268(21)402, (01)00884838085268(21)1042, (01)00884838085268(21)909, (01)00884838085268(21)868, (01)00884838085268(21)931, (01)00884838085268(21)867, (01)00884838099258(21)1563, (01)00884838099258(21)45, (01)00884838085268(21)703425, (01)00884838085268(21)374, (01)00884838085268(21)459, (01)00884838099258(21)2326, (01)00884838099258(21)543, (01)00884838099272(21)159, (01)00884838099258(21)1244, (01)00884838099258(21)3085, (01)00884838099258(21)2196, (01)00884838099258(21)2195, (01)00884838085268(21)1282, (01)00884838085268(21)841, (01)00884838085367(21)295, (01)00884838099272(21)400, (01)00884838085350(21)278, (01)00884838085268(21)703532, (01)00884838099258(21)651, (01)00884838099234(21)80, (01)00884838099234(21)79, (01)00884838099258(21)1459, (01)00884838085367(21)406, (01)00884838099258(21)540, (01)00884838099258(21)285, (01)00884838099258(21)308, (01)00884838085268(21)1436, (01)00884838085268(21)703344, (01)00884838099258(21)351, (01)00884838085268(21)1360, (01)00884838085268(21)1358, (01)00884838085268(21)1359, (01)00884838099258(21)2221, (01)00884838099258(21)1024, (01)00884838085268(21)703389, (01)00884838085268(21)1391, (01)00884838099258(21)1394, (01)00884838099258(21)1381, (01)00884838099258(21)2717, (01)00884838099272(21)757, (01)00884838099241(21)880, (01)00884838085251(21)143, (01)00884838099258(21)2453, (01)00884838099258(21)1111, (01)00884838099265(21)479, (01)00884838099265(21)175, (01)00884838099241(21)665, (01)00884838085268(21)50, (01)00884838099258(21)3509, (01)00884838099258(21)2501, (01)00884838085251(21)661, (01)00884838099258(21)300, (01)00884838099272(21)326, (01)00884838099241(21)188, (01)00884838085268(21)1006, (01)00884838085268(21)581, (01)00884838099258(21)591, (01)00884838099258(21)100, (01)00884838085268(21)1352, (01)00884838085268(21)580, (01)00884838085251(21)824, (01)00884838085268(21)703309, (01)00884838099258(21)140, (01)00884838099241(21)1323, (01)00884838099265(21)474, (01)00884838085367(21)528, (01)00884838099258(21)1310, (01)00884838099258(21)1309, (01)00884838099234(21)157, (01)00884838099258(21)2172, (01)00884838085268(21)703510, (01)00884838085268(21)703383, (01)00884838085251(21)649, (01)00884838099258(21)2720, (01)00884838085268(21)974, (01)00884838099258(21)952, (01)00884838085268(21)1245, (01)00884838085268(21)860, (01)00884838085268(21)396, (01)00884838099258(21)479, (01)00884838085268(21)326, (01)00884838099241(21)367, (01)00884838099258(21)2647, (01)00884838099258(21)735, (01)00884838099258(21)2279, (01)00884838099258(21)616, (01)00884838085268(21)741, (01)00884838099258(21)2132, (01)00884838085268(21)721, (01)00884838085268(21)584, (01)00884838099265(21)205, (01)00884838099258(21)1079, (01)00884838085268(21)216, (01)00884838099258(21)2220, (01)00884838099241(21)52, (01)00884838085251(21)493, (01)00884838099258(21)594, (01)00884838099241(21)371, (01)00884838099258(21)973, (01)00884838099258(21)857, (01)00884838099234(21)24, (01)00884838099258(21)2579, (01)00884838099258(21)1526, (01)00884838099258(21)3165, (01)00884838099272(21)323, (01)00884838099258(21)437, (01)00884838099258(21)370, (01)00884838085251(21)720, (01)00884838085268(21)899, (01)00884838085268(21)703, (01)00884838099258(21)599, (01)00884838099258(21)1702, (01)00884838085268(21)928, (01)00884838085268(21)927, (01)00884838099258(21)3328, (01)00884838099258(21)3333, (01)00884838099258(21)3274, (01)00884838099241(21)892, (01)00884838099265(21)228, (01)00884838099265(21)229, (01)00884838085268(21)981, (01)00884838085268(21)923, (01)00884838085350(21)46, (01)00884838085268(21)865, (01)00884838099241(21)127, (01)00884838085367(21)329, (01)00884838099258(21)251, (01)00884838099272(21)175, (01)00884838099258(21)2859, (01)00884838099258(21)3615, (01)00884838085268(21)1163, (01)00884838099258(21)3172, (01)00884838099258(21)2228, (01)00884838099272(21)427, (01)00884838085367(21)184, (01)00884838099258(21)1956, (01)00884838085268(21)416, (01)00884838099241(21)105, (01)00884838099258(21)2869, (01)00884838099258(21)2939, (01)00884838099234(21)187, (01)00884838099258(21)1299, (01)00884838099241(21)572, (01)00884838099234(21)26, (01)00884838099258(21)125, (01)00884838099258(21)124, (01)00884838099258(21)120, (01)00884838085268(21)717, (01)00884838085268(21)689, (01)00884838085268(21)268, (01)00884838099258(21)2222, (01)00884838099210(21)223, (01)00884838099210(21)187, (01)00884838099258(21)366, (01)00884838099258(21)183, (01)00884838099258(21)2737, (01)00884838099258(21)1041, (01)00884838099258(21)131, (01)00884838099234(21)642, (01)00884838099234(21)431, (01)00884838085367(21)350, (01)00884838099258(21)3258, (01)00884838099258(21)839, (01)00884838099241(21)334, (01)00884838099241(21)104, (01)00884838085367(21)462, (01)00884838099258(21)2278, (01)00884838085268(21)900, (01)00884838085268(21)497, (01)00884838085251(21)402, (01)00884838099241(21)627, (01)00884838099258(21)39, (01)00884838099258(21)999, (01)00884838085282(21)267, (01)00884838099272(21)464, (01)00884838099258(21)785, (01)00884838085268(21)1369, (01)00884838099258(21)2072, (01)00884838099258(21)982, (01)00884838085268(21)424, (01)00884838099241(21)458, (01)00884838099258(21)1221, (01)00884838099258(21)2617, (01)00884838099258(21)945, (01)00884838085268(21)1258, (01)00884838085268(21)1323, (01)00884838099258(21)3207, (01)00884838085367(21)304, (01)00884838085251(21)557, (01)00884838085251(21)520, (01)00884838085251(21)558, (01)00884838085251(21)556, (01)00884838085251(21)521, (01)00884838085268(21)959, (01)00884838085268(21)448, (01)00884838085268(21)447, (01)00884838085251(21)187, (01)00884838099258(21)3039, (01)00884838099258(21)1798, (01)00884838085268(21)1386, (01)00884838085268(21)1385, (01)00884838099258(21)1873, (01)00884838099258(21)1142, (01)00884838099258(21)88, (01)00884838085268(21)852, (01)00884838099272(21)468, (01)00884838099210(21)44, (01)00884838085268(21)1252, (01)00884838099258(21)1215, (01)00884838099258(21)466, (01)00884838085268(21)802, (01)00884838085268(21)915, (01)00884838085350(21)277, (01)00884838099258(21)101, (01)00884838099258(21)1147, (01)00884838099258(21)1954, (01)00884838099258(21)462, (01)00884838099258(21)3303, (01)00884838099272(21)778, (01)00884838099258(21)2729, (01)00884838085367(21)472, (01)00884838085268(21)703379, (01)00884838085268(21)703402, (01)00884838085367(21)470, (01)00884838085367(21)457, (01)00884838085268(21)751, (01)00884838099234(21)29, (01)00884838099272(21)683, (01)00884838085268(21)704281, (01)00884838099258(21)99, (01)00884838085268(21)436, (01)00884838099258(21)3516, (01)00884838099258(21)3176, (01)00884838099258(21)2143, (01)00884838085268(21)703311, (01)00884838085268(21)1402, (01)00884838099258(21)2963, (01)00884838085268(21)910, (01)00884838099258(21)592, (01)00884838085268(21)866, (01)00884838099241(21)1113, (01)00884838099241(21)487, (01)00884838099241(21)463, (01)00884838099258(21)3415, (01)00884838099258(21)3101, (01)00884838099241(21)470, (01)00884838085350(21)115, (01)00884838085367(21)319, (01)00884838085367(21)520, (01)00884838099258(21)2970, (01)00884838085268(21)703418, (01)00884838085367(21)490, (01)00884838085268(21)1242, (01)00884838085268(21)354, (01)00884838085268(21)708, (01)00884838085367(21)243, (01)00884838099258(21)2655, (01)00884838099265(21)340, (01)00884838099265(21)157, (01)00884838085268(21)251, (01)00884838099258(21)188, (01)00884838085251(21)902, (01)00884838085268(21)703501, (01)00884838099258(21)986, (01)00884838099258(21)472, (01)00884838099241(21)96, (01)00884838099258(21)106, (01)00884838085350(21)150, (01)00884838099241(21)616, (01)00884838099241(21)277, (01)00884838099234(21)113, (01)00884838085275(21)96, (01)00884838099258(21)2663, (01)00884838099272(21)143, (01)00884838085268(21)1064, (01)00884838099258(21)3577, (01)00884838099258(21)2369, (01)00884838085268(21)1156, (01)00884838099258(21)1483, (01)00884838099258(21)36, (01)00884838099234(21)122, (01)00884838085268(21)1049, (01)00884838099258(21)626, (01)00884838099272(21)262, (01)00884838099258(21)312, (01)00884838099258(21)2301, (01)00884838085251(21)269, (01)00884838085268(21)344, (01)00884838085268(21)343, (01)00884838085268(21)345, (01)00884838085268(21)342, (01)00884838085367(21)103, (01)00884838099258(21)2130, (01)00884838099258(21)165, (01)00884838099258(21)1551, (01)00884838085268(21)1271, (01)00884838085268(21)479, (01)00884838085268(21)18, (01)00884838085268(21)631, (01)00884838099258(21)3037, (01)00884838099203(21)177, (01)00884838099203(21)115, (01)00884838085367(21)479, (01)00884838099258(21)1793, (01)00884838085268(21)1103, (01)00884838099241(21)1140, (01)00884838099272(21)482, (01)00884838099258(21)1364, (01)00884838099258(21)1363, (01)00884838099272(21)971, (01)00884838099203(21)85, (01)00884838085275(21)98, (01)00884838099258(21)3078, (01)00884838085268(21)1361, (01)00884838099203(21)19, (01)00884838099258(21)72, (01)00884838099258(21)2139, (01)00884838099272(21)391, (01)00884838099258(21)141, (01)00884838085268(21)577, (01)00884838099241(21)211, (01)00884838085350(21)163, (01)00884838085251(21)317, (01)00884838099258(21)956, (01)00884838099258(21)2876, (01)00884838099258(21)3539, (01)00884838099258(21)2152, (01)00884838085251(21)522, (01)00884838085350(21)131, (01)00884838099258(21)627, (01)00884838085367(21)262, (01)00884838085367(21)261, (01)00884838099258(21)3026, (01)00884838099258(21)1383, (01)00884838099258(21)2965, (01)00884838099258(21)2967, (01)00884838085251(21)611, (01)00884838085350(21)71, (01)00884838099272(21)392, (01)00884838099258(21)2000, (01)00884838099258(21)2001, (01)00884838099258(21)1999, (01)00884838099258(21)2425, (01)00884838099258(21)28, (01)00884838099258(21)3444, (01)00884838099241(21)850, (01)00884838099265(21)453, (01)00884838099258(21)1872, (01)00884838085251(21)709, (01)00884838085251(21)677, (01)00884838099241(21)223, (01)00884838099203(21)49, (01)00884838085268(21)1041, (01)00884838099258(21)2199, (01)00884838099258(21)2148, (01)00884838099258(21)2144, (01)00884838085350(21)75, (01)00884838085350(21)132, (01)00884838099258(21)777, (01)00884838085350(21)244, (01)00884838099258(21)46, (01)00884838085350(21)208, (01)00884838085350(21)147, (01)00884838085367(21)277, (01)00884838085350(21)104, (01)00884838085367(21)150, (01)00884838099258(21)177, (01)00884838085268(21)1423, (01)00884838099272(21)173, (01)00884838099258(21)440, (01)00884838085367(21)172, (01)00884838085268(21)29, (01)00884838099241(21)1242, (01)00884838099258(21)1460, (01)00884838099272(21)793, (01)00884838099258(21)1611, (01)00884838085268(21)1043, (01)00884838099241(21)877, (01)00884838099241(21)872, (01)00884838099258(21)1274, (01)00884838099258(21)1301, (01)00884838085268(21)1270, (01)00884838099258(21)884, (01)00884838085268(21)1250, (01)00884838099258(21)2564, (01)00884838099258(21)746, (01)00884838099258(21)364, (01)00884838085367(21)117, (01)00884838099241(21)97, (01)00884838099272(21)147, (01)00884838085268(21)930, (01)00884838085268(21)805, (01)00884838099258(21)2340, (01)00884838099272(21)204, (01)00884838099258(21)3161, (01)00884838099210(21)358, (01)00884838099203(21)171, (01)00884838099258(21)2403, (01)00884838099272(21)467, (01)00884838099258(21)80, (01)00884838099272(21)1017, (01)00884838099241(21)1211, (01)00884838085268(21)1200, (01)00884838099258(21)574, (01)00884838085268(21)921, (01)00884838085251(21)535, (01)00884838085268(21)543, (01)00884838085268(21)366, (01)00884838085268(21)228, (01)00884838085268(21)30, (01)00884838085268(21)588, (01)00884838099258(21)3270, (01)00884838099258(21)2198, (01)00884838099210(21)113, (01)00884838099210(21)112, (01)00884838085367(21)130, (01)00884838085367(21)356, (01)00884838099234(21)127, (01)00884838099272(21)260, (01)00884838099258(21)113, (01)00884838099258(21)644, (01)00884838099265(21)53, (01)00884838099241(21)161, (01)00884838085367(21)143, (01)00884838085367(21)146, (01)00884838099272(21)381, (01)00884838099258(21)169, (01)00884838085251(21)170, (01)00884838099258(21)2328, (01)00884838099258(21)1797, (01)00884838099203(21)153, (01)00884838085268(21)1100, (01)00884838085367(21)84, (01)00884838085268(21)358, (01)00884838085268(21)359, (01)00884838085268(21)355, (01)00884838085268(21)356, (01)00884838085268(21)357, (01)00884838099272(21)912, (01)00884838099258(21)2138, (01)00884838099258(21)250, (01)00884838085268(21)1090, (01)00884838099258(21)1507, (01)00884838099258(21)1509, (01)00884838099241(21)1132, (01)00884838099241(21)386, (01)00884838099258(21)3053, (01)00884838099258(21)3052, (01)00884838099272(21)410, (01)00884838099258(21)1458, (01)00884838099272(21)508, (01)00884838099258(21)435, (01)00884838085268(21)445, (01)00884838085350(21)36, (01)00884838099258(21)3087, (01)00884838085268(21)703314, (01)00884838085268(21)945, (01)00884838099272(21)419, (01)00884838099272(21)313, (01)00884838099258(21)1403, (01)00884838099258(21)194, (01)00884838085268(21)1236, (01)00884838085268(21)1237, (01)00884838085268(21)1232, (01)00884838085268(21)1226, (01)00884838099234(21)6, (01)00884838085367(21)251, (01)00884838099241(21)1123, (01)00884838099241(21)208, (01)00884838099258(21)53, (01)00884838085367(21)469, (01)00884838099272(21)894, (01)00884838099272(21)472, (01)00884838099258(21)2961, (01)00884838099265(21)260, (01)00884838099265(21)259, (01)00884838099258(21)1829, (01)00884838099258(21)1828, (01)00884838099258(21)1830, (01)00884838099258(21)606, (01)00884838085350(21)234, (01)00884838099227(21)271, (01)00884838099227(21)264, (01)00884838099258(21)2417, (01)00884838099227(21)63, (01)00884838099241(21)1061, (01)00884838085268(21)801, (01)00884838099258(21)900, (01)00884838099258(21)332, (01)00884838099258(21)2636, (01)00884838099241(21)361, (01)00884838099258(21)3414, (01)00884838085268(21)742, (01)00884838099258(21)2427, (01)00884838099258(21)517, (01)00884838085268(21)1187, (01)00884838085251(21)723, (01)00884838085268(21)474, (01)00884838085268(21)222, (01)00884838085268(21)70, (01)00884838085268(21)247, (01)00884838085268(21)62, (01)00884838099241(21)258, (01)00884838085251(21)251, (01)00884838099258(21)2219, (01)00884838099241(21)597, (01)00884838099258(21)1354, (01)00884838099258(21)845, (01)00884838085251(21)298, (01)00884838099272(21)318, (01)00884838099258(21)847, (01)00884838099258(21)2131, (01)00884838085268(21)703450, (01)00884838085268(21)537, (01)00884838099272(21)473, (01)00884838099258(21)254, (01)00884838099258(21)843, (01)00884838099258(21)102, (01)00884838099258(21)3579, (01)00884838099258(21)1061, (01)00884838099234(21)185, (01)00884838099234(21)198, (01)00884838099234(21)186, (01)00884838099234(21)124, (01)00884838099234(21)123, (01)00884838085367(21)431, (01)00884838085268(21)48, (01)00884838085251(21)22, (01)00884838099258(21)1499, (01)00884838099258(21)1250, (01)00884838099241(21)727, (01)00884838085251(21)305, (01)00884838085268(21)351, (01)00884838085251(21)725, (01)00884838099258(21)3205, (01)00884838085251(21)728, (01)00884838085268(21)777, (01)00884838099258(21)3690, (01)00884838099258(21)960, (01)00884838099258(21)290, (01)00884838099241(21)1008, (01)00884838099258(21)1336, (01)00884838085251(21)524, (01)00884838099258(21)404, (01)00884838099258(21)2907, (01)00884838099258(21)2903, (01)00884838099258(21)1517, (01)00884838099258(21)1444, (01)00884838099272(21)174, (01)00884838099272(21)383, (01)00884838099258(21)572, (01)00884838099258(21)485, (01)00884838099241(21)1296, (01)00884838099265(21)300, (01)00884838099258(21)622, (01)00884838099258(21)621, (01)00884838099241(21)180, (01)00884838099241(21)1181, (01)00884838099258(21)2288, (01)00884838085268(21)41, (01)00884838099258(21)331, (01)00884838099265(21)268, (01)00884838099265(21)55, (01)00884838085268(21)898, (01)00884838085251(21)290, (01)00884838085268(21)703317, (01)00884838099272(21)689, (01)00884838085268(21)684, (01)00884838085268(21)653, (01)00884838099258(21)468, (01)00884838099258(21)467, (01)00884838099272(21)973, (01)00884838099258(21)3106, (01)00884838099265(21)235, (01)00884838099258(21)974, (01)00884838085282(21)64, (01)00884838085268(21)703334, (01)00884838099265(21)298, (01)00884838099258(21)465, (01)00884838099258(21)1240, (01)00884838099272(21)463, (01)00884838099258(21)1233, (01)00884838085268(21)912, (01)00884838085251(21)441, (01)00884838099272(21)823, (01)00884838099258(21)1235, (01)00884838099258(21)329, (01)00884838099258(21)328, (01)00884838085350(21)141, (01)00884838085350(21)70, (01)00884838085268(21)746, (01)00884838085268(21)703368, (01)00884838099258(21)139, (01)00884838099258(21)159, (01)00884838085268(21)703562, (01)00884838085350(21)26, (01)00884838099258(21)1461, (01)00884838099258(21)1512, (01)00884838099258(21)1511, (01)00884838099258(21)1508, (01)00884838099258(21)593, (01)00884838099258(21)490, (01)00884838085367(21)361, (01)00884838085268(21)54, (01)00884838099272(21)181, (01)00884838099258(21)3593, (01)00884838099258(21)2325, (01)00884838099258(21)2327, (01)00884838085268(21)1231, (01)00884838099258(21)1455, (01)00884838085268(21)731, (01)00884838099258(21)2957, (01)00884838099210(21)367, (01)00884838099258(21)279, (01)00884838099241(21)835, (01)00884838099258(21)955, (01)00884838099258(21)2027, (01)00884838085268(21)998, (01)00884838085268(21)540, (01)00884838085268(21)618, (01)00884838085251(21)384, (01)00884838085268(21)703360, (01)00884838085268(21)1179, (01)00884838085367(21)76, (01)00884838099258(21)3614, (01)00884838085268(21)1283, (01)00884838085268(21)668, (01)00884838099272(21)205, (01)00884838099258(21)253, (01)00884838099234(21)92, (01)00884838099272(21)361, (01)00884838099272(21)59, (01)00884838085268(21)1431, (01)00884838085251(21)480, (01)00884838085251(21)481, (01)00884838085268(21)607, (01)00884838099258(21)1243, (01)00884838099258(21)905, (01)00884838099258(21)1161, (01)00884838099265(21)233, (01)00884838099258(21)3419, (01)00884838099258(21)3448, (01)00884838085367(21)374, (01)00884838085268(21)706, (01)00884838085367(21)270, (01)00884838085268(21)879, (01)00884838085268(21)847, (01)00884838085251(21)360, (01)00884838099234(21)215, (01)00884838099258(21)2367, (01)00884838099241(21)668, (01)00884838099227(21)238, (01)00884838085268(21)703308, (01)00884838099210(21)120, (01)00884838099258(21)1324, (01)00884838099241(21)377, (01)00884838099258(21)2374, (01)00884838099272(21)756, (01)00884838099258(21)307, (01)00884838085251(21)364, (01)00884838099258(21)1959, (01)00884838099258(21)888, (01)00884838099258(21)3483, (01)00884838099258(21)1143, (01)00884838099241(21)882, (01)00884838099258(21)1493, (01)00884838099258(21)64, (01)00884838085251(21)906, (01)00884838085268(21)837, (01)00884838085251(21)294, (01)00884838085268(21)1251, (01)00884838085268(21)703575, (01)00884838099241(21)565, (01)00884838099258(21)130, (01)00884838099241(21)1103, (01)00884838099203(21)108, (01)00884838085367(21)114, (01)00884838085251(21)256, (01)00884838085268(21)417, (01)00884838085268(21)387, (01)00884838085251(21)240, (01)00884838099210(21)356, (01)00884838099258(21)3352, (01)00884838099258(21)1112, (01)00884838099258(21)1650, (01)00884838085268(21)45, (01)00884838085268(21)1316, (01)00884838099258(21)2682, (01)00884838099241(21)196, (01)00884838099258(21)1689, (01)00884838099258(21)116, (01)00884838099258(21)2669, (01)00884838099258(21)3157, (01)00884838085282(21)218, (01)00884838099258(21)1251, (01)00884838099272(21)379, (01)00884838085251(21)603, (01)00884838099258(21)3527, (01)00884838099258(21)3454, (01)00884838099265(21)354, (01)00884838099258(21)3668, (01)00884838085268(21)1393, (01)00884838099241(21)130, (01)00884838099241(21)1185, (01)00884838085350(21)148, (01)00884838085268(21)1290, (01)00884838099258(21)3489, (01)00884838099258(21)607, (01)00884838085367(21)511, (01)00884838099258(21)2437, (01)00884838099272(21)1008, (01)00884838099258(21)2928, (01)00884838099241(21)886, (01)00884838099241(21)887, (01)00884838099241(21)1150, (01)00884838099258(21)975, (01)00884838099258(21)844, (01)00884838085367(21)477, (01)00884838085268(21)334, (01)00884838099258(21)2426, (01)00884838099258(21)1060, (01)00884838085268(21)429, (01)00884838085268(21)320, (01)00884838085268(21)212, (01)00884838099258(21)327, (01)00884838085251(21)289, (01)00884838085268(21)290, (01)00884838099258(21)1776, (01)00884838099258(21)1775, (01)00884838099258(21)3189, (01)00884838099241(21)452, (01)00884838099241(21)453, (01)00884838099258(21)2018, (01)00884838085268(21)703409, (01)00884838099258(21)740, (01)00884838099258(21)747, (01)00884838099258(21)852, (01)00884838099258(21)851, (01)00884838099272(21)266, (01)00884838099265(21)289, (01)00884838085251(21)719, (01)00884838099258(21)1547, (01)00884838085268(21)669, (01)00884838085268(21)239, (01)00884838099258(21)979, (01)00884838099265(21)148, (01)00884838099258(21)862, (01)00884838099258(21)897, (01)00884838085367(21)232, (01)00884838099258(21)3528, (01)00884838099258(21)3471, (01)00884838099258(21)3531, (01)00884838099258(21)3530, (01)00884838099258(21)787, (01)00884838099258(21)2919, (01)00884838099258(21)2197, (01)00884838099258(21)2149, (01)00884838099258(21)176, (01)00884838085268(21)703390, (01)00884838099265(21)435, (01)00884838099265(21)434, (01)00884838099265(21)249, (01)00884838085268(21)629, (01)00884838085251(21)163, (01)00884838099258(21)418, (01)00884838085251(21)215, (01)00884838099258(21)219, (01)00884838099258(21)1998, (01)00884838085268(21)1078, (01)00884838099258(21)1091, (01)00884838085268(21)800, (01)00884838099258(21)1491, (01)00884838099272(21)1002, (01)00884838099258(21)1409, (01)00884838099258(21)582, (01)00884838099258(21)441, (01)00884838085268(21)56, (01)00884838099241(21)269, (01)00884838085268(21)703584, (01)00884838099258(21)84, (01)00884838085367(21)137, (01)00884838099258(21)3664, (01)00884838085251(21)271, (01)00884838099258(21)798, (01)00884838085268(21)470, (01)00884838099241(21)888, (01)00884838085268(21)595, (01)00884838099258(21)823, (01)00884838099258(21)824, (01)00884838099258(21)2506, (01)00884838099258(21)949, (01)00884838099234(21)573, (01)00884838099241(21)1188, (01)00884838099258(21)174, (01)00884838085251(21)788, (01)00884838099227(21)360, (01)00884838099258(21)2926, (01)00884838099258(21)350, (01)00884838099234(21)91, (01)00884838099258(21)3673, (01)00884838099258(21)1468, (01)00884838085268(21)64, (01)00884838099258(21)645, (01)00884838099241(21)567, (01)00884838099258(21)940, (01)00884838085268(21)671, (01)00884838099258(21)271, (01)00884838085268(21)793, (01)00884838085268(21)1381, (01)00884838085268(21)1382, (01)00884838085268(21)1059, (01)00884838099265(21)409, (01)00884838099258(21)76, (01)00884838085268(21)747, (01)00884838099234(21)529, (01)00884838099258(21)1170, (01)00884838099258(21)3573, (01)00884838099258(21)761, (01)00884838085268(21)585, (01)00884838085275(21)108, (01)00884838099234(21)523, (01)00884838085282(21)335, (01)00884838099272(21)398, (01)00884838085268(21)1246, (01)00884838085251(21)412, (01)00884838099258(21)109, (01)00884838099258(21)1288, (01)00884838085251(21)627, (01)00884838085367(21)107, (01)00884838085268(21)663, (01)00884838099265(21)84, (01)00884838099258(21)1982, (01)00884838099258(21)160, (01)00884838099258(21)786, (01)00884838099258(21)34, (01)00884838099258(21)2494, (01)00884838099241(21)966, (01)00884838085268(21)1212, (01)00884838099258(21)2699, (01)00884838099272(21)75, (01)00884838085268(21)703333, (01)00884838099258(21)1000, (01)00884838099241(21)449, (01)00884838099265(21)111, (01)00884838085268(21)1274, (01)00884838085268(21)255, (01)00884838085268(21)455, (01)00884838085251(21)644, (01)00884838085251(21)279, (01)00884838085268(21)637, (01)00884838085251(21)141, (01)00884838099258(21)2557, (01)00884838099272(21)786, (01)00884838085367(21)282, (01)00884838085350(21)133, (01)00884838099258(21)793, (01)00884838085268(21)862, (01)00884838099258(21)3388, (01)00884838099258(21)1353, (01)00884838085268(21)1379, (01)00884838099258(21)581, (01)00884838099241(21)435, (01)00884838099272(21)19, (01)00884838085268(21)1235, (01)00884838099258(21)1827, (01)00884838099272(21)1125, (01)00884838099258(21)2150, (01)00884838099258(21)220, (01)00884838085268(21)810, (01)00884838099258(21)2681, (01)00884838099258(21)225, (01)00884838099258(21)235, (01)00884838099241(21)441, (01)00884838085268(21)1164, (01)00884838099258(21)714, (01)00884838085268(21)227, (01)00884838085268(21)438, (01)00884838085251(21)844, (01)00884838085251(21)876, (01)00884838085268(21)437, (01)00884838085251(21)292, (01)00884838099258(21)2519, (01)00884838099258(21)2518, (01)00884838099258(21)452, (01)00884838085268(21)703345, (01)00884838085251(21)748, (01)00884838099258(21)1213, (01)00884838085367(21)487, (01)00884838085268(21)990, (01)00884838099241(21)669, (01)00884838099234(21)205, (01)00884838099258(21)1126, (01)00884838085251(21)483, (01)00884838085251(21)497, (01)00884838099234(21)20, (01)00884838085367(21)413, (01)00884838099258(21)2156, (01)00884838099258(21)766, (01)00884838085268(21)314, (01)00884838099258(21)1254, (01)00884838099265(21)227, (01)00884838085268(21)613, (01)00884838085350(21)57, (01)00884838085251(21)309, (01)00884838099272(21)213, (01)00884838099272(21)214, (01)00884838099272(21)215, (01)00884838099258(21)522, (01)00884838099258(21)523, (01)00884838099258(21)518, (01)00884838085367(21)337, (01)00884838099241(21)351, (01)00884838099258(21)726, (01)00884838099258(21)1316, (01)00884838099258(21)1314, (01)00884838099258(21)1311, (01)00884838085268(21)427, (01)00884838085367(21)101, (01)00884838099258(21)3299, (01)00884838085268(21)60, (01)00884838099258(21)2770, (01)00884838099241(21)508, (01)00884838099272(21)1007, (01)00884838099258(21)3401, (01)00884838099258(21)3049, (01)00884838099272(21)814, (01)00884838085251(21)635, (01)00884838085251(21)425, (01)00884838085251(21)232, (01)00884838099258(21)2030, (01)00884838085268(21)703486, (01)00884838099258(21)343, (01)00884838099234(21)37, (01)00884838099258(21)1410, (01)00884838099258(21)2733, (01)00884838099258(21)643, (01)00884838085251(21)612, (01)00884838085268(21)327, (01)00884838099258(21)3369, (01)00884838085268(21)602, (01)00884838099258(21)3082, (01)00884838099241(21)750, (01)00884838099272(21)969, (01)00884838099272(21)970, (01)00884838099258(21)3164, (01)00884838085268(21)318, (01)00884838085268(21)937, (01)00884838099258(21)1527, (01)00884838099241(21)164, (01)00884838085367(21)213, (01)00884838085268(21)703515, (01)00884838099258(21)104, (01)0088483808 5268(21)703455, (01)00884838099272(21)469, (01)00884838099258(21)1201, (01)00884838085268(21)703456, (01)00884838085268(21)703457, (01)00884838099258(21)503, (01)00884838085268(21)948, (01)00884838085268(21)942, (01)00884838099272(21)219, (01)00884838099258(21)659, (01)00884838099258(21)1015, (01)00884838099258(21)1335, (01)00884838099272(21)462, (01)00884838099272(21)981, (01)00884838085268(21)1201, (01)00884838085268(21)1247, (01)00884838085367(21)82, (01)00884838085268(21)702, (01)00884838099258(21)185, (01)00884838099258(21)1231, (01)00884838099258(21)944, (01)00884838099241(21)355, (01)00884838099241(21)139, (01)00884838099258(21)196, (01)00884838085251(21)797, (01)00884838099258(21)2752, (01)00884838099241(21)619, (01)00884838099258(21)1247, (01)00884838099258(21)833, (01)00884838099241(21)1236, (01)00884838099258(21)1029, (01)00884838085350(21)128, (01)00884838099258(21)258, (01)00884838099241(21)133, (01)00884838085268(21)333, (01)00884838085251(21)247, (01)00884838099258(21)2843, (01)00884838099241(21)89, (01)00884838099258(21)330, (01)00884838099241(21)370, (01)00884838099272(21)296, (01)00884838085268(21)703388, (01)00884838099258(21)172, (01)00884838085251(21)733, (01)00884838099258(21)372, (01)00884838085268(21)519, (01)00884838099258(21)2955, (01)00884838085367(21)275, (01)00884838099272(21)149, (01)00884838085367(21)265, (01)00884838085268(21)644, (01)00884838085251(21)517, (01)00884838085251(21)162, (01)00884838099241(21)714, (01)00884838099258(21)1477, (01)00884838099241(21)598, (01)00884838099258(21)379, (01)00884838085268(21)845, (01)00884838085268(21)846, (01)00884838099258(21)666, (01)00884838085251(21)923, (01)00884838085251(21)888, (01)00884838085268(21)1415, (01)00884838085268(21)1000, (01)00884838085268(21)814, (01)00884838085350(21)188, (01)00884838085268(21)352, (01)00884838085251(21)204, (01)00884838085251(21)205, (01)00884838085251(21)206, (01)00884838085268(21)284, (01)00884838085268(21)703356, (01)00884838099258(21)722, (01)00884838099258(21)585, (01)00884838085367(21)485, (01)00884838099258(21)1156, (01)00884838099241(21)119, (01)00884838085268(21)703380, (01)00884838085268(21)703381, (01)00884838099258(21)3692, (01)00884838099258(21)3661, (01)00884838099258(21)3662, (01)00884838099258(21)3650, (01)00884838099241(21)1307, (01)00884838099241(21)1306, (01)00884838099241(21)871, (01)00884838099258(21)649, (01)00884838099258(21)875, (01)00884838099258(21)874, (01)00884838085268(21)265, (01)00884838085367(21)459, (01)00884838099258(21)3596, (01)00884838085268(21)679, (01)00884838099241(21)698, (01)00884838099258(21)1787, (01)00884838099241(21)697, (01)00884838085268(21)331, (01)00884838085268(21)332, (01)00884838085268(21)330, (01)00884838085251(21)197, (01)00884838085251(21)195, (01)00884838085268(21)329, (01)00884838085268(21)328, (01)00884838099241(21)1036, (01)00884838099241(21)1035, (01)00884838085251(21)218, (01)00884838099258(21)480, (01)00884838099258(21)495, (01)00884838099258(21)2242, (01)00884838099272(21)704, (01)00884838099241(21)154, (01)00884838099241(21)116, (01)00884838099258(21)173, (01)00884838099241(21)912, (01)00884838099241(21)100, (01)00884838099241(21)99, (01)00884838099241(21)387, (01)00884838099241(21)349, (01)00884838099241(21)136, (01)00884838099258(21)2667, (01)00884838085268(21)1017, (01)00884838085350(21)151, (01)00884838085268(21)1374, (01)00884838085268(21)1039, (01)00884838085268(21)750, (01)00884838099258(21)1478, (01)00884838099241(21)775, (01)00884838099241(21)903, (01)00884838099258(21)2029, (01)00884838099258(21)3505, (01)00884838085268(21)1016, (01)00884838085268(21)783, (01)00884838085251(21)444, (01)00884838085268(21)703685, (01)00884838085268(21)703550, (01)00884838099258(21)3459, (01)00884838085268(21)703551, (01)00884838099241(21)867, (01)00884838085282(21)513, (01)00884838099258(21)3637, (01)00884838110571(21)57, (01)00884838085251(21)986, (01)00884838103276(21)57, (01)00884838099258(21)2772, (01)00884838103276(21)161, (01)00884838085251(21)561, (01)00884838110571(21)54, (01)00884838099227(21)23, (01)00884838099241(21)680, (01)00884838099272(21)573, (01)00884838085268(21)703813, (01)00884838099210(21)262, (01)00884838110564(21)379, (01)00884838085268(21)704301, (01)00884838103276(21)43, (01)00884838085251(21)503, (01)00884838103276(21)179, (01)00884838099203(21)104, (01)00884838099272(21)458, (01)00884838099227(21)127, (01)00884838099227(21)159, (01)00884838085268(21)780, (01)00884838085268(21)704114, (01)00884838085268(21)703719, (01)00884838099227(21)197, (01)00884838099258(21)2937, (01)00884838110564(21)125, (01)00884838099241(21)527, (01)00884838099227(21)149, (01)00884838099241(21)1227, (01)00884838099241(21)629, (01)00884838099258(21)1863, (01)00884838099258(21)780, (01)00884838099241(21)1264, (01)00884838099203(21)169, (01)00884838099241(21)834, (01)00884838099258(21)1622, (01)00884838099258(21)1089, (01)00884838099234(21)659, (01)00884838099272(21)987, (01)00884838099241(21)1226, (01)00884838085268(21)217, (01)00884838085282(21)61, (01)00884838099241(21)837, (01)00884838099241(21)155, (01)00884838099272(21)880, (01)00884838099227(21)54, (01)00884838085251(21)689, (01)00884838099227(21)56, (01)00884838085268(21)476, (01)00884838085282(21)162, (01)00884838085282(21)161, (01)00884838085268(21)815, (01)00884838085251(21)499, (01)00884838085268(21)816, (01)00884838099234(21)145, (01)00884838099258(21)1341, (01)00884838099258(21)352, (01)00884838099272(21)141, (01)00884838099258(21)353, (01)00884838099258(21)257, (01)00884838085268(21)703565, (01)00884838085282(21)297, (01)00884838085268(21)1280, (01)00884838085282(21)169, (01)00884838085275(21)77, (01)00884838085367(21)276, (01)00884838085268(21)214, (01)00884838085251(21)135, (01)00884838099258(21)1568, (01)00884838099227(21)230, (01)00884838099203(21)45, (01)00884838099227(21)192, (01)00884838099272(21)711, (01)00884838099258(21)3464, (01)00884838099203(21)101, (01)00884838085268(21)704304, (01)00884838099258(21)181, (01)00884838099241(21)18, (01)00884838099241(21)17, (01)00884838085268(21)561, (01)00884838099272(21)693, (01)00884838085251(21)589, (01)00884838099272(21)510, (01)00884838099227(21)156, (01)00884838099258(21)1786, (01)00884838085268(21)703338, (01)00884838085251(21)38, (01)00884838085268(21)59, (01)00884838099241(21)804, (01)00884838099234(21)3, (01)00884838099241(21)235, (01)00884838099241(21)477, (01)00884838099241(21)249, (01)00884838099272(21)882, (01)00884838099258(21)623, (01)00884838085251(21)977, (01)00884838099258(21)2852, (01)00884838099227(21)250, (01)00884838085275(21)82, (01)00884838099272(21)1157, (01)00884838099265(21)455, (01)00884838099258(21)2931, (01)00884838099234(21)567, (01)00884838099241(21)1025, (01)00884838099272(21)226, (01)00884838099265(21)365, (01)00884838099241(21)1177, (01)00884838099258(21)3293, (01)00884838099258(21)1764, (01)00884838099241(21)1153, (01)00884838085268(21)719, (01)00884838085251(21)738, (01)00884838099227(21)98, (01)00884838099241(21)1210, (01)00884838099272(21)509, (01)00884838099241(21)167, (01)00884838099203(21)193, (01)00884838099241(21)1293, (01)00884838099241(21)1290, (01)00884838099227(21)31, (01)00884838085251(21)476, (01)00884838085251(21)380, (01)00884838099210(21)37, (01)00884838099241(21)227, (01)00884838099234(21)539, (01)00884838099241(21)965, (01)00884838085268(21)894, (01)00884838085268(21)745, (01)00884838085251(21)453, (01)00884838085282(21)131, (01)00884838085268(21)744, (01)00884838085251(21)24, (01)00884838085251(21)690, (01)00884838099258(21)2742, (01)00884838085268(21)704059, (01)00884838099234(21)645, (01)00884838099272(21)584, (01)00884838099227(21)128, (01)00884838099258(21)3474, (01)00884838099272(21)21, (01)00884838099241(21)999, (01)00884838085251(21)356, (01)00884838099265(21)284, (01)00884838099258(21)1988, (01)00884838099210(21)318, (01)00884838099258(21)1974, (01)00884838099258(21)1851, (01)00884838099241(21)46, (01)00884838099203(21)135, (01)00884838099227(21)81, (01)00884838099258(21)2787, (01)00884838099258(21)2804, (01)00884838099241(21)1131, (01)00884838085268(21)729, (01)00884838099241(21)141, (01)00884838085251(21)343, (01)00884838085251(21)344, (01)00884838085268(21)943, (01)00884838099272(21)1050, (01)00884838099227(21)24, (01)00884838085350(21)68, (01)00884838085251(21)174, (01)00884838085367(21)62, (01)00884838085268(21)302, (01)00884838110564(21)79, (01)00884838099258(21)2933, (01)00884838110564(21)176, (01)00884838099258(21)1632, (01)00884838099258(21)1627, (01)00884838099258(21)1629, (01)00884838099272(21)647, (01)00884838099241(21)1282, (01)00884838099258(21)2777, (01)00884838085268(21)337, (01)00884838085268(21)1035, (01)00884838099258(21)534, (01)00884838085251(21)126, (01)00884838085282(21)479, (01)00884838099227(21)37, (01)00884838099272(21)225, (01)00884838099227(21)320, (01)00884838099234(21)156, (01)00884838110571(21)118, (01)00884838099265(21)362, (01)00884838099241(21)1005, (01)00884838099227(21)179, (01)00884838099258(21)1008, (01)00884838099258(21)1709, (01)00884838099272(21)666, (01)00884838099272(21)303, (01)00884838099272(21)202, (01)00884838099272(21)201, (01)00884838099258(21)497, (01)00884838099272(21)184, (01)00884838085251(21)898, (01)00884838099258(21)3497, (01)00884838085251(21)482, (01)00884838099234(21)308, (01)00884838099241(21)1129, (01)00884838099258(21)1163, (01)00884838099241(21)1115, (01)00884838085268(21)703552, (01)00884838099227(21)270, (01)00884838085282(21)190, (01)00884838099227(21)151, (01)00884838099258(21)578, (01)00884838099272(21)243, (01)00884838085251(21)378, (01)00884838085282(21)268, (01)00884838099227(21)53, (01)00884838099241(21)237, (01)00884838099203(21)51, (01)00884838085268(21)704184, (01)00884838099258(21)2550, (01)00884838099258(21)1705, (01)00884838099258(21)341, (01)00884838099258(21)359, (01)00884838085268(21)703613, (01)00884838110564(21)233, (01)00884838099234(21)333, (01)00884838099227(21)246, (01)00884838099234(21)335, (01)00884838099241(21)348, (01)00884838099272(21)250, (01)00884838099272(21)178, (01)00884838099272(21)179, (01)00884838085282(21)375, (01)00884838085282(21)359, (01)00884838085282(21)220, (01)00884838085282(21)173, (01)00884838099258(21)394, (01)00884838099241(21)287, (01)00884838103276(21)229, (01)00884838099258(21)1943, (01)00884838099258(21)1942, (01)00884838099272(21)1139, (01)00884838099241(21)1325, (01)00884838099210(21)152, (01)00884838099258(21)1802, (01)00884838099203(21)114, (01)00884838099258(21)14, (01)00884838099272(21)674, (01)00884838085251(21)764, (01)00884838085367(21)506, (01)00884838103276(21)249, (01)00884838085367(21)608, (01)00884838085268(21)704242, (01)00884838085268(21)704240, (01)00884838085268(21)703662, (01)00884838099241(21)135, (01)00884838085350(21)233, (01)00884838085367(21)473, (01)00884838099258(21)1216, (01)00884838099258(21)334, (01)00884838085251(21)694, (01)00884838085251(21)693, (01)00884838099272(21)1062, (01)00884838085350(21)111, (01)00884838099258(21)3696, (01)00884838085367(21)417, (01)00884838085251(21)771, (01)00884838085251(21)770, (01)00884838085251(21)773, (01)00884838085251(21)772, (01)00884838085268(21)1340, (01)00884838085251(21)439, (01)00884838085268(21)549, (01)00884838099227(21)84, (01)00884838099272(21)575, (01)00884838099258(21)1557, (01)00884838099227(21)245, (01)00884838099272(21)1121, (01)00884838085268(21)703838, (01)00884838099241(21)417, (01)00884838099241(21)416, (01)00884838099258(21)935, (01)00884838099241(21)414, (01)00884838099227(21)38, (01)00884838099258(21)3129, (01)00884838099227(21)28, (01)00884838099234(21)570, (01)00884838099272(21)950, (01)00884838099258(21)2089, (01)00884838099258(21)1670, (01)00884838099258(21)1686, (01)00884838099272(21)368, (01)00884838099258(21)63, (01)00884838099241(21)132, (01)00884838085275(21)145, (01)00884838099234(21)560, (01)00884838099258(21)2186, (01)00884838099258(21)2546, (01)00884838099258(21)3086, (01)00884838099272(21)861, (01)00884838110564(21)391, (01)00884838099234(21)44, (01)00884838099241(21)814, (01)00884838099272(21)176, (01)00884838085282(21)256, (01)00884838085268(21)1037, (01)00884838099241(21)836, (01)00884838099272(21)712, (01)00884838099272(21)829, (01)00884838099272(21)413, (01)00884838085350(21)255, (01)00884838085251(21)191, (01)00884838085268(21)703823, (01)00884838099210(21)364, (01)00884838099210(21)395, (01)00884838110564(21)296, (01)00884838099272(21)997, (01)00884838099265(21)468, (01)00884838085282(21)290, (01)00884838099241(21)1245, (01)00884838099203(21)36, (01)00884838099227(21)29, (01)00884838099241(21)356, (01)00884838099241(21)810, (01)00884838099241(21)809, (01)00884838099272(21)249, (01)00884838099234(21)569, (01)00884838099227(21)170, (01)00884838099241(21)303, (01)00884838099258(21)3521, (01)00884838099272(21)589, (01)00884838099241(21)1214, (01)00884838099258(21)2513, (01)00884838099234(21)418, (01)00884838099258(21)3105, (01)00884838099227(21)30, (01)00884838085282(21)502, (01)00884838085268(21)703670, (01)00884838085282(21)425, (01)00884838085268(21)1026, (01)00884838085367(21)302, (01)00884838085367(21)314, (01)00884838099241(21)1099, (01)00884838099258(21)2187, (01)00884838099241(21)813, (01)00884838085251(21)938, (01)00884838085251(21)981, (01)00884838085251(21)980, (01)00884838085251(21)982, (01)00884838110564(21)334, (01)00884838099258(21)558, (01)00884838085251(21)307, (01)00884838099272(21)1046, (01)00884838099272(21)576, (01)00884838099234(21)525, (01)00884838110571(21)66, (01)00884838085251(21)628, (01)00884838110571(21)49, (01)00884838099234(21)280, (01)00884838099258(21)942, (01)00884838099258(21)941, (01)00884838085251(21)235, (01)00884838085350(21)43, (01)00884838085268(21)379, (01)00884838085367(21)88, (01)00884838085268(21)377, (01)00884838085268(21)380, (01)00884838099258(21)3473, (01)00884838099241(21)1248, (01)00884838099258(21)3264, (01)00884838099241(21)1232, (01)00884838099258(21)2244, (01)00884838099241(21)1167, (01)00884838099272(21)738, (01)00884838099258(21)3099, (01)00884838099241(21)904, (01)00884838099258(21)2373, (01)00884838099241(21)875, (01)00884838099272(21)708, (01)00884838099258(21)1330, (01)00884838099258(21)2245, (01)00884838099241(21)587, (01)00884838099241(21)812, (01)00884838085268(21)703355, (01)00884838085268(21)703699, (01)00884838099258(21)1369, (01)00884838099258(21)1349, (01)00884838099258(21)1609, (01)00884838099258(21)1350, (01)00884838099241(21)591, (01)00884838099258(21)1326, (01)00884838099258(21)937, (01)00884838099258(21)801, (01)00884838099241(21)322, (01)00884838085268(21)703870, (01)00884838085367(21)468, (01)00884838085268(21)703654, (01)00884838085268(21)703449, (01)00884838085268(21)1405, (01)00884838085251(21)746, (01)00884838085367(21)126, (01)00884838085268(21)395, (01)00884838085251(21)511, (01)00884838085251(21)526, (01)00884838085367(21)211, (01)00884838085268(21)662, (01)00884838085268(21)451, (01)00884838085251(21)266, (01)00884838085350(21)33, (01)00884838085268(21)433, (01)00884838085251(21)265, (01)00884838085251(21)264, (01)00884838085251(21)178, (01)00884838085251(21)179, (01)00884838085268(21)258, (01)00884838085268(21)259, (01)00884838085251(21)139, (01)00884838085268(21)260, (01)00884838099258(21)2052, (01)00884838099203(21)124, (01)00884838099258(21)1970, (01)00884838099241(21)756, (01)00884838085251(21)287, (01)00884838099203(21)197, (01)00884838099272(21)1083, (01)00884838099258(21)776, (01)00884838085268(21)703612, (01)00884838099272(21)1098, (01)00884838099241(21)516, (01)00884838085367(21)422, (01)00884838110564(21)55, (01)00884838099227(21)338, (01)00884838099265(21)56, (01)00884838103276(21)65, (01)00884838099234(21)616, (01)00884838099234(21)387, (01)00884838099272(21)717, (01)00884838099258(21)2571, (01)00884838099258(21)1048, (01)00884838085282(21)89, (01)00884838099258(21)2146, (01)00884838099272(21)697, (01)00884838110564(21)81, (01)00884838085268(21)704300, (01)00884838085268(21)704299, (01)00884838099265(21)232, (01)00884838099272(21)1126, (01)00884838099258(21)149, (01)00884838099258(21)734, (01)00884838099258(21)201, (01)00884838085268(21)1306, (01)00884838085251(21)362, (01)00884838085251(21)361, (01)00884838099265(21)269, (01)00884838099265(21)230, (01)00884838099258(21)3117, (01)00884838099241(21)1016, (01)00884838099265(21)345, (01)00884838110564(21)37, (01)00884838099272(21)665, (01)00884838110564(21)21, (01)00884838099272(21)416, (01)00884838099272(21)977, (01)00884838099272(21)764, (01)00884838099272(21)386, (01)00884838099227(21)324, (01)00884838099258(21)1804, (01)00884838099272(21)595, (01)00884838099258(21)3292, (01)00884838085367(21)454, (01)00884838099258(21)800, (01)00884838099258(21)799, (01)00884838099272(21)859, (01)00884838099265(21)296, (01)00884838085268(21)970, (01)00884838085251(21)1003, (01)00884838099227(21)108, (01)00884838085268(21)1248, (01)00884838099227(21)305, (01)00884838085268(21)449, (01)00884838099258(21)1206, (01)00884838085268(21)822, (01)00884838085251(21)348, (01)00884838085268(21)703999, (01)00884838085268(21)1047, (01)00884838099272(21)1113, (01)00884838099234(21)568, (01)00884838099210(21)146, (01)00884838099241(21)10, (01)00884838085251(21)897, (01)00884838085367(21)410, (01)00884838099258(21)1747, (01)00884838099258(21)1385, (01)00884838099258(21)2405, (01)00884838085275(21)47, (01)00884838099272(21)721, (01)00884838085268(21)1019, (01)00884838085367(21)283, (01)00884838099258(21)2680, (01)00884838099227(21)168, (01)00884838099258(21)1812, (01)00884838085268(21)1241, (01)00884838085268(21)557, (01)00884838099258(21)1204, (01)00884838085268(21)1143, (01)00884838085268(21)752, (01)00884838085251(21)845, (01)00884838099258(21)1529, (01)00884838099241(21)1256, (01)00884838099272(21)1043, (01)00884838099265(21)462, (01)00884838085251(21)960, (01)00884838085268(21)704129, (01)00884838110571(21)154, (01)00884838099258(21)2060, (01)00884838099241(21)582, (01)00884838099258(21)2583, (01)00884838099258(21)3216, (01)00884838099272(21)797, (01)00884838099258(21)3367, (01)00884838099210(21)246, (01)00884838099258(21)3334, (01)00884838099258(21)2598, (01)00884838099210(21)269, (01)00884838099241(21)694, (01)00884838099241(21)537, (01)00884838099210(21)75, (01)00884838110571(21)32, (01)00884838099234(21)471, (01)00884838099258(21)2479, (01)00884838099234(21)428, (01)00884838099241(21)1055, (01)00884838110571(21)7, (01)00884838085367(21)231, (01)00884838099258(21)3502, (01)00884838099272(21)957, (01)00884838099210(21)304, (01)00884838110564(21)135, (01)00884838103276(21)76, (01)00884838099258(21)908, (01)00884838085350(21)22, (01)00884838085367(21)548, (01)00884838085251(21)1036, (01)00884838103276(21)233, (01)00884838085282(21)303, (01)00884838099258(21)2555, (01)00884838099227(21)172, (01)00884838099241(21)576, (01)00884838099227(21)356, (01)00884838099272(21)1006, (01)00884838099258(21)1005, (01)00884838085251(21)967, (01)00884838099241(21)134, (01)00884838085367(21)466, (01)00884838099203(21)129, (01)00884838085268(21)558, (01)00884838085251(21)374, (01)00884838085251(21)375, (01)00884838085251(21)564, (01)00884838085268(21)947, (01)00884838099258(21)3235, (01)00884838085367(21)495, (01)00884838099265(21)374, (01)00884838099234(21)465, (01)00884838099227(21)346, (01)00884838085268(21)1190, (01)00884838085268(21)1196, (01)00884838099272(21)498, (01)00884838085268(21)1172, (01)00884838085350(21)146, (01)00884838085251(21)142, (01)00884838099258(21)2628, (01)00884838099241(21)963, (01)00884838099258(21)2711, (01)00884838099241(21)961, (01)00884838099258(21)1452, (01)00884838099258(21)2044, (01)00884838099241(21)247, (01)00884838099272(21)60, (01)00884838085251(21)469, (01)00884838099258(21)3522, (01)00884838099241(21)1222, (01)00884838099265(21)246, (01)00884838085367(21)207, (01)00884838099234(21)556, (01)00884838085350(21)256, (01)00884838085251(21)934, (01)00884838099241(21)1272, (01)00884838099241(21)1217, (01)00884838099258(21)2025, (01)00884838099258(21)1976, (01)00884838099258(21)1978, (01)00884838099241(21)584, (01)00884838099241(21)7, (01)00884838099272(21)232, (01)00884838085268(21)1319, (01)00884838085367(21)445, (01)00884838085367(21)167, (01)00884838085367(21)537, (01)00884838085268(21)703712, (01)00884838099227(21)62, (01)00884838099241(21)923, (01)00884838099258(21)2576, (01)00884838099265(21)324, (01)00884838099258(21)3163, (01)00884838099272(21)423, (01)00884838099241(21)933, (01)00884838085268(21)1397, (01)00884838085268(21)1398, (01)00884838085367(21)438, (01)00884838099241(21)709, (01)00884838099210(21)316, (01)00884838099272(21)789, (01)00884838099272(21)682, (01)00884838085282(21)263, (01)00884838085282(21)264, (01)00884838085282(21)157, (01)00884838085367(21)289, (01)00884838085367(21)233, (01)00884838085282(21)130, (01)00884838085367(21)173, (01)00884838085282(21)60, (01)00884838099258(21)1151, (01)00884838099203(21)4, (01)00884838099241(21)193, (01)00884838085251(21)565, (01)00884838099258(21)531, (01)00884838099234(21)629, (01)00884838099272(21)698, (01)00884838085350(21)270, (01)00884838099203(21)154, (01)00884838099227(21)236, (01)00884838099227(21)25, (01)00884838085268(21)289, (01)00884838099203(21)106, (01)00884838085268(21)31, (01)00884838099258(21)1861, (01)00884838099241(21)128, (01)00884838099258(21)1164, (01)00884838099272(21)331, (01)00884838085350(21)238, (01)00884838099272(21)1097, (01)00884838099258(21)2561, (01)00884838085268(21)704295, (01)00884838085268(21)994, (01)00884838099258(21)1027, (01)00884838099258(21)1026, (01)00884838099210(21)323, (01)00884838099210(21)343, (01)00884838099272(21)1094, (01)00884838099210(21)345, (01)00884838099210(21)346, (01)00884838099258(21)3347, (01)00884838099210(21)320, (01)00884838099210(21)362, (01)00884838099210(21)338, (01)00884838099210(21)339, (01)00884838099210(21)340, (01)00884838099210(21)344, (01)00884838099210(21)336, (01)00884838099210(21)307, (01)00884838099258(21)2509, (01)00884838099258(21)2588, (01)00884838099210(21)230, (01)00884838099241(21)615, (01)00884838099258(21)1449, (01)00884838099234(21)176, (01)00884838099210(21)174, (01)00884838099241(21)393, (01)00884838099241(21)434, (01)00884838099234(21)102, (01)00884838099234(21)103, (01)00884838085268(21)703702, (01)00884838085367(21)536, (01)00884838085268(21)703655, (01)00884838085251(21)975, (01)00884838085275(21)146, (01)00884838085251(21)929, (01)00884838085251(21)835, (01)00884838085268(21)703326, (01)00884838085251(21)836, (01)00884838085282(21)253, (01)00884838085275(21)80, (01)00884838110564(21)278, (01)00884838099234(21)178, (01)00884838085268(21)703939, (01)00884838099227(21)313, (01)00884838099272(21)378, (01)00884838099227(21)339, (01)00884838085268(21)531, (01)00884838103276(21)232, (01)00884838099210(21)308, (01)00884838099272(21)808, (01)00884838099210(21)182, (01)00884838099234(21)109, (01)00884838099265(21)128, (01)00884838099258(21)765, (01)00884838099234(21)16, (01)00884838085282(21)254, (01)00884838085251(21)681, (01)00884838085367(21)204, (01)00884838099241(21)1165, (01)00884838099241(21)854, (01)00884838099258(21)2349, (01)00884838099258(21)2841, (01)00884838099272(21)1107, (01)00884838099258(21)501, (01)00884838099241(21)550, (01)00884838099258(21)3428, (01)00884838085268(21)703802, (01)00884838099258(21)546, (01)00884838099258(21)167, (01)00884838099234(21)299, (01)00884838099241(21)657, (01)00884838110564(21)390, (01)00884838085251(21)1069, (01)00884838085251(21)1070, (01)00884838085251(21)1071, (01)00884838085251(21)1072, (01)00884838085268(21)703897, (01)00884838099227(21)104, (01)00884838085268(21)704203, (01)00884838085268(21)704266, (01)00884838099258(21)1475, (01)00884838099227(21)321, (01)00884838099234(21)234, (01)00884838110571(21)87, (01)00884838099241(21)1283, (01)00884838099258(21)2586, (01)00884838099227(21)59, (01)00884838099241(21)1196, (01)00884838099258(21)3605, (01)00884838099258(21)1470, (01)00884838099203(21)146, (01)00884838099258(21)3030, (01)00884838099258(21)1690, (01)00884838099210(21)27, (01)00884838099241(21)1258, (01)00884838099272(21)393, (01)00884838085367(21)77, (01)00884838085282(21)257, (01)00884838099203(21)68, (01)00884838085268(21)1389, (01)00884838099265(21)241, (01)00884838099258(21)2496, (01)00884838099258(21)3079, (01)00884838085251(21)893, (01)00884838099265(21)180, (01)00884838099258(21)778, (01)00884838085350(21)187, (01)00884838103276(21)113, (01)00884838099258(21)2541, (01)00884838085268(21)703953, (01)00884838099272(21)949, (01)00884838085251(21)1061, (01)00884838085268(21)703986, (01)00884838099227(21)109, (01)00884838085350(21)74, (01)00884838085251(21)382, (01)00884838099258(21)2464, (01)00884838099234(21)566, (01)00884838099241(21)1212, (01)00884838099210(21)390, (01)00884838099258(21)2585, (01)00884838085268(21)703600, (01)00884838099241(21)612, (01)00884838099241(21)926, (01)00884838099203(21)195, (01)00884838099258(21)2178, (01)00884838099258(21)1668, (01)00884838099258(21)2959, (01)00884838099258(21)3597, (01)00884838099265(21)367, (01)00884838099272(21)718, (01)00884838099258(21)513, (01)00884838099265(21)431, (01)00884838099241(21)555, (01)00884838099241(21)398, (01)00884838099241(21)353, (01)00884838099265(21)450, (01)00884838099272(21)663, (01)00884838099203(21)130, (01)00884838085268(21)570, (01)00884838099241(21)848, (01)00884838099272(21)195, (01)00884838085282(21)280, (01)00884838099258(21)1795, (01)00884838099258(21)1623, (01)00884838099241(21)931, (01)00884838099258(21)429, (01)00884838099227(21)101, (01)00884838085251(21)324, (01)00884838099258(21)3255, (01)00884838099241(21)213, (01)00884838085268(21)703769, (01)00884838099227(21)19, (01)00884838099265(21)460, (01)00884838099258(21)1338, (01)00884838099210(21)56, (01)00884838099241(21)364, (01)00884838099265(21)152, (01)00884838099272(21)1092, (01)00884838085268(21)458, (01)00884838099265(21)292, (01)00884838099272(21)715, (01)00884838099272(21)412, (01)00884838085268(21)962, (01)00884838099272(21)1143, (01)00884838099265(21)145, (01)00884838085367(21)313, (01)00884838085268(21)703374, (01)00884838099210(21)312, (01)00884838085268(21)703710, (01)00884838085367(21)236, (01)00884838085367(21)237, (01)00884838099227(21)155, (01)00884838085268(21)704026, (01)00884838099265(21)124, (01)00884838110564(21)282, (01)00884838110564(21)286, (01)00884838099258(21)3688, (01)00884838110564(21)182, (01)00884838110564(21)179, (01)00884838110564(21)181, (01)00884838099265(21)475, (01)00884838099272(21)775, (01)00884838085268(21)703592, (01)00884838085282(21)373, (01)00884838099258(21)828, (01)00884838085275(21)101, (01)00884838085275(21)102, (01)00884838099258(21)2348, (01)00884838099241(21)910, (01)00884838099241(21)909, (01)00884838099258(21)802, (01)00884838099227(21)99, (01)00884838085282(21)417, (01)00884838085251(21)604, (01)00884838085251(21)713, (01)00884838085350(21)154, (01)00884838085350(21)157, (01)00884838099241(21)918, (01)00884838085367(21)133, (01)00884838085268(21)215, (01)00884838085251(21)114, (01)00884838099272(21)1073, (01)00884838099265(21)306, (01)00884838085367(21)481, (01)00884838085268(21)703384, (01)00884838085268(21)703392, (01)00884838085282(21)9, (01)00884838099272(21)982, (01)00884838085268(21)545, (01)00884838099234(21)508, (01)00884838099203(21)183, (01)00884838103276(21)167, (01)00884838099272(21)282, (01)00884838085268(21)703753, (01)00884838099234(21)571, (01)00884838099210(21)368, (01)00884838099258(21)1502, (01)00884838085367(21)280, (01)00884838099258(21)2515, (01)00884838099258(21)3631, (01)00884838103276(21)290, (01)00884838099241(21)916, (01)00884838099272(21)785, (01)00884838085268(21)703967, (01)00884838099258(21)1044, (01)00884838099265(21)469, (01)00884838099272(21)617, (01)00884838085251(21)849, (01)00884838085268(21)704112, (01)00884838099272(21)958, (01)00884838085268(21)984, (01)00884838085367(21)307, (01)00884838099210(21)172, (01)00884838099234(21)463, (01)00884838099265(21)470, (01)00884838099272(21)222, (01)00884838099241(21)265, (01)00884838099272(21)221, (01)00884838085251(21)326, (01)00884838099241(21)1243, (01)00884838099203(21)187, (01)00884838099258(21)3320, (01)00884838099227(21)285, (01)00884838099241(21)168, (01)00884838099241(21)142, (01)00884838099258(21)3700, (01)00884838099272(21)604, (01)00884838085268(21)703992, (01)00884838099258(21)87, (01)00884838085275(21)43, (01)00884838085251(21)346, (01)00884838085251(21)347, (01)00884838099272(21)430, (01)00884838099265(21)437, (01)00884838099272(21)1067, (01)00884838099272(21)1027, (01)00884838099258(21)1344, (01)00884838085268(21)703807, (01)00884838099258(21)1339, (01)00884838085251(21)26, (01)00884838099258(21)1619, (01)00884838099258(21)294, (01)00884838099265(21)248, (01)00884838099272(21)415, (01)00884838099258(21)2661, (01)00884838099265(21)338, (01)00884838099241(21)518, (01)00884838099272(21)810, (01)00884838099272(21)1114, (01)00884838099258(21)2644, (01)00884838085268(21)1036, (01)00884838085251(21)621, (01)00884838099258(21)3008, (01)00884838099241(21)476, (01)00884838085268(21)703463, (01)00884838099227(21)247, (01)00884838085268(21)871, (01)00884838099272(21)61, (01)00884838099272(21)1147, (01)00884838110564(21)359, (01)00884838110564(21)357, (01)00884838110564(21)358, (01)00884838099272(21)314, (01)00884838099258(21)184, (01)00884838085268(21)703529, (01)00884838099265(21)140, (01)00884838085367(21)577, (01)00884838085367(21)210, (01)00884838099258(21)2238, (01)00884838099258(21)2283, (01)00884838085251(21)553, (01)00884838099272(21)822, (01)00884838099272(21)561, (01)00884838099210(21)137, (01)00884838099241(21)671, (01)00884838099210(21)302, (01)00884838099265(21)465, (01)00884838085367(21)258, (01)00884838103276(21)165, (01)00884838085251(21)325, (01)00884838085350(21)251, (01)00884838085251(21)648, (01)00884838085268(21)1073, (01)00884838085367(21)328, (01)00884838085251(21)1058, (01)00884838099227(21)261, (01)00884838085251(21)1068, (01)00884838085268(21)704154, (01)00884838099258(21)757, (01)00884838099265(21)457, (01)00884838099258(21)720, (01)00884838099272(21)1102, (01)00884838099258(21)2630, (01)00884838099272(21)762, (01)00884838099227(21)298, (01)00884838099272(21)306, (01)00884838085350(21)237, (01)00884838099272(21)359, (01)00884838099272(21)978, (01)00884838099234(21)71, (01)00884838099265(21)314, (01)00884838085350(21)222, (01)00884838085268(21)778, (01)00884838085251(21)339, (01)00884838085251(21)340, (01)00884838099258(21)1581, (01)00884838085268(21)798, (01)00884838085268(21)548, (01)00884838085268(21)704037, (01)00884838099258(21)1138, (01)00884838099272(21)1060, (01)00884838099272(21)972, (01)00884838099227(21)177, (01)00884838099234(21)493, (01)00884838103276(21)80, (01)00884838099241(21)683, (01)00884838099258(21)3457, (01)00884838099227(21)138, (01)00884838085350(21)95, (01)00884838099265(21)83, (01)00884838103276(21)128, (01)00884838099258(21)2732, (01)00884838099241(21)302, (01)00884838099203(21)42, (01)00884838099258(21)1424, (01)00884838099227(21)337, (01)00884838085268(21)933, (01)00884838099258(21)3280, (01)00884838099210(21)328, (01)00884838085251(21)414, (01)00884838099258(21)3653, (01)00884838099258(21)977, (01)00884838099272(21)83, (01)00884838085367(21)325, (01)00884838099241(21)864, (01)00884838099241(21)863, (01)00884838099241(21)1033, (01)00884838085268(21)316, (01)00884838099258(21)1949, (01)00884838099241(21)137, (01)00884838085268(21)704290, (01)00884838085268(21)703669, (01)00884838085251(21)687, (01)00884838085251(21)688, (01)00884838085268(21)234, (01)00884838085251(21)237, (01)00884838099241(21)25, (01)00884838099258(21)2789, (01)00884838099234(21)1, (01)00884838085251(21)958, (01)00884838085367(21)199, (01)00884838085251(21)751, (01)00884838099210(21)35, (01)00884838099203(21)172, (01)00884838099265(21)433, (01)00884838099241(21)1295, (01)00884838099258(21)877, (01)00884838099272(21)199, (01)00884838099258(21)675, (01)00884838085268(21)1429, (01)00884838099272(21)1013, (01)00884838099272(21)920, (01)00884838099210(21)383, (01)00884838085251(21)881, (01)00884838099241(21)482, (01)00884838099272(21)1136, (01)00884838085251(21)698, (01)00884838099272(21)898, (01)00884838099258(21)3081, (01)00884838099234(21)434, (01)00884838099258(21)2696, (01)00884838099241(21)919, (01)00884838099272(21)835, (01)00884838099272(21)836, (01)00884838085350(21)262, (01)00884838085367(21)480, (01)00884838099227(21)157, (01)00884838099258(21)1774, (01)00884838099241(21)693, (01)00884838099227(21)158, (01)00884838085367(21)53, (01)00884838085268(21)313, (01)00884838099258(21)2276, (01)00884838085275(21)117, (01)00884838099272(21)795, (01)00884838099241(21)553, (01)00884838085251(21)610, (01)00884838085350(21)29, (01)00884838085367(21)72, (01)00884838085251(21)20, (01)00884838085268(21)49, (01)00884838099227(21)228, (01)00884838085282(21)277, (01)00884838099227(21)249, (01)00884838099 272(21)671, (01)00884838099258(21)2111, (01)00884838099258(21)3005, (01)00884838099234(21)647, (01)00884838099265(21)461, (01)00884838085268(21)704209, (01)00884838085275(21)2, (01)00884838085268(21)722, (01)00884838099203(21)164, (01)00884838085251(21)846, (01)00884838099241(21)1191, (01)00884838099258(21)1780, (01)00884838099258(21)1154, (01)00884838085268(21)703619, (01)00884838099210(21)96, (01)00884838099272(21)800, (01)00884838099272(21)799, (01)00884838099272(21)899, (01)00884838085367(21)364, (01)00884838085268(21)704142, (01)00884838099258(21)2250, (01)00884838099272(21)905, (01)00884838085367(21)491, (01)00884838085282(21)102, (01)00884838099227(21)61, (01)00884838099265(21)426, (01)00884838085282(21)467, (01)00884838099272(21)944, (01)00884838099272(21)838, (01)00884838099258(21)2307, (01)00884838099234(21)344, (01)00884838099234(21)317, (01)00884838099272(21)451, (01)00884838085367(21)423, (01)00884838099265(21)420, (01)00884838099241(21)868, (01)00884838099234(21)562, (01)00884838099258(21)915, (01)00884838085268(21)418, (01)00884838085268(21)378, (01)00884838085268(21)1053, (01)00884838085251(21)625, (01)00884838099258(21)672, (01)00884838099258(21)3546, (01)00884838099258(21)548, (01)00884838085367(21)509, (01)00884838085350(21)59, (01)00884838085251(21)424, (01)00884838099210(21)382, (01)00884838099241(21)114, (01)00884838103276(21)158, (01)00884838099258(21)1067, (01)00884838099258(21)1194, (01)00884838099258(21)1129, (01)00884838085251(21)800, (01)00884838085350(21)209, (01)00884838085367(21)246, (01)00884838085268(21)825, (01)00884838085268(21)600, (01)00884838085251(21)838, (01)00884838099258(21)1412, (01)00884838099203(21)190, (01)00884838110571(21)55, (01)00884838085367(21)614, (01)00884838110564(21)109, (01)00884838099227(21)287, (01)00884838085251(21)590, (01)00884838103276(21)278, (01)00884838085367(21)142, (01)00884838099258(21)9, (01)00884838099227(21)55, (01)00884838085275(21)149, (01)00884838099258(21)3629, (01)00884838099258(21)3628, (01)00884838099258(21)86, (01)00884838099234(21)230, (01)00884838085251(21)156, (01)00884838110564(21)136, (01)00884838099258(21)791, (01)00884838099241(21)191, (01)00884838099234(21)293, (01)00884838099272(21)1123, (01)00884838099265(21)421, (01)00884838099234(21)552, (01)00884838099265(21)422, (01)00884838099272(21)335, (01)00884838099241(21)409, (01)00884838099272(21)363, (01)00884838099265(21)168, (01)00884838085268(21)452, (01)00884838085251(21)150, (01)00884838099241(21)106, (01)00884838085282(21)461, (01)00884838085268(21)969, (01)00884838085275(21)99, (01)00884838085251(21)762, (01)00884838085282(21)137, (01)00884838099210(21)151, (01)00884838085251(21)485, (01)00884838099227(21)239, (01)00884838099227(21)306, (01)00884838099210(21)138, (01)00884838099272(21)1039, (01)00884838099265(21)158, (01)00884838103276(21)149, (01)00884838085350(21)260, (01)00884838085268(21)1020, (01)00884838085251(21)726, (01)00884838085251(21)727, (01)00884838099258(21)3547, (01)00884838099258(21)2467, (01)00884838085367(21)376, (01)00884838085268(21)1265, (01)00884838085268(21)763, (01)00884838099258(21)1104, (01)00884838085268(21)703545, (01)00884838085251(21)1039, (01)00884838103276(21)194, (01)00884838085268(21)703358, (01)00884838085251(21)847, (01)00884838085251(21)959, (01)00884838099203(21)151, (01)00884838099241(21)506, (01)00884838110571(21)36, (01)00884838099258(21)2560, (01)00884838099258(21)3344, (01)00884838085268(21)1095, (01)00884838085275(21)44, (01)00884838099241(21)232, (01)00884838099258(21)1002, (01)00884838099227(21)215, (01)00884838099258(21)1704, (01)00884838110571(21)133, (01)00884838085251(21)962, (01)00884838099241(21)446, (01)00884838085268(21)703487, (01)00884838099241(21)403, (01)00884838099258(21)916, (01)00884838110564(21)93, (01)00884838085268(21)704092, (01)00884838099272(21)731, (01)00884838099265(21)290, (01)00884838085268(21)461, (01)00884838085268(21)1239, (01)00884838085268(21)765, (01)00884838099258(21)340, (01)00884838099227(21)336, (01)00884838099241(21)1125, (01)00884838085251(21)865, (01)00884838099258(21)2450, (01)00884838099272(21)399, (01)00884838085350(21)191, (01)00884838099203(21)71, (01)00884838099258(21)771, (01)00884838085367(21)539, (01)00884838085268(21)578, (01)00884838085275(21)3, (01)00884838085367(21)257, (01)00884838099258(21)3479, (01)00884838099272(21)869, (01)00884838099272(21)725, (01)00884838099258(21)906, (01)00884838103276(21)44, (01)00884838099241(21)823, (01)00884838099234(21)458, (01)00884838099234(21)457, (01)00884838099265(21)267, (01)00884838099227(21)213, (01)00884838099272(21)701, (01)00884838099272(21)626, (01)00884838085268(21)703572, (01)00884838099272(21)967, (01)00884838099265(21)454, (01)00884838099272(21)623, (01)00884838099265(21)418, (01)00884838099258(21)2592, (01)00884838099203(21)152, (01)00884838103276(21)100, (01)00884838099258(21)744, (01)00884838085268(21)559, (01)00884838085282(21)84, (01)00884838085268(21)560, (01)00884838099272(21)501, (01)00884838099241(21)670, (01)00884838099258(21)1072, (01)00884838085275(21)28, (01)00884838085268(21)982, (01)00884838099227(21)76, (01)00884838099203(21)184, (01)00884838085350(21)274, (01)00884838085367(21)140, (01)00884838085251(21)371, (01)00884838099234(21)369, (01)00884838099241(21)803, (01)00884838099241(21)1284, (01)00884838099272(21)1018, (01)00884838085268(21)703462, (01)00884838099265(21)478, (01)00884838085367(21)369, (01)00884838099258(21)3183, (01)00884838099227(21)216, (01)00884838099258(21)1298, (01)00884838099241(21)579, (01)00884838099227(21)73, (01)00884838085367(21)508, (01)00884838085367(21)97, (01)00884838085268(21)464, (01)00884838085268(21)703583, (01)00884838099258(21)3641, (01)00884838099234(21)561, (01)00884838099258(21)2273, (01)00884838099203(21)147, (01)00884838085251(21)769, (01)00884838099258(21)1368, (01)00884838099258(21)2850, (01)00884838099265(21)438, (01)00884838085268(21)1106, (01)00884838085282(21)510, (01)00884838099272(21)533, (01)00884838099227(21)318, (01)00884838099272(21)677, (01)00884838085367(21)87, (01)00884838085251(21)572, (01)00884838085251(21)570, (01)00884838103276(21)164, (01)00884838099241(21)703, (01)00884838099241(21)705, (01)00884838099265(21)257, (01)00884838099258(21)1707, (01)00884838099272(21)449, (01)00884838085251(21)1031, (01)00884838085268(21)1044, (01)00884838099258(21)3061, (01)00884838099265(21)223, (01)00884838099258(21)2165, (01)00884838099272(21)992, (01)00884838099258(21)3227, (01)00884838099210(21)277, (01)00884838099265(21)366, (01)00884838099210(21)384, (01)00884838099203(21)32, (01)00884838099272(21)119, (01)00884838110571(21)110, (01)00884838110571(21)112, (01)00884838110571(21)111, (01)00884838085268(21)703855, (01)00884838099227(21)353, (01)00884838085251(21)792, (01)00884838099241(21)311, (01)00884838085367(21)303, (01)00884838085268(21)704127, (01)00884838099227(21)65, (01)00884838099265(21)173, (01)00884838099272(21)367, (01)00884838099210(21)363, (01)00884838099272(21)927, (01)00884838099272(21)480, (01)00884838110564(21)153, (01)00884838085268(21)1205, (01)00884838085268(21)1206, (01)00884838085275(21)144, (01)00884838103276(21)240, (01)00884838099272(21)732, (01)00884838099258(21)1522, (01)00884838099227(21)167, (01)00884838085367(21)48, (01)00884838085268(21)1214, (01)00884838099258(21)3660, (01)00884838099258(21)2394, (01)00884838099272(21)974, (01)00884838085367(21)44, (01)00884838099234(21)517, (01)00884838085367(21)301, (01)00884838099210(21)224, (01)00884838099234(21)203, (01)00884838099203(21)105, (01)00884838085367(21)550, (01)00884838085251(21)196, (01)00884838099241(21)776, (01)00884838099272(21)344, (01)00884838099258(21)890, (01)00884838099265(21)43, (01)00884838085282(21)78, (01)00884838099272(21)824, (01)00884838085350(21)218, (01)00884838099272(21)984, (01)00884838099272(21)487, (01)00884838099227(21)41, (01)00884838099265(21)8, (01)00884838085275(21)62, (01)00884838110564(21)247, (01)00884838085282(21)49, (01)00884838085282(21)345, (01)00884838099203(21)182, (01)00884838099227(21)85, (01)00884838099210(21)386, (01)00884838099272(21)919, (01)00884838099227(21)191, (01)00884838099241(21)1104, (01)00884838099258(21)2320, (01)00884838099272(21)605, (01)00884838099241(21)719, (01)00884838099272(21)686, (01)00884838099272(21)523, (01)00884838099258(21)887, (01)00884838099272(21)230, (01)00884838085268(21)703348, (01)00884838085367(21)460, (01)00884838085350(21)231, (01)00884838085282(21)266, (01)00884838085282(21)271, (01)00884838085251(21)683, (01)00884838085282(21)212, (01)00884838085282(21)193, (01)00884838085282(21)213, (01)00884838085350(21)169, (01)00884838085282(21)214, (01)00884838085282(21)180, (01)00884838085282(21)185, (01)00884838099272(21)327, (01)00884838085251(21)541, (01)00884838099234(21)554, (01)00884838099265(21)146, (01)00884838099258(21)687, (01)00884838099272(21)1138, (01)00884838099241(21)480, (01)00884838099272(21)424, (01)00884838085367(21)551, (01)00884838099241(21)113, (01)00884838099227(21)20, (01)00884838099203(21)58, (01)00884838099258(21)2712, (01)00884838099258(21)3466, (01)00884838110564(21)27, (01)00884838110564(21)28, (01)00884838099272(21)684, (01)00884838099203(21)103, (01)00884838085251(21)213, (01)00884838085275(21)119, (01)00884838099241(21)609, (01)00884838099227(21)226, (01)00884838099258(21)2709, (01)00884838085268(21)252, (01)00884838099241(21)822, (01)00884838099234(21)260, (01)00884838099258(21)361, (01)00884838103276(21)246, (01)00884838110564(21)299, (01)00884838099258(21)1474, (01)00884838099227(21)102, (01)00884838099227(21)87, (01)00884838099227(21)40, (01)00884838085275(21)61, (01)00884838099241(21)1095, (01)00884838099227(21)163, (01)00884838099227(21)83, (01)00884838099272(21)477, (01)00884838099234(21)640, (01)00884838099210(21)351, (01)00884838099227(21)316, (01)00884838099272(21)1049, (01)00884838099210(21)373, (01)00884838099210(21)372, (01)00884838099227(21)309, (01)00884838099227(21)310, (01)00884838099227(21)251, (01)00884838099210(21)331, (01)00884838099227(21)268, (01)00884838099227(21)290, (01)00884838099210(21)361, (01)00884838099210(21)209, (01)00884838099210(21)210, (01)00884838099210(21)145, (01)00884838099234(21)108, (01)00884838099234(21)107, (01)00884838099234(21)100, (01)00884838099210(21)119, (01)00884838099234(21)101, (01)00884838099210(21)118, (01)00884838099234(21)99, (01)00884838099272(21)298, (01)00884838099210(21)88, (01)00884838099210(21)102, (01)00884838099210(21)89, (01)00884838099258(21)596, (01)00884838085268(21)703872, (01)00884838099258(21)532, (01)00884838085367(21)478, (01)00884838099210(21)108, (01)00884838085251(21)320, (01)00884838099265(21)414, (01)00884838085282(21)431, (01)00884838099210(21)132, (01)00884838099265(21)276, (01)00884838085367(21)522, (01)00884838085282(21)250, (01)00884838085251(21)1025, (01)00884838099258(21)1306, (01)00884838085282(21)186, (01)00884838085251(21)943, (01)00884838099272(21)809, (01)00884838085251(21)651, (01)00884838099241(21)1137, (01)00884838099258(21)1647, (01)00884838099241(21)1042, (01)00884838099241(21)1043, (01)00884838099258(21)1153, (01)00884838099258(21)57, (01)00884838085282(21)246, (01)00884838099272(21)10, (01)00884838085251(21)722, (01)00884838085268(21)430, (01)00884838085251(21)36, (01)00884838085251(21)242, (01)00884838099241(21)1078, (01)00884838099258(21)2730, (01)00884838099272(21)227, (01)00884838085268(21)997, (01)00884838085251(21)431, (01)00884838099227(21)272, (01)00884838099258(21)2629, (01)00884838099258(21)971, (01)00884838099258(21)2528, (01)00884838085268(21)271, (01)00884838085268(21)312, (01)00884838099272(21)993, (01)00884838099241(21)858, (01)00884838099265(21)301, (01)00884838099272(21)520, (01)00884838099272(21)522, (01)00884838099241(21)519, (01)00884838099272(21)521, (01)00884838099241(21)520, (01)00884838099241(21)524, (01)00884838099241(21)525, (01)00884838099241(21)526, (01)00884838099241(21)521, (01)00884838099241(21)522, (01)00884838099241(21)523, (01)00884838085367(21)568, (01)00884838099241(21)375, (01)00884838099241(21)362, (01)00884838099210(21)48, (01)00884838085251(21)672, (01)00884838085268(21)1338, (01)00884838099258(21)1991, (01)00884838099241(21)747, (01)00884838085268(21)855, (01)00884838099272(21)874, (01)00884838099258(21)2853, (01)00884838085268(21)1169, (01)00884838085367(21)448, (01)00884838099241(21)272, (01)00884838099241(21)81, (01)00884838085251(21)177, (01)00884838099241(21)661, (01)00884838099272(21)259, (01)00884838099258(21)3311, (01)00884838099258(21)2923, (01)00884838099258(21)2864, (01)00884838099258(21)2865, (01)00884838099258(21)2866, (01)00884838099241(21)862, (01)00884838099234(21)318, (01)00884838099241(21)821, (01)00884838099241(21)723, (01)00884838099258(21)1280, (01)00884838099258(21)1281, (01)00884838085251(21)912, (01)00884838085268(21)1030, (01)00884838085251(21)578, (01)00884838085268(21)1120, (01)00884838099272(21)140, (01)00884838099227(21)5, (01)00884838099227(21)6, (01)00884838099227(21)7, (01)00884838099203(21)149, (01)00884838099227(21)218, (01)00884838085350(21)275, (01)00884838085268(21)339, (01)00884838099241(21)140, (01)00884838085367(21)242, (01)00884838099234(21)21, (01)00884838099227(21)36, (01)00884838099234(21)22, (01)00884838099241(21)761, (01)00884838085268(21)1121, (01)00884838099203(21)127, (01)00884838099227(21)75, (01)00884838099241(21)63, (01)00884838085282(21)282, (01)00884838099241(21)109, (01)00884838085268(21)934, (01)00884838099227(21)195, (01)00884838085268(21)1168, (01)00884838099241(21)1003, (01)00884838099258(21)1981, (01)00884838099241(21)1251, (01)00884838099203(21)128, (01)00884838099258(21)2499, (01)00884838099227(21)43, (01)00884838085268(21)1150, (01)00884838085251(21)355, (01)00884838085268(21)699, (01)00884838085251(21)504, (01)00884838099258(21)881, (01)00884838110564(21)140, (01)00884838099241(21)1041, (01)00884838110571(21)45, (01)00884838099258(21)2813, (01)00884838099272(21)628, (01)00884838099265(21)109, (01)00884838085268(21)704199, (01)00884838085251(21)699, (01)00884838099241(21)1271, (01)00884838099272(21)530, (01)00884838099258(21)2566, (01)00884838099241(21)998, (01)00884838099272(21)506, (01)00884838085350(21)127, (01)00884838085251(21)270, (01)00884838099234(21)154, (01)00884838099258(21)1504, (01)00884838099241(21)317, (01)00884838099241(21)316, (01)00884838085251(21)306, (01)00884838099210(21)53, (01)00884838099258(21)3752, (01)00884838099272(21)909, (01)00884838099265(21)5, (01)00884838099210(21)4, (01)00884838085275(21)22, (01)00884838099258(21)62, (01)00884838099265(21)200, (01)00884838099265(21)189, (01)00884838099241(21)78, (01)00884838099265(21)387, (01)00884838099258(21)3004, (01)00884838099241(21)978, (01)00884838085251(21)366, (01)00884838085251(21)190, (01)00884838099258(21)37, (01)00884838099227(21)345, (01)00884838099258(21)3598, (01)00884838099234(21)602, (01)00884838099234(21)574, (01)00884838099234(21)531, (01)00884838099210(21)305, (01)00884838099234(21)78, (01)00884838099234(21)75, (01)00884838099258(21)600, (01)00884838085268(21)703959, (01)00884838099241(21)202, (01)00884838099272(21)142, (01)00884838085282(21)374, (01)00884838085282(21)354, (01)00884838085268(21)703489, (01)00884838099241(21)1276, (01)00884838099265(21)452, (01)00884838099265(21)451, (01)00884838099234(21)158, (01)00884838085268(21)913, (01)00884838085268(21)563, (01)00884838085268(21)304, (01)00884838099258(21)2828, (01)00884838099258(21)2341, (01)00884838099234(21)461, (01)00884838099272(21)733, (01)00884838085268(21)1191, (01)00884838085268(21)567, (01)00884838099258(21)3295, (01)00884838099234(21)370, (01)00884838099258(21)3506, (01)00884838099258(21)3009, (01)00884838099258(21)2921, (01)00884838099258(21)2947, (01)00884838099210(21)80, (01)00884838085268(21)566, (01)00884838099241(21)824, (01)00884838085268(21)916, (01)00884838085268(21)388, (01)00884838099258(21)1570, (01)00884838085268(21)707, (01)00884838099258(21)2260, (01)00884838099258(21)2299, (01)00884838085268(21)1195, (01)00884838085268(21)1193, (01)00884838099258(21)3726, (01)00884838085268(21)704282, (01)00884838085268(21)555, (01)00884838085268(21)554, (01)00884838085268(21)301, (01)00884838099234(21)613, (01)00884838099258(21)3288, (01)00884838099258(21)3665, (01)00884838099258(21)65, (01)00884838085251(21)398, (01)00884838085268(21)704284, (01)00884838085268(21)1083, (01)00884838099258(21)2424, (01)00884838099241(21)915, (01)00884838085251(21)562, (01)00884838085251(21)470, (01)00884838085275(21)29, (01)00884838085282(21)274, (01)00884838085275(21)153, (01)00884838099258(21)775, (01)00884838099258(21)30, (01)00884838085268(21)703966, (01)00884838085268(21)303, (01)00884838099258(21)2818, (01)00884838099258(21)2889, (01)00884838099258(21)3736, (01)00884838085367(21)238, (01)00884838085282(21)127, (01)00884838099258(21)2758, (01)00884838099258(21)2269, (01)00884838099234(21)393, (01)00884838099241(21)711, (01)00884838099234(21)564, (01)00884838099210(21)83, (01)00884838099258(21)2105, (01)00884838099258(21)3238, (01)00884838099241(21)108, (01)00884838099258(21)3411, (01)00884838099258(21)3409, (01)00884838099234(21)482, (01)00884838099234(21)194, (01)00884838099234(21)660, (01)00884838099258(21)3018, (01)00884838085268(21)704101, (01)00884838085268(21)305, (01)00884838099258(21)2263, (01)00884838085268(21)1194, (01)00884838085251(21)489, (01)00884838085251(21)815, (01)00884838099258(21)2101, (01)00884838085282(21)451, (01)00884838085268(21)785, (01)00884838085367(21)68, (01)00884838099234(21)673, (01)00884838099258(21)3500, (01)00884838099258(21)2705, (01)00884838099258(21)3666, (01)00884838099258(21)3681, (01)00884838085268(21)703961, (01)00884838099258(21)1930, (01)00884838099234(21)60, (01)00884838099258(21)3021, (01)00884838099258(21)2129, (01)00884838099210(21)47, (01)00884838099241(21)185, (01)00884838099258(21)3383, (01)00884838099272(21)263, (01)00884838085268(21)1337, (01)00884838085282(21)106, (01)00884838085268(21)993, (01)00884838085282(21)490, (01)00884838085268(21)688, (01)00884838099258(21)1697, (01)00884838099258(21)3113, (01)00884838085268(21)347, (01)00884838099258(21)3467, (01)00884838099258(21)1997, (01)00884838099258(21)2009, (01)00884838099241(21)754, (01)00884838099258(21)1710, (01)00884838099258(21)127, (01)00884838085251(21)147, (01)00884838099258(21)2932, (01)00884838099258(21)2689, (01)00884838099258(21)1392, (01)00884838099272(21)490, (01)00884838099272(21)494, (01)00884838099258(21)1422, (01)00884838099258(21)1399, (01)00884838099258(21)620, (01)00884838099258(21)1429, (01)00884838085268(21)772, (01)00884838099241(21)928, (01)00884838099258(21)2050, (01)00884838099258(21)2049, (01)00884838099258(21)1583, (01)00884838099258(21)2745, (01)00884838085268(21)74, (01)00884838085268(21)632, (01)00884838085251(21)160, (01)00884838099258(21)1, (01)00884838085268(21)383, (01)00884838099258(21)2631, (01)00884838085268(21)735, (01)00884838099258(21)2458, (01)00884838099258(21)2457, (01)00884838099258(21)2193, (01)00884838085251(21)743, (01)00884838099258(21)2738, (01)00884838085282(21)329, (01)00884838085282(21)276, (01)00884838099272(21)633, (01)00884838099272(21)634, (01)00884838099258(21)1586, (01)00884838099210(21)270, (01)00884838099241(21)847, (01)00884838099241(21)846, (01)00884838099258(21)2527, (01)00884838099258(21)2275, (01)00884838099258(21)1837, (01)00884838099272(21)396, (01)00884838099258(21)567, (01)00884838085251(21)152, (01)00884838099258(21)1835, (01)00884838085251(21)759, (01)00884838085251(21)729, (01)00884838085251(21)630, (01)00884838099258(21)2021, (01)00884838099258(21)1865, (01)00884838099258(21)1263, (01)00884838099258(21)1342, (01)00884838085268(21)39, (01)00884838099210(21)21, (01)00884838099234(21)526, (01)00884838085282(21)58, (01)00884838099258(21)686, (01)00884838099258(21)3225, (01)00884838099227(21)116, (01)00884838099241(21)893, (01)00884838099258(21)1678, (01)00884838099258(21)1734, (01)00884838099258(21)1758, (01)00884838085282(21)81, (01)00884838099258(21)3403, (01)00884838099203(21)86, (01)00884838099241(21)1263, (01)00884838099241(21)1262, (01)00884838085251(21)151, (01)00884838099241(21)769, (01)00884838099258(21)3669, (01)00884838099272(21)547, (01)00884838085268(21)1074, (01)00884838085268(21)230, (01)00884838099258(21)3532, (01)00884838099258(21)1597, (01)00884838099258(21)456, (01)00884838099210(21)64, (01)00884838099258(21)1340, (01)00884838099258(21)2013, (01)00884838099258(21)1602, (01)00884838085282(21)200, (01)00884838085268(21)1182, (01)00884838099258(21)2398, (01)00884838085268(21)1341, (01)00884838099203(21)122, (01)00884838099234(21)464, (01)00884838099258(21)3729, (01)00884838099258(21)3753, (01)00884838099258(21)3754, (01)00884838099258(21)3724, (01)00884838099258(21)3501, (01)00884838099258(21)1777, (01)00884838099258(21)2706, (01)00884838085282(21)179, (01)00884838099210(21)22, (01)00884838099258(21)3748, (01)00884838099241(21)755, (01)00884838099272(21)631, (01)00884838099258(21)1584, (01)00884838099258(21)1666, (01)00884838099227(21)18, (01)00884838099234(21)4, (01)00884838085367(21)160, (01)00884838085251(21)334, (01)00884838099258(21)3075, (01)00884838085268(21)293, (01)00884838099258(21)1980, (01)00884838099258(21)1343, (01)00884838099258(21)386, (01)00884838085268(21)936, (01)00884838099234(21)462, (01)00884838085268(21)704258, (01)00884838085275(21)40, (01)00884838085268(21)295, (01)00884838099258(21)2056, (01)00884838099258(21)1327, (01)00884838085268(21)675, (01)00884838085268(21)949, (01)00884838099258(21)3228, (01)00884838099258(21)1711, (01)00884838085268(21)411, (01)00884838085367(21)83, (01)00884838099258(21)2916, (01)00884838099258(21)2124, (01)00884838099258(21)1766, (01)00884838099258(21)632, (01)00884838099258(21)1495, (01)00884838085268(21)701, (01)00884838085268(21)286, (01)00884838085268(21)287, (01)00884838099258(21)2872, (01)00884838099258(21)2871, (01)00884838085251(21)449, (01)00884838085251(21)757, (01)00884838085251(21)429, (01)00884838099210(21)98, (01)00884838099203(21)60, (01)00884838099272(21)514, (01)00884838099272(21)642, (01)00884838099258(21)1806, (01)00884838099258(21)58, (01)00884838099258(21)55, (01)00884838099258(21)2046, (01)00884838085268(21)232, (01)00884838085268(21)760, (01)00884838085268(21)421, (01)00884838099234(21)138, (01)00884838085282(21)249, (01)00884838099258(21)1652, (01)00884838099210(21)29, (01)00884838099210(21)248, (01)00884838085268(21)726, (01)00884838099272(21)505, (01)00884838099258(21)1447, (01)00884838099258(21)256, (01)00884838085268(21)1109, (01)00884838085268(21)1144, (01)00884838099258(21)1535, (01)00884838099258(21)1544, (01)00884838099258(21)1537, (01)00884838099258(21)1543, (01)00884838099258(21)1286, (01)00884838085268(21)434, (01)00884838085268(21)591, (01)00884838099258(21)3570, (01)00884838099258(21)3470, (01)00884838085251(21)506, (01)00884838099241(21)906, (01)00884838085251(21)703, (01)00884838099258(21)3380, (01)00884838099241(21)314, (01)00884838099241(21)313, (01)00884838099258(21)732, (01)00884838099241(21)238, (01)00884838099241(21)239, (01)00884838085282(21)156, (01)00884838099258(21)2837, (01)00884838099241(21)1034, (01)00884838085251(21)149, (01)00884838085268(21)61, (01)00884838099272(21)321, (01)00884838085268(21)961, (01)00884838099258(21)961, (01)00884838085251(21)818, (01)00884838099241(21)430, (01)00884838085268(21)1072, (01)00884838085251(21)647, (01)00884838085367(21)544, (01)00884838099234(21)668, (01)00884838099210(21)389, (01)00884838099234(21)580, (01)00884838099210(21)369, (01)00884838099234(21)496, (01)00884838099210(21)222, (01)00884838099241(21)535, (01)00884838099210(21)134, (01)00884838085282(21)380, (01)00884838085282(21)288, (01)00884838085282(21)289, (01)00884838085282(21)287, (01)00884838085282(21)231, (01)00884838085282(21)233, (01)00884838085282(21)230, (01)00884838085282(21)197, (01)00884838085282(21)176, (01)00884838085282(21)150, (01)00884838085282(21)149, (01)00884838085282(21)44, (01)00884838085268(21)44, (01)00884838085268(21)1185, (01)00884838099241(21)299, (01)00884838099227(21)153, (01)00884838085282(21)184, (01)00884838099241(21)1247, (01)00884838099258(21)367, (01)00884838085268(21)703477, (01)00884838099258(21)707, (01)00884838099258(21)708, (01)00884838099258(21)706, (01)00884838099241(21)363, (01)00884838085282(21)6, (01)00884838110564(21)60, (01)00884838085367(21)230, (01)00884838085268(21)257, (01)00884838099258(21)2635, (01)00884838099210(21)266, (01)00884838099272(21)754, (01)00884838103276(21)66, (01)00884838099241(21)1329, (01)00884838103276(21)135, (01)00884838099272(21)422, (01)00884838099258(21)1785, (01)00884838099258(21)575, (01)00884838099234(21)319, (01)00884838099234(21)312, (01)00884838099234(21)223, (01)00884838099258(21)1862, (01)00884838099272(21)17, (01)00884838085251(21)419, (01)00884838085268(21)661, (01)00884838085275(21)70, (01)00884838085282(21)448, (01)00884838099272(21)1057, (01)00884838099272(21)420, (01)00884838085251(21)328, (01)00884838085275(21)30, (01)00884838099272(21)99, (01)00884838099272(21)247, (01)00884838085268(21)704146, (01)00884838085268(21)703982, (01)00884838099272(21)36, (01)00884838099272(21)432, (01)00884838085251(21)584, (01)00884838085251(21)223, (01)00884838099258(21)2235, (01)00884838099258(21)2441, (01)00884838085251(21)192, (01)00884838099272(21)539, (01)00884838099234(21)98, (01)00884838099258(21)206, (01)00884838085282(21)10, (01)00884838085282(21)82, (01)00884838099265(21)364, (01)00884838085268(21)1440, (01)00884838085268(21)1441, (01)00884838085268(21)1442, (01)00884838085367(21)435, (01)00884838085367(21)434, (01)00884838085350(21)210, (01)00884838085350(21)211, (01)00884838085350(21)212, (01)00884838085350(21)213, (01)00884838085251(21)779, (01)00884838099241(21)40, (01)00884838099258(21)1549, (01)00884838099258(21)1275, (01)00884838099241(21)574, (01)00884838085367(21)360, (01)00884838099241(21)486, (01)00884838085268(21)389, (01)00884838099241(21)354, (01)00884838099241(21)710, (01)00884838085251(21)468, (01)00884838099241(21)279, (01)00884838099241(21)280, (01)00884838099241(21)281, (01)00884838099258(21)783, (01)00884838099203(21)155, (01)00884838099272(21)88, (01)00884838099258(21)604, (01)00884838085268(21)703438, (01)00884838099258(21)2037, (01)00884838099258(21)1728, (01)00884838099210(21)378, (01)00884838099241(21)845, (01)00884838099272(21)245, (01)00884838099203(21)66, (01)00884838099203(21)33, (01)00884838085268(21)703783, (01)00884838085268(21)704211, (01)00884838085268(21)703812, (01)00884838099234(21)623, (01)00884838099241(21)960, (01)00884838085251(21)222, (01)00884838099258(21)2493, (01)00884838099258(21)544, (01)00884838099258(21)2707, (01)00884838085282(21)323, (01)00884838085268(21)1122, (01)00884838099241(21)306, (01)00884838085251(21)200, (01)00884838099258(21)725, (01)00884838099241(21)992, (01)00884838085251(21)262, (01)00884838085251(21)134, (01)00884838085268(21)725, (01)00884838099203(21)55, (01)00884838099203(21)70, (01)00884838099227(21)351, (01)00884838099227(21)352, (01)00884838085268(21)703423, (01)00884838085251(21)855, (01)00884838099210(21)245, (01)00884838085251(21)284, (01)00884838099258(21)2943, (01)00884838085251(21)918, (01)00884838085251(21)919, (01)00884838099241(21)989, (01)00884838085367(21)93, (01)00884838085251(21)277, (01)00884838099272(21)911, (01)00884838099227(21)15, (01)00884838099227(21)14, (01)00884838099210(21)18, (01)00884838099241(21)618, (01)00884838099227(21)120, (01)00884838085251(21)976, (01)00884838099258(21)1762, (01)00884838099258(21)24, (01)00884838085251(21)13, (01)00884838085251(21)14, (01)00884838099272(21)133, (01)00884838099241(21)225, (01)00884838099241(21)226, (01)00884838099258(21)602, (01)00884838099265(21)65, (01)00884838099265(21)66, (01)00884838099272(21)132, (01)00884838085268(21)341, (01)00884838085268(21)367, (01)00884838099258(21)406, (01)00884838085268(21)37, (01)00884838099258(21)932, (01)00884838099258(21)2322, (01)00884838099258(21)1427, (01)00884838099241(21)173, (01)00884838099241(21)212, (01)00884838085251(21)502, (01)00884838099258(21)1562, (01)00884838099258(21)1814, (01)00884838099234(21)475, (01)00884838085268(21)1317, (01)00884838085268(21)1192, (01)00884838099258(21)2134, (01)00884838099258(21)2259, (01)00884838085268(21)1264, (01)00884838099258(21)2261, (01)00884838099258(21)2714, (01)00884838085268(21)575, (01)00884838099272(21)667, (01)00884838099258(21)3135, (01)00884838099258(21)16, (01)00884838085268(21)1202, (01)00884838085268(21)36, (01)00884838099258(21)13, (01)00884838099241(21)8, (01)00884838099272(21)5, (01)00884838099241(21)9, (01)00884838099272(21)1103, (01)00884838099272(21)188, (01)00884838099241(21)157, (01)00884838099272(21)586, (01)00884838099258(21)1779, (01)00884838099258(21)1406, (01)00884838099272(21)14, (01)00884838099258(21)1730, (01)00884838085268(21)309, (01)00884838085268(21)67, (01)00884838085367(21)36, (01)00884838099258(21)816, (01)00884838099258(21)2704, (01)00884838099258(21)2819, (01)00884838099234(21)125, (01)00884838085268(21)703824, (01)00884838099258(21)3676, (01)00884838099272(21)622, (01)00884838099241(21)165, (01)00884838099258(21)93, (01)00884838099241(21)344, (01)00884838085282(21)35, (01)00884838099241(21)1147, (01)00884838099258(21)2404, (01)00884838099241(21)431, (01)00884838099227(21)262, (01)00884838099265(21)398, (01)00884838099241(21)590, (01)00884838085251(21)817, (01)00884838099258(21)2217, (01)00884838099241(21)1083, (01)00884838099272(21)228, (01)00884838085350(21)206, (01)00884838099272(21)4, (01)00884838085251(21)795, (01)00884838085350(21)30, (01)00884838099258(21)2310, (01)00884838099227(21)204, (01)00884838099234(21)218, (01)00884838099241(21)492, (01)00884838099258(21)171, (01)00884838099258(21)413, (01)00884838085268(21)703313, (01)00884838085268(21)1278, (01)00884838085251(21)577, (01)00884838085251(21)574, (01)00884838085251(21)498, (01)00884838085268(21)703544, (01)00884838110571(21)116, (01)00884838099234(21)513, (01)00884838099234(21)442, (01)00884838099227(21)258, (01)00884838099258(21)1125, (01)00884838085275(21)25, (01)00884838085251(21)254, (01)00884838085251(21)255, (01)00884838099258(21)1179, (01)00884838099241(21)64, (01)00884838085251(21)448, (01)00884838099258(21)2159, (01)00884838099258(21)2347, (01)00884838085350(21)23, (01)00884838085251(21)27, (01)00884838099258(21)827, (01)00884838099258(21)573, (01)00884838099210(21)24, (01)00884838099234(21)435, (01)00884838099203(21)123, (01)00884838099227(21)326, (01)00884838099203(21)118, (01)00884838099203(21)61, (01)00884838099227(21)348, (01)00884838099227(21)17, (01)00884838085275(21)19, (01)00884838085268(21)703393, (01)00884838085268(21)703428, (01)00884838085251(21)631, (01)00884838085251(21)632, (01)00884838085282(21)69, (01)00884838099258(21)542, (01)00884838099258(21)48, (01)00884838099234(21)95, (01)00884838099234(21)9, (01)00884838099234(21)557, (01)00884838085282(21)278, (01)00884838085367(21)372, (01)00884838099210(21)335, (01)00884838085282(21)511, (01)00884838099258(21)2190, (01)00884838085268(21)703713, (01)00884838099210(21)177, (01)00884838085268(21)703815, (01)00884838099258(21)3732, (01)00884838085268(21)1372, (01)00884838099258(21)2415, (01)00884838085268(21)703330, (01)00884838085268(21)1347, (01)00884838099258(21)1539, (01)00884838099258(21)463, (01)00884838099258(21)2277, (01)00884838099258(21)1080, (01)00884838099258(21)1348, (01)00884838099258(21)1054, (01)00884838099258(21)658, (01)00884838099258(21)1050, (01)00884838099258(21)1053, (01)00884838099258(21)1052, (01)00884838099258(21)1051, (01)00884838099258(21)383, (01)00884838099258(21)3124, (01)00884838085268(21)958, (01)00884838099258(21)1656, (01)00884838099258(21)1588, (01)00884838099258(21)1566, (01)00884838099258(21)779, (01)00884838099272(21)45, (01)00884838099272(21)316, (01)00884838085268(21)703472, (01)00884838099210(21)11, (01)00884838085251(21)599, (01)00884838099241(21)28, (01)00884838099272(21)164, (01)00884838085268(21)951, (01)00884838085268(21)1009, (01)00884838099258(21)891, (01)00884838099258(21)2749, (01)00884838099258(21)1443, (01)00884838099258(21)1359, (01)00884838085367(21)179, (01)00884838085251(21)370, (01)00884838085367(21)306, (01)00884838085268(21)482, (01)00884838099258(21)1425, (01)00884838085367(21)182, (01)00884838085268(21)1254, (01)00884838085251(21)216, (01)00884838085268(21)300, (01)00884838085268(21)620, (01)00884838085251(21)451, (01)00884838085282(21)39, (01)00884838099272(21)485, (01)00884838085268(21)986, (01)00884838085367(21)583, (01)00884838085268(21)703740, (01)00884838085268(21)493, (01)00884838099258(21)1805, (01)00884838099241(21)274, (01)00884838099210(21)72, (01)00884838085367(21)327, (01)00884838085251(21)428, (01)00884838099258(21)2618, (01)00884838099258(21)2014, (01)00884838099258(21)1593, (01)00 884838099258(21)2305, (01)00884838099241(21)753, (01)00884838099258(21)2002, (01)00884838099258(21)2087, (01)00884838085275(21)120, (01)00884838099258(21)2175, (01)00884838099272(21)448, (01)00884838085268(21)484, (01)00884838099258(21)305, (01)00884838099258(21)2817, (01)00884838099258(21)506, (01)00884838099258(21)3507, (01)00884838099258(21)2751, (01)00884838085268(21)988, (01)00884838099258(21)1854, (01)00884838099258(21)1645, (01)00884838099258(21)1345, (01)00884838099258(21)2160, (01)00884838099258(21)311, (01)00884838099258(21)1387, (01)00884838099258(21)1433, (01)00884838099272(21)507, (01)00884838085251(21)467, (01)00884838099258(21)2575, (01)00884838085268(21)1082, (01)00884838099258(21)2055, (01)00884838099258(21)98, (01)00884838099258(21)1677, (01)00884838099258(21)1691, (01)00884838085251(21)450, (01)00884838099272(21)97, (01)00884838099272(21)351, (01)00884838099258(21)2788, (01)00884838085268(21)704171, (01)00884838099258(21)1393, (01)00884838099272(21)497, (01)00884838099258(21)1402, (01)00884838099258(21)1391, (01)00884838099258(21)1196, (01)00884838099234(21)577, (01)00884838099234(21)548, (01)00884838099258(21)2605, (01)00884838099234(21)174, (01)00884838099258(21)1092, (01)00884838099203(21)44, (01)00884838085268(21)703470, (01)00884838085275(21)121, (01)00884838085282(21)66, (01)00884838085268(21)542, (01)00884838085282(21)45, (01)00884838085282(21)40, (01)00884838099258(21)1351, (01)00884838099258(21)302, (01)00884838085268(21)589, (01)00884838099258(21)2287, (01)00884838085268(21)552, (01)00884838085275(21)18, (01)00884838099258(21)3287, (01)00884838085268(21)704277, (01)00884838099258(21)77, (01)00884838099258(21)605, (01)00884838085268(21)700, (01)00884838085268(21)486, (01)00884838085268(21)235, (01)00884838099258(21)3167, (01)00884838099258(21)1700, (01)00884838099258(21)2987, (01)00884838099258(21)2534, (01)00884838099258(21)145, (01)00884838085282(21)125, (01)00884838099258(21)3212, (01)00884838085268(21)348, (01)00884838085282(21)484, (01)00884838099258(21)1799, (01)00884838085268(21)703867, (01)00884838099258(21)1815, (01)00884838085251(21)102, (01)00884838099258(21)3416, (01)00884838085251(21)662, (01)00884838099258(21)35, (01)00884838085268(21)565, (01)00884838085251(21)189, (01)00884838085268(21)322, (01)00884838099265(21)196, (01)00884838099227(21)141, (01)00884838085251(21)633, (01)00884838085268(21)1123, (01)00884838099258(21)2905, (01)00884838099241(21)820, (01)00884838099241(21)817, (01)00884838099258(21)2069, (01)00884838099241(21)425, (01)00884838099241(21)424, (01)00884838099258(21)81, (01)00884838085367(21)244, (01)00884838085268(21)819, (01)00884838085282(21)55, (01)00884838085282(21)36, (01)00884838085282(21)94, (01)00884838099210(21)244, (01)00884838099258(21)519, (01)00884838099258(21)368, (01)00884838085268(21)703749, (01)00884838099258(21)2836, (01)00884838099258(21)563, (01)00884838085268(21)782, (01)00884838099258(21)1398, (01)00884838085367(21)382, (01)00884838085268(21)25, (01)00884838099258(21)2016, (01)00884838099258(21)1699, (01)00884838099203(21)185, (01)00884838085282(21)497, (01)00884838099210(21)9, (01)00884838085268(21)1154, (01)00884838085251(21)512, (01)00884838085350(21)265, (01)00884838085282(21)104, (01)00884838085268(21)703965, (01)00884838085282(21)155, (01)00884838099210(21)66, (01)00884838099265(21)372, (01)00884838099203(21)74, (01)00884838099258(21)2715, (01)00884838085367(21)192, (01)00884838085268(21)704239, (01)00884838085268(21)703993, (01)00884838085251(21)175, (01)00884838110564(21)341, (01)00884838099241(21)1237, (01)00884838099258(21)2950, (01)00884838103276(21)175, (01)00884838099272(21)685, (01)00884838099258(21)2608, (01)00884838085367(21)428, (01)00884838085251(21)796, (01)00884838099258(21)727, (01)00884838085251(21)459, (01)00884838085251(21)460, (01)00884838099241(21)819, (01)00884838099241(21)818, (01)00884838103276(21)99, (01)00884838085268(21)703542, (01)00884838085282(21)422, (01)00884838099203(21)95, (01)00884838099234(21)486, (01)00884838099234(21)487, (01)00884838099258(21)2692, (01)00884838099258(21)2953, (01)00884838099258(21)3236, (01)00884838085268(21)1001, (01)00884838099234(21)84, (01)00884838099258(21)2998, (01)00884838099258(21)3634, (01)00884838085251(21)718, (01)00884838085350(21)190, (01)00884838085268(21)704173, (01)00884838099203(21)205, (01)00884838099258(21)3545, (01)00884838085282(21)132, (01)00884838099203(21)13, (01)00884838085268(21)71, (01)00884838099234(21)454, (01)00884838099241(21)1207, (01)00884838085367(21)432, (01)00884838099272(21)771, (01)00884838110571(21)59, (01)00884838110564(21)192, (01)00884838099241(21)1301, (01)00884838099241(21)405, (01)00884838085268(21)703445, (01)00884838099258(21)3249, (01)00884838099258(21)2267, (01)00884838099234(21)346, (01)00884838099241(21)976, (01)00884838099241(21)975, (01)00884838099241(21)974, (01)00884838085251(21)891, (01)00884838085251(21)464, (01)00884838085251(21)462, (01)00884838099210(21)50, (01)00884838099241(21)635, (01)00884838085251(21)768, (01)00884838085268(21)551, (01)00884838099258(21)3286, (01)00884838085350(21)268, (01)00884838085367(21)215, (01)00884838085251(21)263, (01)00884838085268(21)550, (01)00884838099265(21)293, (01)00884838099265(21)315, (01)00884838099258(21)639, (01)00884838099258(21)882, (01)00884838085251(21)944, (01)00884838099258(21)1840, (01)00884838099241(21)924, (01)00884838099258(21)829, (01)00884838099241(21)217, (01)00884838085251(21)669, (01)00884838099258(21)2812, (01)00884838099258(21)204, (01)00884838099234(21)396, (01)00884838110564(21)275, (01)00884838099258(21)8, (01)00884838099258(21)3335, (01)00884838085367(21)447, (01)00884838099258(21)3111, (01)00884838099258(21)2835, (01)00884838085268(21)703526, (01)00884838085268(21)1186, (01)00884838099258(21)913, (01)00884838099258(21)2651, (01)00884838099272(21)790, (01)00884838099258(21)2510, (01)00884838085251(21)112, (01)00884838085251(21)23, (01)00884838085251(21)452, (01)00884838085367(21)399, (01)00884838085268(21)740, (01)00884838085367(21)338, (01)00884838085268(21)1104, (01)00884838099227(21)103, (01)00884838099258(21)2606, (01)00884838099258(21)3019, (01)00884838099258(21)1289, (01)00884838085268(21)704085, (01)00884838085268(21)704079, (01)00884838099272(21)1076, (01)00884838085282(21)67, (01)00884838099241(21)381, (01)00884838085268(21)703485, (01)00884838085251(21)338, (01)00884838099258(21)3549, (01)00884838085268(21)703500, (01)00884838085268(21)547, (01)00884838099258(21)2556, (01)00884838099258(21)911, (01)00884838099241(21)395, (01)00884838099258(21)3250, (01)00884838099258(21)2103, (01)00884838085268(21)703935, (01)00884838099272(21)543, (01)00884838099258(21)1620, (01)00884838099258(21)1600, (01)00884838099258(21)2071, (01)00884838099258(21)2028, (01)00884838085268(21)371, (01)00884838099258(21)2099, (01)00884838099258(21)2076, (01)00884838099258(21)2011, (01)00884838099272(21)66, (01)00884838099258(21)111, (01)00884838099258(21)1587, (01)00884838085268(21)835, (01)00884838099258(21)1644, (01)00884838099258(21)259, (01)00884838099258(21)18, (01)00884838099258(21)1580, (01)00884838099272(21)542, (01)00884838099258(21)1536, (01)00884838085268(21)703436, (01)00884838085367(21)339, (01)00884838099272(21)551, (01)00884838099258(21)2330, (01)00884838099258(21)1936, (01)00884838099258(21)1968, (01)00884838099272(21)550, (01)00884838099258(21)248, (01)00884838099258(21)926, (01)00884838099258(21)1957, (01)00884838099234(21)39, (01)00884838099272(21)484, (01)00884838085275(21)110, (01)00884838085268(21)1203, (01)00884838099258(21)1596, (01)00884838099258(21)1603, (01)00884838099272(21)504, (01)00884838099258(21)1462, (01)00884838099258(21)249, (01)00884838099203(21)99, (01)00884838099210(21)263, (01)00884838099234(21)339, (01)00884838099272(21)714, (01)00884838099234(21)172, (01)00884838085282(21)183, (01)00884838085367(21)316, (01)00884838099227(21)355, (01)00884838099241(21)111, (01)00884838099203(21)34, (01)00884838099227(21)113, (01)00884838099227(21)110, (01)00884838099258(21)264, (01)00884838099241(21)691, (01)00884838085268(21)485, (01)00884838085251(21)367, (01)00884838099272(21)1054, (01)00884838099258(21)3230, (01)00884838099258(21)496, (01)00884838099234(21)217, (01)00884838099258(21)2073, (01)00884838085282(21)439, (01)00884838099265(21)254, (01)00884838099272(21)366, (01)00884838085251(21)440, (01)00884838085268(21)703405, (01)00884838099227(21)315, (01)00884838099272(21)18, (01)00884838085268(21)704289, (01)00884838085268(21)596, (01)00884838099258(21)545, (01)00884838085367(21)555, (01)00884838099258(21)1919, (01)00884838085282(21)70, (01)00884838099210(21)65, (01)00884838099258(21)1538, (01)00884838099227(21)21, (01)00884838099234(21)191, (01)00884838085268(21)803, (01)00884838085251(21)490, (01)00884838099258(21)3701, (01)00884838099241(21)815, (01)00884838099241(21)436, (01)00884838099241(21)438, (01)00884838099241(21)437, (01)00884838085268(21)703806, (01)00884838085367(21)552, (01)00884838099241(21)35, (01)00884838099241(21)830, (01)00884838085268(21)1052, (01)00884838099258(21)803, (01)00884838099258(21)3600, (01)00884838099272(21)1140, (01)00884838099241(21)1327, (01)00884838085268(21)703804, (01)00884838085367(21)570, (01)00884838085268(21)703906, (01)00884838085268(21)703596, (01)00884838099272(21)1048, (01)00884838099203(21)142, (01)00884838099272(21)893, (01)00884838099203(21)39, (01)00884838099227(21)114, (01)00884838085268(21)1181, (01)00884838085268(21)224, (01)00884838099258(21)1667, (01)00884838085268(21)703937, (01)00884838099272(21)1152, (01)00884838099272(21)1132, (01)00884838085268(21)703972, (01)00884838099258(21)808, (01)00884838085251(21)181, (01)00884838085367(21)500, (01)00884838085367(21)347, (01)00884838085367(21)148, (01)00884838099241(21)938, (01)00884838085251(21)250, (01)00884838085251(21)145, (01)00884838099272(21)217, (01)00884838085268(21)1288, (01)00884838085251(21)920, (01)00884838099241(21)331, (01)00884838085251(21)705, (01)00884838099258(21)3512, (01)00884838110571(21)81, (01)00884838099265(21)312, (01)00884838099258(21)2431, (01)00884838099272(21)752, (01)00884838103276(21)112, (01)00884838099258(21)3318, (01)00884838085282(21)446, (01)00884838103276(21)34, (01)00884838099258(21)2064, (01)00884838103276(21)79, (01)00884838099258(21)2416, (01)00884838103276(21)130, (01)00884838085268(21)703954, (01)00884838099272(21)1088, (01)00884838099234(21)327, (01)00884838099227(21)161, (01)00884838099241(21)79, (01)00884838099258(21)1184, (01)00884838099241(21)1267, (01)00884838099272(21)47, (01)00884838085268(21)1183, (01)00884838099258(21)1169, (01)00884838085251(21)50, (01)00884838085251(21)49, (01)00884838099258(21)729, (01)00884838099227(21)194, (01)00884838099258(21)2885, (01)00884838099241(21)295, (01)00884838099241(21)65, (01)00884838085251(21)617, (01)00884838099258(21)861, (01)00884838099258(21)2713, (01)00884838099272(21)544, (01)00884838099258(21)3150, (01)00884838085268(21)1217, (01)00884838085282(21)50, (01)00884838085367(21)134, (01)00884838085251(21)336, (01)00884838085251(21)226, (01)00884838085268(21)368, (01)00884838085367(21)38, (01)00884838103276(21)48, (01)00884838099241(21)234, (01)00884838085251(21)761, (01)00884838099241(21)1246, (01)00884838099227(21)105, (01)00884838099203(21)144, (01)00884838099241(21)796, (01)00884838099234(21)52, (01)00884838099210(21)353, (01)00884838085251(21)221, (01)00884838099258(21)146, (01)00884838099258(21)152, (01)00884838085251(21)220, (01)00884838099258(21)3191, (01)00884838099272(21)148, (01)00884838085268(21)1314, (01)00884838099234(21)575, (01)00884838099234(21)697, (01)00884838110564(21)127, (01)00884838110571(21)89, (01)00884838103276(21)52, (01)00884838099258(21)2767, (01)00884838099258(21)2768, (01)00884838085367(21)560, (01)00884838110564(21)108, (01)00884838099258(21)3751, (01)00884838085268(21)704262, (01)00884838099210(21)278, (01)00884838099227(21)171, (01)00884838110564(21)254, (01)00884838110564(21)255, (01)00884838110564(21)24, (01)00884838103276(21)141, (01)00884838099258(21)1411, (01)00884838085268(21)703948, (01)00884838085251(21)1037, (01)00884838099265(21)343, (01)00884838099234(21)196, (01)00884838099258(21)3648, (01)00884838099272(21)906, (01)00884838099272(21)408, (01)00884838099258(21)1043, (01)00884838099241(21)337, (01)00884838085251(21)799, (01)00884838099241(21)1075, (01)00884838099258(21)3197, (01)00884838085268(21)703556, (01)00884838085268(21)687, (01)00884838099258(21)768, (01)00884838085251(21)839, (01)00884838099258(21)2084, (01)00884838085367(21)535, (01)00884838085367(21)158, (01)00884838099258(21)449, (01)00884838099258(21)448, (01)00884838099241(21)270, (01)00884838099241(21)663, (01)00884838099272(21)289, (01)00884838085268(21)902, (01)00884838099272(21)548, (01)00884838099258(21)842, (01)00884838099241(21)468, (01)00884838085251(21)354, (01)00884838099258(21)3768, (01)00884838099258(21)1737, (01)00884838085251(21)154, (01)00884838085251(21)153, (01)00884838085251(21)957, (01)00884838099234(21)677, (01)00884838099234(21)676, (01)00884838099210(21)82, (01)00884838099258(21)520, (01)00884838085268(21)1422, (01)00884838085282(21)138, (01)00884838085251(21)471, (01)00884838085268(21)571, (01)00884838085275(21)17, (01)00884838085275(21)16, (01)00884838085275(21)14, (01)00884838099258(21)1875, (01)00884838099241(21)376, (01)00884838099241(21)451, (01)00884838099241(21)450, (01)00884838099272(21)299, (01)00884838099272(21)238, (01)00884838085251(21)357, (01)00884838099258(21)1085, (01)00884838085251(21)132, (01)00884838085251(21)133, (01)00884838085251(21)130, (01)00884838099265(21)201, (01)00884838099234(21)406, (01)00884838085268(21)278, (01)00884838099258(21)3535, (01)00884838099258(21)3533, (01)00884838099234(21)436, (01)00884838099234(21)118, (01)00884838099258(21)2591, (01)00884838099234(21)415, (01)00884838099272(21)404, (01)00884838099258(21)1463, (01)00884838099272(21)401, (01)00884838099272(21)283, (01)00884838099272(21)284, (01)00884838085268(21)703754, (01)00884838085282(21)379, (01)00884838085282(21)321, (01)00884838085367(21)518, (01)00884838085282(21)293, (01)00884838085282(21)292, (01)00884838085367(21)318, (01)00884838085268(21)457, (01)00884838085367(21)250, (01)00884838099241(21)49, (01)00884838085251(21)261, (01)00884838099272(21)827, (01)00884838099227(21)173, (01)00884838099227(21)174, (01)00884838099234(21)114, (01)00884838099258(21)1933, (01)00884838099258(21)1703, (01)00884838099258(21)423, (01)00884838099210(21)188, (01)00884838085282(21)151, (01)00884838085282(21)237, (01)00884838085251(21)695, (01)00884838099272(21)833, (01)00884838099265(21)320, (01)00884838085367(21)118, (01)00884838099258(21)193, (01)00884838085251(21)158, (01)00884838085268(21)703569, (01)00884838099210(21)68, (01)00884838085251(21)507, (01)00884838099265(21)392, (01)00884838099203(21)82, (01)00884838099241(21)505, (01)00884838110564(21)361, (01)00884838085268(21)704081, (01)00884838099241(21)650, (01)00884838085268(21)703854, (01)00884838099258(21)1963, (01)00884838099234(21)452, (01)00884838085251(21)854, (01)00884838099234(21)53, (01)00884838085268(21)704103, (01)00884838110564(21)20, (01)00884838099258(21)2097, (01)00884838085268(21)1116, (01)00884838085367(21)336, (01)00884838099258(21)31, (01)00884838085268(21)478, (01)00884838085282(21)5, (01)00884838110564(21)306, (01)00884838099203(21)145, (01)00884838099241(21)1059, (01)00884838099234(21)216, (01)00884838085367(21)185, (01)00884838085251(21)477, (01)00884838085268(21)682, (01)00884838099258(21)1033, (01)00884838085282(21)520, (01)00884838085282(21)519, (01)00884838085268(21)704019, (01)00884838099241(21)1259, (01)00884838099234(21)761, (01)00884838099241(21)1050, (01)00884838099258(21)1133, (01)00884838099258(21)1071, (01)00884838099241(21)465, (01)00884838099265(21)38, (01)00884838099241(21)68, (01)00884838085367(21)206, (01)00884838085268(21)703995, (01)00884838099234(21)179, (01)00884838085282(21)208, (01)00884838085268(21)1119, (01)00884838085268(21)960, (01)00884838085275(21)53, (01)00884838085268(21)824, (01)00884838085350(21)203, (01)00884838099241(21)383, (01)00884838099258(21)866, (01)00884838099241(21)20, (01)00884838085268(21)1257, (01)00884838099258(21)1800, (01)00884838099258(21)2057, (01)00884838085268(21)523, (01)00884838085268(21)522, (01)00884838099258(21)3259, (01)00884838099258(21)358, (01)00884838099258(21)3223, (01)00884838085268(21)839, (01)00884838085268(21)703424, (01)00884838099272(21)84, (01)00884838099203(21)59, (01)00884838099227(21)64, (01)00884838099241(21)246, (01)00884838099241(21)245, (01)00884838099258(21)2224, (01)00884838085367(21)50, (01)00884838099272(21)692, (01)00884838099258(21)299, (01)00884838099258(21)3118, (01)00884838085282(21)462, (01)00884838085268(21)704023, (01)00884838099210(21)51, (01)00884838099203(21)77, (01)00884838099227(21)49, (01)00884838110564(21)328, (01)00884838099258(21)1090, (01)00884838099227(21)71, (01)00884838099234(21)398, (01)00884838099272(21)895, (01)00884838099258(21)489, (01)00884838099241(21)1261, (01)00884838099227(21)91, (01)00884838099203(21)180, (01)00884838099265(21)423, (01)00884838085268(21)704119, (01)00884838099258(21)378, (01)00884838085268(21)1015, (01)00884838099210(21)5, (01)00884838099241(21)222, (01)00884838099227(21)187, (01)00884838099227(21)166, (01)00884838099234(21)88, (01)00884838099234(21)31, (01)00884838085367(21)489, (01)00884838085268(21)703614, (01)00884838085251(21)785, (01)00884838099241(21)374, (01)00884838099272(21)118, (01)00884838085268(21)1089, (01)00884838099241(21)930, (01)00884838099258(21)1088, (01)00884838099258(21)208, (01)00884838099241(21)794, (01)00884838099241(21)795, (01)00884838099241(21)793, (01)00884838099241(21)849, (01)00884838099258(21)407, (01)00884838085367(21)521, (01)00884838085251(21)789, (01)00884838085268(21)1175, (01)00884838099258(21)1791, (01)00884838099265(21)14, (01)00884838085268(21)748, (01)00884838085268(21)781, (01)00884838085268(21)1351, (01)00884838099272(21)924, (01)00884838099241(21)712, (01)00884838099272(21)224, (01)00884838099272(21)233, (01)00884838085251(21)404, (01)00884838099210(21)76, (01)00884838085275(21)103, (01)00884838099272(21)1087, (01)00884838099258(21)2086, (01)00884838085251(21)877, (01)00884838099210(21)243, (01)00884838099258(21)537, (01)00884838099241(21)861, (01)00884838099227(21)210, (01)00884838099203(21)126, (01)00884838085251(21)806, (01)00884838085275(21)127, (01)00884838099234(21)331, (01)00884838085275(21)97, (01)00884838099265(21)350, (01)00884838099272(21)579, (01)00884838099258(21)2760, (01)00884838085282(21)342, (01)00884838099234(21)297, (01)00884838099241(21)1291, (01)00884838099258(21)175, (01)00884838085268(21)1076, (01)00884838099272(21)515, (01)00884838085350(21)138, (01)00884838099258(21)1742, (01)00884838085251(21)422, (01)00884838099227(21)219, (01)00884838099227(21)222, (01)00884838099234(21)392, (01)00884838099234(21)374, (01)00884838099241(21)325, (01)00884838085251(21)194, (01)00884838085268(21)1129, (01)00884838099241(21)908, (01)00884838099241(21)907, (01)00884838099265(21)243, (01)00884838099258(21)550, (01)00884838085251(21)802, (01)00884838085251(21)801, (01)00884838085350(21)34, (01)00884838099258(21)3285, (01)00884838085367(21)37, (01)00884838099241(21)859, (01)00884838085251(21)140, (01)00884838099241(21)605, (01)00884838085268(21)532, (01)00884838099258(21)2456, (01)00884838099241(21)61, (01)00884838085282(21)57, (01)00884838085367(21)79, (01)00884838099241(21)544, (01)00884838099272(21)455, (01)00884838085251(21)878, (01)00884838085251(21)608, (01)00884838085268(21)703347, (01)00884838099203(21)97, (01)00884838085251(21)952, (01)00884838099241(21)1289, (01)00884838099241(21)1285, (01)00884838099227(21)273, (01)00884838099258(21)1304, (01)00884838099203(21)132, (01)00884838099258(21)2476, (01)00884838110564(21)129, (01)00884838085268(21)1166, (01)00884838099258(21)1984, (01)00884838099227(21)205, (01)00884838099258(21)752, (01)00884838103276(21)131, (01)00884838110564(21)417, (01)00884838085268(21)704030, (01)00884838099272(21)946, (01)00884838099272(21)599, (01)00884838099258(21)2975, (01)00884838099234(21)503, (01)00884838085268(21)704126, (01)00884838099234(21)412, (01)00884838085268(21)489, (01)00884838099258(21)2429, (01)00884838099258(21)1114, (01)00884838099227(21)280, (01)00884838099258(21)1545, (01)00884838110564(21)457, (01)00884838099203(21)79, (01)00884838099258(21)1119, (01)00884838099258(21)2962, (01)00884838085251(21)10, (01)00884838099241(21)851, (01)00884838099272(21)726, (01)00884838099241(21)1201, (01)00884838099241(21)921, (01)00884838085367(21)529, (01)00884838099272(21)676, (01)00884838099241(21)589, (01)00884838085268(21)956, (01)00884838099227(21)300, (01)00884838099203(21)91, (01)00884838099203(21)161, (01)00884838085282(21)222, (01)00884838099234(21)385, (01)00884838099258(21)2237, (01)00884838085251(21)123, (01)00884838099241(21)1100, (01)00884838099258(21)987, (01)00884838099258(21)943, (01)00884838085251(21)319, (01)00884838099258(21)2102, (01)00884838085367(21)111, (01)00884838099210(21)139, (01)00884838099258(21)2423, (01)00884838099258(21)505, (01)00884838085251(21)805, (01)00884838085251(21)686, (01)00884838085268(21)569, (01)00884838099258(21)3193, (01)00884838085251(21)397, (01)00884838085251(21)127, (01)00884838085251(21)128, (01)00884838085367(21)254, (01)00884838099203(21)102, (01)00884838099258(21)1331, (01)00884838099234(21)343, (01)00884838085251(21)253, (01)00884838099241(21)894, (01)00884838085350(21)53, (01)00884838110564(21)62, (01)00884838099241(21)273, (01)00884838085251(21)594, (01)00884838099203(21)52, (01)00884838085282(21)391, (01)00884838099234(21)488, (01)00884838099227(21)256, (01)00884838099210(21)319, (01)00884838099234(21)427, (01)00884838099272(21)395, (01)00884838099258(21)3441, (01)00884838099203(21)120, (01)00884838099258(21)3232, (01)00884838099258(21)3252, (01)00884838099258(21)2927, (01)00884838085251(21)988, (01)00884838099227(21)335, (01)00884838099241(21)294, (01)00884838085268(21)432, (01)00884838099258(21)2652, (01)00884838099258(21)2741, (01)00884838085367(21)427, (01)00884838099258(21)3498, (01)00884838099241(21)806, (01)00884838099241(21)800, (01)00884838099234(21)389, (01)00884838099258(21)555, (01)00884838099258(21)7, (01)00884838099258(21)3001, (01)00884838099265(21)393, (01)00884838099241(21)333, (01)00884838099227(21)121, (01)00884838085268(21)874, (01)00884838110564(21)343, (01)00884838099241(21)4, (01)00884838085251(21)301, (01)00884838085251(21)300, (01)00884838099241(21)1195, (01)00884838099258(21)2007, (01)00884838099241(21)118, (01)00884838099241(21)1197, (01)00884838099241(21)1135, (01)00884838099272(21)261, (01)00884838099272(21)1, (01)00884838099241(21)429, (01)00884838099241(21)720, (01)00884838099241(21)264, (01)00884838099258(21)133, (01)00884838099258(21)349, (01)00884838085268(21)703819, (01)00884838099241(21)707, (01)00884838085251(21)332, (01)00884838085251(21)35, (01)00884838099258(21)2811, (01)00884838085367(21)394, (01)00884838110564(21)356, (01)00884838085268(21)704036, (01)00884838099241(21)606, (01)00884838099272(21)444, (01)00884838085367(21)71, (01)00884838085268(21)446, (01)00884838085268(21)914, (01)00884838099234(21)349, (01)00884838099241(21)43, (01)00884838099227(21)80, (01)00884838099241(21)392, (01)00884838085251(21)563, (01)00884838085251(21)341, (01)00884838085268(21)703690, (01)00884838085282(21)98, (01)00884838085251(21)316, (01)00884838085251(21)314, (01)00884838099241(21)759, (01)00884838099258(21)933, (01)00884838099241(21)418, (01)00884838099272(21)826, (01)00884838099258(21)3372, (01)00884838099258(21)878, (01)00884838099272(21)93, (01)00884838099234(21)536, (01)00884838099227(21)123, (01)00884838085350(21)113, (01)00884838085251(21)496, (01)00884838099258(21)186, (01)00884838085251(21)641, (01)00884838099258(21)2151, (01)00884838099227(21)243, (01)00884838085268(21)703597, (01)00884838085268(21)703866, (01)00884838085251(21)953, (01)00884838085268(21)703611, (01)00884838099210(21)31, (01)00884838099241(21)1280, (01)00884838099272(21)956, (01)00884838099272(21)1077, (01)00884838099265(21)373, (01)00884838085268(21)1080, (01)00884838085251(21)30, (01)00884838099234(21)214, (01)00884838099227(21)344, (01)00884838085268(21)703375, (01)00884838085251(21)950, (01)00884838085282(21)441, (01)00884838110564(21)428, (01)00884838099210(21)193, (01)00884838103276(21)189, (01)00884838085282(21)320, (01)00884838110564(21)215, (01)00884838103276(21)37, (01)00884838085268(21)703648, (01)00884838085268(21)703709, (01)00884838103276(21)83, (01)00884838085251(21)331, (01)00884838099258(21)2324, (01)00884838099272(21)355, (01)00884838099272(21)320, (01)00884838085268(21)787, (01)00884838085367(21)47, (01)00884838099258(21)3065, (01)00884838085282(21)494, (01)00884838110564(21)66, (01)00884838085268(21)704207, (01)00884838103276(21)223, (01)00884838099258(21)2574, (01)00884838085268(21)703728, (01)00884838085251(21)652, (01)00884838099241(21)179, (01)00884838099258(21)635, (01)00884838099265(21)99, (01)00884838099272(21)710, (01)00884838099265(21)456, (01)00884838099265(21)206, (01)00884838099241(21)445, (01)00884838099210(21)106, (01)00884838085268(21)703826, (01)00884838085367(21)534, (01)00884838099258(21)2538, (01)00884838085367(21)292, (01)00884838099241(21)556, (01)00884838099241(21)215, (01)00884838085367(21)497, (01)00884838099258(21)60, (01)00884838099258(21)19, (01)00884838099241(21)1027, (01)00884838099258(21)2809, (01)00884838099265(21)4, (01)00884838099241(21)21, (01)00884838099265(21)442, (01)00884838099241(21)442, (01)00884838085268(21)381, (01)00884838099258(21)6, (01)00884838085268(21)1111, (01)00884838099241(21)807, (01)00884838099258(21)2157, (01)00884838099241(21)791, (01)00884838099258(21)1366, (01)00884838099258(21)1365, (01)00884838099258(21)551, (01)00884838085251(21)1005, (01)00884838085251(21)349, (01)00884838099241(21)76, (01)00884838099272(21)435, (01)00884838099234(21)232, (01)00884838099258(21)32, (01)00884838085350(21)254, (01)00884838099241(21)922, (01)00884838085268(21)630, (01)00884838099272(21)1072, (01)00884838099234(21)500, (01)00884838110564(21)98, (01)00884838110564(21)48, (01)00884838099258(21)1760, (01)00884838103276(21)139, (01)00884838103276(21)87, (01)00884838110564(21)235, (01)00884838110571(21)120, (01)00884838099241(21)1275, (01)00884838110571(21)42, (01)00884838099258(21)3319, (01)00884838085367(21)591, (01)00884838103276(21)61, (01)00884838110564(21)378, (01)00884838110564(21)145, (01)00884838103276(21)259, (01)00884838110564(21)401, (01)00884838110564(21)402, (01)00884838099272(21)820, (01)00884838103276(21)218, (01)00884838085268(21)703794, (01)00884838110564(21)488, (01)00884838103276(21)90, (01)00884838099258(21)2268, (01)00884838099258(21)1272, (01)00884838085268(21)638, (01)00884838085367(21)35, (01)00884838099258(21)2665, (01)00884838099241(21)427, (01)00884838099272(21)749, (01)00884838085268(21)826, (01)00884838085268(21)901, (01)00884838099265(21)327, (01)00884838099241(21)257, (01)00884838085282(21)79, (01)00884838099241(21)266, (01)00884838099265(21)447, (01)00884838099272(21)235, (01)00884838099265(21)96, (01)00884838099258(21)460, (01)00884838085282(21)65, (01)00884838099234(21)713, (01)00884838099258(21)2646, (01)00884838099234(21)622, (01)00884838099272(21)846, (01)00884838099227(21)308, (01)00884838099258(21)2781, (01)00884838085282(21)87, (01)00884838099210(21)392, (01)00884838099272(21)727, (01)00884838085268(21)703607, (01)00884838085268(21)703574, (01)00884838085251(21)236, (01)00884838099258(21)3412, (01)00884838099258(21)990, (01)00884838099265(21)48, (01)00884838099258(21)355, (01)00884838099265(21)47, (01)00884838099265(21)52, (01)00884838085251(21)979, (01)00884838099258(21)2779, (01)00884838099234(21)727, (01)00884838085350(21)140, (01)00884838099258(21)1446, (01)00884838085282(21)109, (01)00884838099234(21)469, (01)00884838085251(21)831, (01)00884838099258(21)2477, (01)00884838099227(21)22, (01)00884838099241(21)1336, (01)00884838099241(21)1268, (01)00884838099258(21)880, (01)00884838099258(21)200, (01)00884838085268(21)1165, (01)00884838085282(21)75, (01)00884838099258(21)2536, (01)00884838085268(21)762, (01)00884838099227(21)140, (01)00884838085350(21)142, (01)00884838085350(21)143, (01)00884838085350(21)144, (01)00884838085367(21)293, (01)00884838099272(21)1026, (01)00884838085282(21)46, (01)00884838085282(21)101, (01)00884838099234(21)588, (01)00884838099265(21)446, (01)00884838099258(21)716, (01)00884838085268(21)703743, (01)00884838085350(21)266, (01)00884838085268(21)703632, (01)00884838099258(21)1173, (01)00884838099265(21)413, (01)00884838099258(21)3204, (01)00884838085268(21)703913, (01)00884838099234(21)365, (01)00884838110564(21)336, (01)00884838085282(21)311, (01)00884838099258(21)1684, (01)00884838085282(21)442, (01)00884838085268(21)703889, (01)00884838099234(21)443, (01)00884838085282(21)485, (01)00884838099272(21)621, (01)00884838103276(21)188, (01)00884838110564(21)444, (01)00884838099272(21)411, (01)00884838110564(21)431, (01)00884838110564(21)115, (01)00884838103276(21)105, (01)00884838110571(21)47, (01)00884838110571(21)48, (01)00884838110571(21)31, (01)00884838085268(21)703700, (01)00884838085268(21)704246, (01)00884838085268(21)704054, (01)00884838110571(21)187, (01)00884838110571(21)178, (01)00884838110571(21)26, (01)00884838085268(21)703606, (01)00884838085282(21)121, (01)00884838099258(21)1498, (01)00884838099234(21)510, (01)00884838099258(21)2643, (01)00884838099258(21)410, (01)00884838099258(21)2487, (01)00884838099234(21)751, (01)00884838099234(21)757, (01)00884838085268(21)703473, (01)00884838085282(21)68, (01)00884838099272(21)1144, (01)00884838099272(21)1129, (01)00884838099258(21)504, (01)00884838099258(21)2658, (01)00884838085251(21)501, (01)00884838099234(21)758, (01)00884838099272(21)1041, (01)00884838099258(21)2797, (01)00884838099258(21)2445, (01)00884838099258(21)1075, (01)00884838099210(21)287, (01)00884838099234(21)692, (01)00884838099234(21)663, (01)00884838099258(21)459, (01)00884838099241(21)228, (01)00884838099258(21)2285, (01)00884838099258(21)2444, (01)00884838099234(21)747, (01)00884838099258(21)1076, (01)00884838099258(21)2447, (01)00884838099272(21)777, (01)00884838099210(21)178, (01)00884838099258(21)2286, (01)00884838099234(21)705, (01)00884838099241(21)460, (01)00884838099234(21)414, (01)00884838099234(21)746, (01)00884838099234(21)712, (01)00884838099234(21)754, (01)00884838099234(21)728, (01)00884838099234(21)621, (01)00884838085251(21)841, (01)00884838099272(21)1053, (01)00884838099234(21)752, (01)00884838085268(21)703346, (01)00884838099241(21)554, (01)00884838099272(21)274, (01)00884838099234(21)710, (01)00884838099272(21)486, (01)00884838099234(21)715, (01)00884838099234(21)740, (01)00884838099258(21)2719, (01)00884838085251(21)842, (01)00884838099258(21)2535, (01)00884838085251(21)905, (01)00884838099234(21)726, (01)00884838099258(21)1688, (01)00884838099258(21)2360, (01)00884838099210(21)117, (01)00884838099234(21)766, (01)00884838099258(21)1681, (01)00884838099210(21)267, (01)00884838099234(21)717, (01)00884838099234(21)90, (01)00884838099234(21)706, (01)00884838099234(21)756, (01)00884838099234(21)739, (01)00884838099234(21)417, (01)00884838099234(21)709, (01)00884838099272(21)1015, (01)00884838099272(21)1014, (01)00884838099210(21)92, (01)00884838099258(21)2674, (01)00884838099234(21)498, (01)00884838085268(21)521, (01)00884838099210(21)324, (01)00884838099234(21)729, (01)00884838099234(21)748, (01)00884838099234(21)743, (01)00884838099234(21)130, (01)00884838099234(21)736, (01)00884838099234(21)753, (01)00884838099227(21)90, (01)00884838099265(21)302, (01)00884838085251(21)682, (01)00884838099210(21)213, (01)00884838099258(21)2485, (01)00884838099258(21)3655, (01)00884838099234(21)737, (01)00884838099272(21)503, (01)00884838085251(21)843, (01)00884838099234(21)707, (01)00884838099227(21)142, (01)00884838099258(21)2570, (01)00884838099234(21)714, (01)00884838085282(21)384, (01)00884838099234(21)738, (01)00884838099258(21)2362, (01)00884838099241(21)194, (01)00884838099234(21)718, (01)00884838099234(21)755, (01)00884838099272(21)1071, (01)00884838099234(21)711, (01)00884838099258(21)1715, (01)00884838099258(21)2780, (01)00884838099258(21)2478, (01)00884838099234(21)716, (01)00884838099234(21)750, (01)00884838099258(21)638, (01)00884838099227(21)69, (01)00884838099234(21)708, (01)00884838099234(21)744, (01)00884838099272(21)431, (01)00884838099272(21)481, (01)00884838099234(21)704, (01)00884838099234(21)284, (01)00884838085251(21)830, (01)00884838099210(21)281, (01)00884838099234(21)742, (01)00884838099234(21)725, (01)00884838099210(21)229, (01)00884838085251(21)323, (01)00884838099258(21)412, (0 1)00884838099258(21)1713, (01)00884838099234(21)565, (01)00884838099234(21)745, (01)00884838099241(21)954, (01)00884838099234(21)73, (01)00884838099227(21)200, (01)00884838099234(21)474, (01)00884838099272(21)460, (01)00884838099258(21)2191, (01)00884838099234(21)724, (01)00884838099265(21)195, (01)00884838085268(21)703750, (01)00884838099234(21)735, (01)00884838099234(21)749, (01)00884838099234(21)703, (01)00884838099258(21)1714, (01)00884838099241(21)956, (01)00884838099241(21)174, (01)00884838085251(21)315, (01)00884838099234(21)741, (01)00884838099234(21)201, (01)00884838099234(21)407, (01)00884838099210(21)215, (01)00884838099241(21)1151, (01)00884838099265(21)186, (01)00884838085251(21)680, (01)00884838099241(21)684, (01)00884838099210(21)300, (01)00884838099227(21)72, (01)00884838099265(21)448, (01)00884838099272(21)765, (01)00884838085367(21)227, (01)00884838085350(21)60, (01)00884838099258(21)2915, (01)00884838099227(21)362, (01)00884838099272(21)194, (01)00884838085367(21)359, (01)00884838085268(21)966, (01)00884838085268(21)703638, (01)00884838099258(21)2539, (01)00884838085268(21)703990, (01)00884838099210(21)268, (01)00884838085268(21)703834, (01)00884838085251(21)365, (01)00884838099234(21)397, (01)00884838099265(21)242, (01)00884838085268(21)703727, (01)00884838099272(21)696, (01)00884838085268(21)586, (01)00884838085268(21)965, (01)00884838085268(21)703875, (01)00884838099258(21)3379, (01)00884838099272(21)1099, (01)00884838099203(21)162, (01)00884838099234(21)104, (01)00884838099234(21)202, (01)00884838099241(21)563, (01)00884838099241(21)564, (01)00884838099241(21)559, (01)00884838099241(21)558, (01)00884838099258(21)1217, (01)00884838099258(21)1226, (01)00884838099258(21)1219, (01)00884838099258(21)1222, (01)00884838099258(21)1319, (01)00884838099258(21)1218, (01)00884838099241(21)724, (01)00884838085268(21)1081, (01)00884838099241(21)1077, (01)00884838099258(21)1952, (01)00884838099241(21)702, (01)00884838099258(21)20, (01)00884838099234(21)690, (01)00884838099241(21)1092, (01)00884838085268(21)498, (01)00884838110564(21)348, (01)00884838099258(21)1921, (01)00884838103276(21)23, (01)00884838085251(21)311, (01)00884838099241(21)475, (01)00884838099241(21)474, (01)00884838085251(21)312, (01)00884838099258(21)2271, (01)00884838099227(21)242, (01)00884838099241(21)48, (01)00884838085268(21)789, (01)00884838099241(21)779, (01)00884838099258(21)719, (01)00884838099265(21)167, (01)00884838099241(21)422, (01)00884838099272(21)25, (01)00884838085350(21)97, (01)00884838085268(21)728, (01)00884838099258(21)756, (01)00884838099227(21)323, (01)00884838099241(21)583, (01)00884838099241(21)573, (01)00884838099258(21)1273, (01)00884838099203(21)64, (01)00884838099241(21)413, (01)00884838085275(21)143, (01)00884838099234(21)46, (01)00884838099203(21)75, (01)00884838085268(21)291, (01)00884838099241(21)995, (01)00884838099210(21)104, (01)00884838099227(21)331, (01)00884838099258(21)1188, (01)00884838099258(21)1203, (01)00884838099258(21)614, (01)00884838099241(21)541, (01)00884838099258(21)896, (01)00884838099258(21)895, (01)00884838099258(21)1035, (01)00884838085282(21)228, (01)00884838085251(21)282, (01)00884838085251(21)283, (01)00884838085367(21)527, (01)00884838085367(21)387, (01)00884838099203(21)41, (01)00884838085275(21)41, (01)00884838099241(21)600, (01)00884838099227(21)136, (01)00884838099234(21)171, (01)00884838085367(21)168, (01)00884838085251(21)602, (01)00884838099265(21)187, (01)00884838099234(21)34, (01)00884838085268(21)703693, (01)00884838085268(21)703482, (01)00884838085282(21)252, (01)00884838085367(21)128, (01)00884838085282(21)91, (01)00884838085367(21)129, (01)00884838099272(21)438, (01)00884838099241(21)183, (01)00884838099258(21)508, (01)00884838085367(21)396, (01)00884838085268(21)703533, (01)00884838085268(21)73, (01)00884838099258(21)3481, (01)00884838099258(21)804, (01)00884838085268(21)967, (01)00884838085282(21)74, (01)00884838085282(21)73, (01)00884838099258(21)883, (01)00884838099241(21)6, (01)00884838085282(21)124, (01)00884838085282(21)191, (01)00884838085282(21)122, (01)00884838099272(21)275, (01)00884838099241(21)1190, (01)00884838085268(21)475, (01)00884838099227(21)269, (01)00884838099265(21)256, (01)00884838099241(21)700, (01)00884838099258(21)1794, (01)00884838099265(21)255, (01)00884838099241(21)699, (01)00884838099241(21)701, (01)00884838099258(21)399, (01)00884838085268(21)1008, (01)00884838099210(21)141, (01)00884838085282(21)421, (01)00884838085275(21)104, (01)00884838099203(21)84, (01)00884838099272(21)552, (01)00884838099241(21)664, (01)00884838099227(21)137, (01)00884838099241(21)876, (01)00884838085251(21)203, (01)00884838085268(21)340, (01)00884838085251(21)202, (01)00884838099258(21)2126, (01)00884838099258(21)2724, (01)00884838099258(21)2725, (01)00884838085268(21)963, (01)00884838099241(21)419, (01)00884838099241(21)666, (01)00884838085282(21)119, (01)00884838085282(21)309, (01)00884838099258(21)211, (01)00884838099258(21)648, (01)00884838099241(21)12, (01)00884838099241(21)243, (01)00884838085251(21)40, (01)00884838099227(21)77, (01)00884838099241(21)1017, (01)00884838099241(21)1037, (01)00884838099258(21)337, (01)00884838099258(21)227, (01)00884838085367(21)515, (01)00884838085268(21)703560, (01)00884838085350(21)45, (01)00884838099234(21)70, (01)00884838099241(21)283, (01)00884838085268(21)704261, (01)00884838099272(21)352, (01)00884838099227(21)89, (01)00884838085275(21)116, (01)00884838085268(21)704067, (01)00884838099227(21)294, (01)00884838099241(21)1286, (01)00884838085268(21)861, (01)00884838085350(21)107, (01)00884838099258(21)3386, (01)00884838099258(21)3387, (01)00884838099258(21)1594, (01)00884838099258(21)1595, (01)00884838099241(21)30, (01)00884838099258(21)1313, (01)00884838099234(21)680, (01)00884838085282(21)400, (01)00884838085282(21)251, (01)00884838099272(21)87, (01)00884838085251(21)674, (01)00884838085268(21)1135, (01)00884838085251(21)676, (01)00884838085251(21)673, (01)00884838085251(21)675, (01)00884838085268(21)349, (01)00884838099203(21)131, (01)00884838099241(21)502, (01)00884838099203(21)94, (01)00884838099258(21)1030, (01)00884838085251(21)609, (01)00884838099258(21)2621, (01)00884838099234(21)657, (01)00884838099241(21)790, (01)00884838085268(21)784, (01)00884838085268(21)221, (01)00884838085251(21)295, (01)00884838085268(21)704274, (01)00884838085268(21)703602, (01)00884838099258(21)795, (01)00884838099258(21)1297, (01)00884838099241(21)778, (01)00884838099241(21)22, (01)00884838085268(21)425, (01)00884838099241(21)1239, (01)00884838099241(21)1117, (01)00884838099258(21)2451, (01)00884838099258(21)2098, (01)00884838099241(21)540, (01)00884838099258(21)2615, (01)00884838099265(21)385, (01)00884838085268(21)831, (01)00884838085268(21)832, (01)00884838085350(21)112, (01)00884838085251(21)368, (01)00884838085251(21)369, (01)00884838099241(21)169, (01)00884838099241(21)485, (01)00884838099265(21)30, (01)00884838085251(21)615, (01)00884838099272(21)630, (01)00884838099265(21)384, (01)00884838099258(21)2938, (01)00884838099258(21)66, (01)00884838099258(21)67, (01)00884838099258(21)223, (01)00884838099241(21)26, (01)00884838099258(21)2936, (01)00884838085251(21)618, (01)00884838085350(21)134, (01)00884838085268(21)346, (01)00884838099234(21)229, (01)00884838099241(21)787, (01)00884838099258(21)1871, (01)00884838099258(21)2489, (01)00884838099258(21)2490, (01)00884838099227(21)181, (01)00884838099258(21)198, (01)00884838099241(21)507, (01)00884838085268(21)985, (01)00884838085251(21)310, (01)00884838099258(21)848, (01)00884838085268(21)712, (01)00884838099241(21)396, (01)00884838085367(21)486, (01)00884838099258(21)1763, (01)00884838085367(21)543, (01)00884838103276(21)30, (01)00884838099258(21)3427, (01)00884838099234(21)259, (01)00884838085268(21)283, (01)00884838085268(21)319, (01)00884838085268(21)834, (01)00884838085268(21)833, (01)00884838085367(21)69, (01)00884838099234(21)332, (01)00884838099258(21)3025, (01)00884838099241(21)382, (01)00884838085251(21)199, (01)00884838099258(21)1881, (01)00884838085282(21)480, (01)00884838099258(21)1831, (01)00884838085251(21)455, (01)00884838085367(21)453, (01)00884838099272(21)1047, (01)00884838085251(21)668, (01)00884838085268(21)583, (01)00884838085251(21)352, (01)00884838085268(21)703506, (01)00884838099234(21)643, (01)00884838099227(21)234, (01)00884838099258(21)634, (01)00884838099241(21)176, (01)00884838103276(21)132, (01)00884838085282(21)387, (01)00884838103276(21)96, (01)00884838099258(21)2253, (01)00884838099272(21)941, (01)00884838099241(21)986, (01)00884838099234(21)501, (01)00884838085251(21)280, (01)00884838085251(21)193, (01)00884838085268(21)233, (01)00884838085268(21)66, (01)00884838085268(21)773, (01)00884838099265(21)321, (01)00884838099234(21)286, (01)00884838085251(21)667, (01)00884838099258(21)1302, (01)00884838099258(21)1303, (01)00884838099258(21)1040, (01)00884838099258(21)2272, (01)00884838099210(21)133, (01)00884838099241(21)685, (01)00884838099241(21)682, (01)00884838099265(21)459, (01)00884838099241(21)1141, (01)00884838099258(21)3196, (01)00884838099265(21)169, (01)00884838085268(21)996, (01)00884838085268(21)704144, (01)00884838099227(21)95, (01)00884838099258(21)2230, (01)00884838099272(21)349, (01)00884838099258(21)2231, (01)00884838099272(21)345, (01)00884838099258(21)2480, (01)00884838085275(21)151, (01)00884838085251(21)183, (01)00884838099258(21)1855, (01)00884838099258(21)2153, (01)00884838099241(21)721, (01)00884838099265(21)135, (01)00884838099265(21)134, (01)00884838099227(21)2, (01)00884838099258(21)1232, (01)00884838099258(21)3142, (01)00884838085268(21)704167, (01)00884838085268(21)704168, (01)00884838099241(21)91, (01)00884838099234(21)306, (01)00884838085251(21)749, (01)00884838085251(21)692, (01)00884838085367(21)174, (01)00884838099241(21)773, (01)00884838099241(21)378, (01)00884838099234(21)610, (01)00884838085268(21)1223, (01)00884838099241(21)641, (01)00884838099241(21)640, (01)00884838085251(21)32, (01)00884838099241(21)1021, (01)00884838085251(21)241, (01)00884838099258(21)236, (01)00884838099258(21)1332, (01)00884838099272(21)429, (01)00884838099210(21)142, (01)00884838099258(21)1105, (01)00884838099258(21)416, (01)00884838085367(21)484, (01)00884838085367(21)483, (01)00884838099210(21)105, (01)00884838099258(21)2460, (01)00884838099258(21)2530, (01)00884838099241(21)805, (01)00884838085367(21)334, (01)00884838110564(21)120, (01)00884838085282(21)269, (01)00884838099227(21)111, (01)00884838085251(21)935, (01)00884838099234(21)117,  (01)00884838099241(21)304, (01)00884838099258(21)2845, (01)00884838099272(21)854, (01)00884838099258(21)1355, (01)00884838099258(21)1676, (01)00884838099258(21)1205, (01)00884838099241(21)443, (01)00884838085268(21)1210, (01)00884838085251(21)505, (01)00884838085268(21)642, (01)00884838085251(21)407, (01)00884838085268(21)698, (01)00884838085268(21)643, (01)00884838085251(21)408, (01)00884838099258(21)1761, (01)00884838099258(21)1360, (01)00884838099234(21)646, (01)00884838099241(21)562, (01)00884838099258(21)1180, (01)00884838085268(21)435, (01)00884838099272(21)1009, (01)00884838085268(21)704070, (01)00884838099241(21)259, (01)00884838099241(21)115, (01)00884838085275(21)93, (01)00884838099272(21)888, (01)00884838099258(21)1633, (01)00884838099258(21)3554, (01)00884838085268(21)1077, (01)00884838099241(21)985, (01)00884838085268(21)882, (01)00884838099258(21)3234, (01)00884838099258(21)85, (01)00884838085367(21)425, (01)00884838085251(21)752, (01)00884838085251(21)157, (01)00884838085251(21)138, (01)00884838099258(21)2983, (01)00884838085251(21)461, (01)00884838085251(21)463, (01)00884838099272(21)369, (01)00884838099258(21)872, (01)00884838099241(21)159, (01)00884838099241(21)160, (01)00884838085268(21)703349, (01)00884838085251(21)554, (01)00884838085251(21)423, (01)00884838099241(21)994, (01)00884838099258(21)985, (01)00884838099272(21)557, (01)00884838099258(21)898, (01)00884838099258(21)59, (01)00884838085268(21)477, (01)00884838099258(21)2446, (01)00884838099227(21)221, (01)00884838085282(21)111, (01)00884838099258(21)371, (01)00884838099258(21)3268, (01)00884838099258(21)1864, (01)00884838099234(21)13, (01)00884838085251(21)252, (01)00884838085268(21)703841, (01)00884838099258(21)3141, (01)00884838099258(21)647, (01)00884838099258(21)3366, (01)00884838099227(21)293, (01)00884838099241(21)1186, (01)00884838099258(21)2589, (01)00884838085282(21)11, (01)00884838099258(21)156, (01)00884838099241(21)170, (01)00884838099258(21)1182, (01)00884838099258(21)2678, (01)00884838099258(21)1809, (01)00884838099258(21)1017, (01)00884838099241(21)82, (01)00884838099258(21)347, (01)00884838085251(21)650, (01)00884838085367(21)181, (01)00884838085251(21)426, (01)00884838085251(21)169, (01)00884838085367(21)166, (01)00884838099265(21)287, (01)00884838085268(21)450, (01)00884838099234(21)395, (01)00884838099234(21)429, (01)00884838085251(21)442, (01)00884838085251(21)351, (01)00884838099234(21)12, (01)00884838085251(21)750, (01)00884838085367(21)408, (01)00884838099258(21)2345, (01)00884838085251(21)968, (01)00884838099203(21)96, (01)00884838099258(21)1420, (01)00884838085251(21)210, (01)00884838085251(21)211, (01)00884838085268(21)400, (01)00884838085251(21)207, (01)00884838085268(21)399, (01)00884838085251(21)208, (01)00884838085251(21)209, (01)00884838099258(21)1948, (01)00884838099241(21)199, (01)00884838085367(21)415, (01)00884838085268(21)703629, (01)00884838110564(21)421, (01)00884838099272(21)585, (01)00884838099272(21)596, (01)00884838099227(21)165, (01)00884838099258(21)2017, (01)00884838099258(21)1590, (01)00884838085367(21)141, (01)00884838099258(21)2590, (01)00884838110564(21)17, (01)00884838103276(21)107, (01)00884838099272(21)342, (01)00884838099258(21)1660, (01)00884838099203(21)38, (01)00884838099241(21)271, (01)00884838099227(21)342, (01)00884838099258(21)2844, (01)00884838085251(21)623, (01)00884838085251(21)629, (01)00884838085251(21)622, (01)00884838099241(21)988, (01)00884838085251(21)129, (01)00884838099227(21)1, (01)00884838099265(21)79, (01)00884838099241(21)481, (01)00884838099241(21)828, (01)00884838099272(21)457, (01)00884838099265(21)103, (01)00884838099258(21)2710, (01)00884838085367(21)365, (01)00884838099258(21)3540, (01)00884838099258(21)3558, (01)00884838099258(21)1523, (01)00884838085251(21)529, (01)00884838085268(21)42, (01)00884838085251(21)144, (01)00884838099227(21)12, (01)00884838099227(21)16, (01)00884838099241(21)860, (01)00884838099227(21)257, (01)00884838099258(21)1434, (01)00884838099210(21)253, (01)00884838099210(21)252, (01)00884838099210(21)251, (01)00884838099210(21)250, (01)00884838099258(21)991, (01)00884838099234(21)56, (01)00884838085251(21)401, (01)00884838085251(21)400, (01)00884838085268(21)639, (01)00884838085251(21)411, (01)00884838103276(21)171, (01)00884838099227(21)330, (01)00884838099227(21)67, (01)00884838110571(21)63, (01)00884838085251(21)259, (01)00884838099241(21)278, (01)00884838085350(21)100, (01)00884838085350(21)101, (01)00884838099241(21)219, (01)00884838085251(21)457, (01)00884838099258(21)569, (01)00884838085268(21)703412, (01)00884838085282(21)48, (01)00884838099272(21)645, (01)00884838085251(21)579, (01)00884838085268(21)1272, (01)00884838099234(21)654, (01)00884838085251(21)926, (01)00884838099241(21)461, (01)00884838099234(21)460, (01)00884838110571(21)147, (01)00884838085268(21)703973, (01)00884838085268(21)704110, (01)00884838099258(21)2816, (01)00884838085268(21)703701, (01)00884838099234(21)601, (01)00884838099234(21)579, (01)00884838085275(21)141, (01)00884838099258(21)2312, (01)00884838099227(21)340, (01)00884838099203(21)48, (01)00884838085275(21)115, (01)00884838110564(21)12, (01)00884838099241(21)549, (01)00884838099241(21)802, (01)00884838099227(21)260, (01)00884838099234(21)57, (01)00884838103276(21)118, (01)00884838099258(21)3563, (01)00884838085251(21)597, (01)00884838099227(21)148, (01)00884838099241(21)66, (01)00884838085251(21)899, (01)00884838099203(21)31, (01)00884838099241(21)552, (01)00884838099241(21)298, (01)00884838099241(21)285, (01)00884838099241(21)497, (01)00884838085251(21)978, (01)00884838085251(21)322, (01)00884838099258(21)4, (01)00884838085251(21)418, (01)00884838085251(21)173, (01)00884838099265(21)214, (01)00884838085251(21)670, (01)00884838085350(21)168, (01)00884838085251(21)637, (01)00884838085251(21)909, (01)00884838099258(21)533, (01)00884838085268(21)703628, (01)00884838085282(21)147, (01)00884838099234(21)86, (01)00884838099203(21)92, (01)00884838085268(21)1153, (01)00884838085268(21)385, (01)00884838099258(21)270, (01)00884838099234(21)665, (01)00884838085251(21)601, (01)00884838099227(21)281, (01)00884838099241(21)962, (01)00884838099258(21)3667, (01)00884838099234(21)628, (01)00884838085282(21)509, (01)00884838085282(21)454, (01)00884838085268(21)791, (01)00884838099227(21)284, (01)00884838099227(21)248, (01)00884838085268(21)1233, (01)00884838099258(21)3557, (01)00884838099241(21)534, (01)00884838099258(21)2296, (01)00884838099227(21)333, (01)00884838110564(21)412, (01)00884838099265(21)395, (01)00884838099272(21)670, (01)00884838099258(21)2107, (01)00884838085251(21)37, (01)00884838099258(21)3766, (01)00884838085282(21)478, (01)00884838099272(21)739, (01)00884838099241(21)1073, (01)00884838099258(21)2204, (01)00884838099272(21)736, (01)00884838085268(21)703955, (01)00884838085251(21)1035, (01)00884838085251(21)1033, (01)00884838085251(21)227, (01)00884838099272(21)1127, (01)00884838085350(21)235, (01)00884838085282(21)262, (01)00884838110564(21)65, (01)00884838099210(21)183, (01)00884838085282(21)491, (01)00884838103276(21)280, (01)00884838099210(21)375, (01)00884838103276(21)21, (01)00884838085268(21)703665, (01)00884838099258(21)3721, (01)00884838099258(21)3424, (01)00884838099272(21)963, (01)00884838085268(21)1367, (01)00884838085282(21)133, (01)00884838110564(21)217, (01)00884838085367(21)471, (01)00884838110564(21)351, (01)00884838110564(21)339, (01)00884838085251(21)1009, (01)00884838110564(21)95, (01)00884838110564(21)337, (01)00884838085367(21)540, (01)00884838110564(21)364, (01)00884838099258(21)1673, (01)00884838110571(21)73, (01)00884838085268(21)703899, (01)00884838103276(21)174, (01)00884838099272(21)527, (01)00884838099272(21)251, (01)00884838110571(21)83, (01)00884838110571(21)149, (01)00884838085268(21)704202, (01)00884838099258(21)2351, (01)00884838099258(21)2356, (01)00884838099258(21)2357, (01)00884838099258(21)3523, (01)00884838099272(21)1115, (01)00884838103276(21)284, (01)00884838110571(21)99, (01)00884838099258(21)1986, (01)00884838085268(21)703828, (01)00884838085268(21)703593, (01)00884838099241(21)613, (01)00884838110564(21)360, (01)00884838099258(21)1665, (01)00884838085268(21)703641, (01)00884838099258(21)1824, (01)00884838110564(21)210, (01)00884838099258(21)1655, (01)00884838099234(21)257, (01)00884838085268(21)703711, (01)00884838085251(21)1008, (01)00884838103276(21)220, (01)00884838099265(21)427, (01)00884838099272(21)1000, (01)00884838099258(21)3263, (01)00884838099210(21)311, (01)00884838099258(21)3716, (01)00884838110571(21)162, (01)00884838099234(21)607, (01)00884838099234(21)608, (01)00884838110564(21)188, (01)00884838099258(21)3317, (01)00884838085268(21)703856, (01)00884838085268(21)703842, (01)00884838085268(21)703651, (01)00884838110564(21)203, (01)00884838099258(21)3612, (01)00884838099258(21)2887, (01)00884838099258(21)3586, (01)00884838099258(21)2061, (01)00884838110564(21)476, (01)00884838099258(21)3718, (01)00884838085268(21)704186, (01)00884838099272(21)965, (01)00884838099258(21)1927, (01)00884838110564(21)419, (01)00884838099258(21)2402, (01)00884838099258(21)2735, (01)00884838099258(21)2862, (01)00884838099258(21)2734, (01)00884838110564(21)295, (01)00884838110564(21)236, (01)00884838110571(21)46, (01)00884838085268(21)704269, (01)00884838085367(21)584, (01)00884838110564(21)308, (01)00884838085367(21)553, (01)00884838099210(21)228, (01)00884838085268(21)704074, (01)00884838110564(21)45, (01)00884838110564(21)44, (01)00884838103276(21)288, (01)00884838110571(21)91, (01)00884838099258(21)2904, (01)00884838103276(21)147, (01)00884838099258(21)1553, (01)00884838099258(21)3109, (01)00884838103276(21)133, (01)00884838099272(21)553, (01)00884838099258(21)2380, (01)00884838085282(21)476, (01)00884838099258(21)3789, (01)00884838085251(21)1047, (01)00884838110564(21)319, (01)00884838099258(21)1413, (01)00884838099234(21)430, (01)00884838099210(21)322, (01)00884838110564(21)307, (01)00884838110564(21)350, (01)00884838099234(21)360, (01)00884838103276(21)252, (01)00884838099241(21)1250, (01)00884838085251(21)984, (01)00884838085350(21)272, (01)00884838099234(21)650, (01)00884838099234(21)651, (01)00884838085268(21)704108, (01)00884838110564(21)304, (01)00884838085268(21)703821, (01)00884838085268(21)704134, (01)00884838099258(21)1127, (01)00884838110564(21)342, (01)00884838085282(21)337, (01)00884838085268(21)704058, (01)00884838085251(21)989, (01)00884838085367(21)602, (01)00884838085268(21)703586, (01)00884838110571(21)142, (01)00884838110571(21)143, (01)00884838110564(21)164, (01)00884838085268(21)704029, (01)00884838099258(21)2123, (01)00884838110564(21)124, (01)00884838099258(21)3437, (01)00884838110564(21)159, (01)00884838110564(21)317, (01)00884838110564(21)318, (01)00884838099258(21)1732, (01)00884838085282(21)275, (01)00884838085367(21)240, (01)00884838110571(21)39, (01)00884838110571(21)40, (01)00884838099272(21)1160, (01)00884838110571(21)130, (01)00884838103276(21)38, (01)00884838085268(21)703979, (01)00884838110564(21)74, (01)00884838099258(21)1312, (01)00884838110564(21)64, (01)00884838103276(21)243, (01)00884838099234(21)591, (01)00884838099234(21)590, (01)00884838110564(21)189, (01)00884838099234(21)595, (01)00884838085268(21)703777, (01)00884838110564(21)71, (01)00884838099258(21)1193, (01)00884838099210(21)189, (01)00884838110564(21)410, (01)00884838085282(21)468, (01)00884838099258(21)3241, (01)00884838099258(21)3240, (01)00884838103276(21)177, (01)00884838103276(21)178, (01)00884838099258(21)1771, (01)00884838110564(21)38, (01)00884838110564(21)387, (01)00884838085251(21)987, (01)00884838099258(21)1448, (01)00884838110564(21)228, (01)00884838110564(21)272, (01)00884838085268(21)703741, (01)00884838110564(21)54, (01)00884838099258(21)3737, (01)00884838110564(21)19, (01)00884838085268(21)704174, (01)00884838085282(21)412, (01)00884838099234(21)352, (01)00884838103276(21)150, (01)00884838085268(21)704041, (01)00884838085251(21)907, (01)00884838110564(21)26, (01)00884838099258(21)2791, (01)00884838099258(21)1788, (01)00884838085367(21)617, (01)00884838110564(21)105, (01)00884838110564(21)344, (01)00884838110571(21)34, (01)00884838099241(21)917, (01)00884838099258(21)1144, (01)00884838110564(21)260, (01)00884838110564(21)261, (01)00884838110564(21)259, (01)00884838085251(21)1081, (01)00884838085268(21)704237, (01)00884838110571(21)145, (01)00884838085268(21)704191, (01)00884838110571(21)166, (01)00884838099272(21)1042, (01)00884838099272(21)699, (01)00884838085282(21)414, (01)00884838085268(21)703650, (01)00884838110571(21)105, (01)00884838110564(21)268, (01)00884838099258(21)2632, (01)00884838085268(21)703646, (01)00884838099258(21)1500, (01)00884838085268(21)703846, (01)00884838110564(21)392, (01)00884838110571(21)127, (01)00884838110564(21)469, (01)00884838099258(21)3436, (01)00884838110564(21)316, (01)00884838085350(21)58, (01)00884838099258(21)3583, (01)00884838099258(21)2334, (01)00884838085282(21)405, (01)00884838085268(21)703945, (01)00884838085268(21)704021, (01)00884838110564(21)80, (01)00884838099241(21)1337, (01)00884838085251(21)1004, (01)00884838085268(21)703817, (01)00884838103276(21)69, (01)00884838110564(21)427, (01)00884838110564(21)265, (01)00884838099258(21)2956, (01)00884838085268(21)703908, (01)00884838085268(21)703909, (01)00884838099272(21)1064, (01)00884838099258(21)3226, (01)00884838099258(21)1513, (01)00884838099234(21)329, (01)00884838099258(21)3633, (01)00884838099258(21)3635, (01)00884838085268(21)704031, (01)00884838085268(21)703726, (01)00884838103276(21)148, (01)00884838110564(21)382, (01)00884838110564(21)70, (01)00884838099258(21)1198, (01)00884838085268(21)704271, (01)00884838099258(21)2660, (01)00884838085268(21)703756, (01)00884838085268(21)703618, (01)00884838099258(21)3517, (01)00884838085367(21)604, (01)00884838099258(21)1819, (01)00884838099258(21)1440, (01)00884838099258(21)1285, (01)00884838099265(21)116, (01)00884838099258(21)2650, (01)00884838099258(21)1178, (01)00884838085268(21)704060, (01)00884838099258(21)1607, (01)00884838099272(21)574, (01)00884838110564(21)425, (01)00884838085251(21)1086, (01)00884838085268(21)704120, (01)00884838085268(21)704017, (01)00884838085282(21)396, (01)00884838099234(21)197, (01)00884838110571(21)119, (01)00884838099258(21)1418, (01)00884838099258(21)2471, (01)00884838099272(21)741, (01)00884838110571(21)156, (01)00884838110564(21)333, (01)00884838099272(21)748, (01)00884838110564(21)190, (01)00884838103276(21)210, (01)00884838099234(21)366, (01)00884838110571(21)84, (01)00884838110571(21)114, (01)00884838085268(21)703816, (01)00884838099258(21)3115, (01)00884838099258(21)1631, (01)00884838099258(21)2115, (01)00884838103276(21)207, (01)00884838103276(21)98, (01)00884838085268(21)703988, (01)00884838103276(21)211, (01)00884838085282(21)247, (01)00884838110564(21)225, (01)00884838085268(21)703573, (01)00884838099258(21)2875, (01)00884838110571(21)141, (01)00884838110564(21)186, (01)00884838099258(21)2666, (01)00884838103276(21)127, (01)00884838085268(21)703800, (01)00884838085268(21)703997, (01)00884838099234(21)345, (01)00884838085282(21)443, (01)00884838099258(21)2212, (01)00884838099258(21)1599, (01)00884838085282(21)310, (01)00884838085367(21)593, (01)00884838085268(21)704073, (01)00884838099258(21)2634, (01)00884838099272(21)740, (01)00884838085268(21)703674, (01)00884838110564(21)180, (01)00884838110564(21)405, (01)00884838099241(21)1270, (01)00884838099234(21)627, (01)00884838110564(21)352, (01)00884838103276(21)222, (01)00884838085350(21)47, (01)00884838085268(21)704291, (01)00884838099272(21)1137, (01)00884838099258(21)2274, (01)00884838099258(21)1389, (01)00884838099272(21)900, (01)00884838099272(21)901, (01)00884838103276(21)142, (01)00884838103276(21)143, (01)00884838103276(21)144, (01)00884838099258(21)2941, (01)00884838099258(21)2940, (01)00884838085282(21)406, (01)00884838085251(21)1085, (01)00884838085251(21)1084, (01)00884838099258(21)1457, (01)00884838110571(21)43, (01)00884838110571(21)44, (01)00884838085268(21)1343, (01)00884838099272(21)887, (01)00884838110564(21)426, (01)00884838085268(21)703673, (01)00884838099210(21)333, (01)00884838099210(21)297, (01)00884838085268(21)703591, (01)00884838085367(21)589, (01)00884838085251(21)1048, (01)00884838099258(21)1489, (01)00884838099241(21)1335, (01)00884838085251(21)1059, (01)00884838110564(21)281, (01)00884838099258(21)2085, (01)00884838099258(21)1624, (01)00884838085268(21)704002, (01)00884838110564(21)202, (01)00884838085268(21)703735, (01)00884838085268(21)703733, (01)00884838103276(21)198, (01)00884838110564(21)349, (01)00884838099258(21)1497, (01)00884838099272(21)632, (01)00884838110564(21)453, (01)00884838110564(21)149, (01)00884838099234(21)302, (01)00884838103276(21)71, (01)00884838110571(21)106, (01)00884838085282(21)517, (01)00884838099258(21)1946, (01)00884838099258(21)3551, (01)00884838110571(21)76, (01)00884838099227(21)199, (01)00884838085268(21)703940, (01)00884838085251(21)1034, (01)00884838110564(21)117, (01)00884838110564(21)388, (01)00884838099258(21)1971, (01)00884838085251(21)1080, (01)00884838085268(21)703757, (01)00884838085282(21)394, (01)00884838110564(21)52, (01)00884838110564(21)500, (01)00884838099234(21)166, (01)00884838103276(21)146, (01)00884838085268(21)703751, (01)00884838103276(21)241, (01)00884838099258(21)1990, (01)00884838085282(21)432, (01)00884838099258(21)3687, (01)00884838099258(21)3002, (01)00884838110564(21)175, (01)00884838103276(21)206, (01)00884838099258(21)2796, (01)00884838099241(21)611, (01)00884838099258(21)1445, (01)00884838099258(21)2024, (01)00884838085268(21)703883, (01)00884838099241(21)1122, (01)00884838099258(21)3056, (01)00884838099258(21)3645, (01)00884838110564(21)187, (01)00884838110571(21)144, (01)00884838110564(21)162, (01)00884838110564(21)270, (01)00884838103276(21)152, (01)00884838110571(21)28, (01)00884838099258(21)2225, (01)00884838099258(21)2702, (01)00884838085268(21)704139, (01)00884838110564(21)11, (01)00884838085268(21)704157, (01)00884838110564(21)219, (01)00884838099210(21)190, (01)00884838099258(21)1674, (01)00884838099234(21)664, (01)00884838103276(21)279, (01)00884838085268(21)704293, (01)00884838103276(21)262, (01)00884838103276(21)247, (01)00884838110571(21)95, (01)00884838099258(21)2282, (01)00884838099234(21)432, (01)00884838110571(21)115, (01)00884838085268(21)703877, (01)00884838085268(21)704279, (01)00884838099258(21)3102, (01)00884838099258(21)3140, (01)00884838099258(21)2408, (01)00884838099258(21)2822, (01)00884838099258(21)1941, (01)00884838103276(21)157, (01)00884838110564(21)371, (01)00884838110564(21)269, (01)00884838085367(21)526, (01)00884838099241(21)1124, (01)00884838099258(21)2607, (01)00884838099258(21)1825, (01)00884838099258(21)3128, (01)00884838099258(21)3552, (01)00884838099241(21)1274, (01)00884838099272(21)1112, (01)00884838099241(21)1300, (01)00884838099258(21)2387, (01)00884838099258(21)2388, (01)00884838099258(21)1520, (01)00884838099241(21)692, (01)00884838085251(21)875, (01)00884838099258(21)1696, (01)00884838085268(21)704280, (01)00884838110564(21)157, (01)00884838099258(21)2216, (01)00884838110571(21)131, (01)00884838099258(21)1708, (01)00884838099258(21)2894, (01)00884838085268(21)704024, (01)00884838099258(21)3626, (01)00884838099272(21)1156, (01)00884838103276(21)40, (01)00884838099272(21)737, (01)00884838099234(21)361, (01)00884838085268(21)703912, (01)00884838085268(21)703911, (01)00884838085251(21)1026, (01)00884838099258(21)2019, (01)00884838099258(21)2020, (01)00884838099258(21)3103, (01)00884838110564(21)290, (01)00884838085268(21)703605, (01)00884838085268(21)703797, (01)00884838110564(21)102, (01)00884838099258(21)3781, (01)00884838099241(21)766, (01)00884838110571(21)9, (01)00884838099234(21)305, (01)00884838085268(21)690, (01)00884838099258(21)3054, (01)00884838099234(21)542, (01)00884838099272(21)1131, (01)00884838099258(21)1729, (01)00884838110564(21)244, (01)00884838085268(21)703969, (01)00884838110564(21)94, (01)00884838099258(21)3765, (01)00884838099258(21)3599, (01)00884838099234(21)313, (01)00884838085268(21)704022, (01)00884838099258(21)3431, (01)00884838099258(21)3080, (01)00884838099258(21)2821, (01)00884838110564(21)389, (01)00884838085282(21)504, (01)00884838110564(21)110, (01)00884838103276(21)17, (01)00884838103276(21)204, (01)00884838085268(21)704039, (01)00884838085268(21)703471, (01)00884838085251(21)1077, (01)00884838085268(21)704095, (01)00884838099258(21)3601, (01)00884838099258(21)2407, (01)00884838085268(21)703768, (01)00884838085268(21)703779, (01)00884838110564(21)442, (01)00884838099258(21)1548, (01)00884838099265(21)439, (01)00884838103276(21)260, (01)00884838099258(21)2116, (01)00884838099272(21)1148, (01)00884838085268(21)703663, (01)00884838085268(21)704138, (01)00884838110564(21)380, (01)00884838110564(21)23, (01)00884838103276(21)275, (01)00884838099258(21)1979, (01)00884838085282(21)457, (01)00884838110571(21)14, (01)00884838110571(21)13, (01)00884838103276(21)187, (01)00884838085268(21)703844, (01)00884838099258(21)2774, (01)00884838099258(21)2773, (01)00884838099258(21)2492, (01)00884838099241(21)996, (01)00884838099258(21)2491, (01)00884838099241(21)997, (01)00884838085367(21)559, (01)00884838085268(21)703604, (01)00884838099258(21)2664, (01)00884838110571(21)101, (01)00884838110571(21)165, (01)00884838099258(21)1202, (01)00884838099272(21)1111, (01)00884838110571(21)172, (01)00884838085268(21)703830, (01)00884838110564(21)9, (01)00884838085268(21)704247, (01)00884838110564(21)208, (01)00884838110564(21)395, (01)00884838103276(21)208, (01)00884838085268(21)703707, (01)00884838110564(21)209, (01)00884838085268(21)703731, (01)00884838085268(21)703734, (01)00884838099258(21)2372, (01)00884838085268(21)704286, (01)00884838099258(21)1836, (01)00884838099258(21)2213, (01)00884838085268(21)704008, (01)00884838110564(21)458, (01)00884838099234(21)620, (01)00884838099258(21)3041, (01)00884838085268(21)704221, (01)00884838099258(21)1416, (01)00884838110564(21)168, (01)00884838099241(21)900, (01)00884838099258(21)2333, (01)00884838099241(21)652, (01)00884838099258(21)1754, (01)00884838099241(21)614, (01)008848 38099234(21)220, (01)00884838110564(21)204, (01)00884838103276(21)46, (01)00884838085268(21)703799, (01)00884838085251(21)716, (01)00884838085251(21)717, (01)00884838099258(21)2040, (01)00884838110564(21)114, (01)00884838099258(21)1658, (01)00884838099258(21)2169, (01)00884838099234(21)325, (01)00884838099258(21)1756, (01)00884838110564(21)466, (01)00884838110564(21)463, (01)00884838099241(21)517, (01)00884838099258(21)1152, (01)00884838103276(21)273, (01)00884838110564(21)447, (01)00884838110571(21)16, (01)00884838110571(21)17, (01)00884838099272(21)907, (01)00884838110564(21)253, (01)00884838099258(21)3611, (01)00884838099258(21)3430, (01)00884838099272(21)662, (01)00884838110571(21)128, (01)00884838110571(21)51, (01)00884838099258(21)3058, (01)00884838099234(21)606, (01)00884838099272(21)825, (01)00884838099258(21)2726, (01)00884838110564(21)63, (01)00884838085268(21)703724, (01)00884838085282(21)426, (01)00884838085268(21)703576, (01)00884838085282(21)304, (01)00884838099258(21)3421, (01)00884838103276(21)274, (01)00884838099258(21)3663, (01)00884838099241(21)1172, (01)00884838099258(21)3434, (01)00884838099258(21)3254, (01)00884838110571(21)135, (01)00884838110571(21)136, (01)00884838099258(21)2762, (01)00884838099241(21)990, (01)00884838099241(21)991, (01)00884838099258(21)2761, (01)00884838099258(21)2763, (01)00884838099258(21)2808, (01)00884838099258(21)2807, (01)00884838099258(21)1292, (01)00884838110564(21)310, (01)00884838103276(21)271, (01)00884838099234(21)545, (01)00884838099241(21)1176, (01)00884838085268(21)704009, (01)00884838085282(21)469, (01)00884838099234(21)405, (01)00884838110564(21)35, (01)00884838103276(21)235, (01)00884838099210(21)313, (01)00884838110564(21)155, (01)00884838099241(21)1310, (01)00884838110571(21)79, (01)00884838085251(21)1015, (01)00884838099258(21)1384, (01)00884838099258(21)2173, (01)00884838110571(21)56, (01)00884838085282(21)340, (01)00884838099258(21)1822, (01)00884838099258(21)1641, (01)00884838099234(21)764, (01)00884838085268(21)704069, (01)00884838085268(21)703840, (01)00884838099258(21)3110, (01)00884838099241(21)1331, (01)00884838099272(21)1075, (01)00884838110564(21)485, (01)00884838099210(21)169, (01)00884838110564(21)199, (01)00884838110564(21)201, (01)00884838099258(21)1291, (01)00884838099241(21)1253, (01)00884838099241(21)1254, (01)00884838103276(21)234, (01)00884838110564(21)161, (01)00884838099234(21)514, (01)00884838099258(21)1442, (01)00884838110564(21)468, (01)00884838110564(21)76, (01)00884838085268(21)703708, (01)00884838099258(21)3186, (01)00884838103276(21)91, (01)00884838099258(21)3742, (01)00884838099241(21)957, (01)00884838099272(21)744, (01)00884838099258(21)2624, (01)00884838099258(21)2545, (01)00884838103276(21)162, (01)00884838085268(21)704080, (01)00884838085268(21)703998, (01)00884838099258(21)2697, (01)00884838110564(21)126, (01)00884838110571(21)163, (01)00884838085268(21)704040, (01)00884838085268(21)704012, (01)00884838103276(21)277, (01)00884838099258(21)3683, (01)00884838085268(21)703603, (01)00884838099258(21)1532, (01)00884838099265(21)107, (01)00884838099258(21)2695, (01)00884838110564(21)211, (01)00884838099272(21)877, (01)00884838085268(21)704130, (01)00884838099234(21)433, (01)00884838085282(21)393, (01)00884838085268(21)703786, (01)00884838103276(21)97, (01)00884838099258(21)2118, (01)00884838099258(21)2823, (01)00884838099258(21)2825, (01)00884838099258(21)2337, (01)00884838099258(21)2338, (01)00884838085268(21)704131, (01)00884838103276(21)153, (01)00884838099234(21)342, (01)00884838099234(21)341, (01)00884838099210(21)219, (01)00884838085268(21)704078, (01)00884838085268(21)704077, (01)00884838099258(21)2290, (01)00884838110571(21)30, (01)00884838099234(21)184, (01)00884838085282(21)343, (01)00884838110564(21)287, (01)00884838099258(21)1749, (01)00884838110564(21)365, (01)00884838085268(21)703630, (01)00884838085268(21)704066, (01)00884838099258(21)2109, (01)00884838085268(21)703983, (01)00884838110564(21)207, (01)00884838099258(21)1439, (01)00884838110571(21)125, (01)00884838110571(21)61, (01)00884838110571(21)121, (01)00884838099272(21)549, (01)00884838085268(21)703507, (01)00884838110564(21)377, (01)00884838103276(21)276, (01)00884838103276(21)245, (01)00884838099272(21)1161, (01)00884838099241(21)1320, (01)00884838099272(21)1134, (01)00884838110571(21)97, (01)00884838110571(21)169, (01)00884838110571(21)41, (01)00884838085251(21)1045, (01)00884838103276(21)106, (01)00884838110564(21)85, (01)00884838085268(21)704001, (01)00884838099258(21)2639, (01)00884838103276(21)173, (01)00884838099258(21)1130, (01)00884838099258(21)2766, (01)00884838085268(21)703645, (01)00884838110564(21)406, (01)00884838085268(21)703919, (01)00884838085268(21)704196, (01)00884838085268(21)704197, (01)00884838085268(21)704198, (01)00884838099258(21)1992, (01)00884838099258(21)1790, (01)00884838099258(21)3773, (01)00884838099272(21)1023, (01)00884838099258(21)1195, (01)00884838085367(21)612, (01)00884838099258(21)2249, (01)00884838110571(21)20, (01)00884838099241(21)781, (01)00884838099241(21)780, (01)00884838110571(21)33, (01)00884838085251(21)947, (01)00884838085251(21)948, (01)00884838085251(21)946, (01)00884838085251(21)945, (01)00884838085251(21)942, (01)00884838085251(21)940, (01)00884838099258(21)2890, (01)00884838110564(21)321, (01)00884838110564(21)227, (01)00884838099234(21)425, (01)00884838099234(21)424, (01)00884838085268(21)704088, (01)00884838103276(21)258, (01)00884838085268(21)703616, (01)00884838110564(21)451, (01)00884838085268(21)704201, (01)00884838110564(21)396, (01)00884838099258(21)2469, (01)00884838099258(21)2065, (01)00884838099258(21)2067, (01)00884838099272(21)648, (01)00884838110564(21)276, (01)00884838099241(21)1304, (01)00884838103276(21)120, (01)00884838085268(21)704133, (01)00884838110564(21)327, (01)00884838099258(21)2826, (01)00884838099241(21)688, (01)00884838099241(21)687, (01)00884838099258(21)1752, (01)00884838099241(21)686, (01)00884838099258(21)1755, (01)00884838099241(21)689, (01)00884838110564(21)226, (01)00884838099234(21)426, (01)00884838099258(21)2677, (01)00884838110564(21)169, (01)00884838099241(21)1288, (01)00884838099241(21)1287, (01)00884838099234(21)497, (01)00884838085268(21)703796, (01)00884838099241(21)1171, (01)00884838110564(21)422, (01)00884838110564(21)156, (01)00884838110564(21)258, (01)00884838085268(21)703863, (01)00884838085268(21)703860, (01)00884838085268(21)703859, (01)00884838085268(21)703857, (01)00884838085268(21)703858, (01)00884838110564(21)170, (01)00884838110571(21)132, (01)00884838099241(21)1332, (01)00884838103276(21)45, (01)00884838085268(21)703780, (01)00884838085367(21)450, (01)00884838085268(21)704166, (01)00884838085268(21)703640, (01)00884838085268(21)703642, (01)00884838085268(21)704093, (01)00884838099234(21)630, (01)00884838085268(21)704254, (01)00884838099241(21)634, (01)00884838085268(21)704259, (01)00884838099241(21)1198, (01)00884838085268(21)703680, (01)00884838103276(21)92, (01)00884838099234(21)611, (01)00884838085268(21)703878, (01)00884838085251(21)1052, (01)00884838085268(21)703901, (01)00884838110564(21)407, (01)00884838099258(21)2382, (01)00884838110564(21)415, (01)00884838099258(21)3798, (01)00884838085367(21)556, (01)00884838110564(21)294, (01)00884838085268(21)704020, (01)00884838085268(21)703884, (01)00884838099258(21)2785, (01)00884838099241(21)979, (01)00884838085268(21)703958, (01)00884838110571(21)77, (01)00884838103276(21)226, (01)00884838099258(21)1438, (01)00884838099258(21)3689, (01)00884838110564(21)101, (01)00884838099272(21)831, (01)00884838110571(21)68, (01)00884838099258(21)1996, (01)00884838099258(21)1844, (01)00884838099258(21)1848, (01)00884838085268(21)704091, (01)00884838099258(21)1989, (01)00884838110564(21)229, (01)00884838085367(21)557, (01)00884838085268(21)1376, (01)00884838103276(21)254, (01)00884838099234(21)258, (01)00884838099258(21)1841, (01)00884838103276(21)213, (01)00884838110564(21)146, (01)00884838085268(21)704263, (01)00884838110564(21)434, (01)00884838110564(21)14, (01)00884838110571(21)175, (01)00884838099258(21)2059, (01)00884838103276(21)224, (01)00884838110571(21)65, (01)00884838099234(21)183, (01)00884838085268(21)977, (01)00884838085268(21)1093, (01)00884838099258(21)1121, (01)00884838099258(21)3625, (01)00884838099258(21)3731, (01)00884838099234(21)195, (01)00884838099265(21)477, (01)00884838099265(21)133, (01)00884838099258(21)772, (01)00884838099258(21)773, (01)00884838103276(21)95, (01)00884838085268(21)703975, (01)00884838099241(21)1334, (01)00884838085251(21)971, (01)00884838099241(21)843, (01)00884838110564(21)88, (01)00884838099258(21)1950, (01)00884838110564(21)291, (01)00884838110564(21)99, (01)00884838110571(21)158, (01)00884838085367(21)574, (01)00884838099258(21)2295, (01)00884838110571(21)98, (01)00884838110571(21)67, (01)00884838099234(21)314, (01)00884838099258(21)2381, (01)00884838110564(21)89, (01)00884838099234(21)555, (01)00884838110564(21)273, (01)00884838085268(21)703688, (01)00884838099258(21)3432, (01)00884838085268(21)704113, (01)00884838085268(21)704090, (01)00884838085268(21)704147, (01)00884838110564(21)340, (01)00884838103276(21)54, (01)00884838099234(21)626, (01)00884838085268(21)703571, (01)00884838085268(21)703657, (01)00884838099258(21)2683, (01)00884838085251(21)1063, (01)00884838085251(21)1062, (01)00884838085282(21)458, (01)00884838085251(21)994, (01)00884838085282(21)322, (01)00884838110564(21)68, (01)00884838099241(21)651, (01)00884838099258(21)3703, (01)00884838099258(21)3710, (01)00884838099272(21)1135, (01)00884838099241(21)678, (01)00884838085268(21)703547, (01)00884838099258(21)2092, (01)00884838110564(21)151, (01)00884838099258(21)2023, (01)00884838099258(21)2053, (01)00884838099258(21)1166, (01)00884838085268(21)703771, (01)00884838085251(21)701, (01)00884838099258(21)1070, (01)00884838110564(21)141, (01)00884838085282(21)444, (01)00884838110564(21)221, (01)00884838085268(21)703703, (01)00884838099258(21)2114, (01)00884838099258(21)3743, (01)00884838099258(21)3744, (01)00884838085282(21)508, (01)00884838110564(21)33, (01)00884838099258(21)2200, (01)00884838103276(21)126, (01)00884838085268(21)703887, (01)00884838099258(21)3783, (01)00884838099258(21)3686, (01)00884838099241(21)1317, (01)00884838085251(21)993, (01)00884838085268(21)703781, (01)00884838099258(21)3787, (01)00884838110564(21)41, (01)00884838099258(21)1849, (01)00884838085268(21)703923, (01)00884838099258(21)2428, (01)00884838099258(21)1503, (01)00884838099234(21)282, (01)00884838099272(21)751, (01)00884838085268(21)704145, (01)00884838085268(21)704215, (01)00884838085268(21)704111, (01)00884838110571(21)11, (01)00884838085268(21)703805, (01)00884838110571(21)88, (01)00884838099258(21)3695, (01)00884838099258(21)1846, (01)00884838099258(21)1847, (01)00884838099258(21)3024, (01)00884838099258(21)3585, (01)00884838085268(21)704217, (01)00884838099272(21)1061, (01)00884838085268(21)704124, (01)00884838085268(21)703585, (01)00884838085282(21)445, (01)00884838085268(21)704094, (01)00884838099258(21)2497, (01)00884838085268(21)703759, (01)00884838099210(21)247, (01)00884838110571(21)155, (01)00884838099258(21)2313, (01)00884838085268(21)703639, (01)00884838085282(21)395, (01)00884838099258(21)2470, (01)00884838085268(21)703647, (01)00884838110564(21)289, (01)00884838110564(21)59, (01)00884838110571(21)126, (01)00884838110564(21)246, (01)00884838110571(21)71, (01)00884838099258(21)1983, (01)00884838103276(21)216, (01)00884838085268(21)703627, (01)00884838099234(21)402, (01)00884838085268(21)704064, (01)00884838099234(21)490, (01)00884838099258(21)2379, (01)00884838099258(21)2934, (01)00884838110564(21)83, (01)00884838099258(21)3780, (01)00884838085268(21)704050, (01)00884838085251(21)620, (01)00884838110564(21)284, (01)00884838110564(21)10, (01)00884838110564(21)363, (01)00884838085268(21)704255, (01)00884838110564(21)177, (01)00884838099234(21)439, (01)00884838099234(21)421, (01)00884838099272(21)1036, (01)00884838110564(21)40, (01)00884838110564(21)43, (01)00884838110564(21)39, (01)00884838085268(21)703803, (01)00884838110564(21)326, (01)00884838110564(21)331, (01)00884838103276(21)19, (01)00884838099241(21)782, (01)00884838099258(21)2066, (01)00884838110571(21)94, (01)00884838099258(21)1972, (01)00884838099258(21)2755, (01)00884838110564(21)57, (01)00884838085251(21)1075, (01)00884838085367(21)603, (01)00884838085268(21)704175, (01)00884838099234(21)440, (01)00884838085268(21)703697, (01)00884838099234(21)446, (01)00884838085251(21)1046, (01)00884838099258(21)3494, (01)00884838110571(21)148, (01)00884838085367(21)601, (01)00884838085268(21)704181, (01)00884838085268(21)703891, (01)00884838085268(21)703894, (01)00884838085268(21)703895, (01)00884838085268(21)1263, (01)00884838085268(21)704180, (01)00884838110564(21)212, (01)00884838110571(21)92, (01)00884838085268(21)703918, (01)00884838085275(21)137, (01)00884838099234(21)338, (01)00884838085268(21)704007, (01)00884838085268(21)703921, (01)00884838085268(21)703789, (01)00884838110571(21)123, (01)00884838085268(21)704075, (01)00884838110564(21)122, (01)00884838103276(21)215, (01)00884838099258(21)1630, (01)00884838085268(21)703827, (01)00884838099234(21)472, (01)00884838110564(21)178, (01)00884838099258(21)2543, (01)00884838085282(21)385, (01)00884838085268(21)703587, (01)00884838085251(21)970, (01)00884838099258(21)2292, (01)00884838085282(21)505, (01)00884838110564(21)367, (01)00884838099258(21)3566, (01)00884838099258(21)2992, (01)00884838099258(21)1746, (01)00884838099272(21)1104, (01)00884838099258(21)3402, (01)00884838099258(21)3114, (01)00884838099272(21)830, (01)00884838099258(21)1675, (01)00884838103276(21)93, (01)00884838103276(21)24, (01)00884838085268(21)704056, (01)00884838103276(21)129, (01)00884838103276(21)74, (01)00884838110564(21)323, (01)00884838099258(21)2256, (01)00884838110571(21)27, (01)00884838085268(21)704122, (01)00884838110564(21)69, (01)00884838110564(21)191, (01)00884838103276(21)219, (01)00884838110564(21)50, (01)00884838085268(21)703876, (01)00884838110571(21)24, (01)00884838099234(21)662, (01)00884838110564(21)13, (01)00884838085268(21)703989, (01)00884838099234(21)689, (01)00884838110571(21)139, (01)00884838110564(21)61, (01)00884838085268(21)704046, (01)00884838085268(21)703950, (01)00884838085282(21)472, (01)00884838110564(21)194, (01)00884838099272(21)860, (01)00884838085268(21)703679, (01)00884838103276(21)88, (01)00884838110564(21)200, (01)00884838085268(21)703924, (01)00884838110564(21)250, (01)00884838110564(21)251, (01)00884838099258(21)1843, (01)00884838110564(21)245, (01)00884838085251(21)985, (01)00884838085251(21)798, (01)00884838085367(21)609, (01)00884838085282(21)514, (01)00884838085367(21)616, (01)00884838099258(21)2158, (01)00884838085268(21)704223, (01)00884838110564(21)34, (01)00884838099241(21)987, (01)00884838099258(21)2798, (01)00884838085268(21)703851, (01)00884838085268(21)704051, (01)00884838085268(21)703934, (01)00884838099258(21)1135, (01)00884838099258(21)1415, (01)00884838099258(21)2113, (01)00884838099241(21)825, (01)00884838099241(21)798, (01)00884838103276(21)86, (01)00884838103276(21)85, (01)00884838099258(21)1293, (01)00884838099258(21)2117, (01)00884838085251(21)1074, (01)00884838110564(21)381, (01)00884838103276(21)170, (01)00884838085268(21)1287, (01)00884838085268(21)704216, (01)00884838099258(21)1390, (01)00884838099258(21)2824, (01)00884838085268(21)704267, (01)00884838103276(21)55, (01)00884838099258(21)2701, (01)00884838099203(21)78, (01)00884838099258(21)3702, (01)00884838099258(21)1123, (01)00884838099258(21)2236, (01)00884838110564(21)132, (01)00884838085268(21)1318, (01)00884838099258(21)2815, (01)00884838099258(21)2794, (01)00884838110564(21)404, (01)00884838110564(21)325, (01)00884838099272(21)798, (01)00884838085268(21)704177, (01)00884838085268(21)703825, (01)00884838099258(21)3693, (01)00884838085282(21)333, (01)00884838099210(21)282, (01)00884838110564(21)32, (01)00884838110564(21)205, (01)00884838103276(21)102, (01)00884838110564(21)82, (01)00884838099234(21)633, (01)00884838099258(21)1744, (01)00884838099258(21)3791, (01)00884838099258(21)3243, (01)00884838099234(21)320, (01)00884838099258(21)2461, (01)00884838110571(21)29, (01)00884838099258(21)2110, (01)00884838099241(21)797, (01)00884838085268(21)704115, (01)00884838099241(21)1302, (01)00884838110571(21)38, (01)00884838085282(21)236, (01)00884838103276(21)125, (01)00884838103276(21)124, (01)00884838103276(21)123, (01)00884838099241(21)1128, (01)00884838085268(21)704044, (01)00884838110571(21)107, (01)00884838099241(21)604, (01)00884838099258(21)2610, (01)00884838085251(21)710, (01)00884838110571(21)122, (01)00884838110571(21)25, (01)00884838110564(21)332, (01)00884838099258(21)1818, (01)00884838110571(21)10, (01)00884838099258(21)2041, (01)00884838099234(21)281, (01)00884838099272(21)603, (01)00884838085268(21)704038, (01)00884838103276(21)151, (01)00884838085282(21)471, (01)00884838085282(21)401, (01)00884838085282(21)234, (01)00884838110564(21)222, (01)00884838103276(21)39, (01)00884838085268(21)703809, (01)00884838110564(21)311, (01)00884838110564(21)312, (01)00884838110564(21)174, (01)00884838110564(21)138, (01)00884838099210(21)217, (01)00884838099272(21)577, (01)00884838110564(21)121, (01)00884838099272(21)938, (01)00884838085268(21)703836, (01)00884838085251(21)1011, (01)00884838085268(21)703932, (01)00884838110564(21)134, (01)00884838099258(21)2091, (01)00884838099258(21)2304, (01)00884838099258(21)3609, (01)00884838099258(21)2393, (01)00884838085282(21)503, (01)00884838103276(21)47, (01)00884838085367(21)597, (01)00884838110564(21)300, (01)00884838099258(21)2694, (01)00884838099210(21)143, (01)00884838099258(21)1833, (01)00884838099234(21)480, (01)00884838099258(21)2090, (01)00884838110571(21)93, (01)00884838103276(21)101, (01)00884838085268(21)703661, (01)00884838099210(21)349, (01)00884838099258(21)2463, (01)00884838110564(21)90, (01)00884838085268(21)703676, (01)00884838099272(21)886, (01)00884838099272(21)1066, (01)00884838110564(21)96, (01)00884838099241(21)1322, (01)00884838099241(21)1321, (01)00884838099258(21)2601, (01)00884838099258(21)2602, (01)00884838085251(21)1065, (01)00884838085251(21)1066, (01)00884838085251(21)1064, (01)00884838085251(21)1067, (01)00884838085268(21)704152, (01)00884838110571(21)75, (01)00884838110564(21)231, (01)00884838099203(21)186, (01)00884838085268(21)703831, (01)00884838099234(21)631, (01)00884838099258(21)2468, (01)00884838103276(21)209, (01)00884838099210(21)225, (01)00884838085268(21)704132, (01)00884838099272(21)773, (01)00884838099272(21)939, (01)00884838085268(21)703548, (01)00884838099234(21)364, (01)00884838099272(21)1128, (01)00884838103276(21)41, (01)00884838103276(21)42, (01)00884838085268(21)704200, (01)00884838085268(21)703880, (01)00884838085268(21)703666, (01)00884838085268(21)703696, (01)00884838110564(21)123, (01)00884838085268(21)703730, (01)00884838110564(21)424, (01)00884838085268(21)703837, (01)00884838099258(21)2898, (01)00884838099258(21)2966, (01)00884838099241(21)935, (01)00884838099234(21)581, (01)00884838099258(21)1437, (01)00884838103276(21)53, (01)00884838085268(21)1167, (01)00884838103276(21)250, (01)00884838085268(21)704270, (01)00884838099258(21)3439, (01)00884838099258(21)2147, (01)00884838085268(21)703798, (01)00884838099258(21)1419, (01)00884838085268(21)703818, (01)00884838085282(21)407, (01)00884838085282(21)470, (01)00884838099258(21)2136, (01)00884838099234(21)182, (01)00884838110564(21)298, (01)00884838099241(21)838, (01)00884838110564(21)264, (01)00884838099234(21)448, (01)00884838099258(21)3116, (01)00884838085275(21)73, (01)00884838099258(21)1858, (01)00884838099258(21)1945, (01)00884838099272(21)1065, (01)00884838110564(21)172, (01)00884838103276(21)239, (01)00884838099258(21)2551, (01)00884838103276(21)248, (01)00884838085367(21)605, (01)00884838110571(21)113, (01)00884838099258(21)1947, (01)00884838099234(21)283, (01)00884838099234(21)291, (01)00884838110564(21)160, (01)00884838085268(21)704096, (01)00884838085268(21)704151, (01)00884838085282(21)498, (01)00884838099272(21)1038, (01)00884838099258(21)1423, (01)00884838099210(21)162, (01)00884838085251(21)1000, (01)00884838085268(21)703850, (01)00884838099258(21)3799, (01)00884838099258(21)2205, (01)00884838085251(21)1056, (01)00884838099241(21)1218, (01)00884838099227(21)183, (01)00884838110564(21)335, (01)00884838085251(21)992, (01)00884838099258(21)1770, (01)00884838103276(21)108, (01)00884838099272(21)716, (01)00884838099258(21)1944, (01)00884838085268(21)703725, (01)00884838085268(21)704178, (01)00884838099258(21)1361, (01)00884838085268(21)704161, (01)00884838099258(21)3630, (01)00884838099258(21)2897, (01)00884838110564(21)75, (01)00884838085268(21)703871, (01)00884838099258(21)2942, (01)00884838110564(21)91, (01)00884838099241(21)1225, (01)00884838099258(21)3260, (01)00884838110564(21)374, (01)00884838085251(21)663, (01)00884838085268(21)703737, (01)00884838099265(21)313, (01)00884838099258(21)3705, (01)00884838099258(21)3704, (01)00884838099272(21)1133, (01)00884838099258(21)3706, (01)00884838085268(21)704187, (01)00884838099258(21)1417, (01)00884838110564(21)292, (01)00884838099258(21)2179, (01)00884838110571(21)18, (01)00884838099234(21)473, (01)00884838110564(21)369, (01)00884838099241(21)811, (01)00884838085282(21)413, (01)00884838085268(21)704014, (01)00884838085282(21)7, (01)00884838099234(21)321, (01)00884838085268(21)703977, (01)00884838110571(21)104, (01)00884838099258(21)1857, (01)00884838085268(21)704082, (01)00884838085268(21)704089, (01)00884838099258(21)1441, (01)00884838085268(21)703888, (01)00884838085282(21)440, (01)00884838099272(21)979, (01)00884838099258(21)1748, (01)00884838099258(21)1987, (01)00884838110564(21)414, (01)00884838099234(21)476, (01)00884838085268(21)703928, (01)00884838085268(21)703930, (01)00884838085268(21)703929, (01)00884838085367(21)575, (01)00884838085268(21)703931, (01)00884838099258(21)2318, (01)00884838099234(21)681, (01)00884838099258(21)1842, (01)00884838103276(21)119, (01)00884838110564(21)501, (01)00884838110564(21)223, (01)00884838085268(21)703974, (01)00884838103276(21)205, (01)00884838110564(21)18, (01)00884838099258(21)2922, (01)00884838099258(21)1820, (01)00884838099234(21)409, (01)00884838085268(21)703881, (01)00884838085367(21)516, (01)00884838085268(21)704252, (01)00884838085367(21)624, (01)00884838085282(21)463, (01)00884838085268(21)703941, (01)00884838103276(21)197, (01)00884838099258(21)2392, (01)00884838085268(21)703996, (01)00884838085268(21)703917, (01)00884838103276(21)82, (01)00884838099234(21)535, (01)00884838085268(21)703667, (01)00884838085268(21)703668, (01)00884838099241(21)1328, (01)00884838110571(21)117, (01)00884838099258(21)1531, (01)00884838085268(21)704153, (01)00884838110564(21)87, (01)00884838110564(21)86, (01)00884838103276(21)33, (01)00884838085367(21)599, (01)00884838103276(21)20, (01)00884838099258(21)3059, (01)00884838110571(21)102, (01)00884838099241(21)762, (01)00884838110564(21)418, (01)00884838110571(21)72, (01)00884838110564(21)49, (01)00884838099258(21)2516, (01)00884838085251(21)1057, (01)00884838099258(21)2093, (01)00884838110564(21)484, (01)00884838085367(21)563, (01)00884838085268(21)704169, (01)00884838085282(21)330, (01)00884838085268(21)703617, (01)00884838085268(21)704219, (01)00884838085268(21)704275, (01)00884838099258(21)1464, (01)00884838085268(21)703625, (01)00884838085282(21)475, (01)00884838085268(21)703660, (01)00884838110564(21)385, (01)00884838085268(21)703949, (01)00884838085268(21)704121, (01)00884838085275(21)150, (01)00884838110564(21)166, (01)00884838110564(21)72, (01)00884838110564(21)274, (01)00884838099258(21)3203, (01)00884838099258(21)1867, (01)00884838103276(21)231, (01)00884838085268(21)704104, (01)00884838099258(21)2095, (01)00884838099241(21)758, (01)00884838085268(21)704128, (01)00884838085282(21)473, (01)00884838085268(21)704065, (01)00884838099234(21)376, (01)00884838085268(21)704192, (01)00884838099272(21)672, (01)00884838085282(21)501, (01)00884838099258(21)2411, (01)00884838099258(21)1191, (01)00884838099241(21)1330, (01)00884838099241(21)1057, (01)00884838099258(21)2893, (01)00884838110564(21)47, (01)00884838110564(21)454, (01)00884838085282(21)308, (01)00884838099241(21)763, (01)00884838099241(21)764, (01)00884838099241(21)765, (01)00884838099241(21)767, (01)00884838110564(21)242, (01)00884838085268(21)1228, (01)00884838099234(21)519, (01)00884838085268(21)704116, (01)00884838085282(21)225, (01)00884838085268(21)704102, (01)00884838085268(21)703898, (01)00884838110571(21)21, (01)00884838099258(21)2094, (01)00884838099258(21)2410, (01)00884838099258(21)2412, (01)00884838099265(21)346, (01)00884838110564(21)324, (01)00884838110564(21)144, (01)00884838110564(21)486, (01)00884838085268(21)703636, (01)00884838099210(21)185, (01)00884838099258(21)2012, (01)00884838085268(21)703843, (01)00884838085282(21)483, (01)00884838085268(21)704264, (01)00884838110564(21)480, (01)00884838085268(21)703691, (01)00884838085268(21)703692, (01)00884838085282(21)474, (01)00884838099258(21)3694, (01)00884838085268(21)703664, (01)00884838085268(21)703555, (01)00884838110564(21)355, (01)00884838085268(21)703601, (01)00884838103276(21)63, (01)00884838085268(21)703633, (01)00884838099258(21)3429, (01)00884838110564(21)158, (01)00884838085268(21)704218, (01)00884838110564(21)234, (01)00884838099258(21)3632, (01)00884838099258(21)2203, (01)00884838085268(21)704109, (01)00884838099272(21)1158, (01)00884838099210(21)166, (01)00884838085268(21)1224, (01)00884838085268(21)1225, (01)00884838110564(21)345, (01)00884838110571(21)35, (01)00884838103276(21)281, (01)00884838085268(21)704125, (01)00884838110571(21)151, (01)00884838099241(21)953, (01)00884838085251(21)964, (01)00884838085251(21)965, (01)00884838085268(21)703684, (01)00884838099272(21)1109, (01)00884838099234(21)655, (01)00884838099258(21)1920, (01)00884838085268(21)703742, (01)00884838110571(21)86, (01)00884838099258(21)2914, (01)00884838099210(21)161, (01)00884838085268(21)703723, (01)00884838110564(21)22, (01)00884838103276(21)261, (01)00884838099234(21)384, (01)00884838110564(21)224, (01)00884838085268(21)704148, (01)00884838110564(21)197, (01)00884838103276(21)138, (01)00884838085282(21)344, (01)00884838099258(21)1977, (01)00884838099258(21)1695, (01)00884838085350(21)273, (01)00884838099272(21)903, (01)00884838085268(21)703681, (01)00884838085251(21)1043, (01)00884838099258(21)1721, (01)00884838110571(21)157, (01)00884838110571(21)85, (01)00884838099258(21)1740, (01)00884838085251(21)735, (01)00884838085251(21)736, (01)00884838110571(21)146, (01)00884838110564(21)206, (01)00884838110564(21)198, (01)00884838099258(21)3220, (01)00884838110571(21)69, (01)00884838110571(21)64, (01)00884838099258(21)1610, (01)00884838099234(21)534, (01)00884838085268(21)703778, (01)00884838099258(21)2033, (01)00884838110571(21)8, (01)00884838099258(21)2418, (01)00884838110571(21)161, (01)00884838085367(21)590, (01)00884838085268(21)703635, (01)00884838085268(21)704206, (01)00884838099241(21)644, (01)00884838110564(21)329, (01)00884838099272(21)669, (01)00884838103276(21)255, (01)00884838110571(21)174, (01)00884838085268(21)703910, (01)00884838085268(21)703553, (01)00884838085268(21)703946, (01)00884838099258(21)2108, (01)00884838099234(21)362, (01)00884838110564(21)249, (01)00884838110564(21)58, (01)00884838085268(21)704265, (01)00884838085268(21)704189, (01)00884838085282(21)389, (01)00884838110564(21)133, (01)00884838099272(21)747, (01)00884838099258(21)1136, (01)00884838099258(21)2206, (01)00884838085282(21)492, (01)00884838110571(21)62, (01)00884838110564(21)263, (01)00884838103276(21)212, (01)00884838099234(21)520, (01)00884838085268(21)703951, (01)00884838085251(21)820, (01)00884838085251(21)819, (01)00884838099234(21)489, (01)00884838110564(21)368, (01)00884838099241(21)1169, (01)00884838085268(21)704098, (01)00884838110571(21)137, (01)00884838085282(21)324, (01)00884838085282(21)265, (01)00884838099258(21)3182, (01)00884838099258(21)1183, (01)00884838085268(21)703539, (01)00884838103276(21)84, (01)00884838085268(21)703541, (01)00884838099258(21)1616, (01)00884838110564(21)280, (01)00884838110564(21)193, (01)00884838099258(21)1414, (01)00884838085367(21)198, (01)00884838099241(21)1228, (01)00884838099241(21)421, (01)00884838085251(21)272, (01)00884838085268(21)703678, (01)00884838110571(21)70, (01)00884838099265(21)476, (01)00884838085268(21)1034, (01)00884838099265(21)472, (01)00884838099265(21)471, (01)00884838085251(21)1049, (01)00884838099258(21)1451, (01)00884838099272(21)1124, (01)00884838085268(21)703900, (01)00884838099258(21)2895, (01)00884838110564(21)118, (01)00884838085282(21)235, (01)00884838099272(21)29, (01)00884838085282(21)120, (01)00884838103276(21)145, (01)00884838110564(21)137, (01)00884838110564(21)303, (01)00884838110564(21)375, (01)00884838099234(21)592, (01)00884838110564(21)439, (01)00884838085251(21)856, (01)00884838085282(21)496, (01)00884838099258(21)2827, (01)00884838103276(21)221, (01)00884838085251(21)1073, (01)00884838085268(21)704162, (01)00884838085251(21)1079, (01)00884838085268(21)704159, (01)00884838085367(21)600, (01)00884838085268(21)703893, (01)00884838110564(21)302, (01)00884838099234(21)303, (01)00884838099258(21)1769, (01)00884838099265(21)251, (01)00884838099265(21)253, (01)00884838099241(21)662, (01)00884838099241(21)655, (01)00884838099258(21)3604, (01)00884838099241(21)874, (01)00884838099241(21)873, (01)00884838099258(21)29, (01)00884838099258(21)52, (01)00884838085251(21)714, (01)00884838085350(21)176, (01)00884838085251(21)712, (01)00884838099210(21)16, (01)00884838085251(21)732, (01)00884838099241(21)92, (01)00884838085282(21)427, (01)00884838085268(21)703761, (01)00884838099258(21)1591, (01)00884838099258(21)1042, (01)00884838085268(21)756, (01)00884838099258(21)2251, (01)00884838099258(21)71, (01)00884838085282(21)341, (01)00884838085268(21)786, (01)00884838099203(21)167, (01)00884838099203(21)133, (01)00884838099203(21)26, (01)00884838099203(21)23, (01)00884838099241(21)1133, (01)00884838099241(21)1136, (01)00884838099203(21)40, (01)00884838085251(21)884, (01)00884838085251(21)567, (01)00884838085268(21)390, (01)00884838099272(21)817, (01)00884838099227(21)106, (01)00884838099272(21)995, (01)00884838085268(21)703845, (01)00884838099258(21)424, (01)00884838085251(21)774, (01)00884838085251(21)381, (01)00884838099203(21)10, (01)00884838085275(21)128, (01)00884838099241(21)632, (01)00884838099241(21)792, (01)00884838099258(21)1778, (01)00884838085251(21)233, (01)00884838099234(21)413, (01)00884838099227(21)253, (01)00884838099241(21)218, (01)00884838085268(21)703385, (01)00884838099227(21)145, (01)00884838099265(21)441, (01)00884838099258(21)1962, (01)00884838099272(21)625, (01)00884838099265(21)326, (01)00884838085251(21)638, (01)00884838099241(21)784, (01)00884838085251(21)318, (01)00884838099258(21)2851, (01)00884838099241(21)760, (01)00884838085367(21)228, (01)00884838099272(21)439, (01)00884838099241(21)543, (01)00884838099241(21)542, (01)00884838099258(21)671, (01)00884838085268(21)364, (01)00884838099241(21)895, (01)00884838085367(21)380, (01)00884838099272(21)297, (01)00884838099272(21)743, (01)00884838085268(21)398, (01)00884838085367(21)420, (01)00884838085282(21)100, (01)00884838085367(21)397, (01)00884838099258(21)2863, (01)00884838099241(21)426, (01)00884838085275(21)105, (01)00884838099258(21)2264, (01)00884838099258(21)2015, (01)00884838085282(21)51, (01)00884838085268(21)703653, (01)00884838099258(21)2525, (01)00884838099258(21)3452, (01)00884838099241(21)1160, (01)00884838099258(21)2792, (01)00884838099241(21)195, (01)00884838099272(21)27, (01)00884838085367(21)212, (01)00884838110571(21)150, (01)00884838110564(21)195, (01)00884838085251(21)593, (01)00884838085268(21)686, (01)00884838085350(21)64, (01)00884838085268(21)703351, (01)00884838085268(21)1046, (01)00884838085350(21)156, (01)00884838085251(21)624, (01)00884838099258(21)2964, (01)00884838099241(21)255, (01)00884838099272(21)131, (01)00884838085367(21)475, (01)00884838085251(21)776, (01)00884838085251(21)775, (01)00884838099258(21)310, (01)00884838099258(21)1137, (01)00884838099258(21)1432, (01)00884838099258(21)209, (01)00884838099258(21)1372, (01)00884838099258(21)1408, (01)00884838085275(21)4, (01)00884838099258(21)1649, (01)00884838099258(21)631, (01)00884838099234(21)63, (01)00884838099258(21)629, (01)00884838085268(21)1255, (01)00884838085367(21)180, (01)00884838099258(21)2997, (01)00884838099258(21)3511, (01)00884838099258(21)2300, (01)00884838085268(21)553, (01)00884838085268(21)703441, (01)00884838099272(21)644, (01)00884838099272(21)643, (01)008 84838099241(21)42, (01)00884838099258(21)38, (01)00884838085268(21)767, (01)00884838099258(21)1333, (01)00884838099272(21)563, (01)00884838099258(21)1969, (01)00884838085268(21)667, (01)00884838085268(21)665, (01)00884838085268(21)1045, (01)00884838099258(21)1589, (01)00884838099258(21)11, (01)00884838099272(21)640, (01)00884838099258(21)2208, (01)00884838099241(21)839, (01)00884838099272(21)639, (01)00884838085268(21)317, (01)00884838099258(21)304, (01)00884838099258(21)303, (01)00884838099203(21)203, (01)00884838099203(21)30, (01)00884838099227(21)164, (01)00884838099241(21)708, (01)00884838099258(21)782, (01)00884838085268(21)1261, (01)00884838099227(21)241, (01)00884838099258(21)3785, (01)00884838099258(21)3797, (01)00884838099258(21)3796, (01)00884838099258(21)3782, (01)00884838099234(21)765, (01)00884838099258(21)3778, (01)00884838099258(21)3774, (01)00884838099234(21)730, (01)00884838099234(21)732, (01)00884838099234(21)685, (01)00884838099234(21)722, (01)00884838099234(21)723, (01)00884838099203(21)207, (01)00884838099203(21)208, (01)00884838099234(21)720, (01)00884838099258(21)3763, (01)00884838099234(21)719, (01)00884838099234(21)702, (01)00884838099234(21)701, (01)00884838099258(21)3734, (01)00884838099234(21)691, (01)00884838099258(21)3749, (01)00884838099234(21)688, (01)00884838099210(21)396, (01)00884838099258(21)3567, (01)00884838099258(21)3445, (01)00884838099210(21)374, (01)00884838099210(21)370, (01)00884838099241(21)536, (01)00884838099241(21)254, (01)00884838099258(21)132, (01)00884838085282(21)388, (01)00884838099241(21)1108, (01)00884838099272(21)945, (01)00884838085268(21)703715, (01)00884838099210(21)280, (01)00884838099241(21)914, (01)00884838099241(21)41, (01)00884838099258(21)690, (01)00884838085268(21)391, (01)00884838099272(21)22, (01)00884838085282(21)118, (01)00884838099258(21)1148, (01)00884838099258(21)530, (01)00884838099272(21)203, (01)00884838099258(21)499, (01)00884838085282(21)62, (01)00884838099203(21)76, (01)00884838110564(21)103, (01)00884838099234(21)35, (01)00884838099258(21)653, (01)00884838085251(21)21, (01)00884838099258(21)660, (01)00884838085251(21)267, (01)00884838085251(21)268, (01)00884838085268(21)1253, (01)00884838099234(21)512, (01)00884838099258(21)864, (01)00884838099272(21)333, (01)00884838099241(21)359, (01)00884838099241(21)360, (01)00884838099241(21)408, (01)00884838099258(21)921, (01)00884838085367(21)414, (01)00884838085367(21)268, (01)00884838085367(21)269, (01)00884838085251(21)487, (01)00884838099241(21)1019, (01)00884838099258(21)972, (01)00884838085268(21)1427, (01)00884838099258(21)2401, (01)00884838099258(21)74, (01)00884838099241(21)642, (01)00884838099258(21)1579, (01)00884838099241(21)90, (01)00884838099272(21)50, (01)00884838085367(21)74, (01)00884838099241(21)643, (01)00884838099258(21)3440, (01)00884838099258(21)1521, (01)00884838099258(21)2542, (01)00884838085268(21)703414, (01)00884838099234(21)404, (01)00884838085268(21)693, (01)00884838085367(21)562, (01)00884838099258(21)3072, (01)00884838099241(21)288, (01)00884838099241(21)290, (01)00884838099241(21)286, (01)00884838085268(21)1366, (01)00884838099258(21)737, (01)00884838099258(21)2684, (01)00884838099258(21)1016, (01)00884838099241(21)80, (01)00884838099241(21)74, (01)00884838099241(21)72, (01)00884838099258(21)68, (01)00884838085251(21)421, (01)00884838099258(21)2693, (01)00884838099258(21)1601, (01)00884838099258(21)1621, (01)00884838085367(21)115, (01)00884838099258(21)1598, (01)00884838099234(21)74, (01)00884838099241(21)1184, (01)00884838099258(21)3284, (01)00884838099258(21)3160, (01)00884838099258(21)938, (01)00884838099258(21)2397, (01)00884838099241(21)955, (01)00884838099258(21)3508, (01)00884838099241(21)1063, (01)00884838085251(21)760, (01)00884838099265(21)411, (01)00884838099227(21)52, (01)00884838085251(21)542, (01)00884838099203(21)50, (01)00884838099203(21)98, (01)00884838099241(21)423, (01)00884838099234(21)437, (01)00884838099227(21)26, (01)00884838085275(21)50, (01)00884838099227(21)233, (01)00884838099227(21)301, (01)00884838099203(21)107, (01)00884838085275(21)129, (01)00884838099203(21)158, (01)00884838085282(21)123, (01)00884838099210(21)249, (01)00884838085268(21)456, (01)00884838085275(21)72, (01)00884838085275(21)38, (01)00884838099227(21)211, (01)00884838099227(21)154, (01)00884838099203(21)157, (01)00884838099203(21)62, (01)00884838085275(21)21, (01)00884838099241(21)706, (01)00884838099227(21)217, (01)00884838099227(21)332, (01)00884838099258(21)2142, (01)00884838099227(21)334, (01)00884838085275(21)78, (01)00884838099203(21)29, (01)00884838085275(21)23, (01)00884838099227(21)44, (01)00884838099227(21)198, (01)00884838099241(21)1252, (01)00884838099234(21)614, (01)00884838085275(21)76, (01)00884838085251(21)614, (01)00884838099241(21)291, (01)00884838085275(21)63, (01)00884838099227(21)297, (01)00884838099203(21)194, (01)00884838099258(21)717, (01)00884838085251(21)864, (01)00884838099203(21)73, (01)00884838099203(21)25, (01)00884838099241(21)483, (01)00884838085275(21)92, (01)00884838099227(21)206, (01)00884838085251(21)148, (01)00884838085275(21)69, (01)00884838099272(21)813, (01)00884838099258(21)2999, (01)00884838085367(21)349, (01)00884838085268(21)941, (01)00884838085251(21)837, (01)00884838099227(21)292, (01)00884838085282(21)177, (01)00884838099258(21)3076, (01)00884838099241(21)241, (01)00884838099241(21)1007, (01)00884838099258(21)2848, (01)00884838099227(21)130, (01)00884838099241(21)625, (01)00884838085268(21)392, (01)00884838099258(21)528, (01)00884838099258(21)529, (01)00884838099227(21)58, (01)00884838085268(21)703505, (01)00884838085251(21)458, (01)00884838085268(21)544, (01)00884838085251(21)596, (01)00884838085251(21)377, (01)00884838085268(21)323, (01)00884838085282(21)47, (01)00884838099265(21)368, (01)00884838099241(21)656, (01)00884838085350(21)276, (01)00884838099241(21)774, (01)00884838099241(21)751, (01)00884838099272(21)198, (01)00884838099241(21)630, (01)00884838099258(21)293, (01)00884838099203(21)15, (01)00884838085275(21)59, (01)00884838099203(21)119, (01)00884838099272(21)168, (01)00884838085268(21)610, (01)00884838099234(21)656, (01)00884838085268(21)950, (01)00884838085367(21)291, (01)00884838099258(21)1155, (01)00884838085275(21)51, (01)00884838099203(21)140, (01)00884838099234(21)546, (01)00884838099227(21)203, (01)00884838099258(21)633, (01)00884838099241(21)531, (01)00884838099227(21)47, (01)00884838099272(21)166, (01)00884838085251(21)281, (01)00884838099234(21)14, (01)00884838099258(21)1506, (01)00884838099241(21)941, (01)00884838099241(21)939, (01)00884838099258(21)2361, (01)00884838085268(21)576, (01)00884838099272(21)766, (01)00884838099203(21)166, (01)00884838099258(21)3381, (01)00884838085268(21)556, (01)00884838085282(21)272, (01)00884838099234(21)336, (01)00884838099234(21)200, (01)00884838099210(21)115, (01)00884838099272(21)1163, (01)00884838099210(21)321, (01)00884838085282(21)110, (01)00884838085367(21)519, (01)00884838085268(21)704150, (01)00884838099272(21)16, (01)00884838085268(21)703658, (01)00884838085367(21)234, (01)00884838099203(21)57, (01)00884838099258(21)2481, (01)00884838099265(21)333, (01)00884838099203(21)27, (01)00884838085268(21)704296, (01)00884838099272(21)580, (01)00884838099241(21)1024, (01)00884838099241(21)478, (01)00884838085268(21)533, (01)00884838085282(21)59, (01)00884838085282(21)167, (01)00884838099210(21)211, (01)00884838099272(21)384, (01)00884838085251(21)659, (01)00884838099265(21)252, (01)00884838099210(21)168, (01)00884838099234(21)357, (01)00884838099234(21)356, (01)00884838085268(21)704135, (01)00884838099258(21)1261, (01)00884838110564(21)353, (01)00884838110564(21)443, (01)00884838085282(21)174, (01)00884838099258(21)3426, (01)00884838099258(21)2343, (01)00884838099258(21)2106, (01)00884838085268(21)1004, (01)00884838099258(21)541, (01)00884838085268(21)703362, (01)00884838099203(21)2, (01)00884838099258(21)2877, (01)00884838099210(21)127, (01)00884838099241(21)1143, (01)00884838099258(21)2638, (01)00884838099241(21)645, (01)00884838099241(21)647, (01)00884838099241(21)648, (01)00884838099241(21)646, (01)00884838099210(21)261, (01)00884838099258(21)1325, (01)00884838085275(21)139, (01)00884838099210(21)54, (01)00884838099258(21)2676, (01)00884838099227(21)107, (01)00884838099241(21)786, (01)00884838099258(21)205, (01)00884838085282(21)165, (01)00884838099227(21)93, (01)00884838085275(21)130, (01)00884838099258(21)221, (01)00884838099258(21)1772, (01)00884838099258(21)2430, (01)00884838099227(21)125, (01)00884838110564(21)196, (01)00884838085268(21)704149, (01)00884838099265(21)351, (01)00884838099272(21)755, (01)00884838099272(21)1022, (01)00884838085251(21)784, (01)00884838099265(21)283, (01)00884838099258(21)509, (01)00884838099265(21)87, (01)00884838099272(21)207, (01)00884838099203(21)47, (01)00884838085251(21)991, (01)00884838099227(21)240, (01)00884838099258(21)3006, (01)00884838085367(21)219, (01)00884838085251(21)528, (01)00884838085268(21)876, (01)00884838085268(21)285, (01)00884838085350(21)52, (01)00884838099265(21)310, (01)00884838099265(21)309, (01)00884838099258(21)1826, (01)00884838099272(21)513, (01)00884838085367(21)195, (01)00884838099265(21)349, (01)00884838085367(21)532, (01)00884838099258(21)1878, (01)00884838099258(21)387, (01)00884838099258(21)2375, (01)00884838085268(21)704034, (01)00884838110571(21)80, (01)00884838085268(21)1332, (01)00884838085268(21)703873, (01)00884838099241(21)660, (01)00884838099272(21)598, (01)00884838085367(21)533, (01)00884838099258(21)3606, (01)00884838085268(21)703321, (01)00884838099258(21)1034, (01)00884838099241(21)407, (01)00884838099203(21)16, (01)00884838085268(21)369, (01)00884838099258(21)2946, (01)00884838085268(21)370, (01)00884838085367(21)60, (01)00884838099265(21)1, (01)00884838099227(21)122, (01)00884838099227(21)118, (01)00884838085282(21)452, (01)00884838099203(21)141, (01)00884838099234(21)394, (01)00884838085350(21)232, (01)00884838099258(21)970, (01)00884838099258(21)3636, (01)00884838085350(21)54, (01)00884838085268(21)546, (01)00884838085350(21)55, (01)00884838085282(21)41, (01)00884838085282(21)168, (01)00884838085268(21)704010, (01)00884838099258(21)597, (01)00884838085350(21)165, (01)00884838099234(21)7, (01)00884838099241(21)2, (01)00884838085268(21)254, (01)00884838099272(21)491, (01)00884838085268(21)812, (01)00884838099241(21)840, (01)00884838099227(21)112, (01)00884838085251(21)44, (01)00884838085268(21)77, (01)00884838085251(21)42, (01)00884838085251(21)45, (01)00884838085350(21)25, (01)00884838085251(21)41, (01)00884838085251(21)43, (01)00884838085268(21)703369, (01)00884838099227(21)46, (01)00884838099272(21)896, (01)00884838099258(21)1671, (01)00884838099241(21)24, (01)00884838099272(21)54, (01)00884838099227(21)176, (01)00884838099272(21)849, (01)00884838099258(21)2339, (01)00884838099272(21)636, (01)00884838099241(21)674, (01)00884838099258(21)1585, (01)00884838099210(21)14, (01)00884838099258(21)1834, (01)00884838099258(21)1572, (01)00884838099241(21)358, (01)00884838099258(21)822, (01)00884838085268(21)727, (01)00884838110564(21)139, (01)00884838110564(21)408, (01)00884838103276(21)36, (01)00884838085350(21)271, (01)00884838085367(21)398, (01)00884838085275(21)154, (01)00884838085268(21)704057, (01)00884838103276(21)256, (01)00884838085282(21)447, (01)00884838085367(21)618, (01)00884838085268(21)1357, (01)00884838099241(21)1079, (01)00884838099258(21)2892, (01)00884838099265(21)220, (01)00884838099265(21)123, (01)00884838085350(21)35, (01)00884838085251(21)466, (01)00884838085251(21)704, (01)00884838099227(21)92, (01)00884838085251(21)219, (01)00884838099258(21)1876, (01)00884838085268(21)1133, (01)00884838085282(21)148, (01)00884838099258(21)2079, (01)00884838099227(21)252, (01)00884838085251(21)28, (01)00884838085251(21)29, (01)00884838099234(21)504, (01)00884838085367(21)549, (01)00884838099227(21)184, (01)00884838085268(21)703459, (01)00884838099227(21)57, (01)00884838099241(21)110, (01)00884838099258(21)243, (01)00884838085282(21)182, (01)00884838099234(21)190, (01)00884838099234(21)377, (01)00884838085282(21)112, (01)00884838099241(21)309, (01)00884838099210(21)38, (01)00884838099258(21)207, (01)00884838099241(21)649, (01)00884838099241(21)102, (01)00884838099241(21)315, (01)00884838099272(21)82, (01)00884838085282(21)302, (01)00884838099258(21)180, (01)00884838099234(21)383, (01)00884838085268(21)1065, (01)00884838099258(21)3126, (01)00884838099258(21)664, (01)00884838085251(21)217, (01)00884838099258(21)3180, (01)00884838099258(21)1664, (01)00884838099234(21)193, (01)00884838099258(21)3010, (01)00884838099234(21)94, (01)00884838099234(21)89, (01)00884838085268(21)1199, (01)00884838099234(21)93, (01)00884838085275(21)75, (01)00884838099258(21)155, (01)00884838099272(21)679, (01)00884838099227(21)202, (01)00884838099258(21)899, (01)00884838099210(21)10, (01)00884838085268(21)703944, (01)00884838099272(21)330, (01)00884838099258(21)1001, (01)00884838099210(21)67, (01)00884838085268(21)1299, (01)00884838085282(21)86, (01)00884838099227(21)100, (01)00884838099258(21)1868, (01)00884838099234(21)310, (01)00884838099234(21)309, (01)00884838099258(21)3553, (01)00884838085282(21)317, (01)00884838099241(21)1109, (01)00884838099227(21)115, (01)00884838099210(21)165, (01)00884838099258(21)3316, (01)00884838099258(21)3315, (01)00884838085251(21)260, (01)00884838099210(21)272, (01)00884838099210(21)326, (01)00884838085282(21)198, (01)00884838099234(21)494, (01)00884838099210(21)173, (01)00884838099227(21)133, (01)00884838099265(21)184, (01)00884838099258(21)2043, (01)00884838099241(21)1235, (01)00884838099272(21)417, (01)00884838099258(21)655, (01)00884838099272(21)892, (01)00884838085268(21)1388, (01)00884838099241(21)38, (01)00884838085251(21)103, (01)00884838099241(21)673, (01)00884838085367(21)46, (01)00884838085268(21)529, (01)00884838085251(21)430, (01)00884838099258(21)1007, (01)00884838099241(21)1066, (01)00884838099241(21)925, (01)00884838099241(21)940, (01)00884838085251(21)342, (01)00884838099241(21)934, (01)00884838099272(21)278, (01)00884838099272(21)277, (01)00884838085251(21)184, (01)00884838099241(21)201, (01)00884838099241(21)1229, (01)00884838099258(21)1162, (01)00884838099258(21)3036, (01)00884838099258(21)23, (01)00884838099234(21)444, (01)00884838085367(21)96, (01)00884838085268(21)952, (01)00884838099241(21)121, (01)00884838099241(21)122, (01)00884838085268(21)676, (01)00884838085275(21)27, (01)00884838099227(21)350, (01)00884838099241(21)1189, (01)00884838099234(21)199, (01)00884838099227(21)263, (01)00884838099241(21)902, (01)00884838085251(21)778, (01)00884838099210(21)90, (01)00884838085275(21)24, (01)00884838099203(21)143, (01)00884838099227(21)3, (01)00884838099258(21)3177, (01)00884838085251(21)547, (01)00884838099241(21)1116, (01)00884838085251(21)879, (01)00884838085251(21)586, (01)00884838099272(21)152, (01)00884838099258(21)3014, (01)00884838099258(21)3013, (01)00884838085268(21)703720, (01)00884838085268(21)703892, (01)00884838099227(21)32, (01)00884838099241(21)123, (01)00884838099272(21)720, (01)00884838099210(21)109, (01)00884838099258(21)1672, (01)00884838099272(21)385, (01)00884838099241(21)34, (01)00884838099241(21)1206, (01)00884838099227(21)175, (01)00884838099272(21)641, (01)00884838099234(21)667, (01)00884838110564(21)455, (01)00884838085251(21)159, (01)00884838110564(21)173, (01)00884838099258(21)3340, (01)00884838099234(21)391, (01)00884838099272(21)1020, (01)00884838099272(21)55, (01)00884838099241(21)250, (01)00884838085268(21)694, (01)00884838085268(21)336, (01)00884838085350(21)202, (01)00884838099234(21)698, (01)00884838099234(21)532, (01)00884838099234(21)408, (01)00884838085251(21)363, (01)00884838099234(21)83, (01)00884838099227(21)229, (01)00884838099272(21)742, (01)00884838099265(21)258, (01)00884838099258(21)3266, (01)00884838099258(21)774, (01)00884838099272(21)305, (01)00884838085251(21)896, (01)00884838099258(21)2593, (01)00884838085251(21)642, (01)00884838085367(21)416, (01)00884838099210(21)13, (01)00884838085268(21)410, (01)00884838099258(21)2164, (01)00884838099234(21)143, (01)00884838099265(21)237, (01)00884838085251(21)472, (01)00884838085268(21)75, (01)00884838085268(21)703985, (01)00884838099272(21)832, (01)00884838099227(21)304, (01)00884838099234(21)518, (01)00884838099203(21)35, (01)00884838099241(21)172, (01)00884838099227(21)124, (01)00884838099241(21)369, (01)00884838099272(21)1059, (01)00884838099241(21)175, (01)00884838099241(21)1298, (01)00884838099203(21)168, (01)00884838085350(21)149, (01)00884838085268(21)338, (01)00884838085251(21)39, (01)00884838099234(21)85, (01)00884838099258(21)2703, (01)00884838085350(21)243, (01)00884838085251(21)172, (01)00884838099241(21)1056, (01)00884838085268(21)906, (01)00884838099258(21)3594, (01)00884838099272(21)947, (01)00884838099241(21)490, (01)00884838099272(21)409, (01)00884838085268(21)1348, (01)00884838099258(21)3261, (01)00884838085251(21)782, (01)00884838085268(21)1349, (01)00884838099258(21)510, (01)00884838099258(21)610, (01)00884838085268(21)72, (01)00884838099241(21)1014, (01)00884838099241(21)1013, (01)00884838085268(21)703579, (01)00884838085282(21)378, (01)00884838085282(21)206, (01)00884838099234(21)483, (01)00884838099203(21)83, (01)00884838099203(21)163, (01)00884838099272(21)902, (01)00884838099258(21)1733, (01)00884838099234(21)449, (01)00884838099210(21)292, (01)00884838099210(21)288, (01)00884838099210(21)290, (01)00884838099210(21)294, (01)00884838099210(21)289, (01)00884838099210(21)291, (01)00884838099210(21)295, (01)00884838085282(21)136, (01)00884838099227(21)307, (01)00884838099258(21)2814, (01)00884838099241(21)1026, (01)00884838085282(21)312, (01)00884838085282(21)143, (01)00884838099203(21)202, (01)00884838099258(21)375, (01)00884838085268(21)703810, (01)00884838085275(21)94, (01)00884838099272(21)1118, (01)00884838085268(21)704303, (01)00884838099241(21)832, (01)00884838099234(21)337, (01)00884838099241(21)578, (01)00884838099241(21)473, (01)00884838099241(21)138, (01)00884838099258(21)292, (01)00884838099227(21)220, (01)00884838103276(21)59, (01)00884838110564(21)143, (01)00884838099234(21)438, (01)00884838099258(21)1617, (01)00884838099227(21)79, (01)00884838099272(21)867, (01)00884838085268(21)704172, (01)00884838103276(21)196, (01)00884838085367(21)249, (01)00884838099241(21)462, (01)00884838085268(21)703694, (01)00884838110564(21)29, (01)00884838099227(21)60, (01)00884838103276(21)50, (01)00884838085367(21)531, (01)00884838099203(21)67, (01)00884838099227(21)311, (01)00884838099203(21)198, (01)00884838085282(21)403, (01)00884838110564(21)239, (01)00884838085268(21)703874, (01)00884838099234(21)453, (01)00884838103276(21)110, (01)00884838085275(21)100, (01)00884838099227(21)147, (01)00884838099227(21)146, (01)00884838103276(21)272, (01)00884838085251(21)961, (01)00884838085268(21)764, (01)00884838085268(21)1188, (01)00884838099258(21)1139, (01)00884838099272(21)627, (01)00884838099258(21)3595, (01)00884838085268(21)524, (01)00884838099258(21)2396, (01)00884838099241(21)841, (01)00884838099241(21)842, (01)00884838099241(21)1159, (01)00884838099258(21)2026, (01)00884838099227(21)319, (01)00884838085251(21)33, (01)00884838099272(21)872, (01)00884838099241(21)808, (01)00884838099258(21)2163, (01)00884838085251(21)643, (01)00884838085268(21)490, (01)00884838099272(21)418, (01)00884838099272(21)489, (01)00884838099258(21)1367, (01)00884838085251(21)581, (01)00884838099234(21)106, (01)00884838085268(21)759, (01)00884838099258(21)1530, (01)00884838099258(21)1626, (01)00884838085268(21)704015, (01)00884838110564(21)502, (01)00884838085367(21)426, (01)00884838099258(21)122, (01)00884838099272(21)231, (01)00884838099241(21)305, (01)00884838099258(21)3271, (01)00884838085251(21)552, (01)00884838085251(21)525, (01)00884838099258(21)2081, (01)00884838099258(21)2082, (01)00884838099258(21)2080, (01)00884838099241(21)530, (01)00884838099210(21)186, (01)00884838085282(21)386, (01)00884838085268(21)703675, (01)00884838085282(21)397, (01)00884838085282(21)500, (01)00884838103276(21)199, (01)00884838099234(21)311, (01)00884838099258(21)2443, (01)00884838099241(21)1341, (01)00884838099241(21)1340, (01)00884838099241(21)1338, (01)00884838110564(21)15, (01)00884838110564(21)512, (01)00884838110564(21)163, (01)00884838085268(21)703970, (01)00884838099258(21)2597, (01)00884838099258(21)2048, (01)00884838110564(21)437, (01)00884838110564(21)111, (01)00884838099210(21)181, (01)00884838110571(21)82, (01)00884838099258(21)2584, (01)00884838085268(21)704117, (01)00884838085268(21)704084, (01)00884838085268(21)703956, (01)00884838085268(21)704213, (01)00884838110564(21)479, (01)00884838099234(21)324, (01)00884838110564(21)252, (01)00884838099272(21)1080, (01)00884838099234(21)639, (01)00884838099234(21)353, (01)00884838099258(21)3339, (01)00884838099227(21)188, (01)00884838099227(21)74, (01)00884838099227(21)9, (01)00884838099227(21)8, (01)00884838099241(21)1074, (01)00884838085282(21)226, (01)00884838099272(21)407, (01)00884838099241(21)515, (01)00884838085367(21)162, (01)00884838099265(21)234, (01)00884838085251(21)510, (01)00884838099258(21)237, (01)00884838099203(21)134, (01)00884838099258(21)978, (01)00884838085282(21)408, (01)00884838085367(21)290, (01)00884838085268(21)467, (01)00884838099210(21)107, (01)00884838099227(21)94, (01)00884838099227(21)134, (01)00884838099258(21)1401, (01)00884838099258(21)758, (01)00884838099265(21)218, (01)00884838085251(21)857, (01)00884838099241(21)1011, (01)00884838099241(21)1010, (01)00884838085268(21)276, (01)00884838085282(21)43, (01)00884838099258(21)3576, (01)00884838099241(21)768, (01)00884838099241(21)620, (01)00884838099227(21)82, (01)00884838099265(21)149, (01)00884838085268(21)670, (01)00884838099241(21)857, (01)00884838085268(21)76, (01)00884838099241(21)913, (01)00884838085251(21)731, (01)00884838099234(21)459, (01)00884838099203(21)5, (01)00884838099258(21)2611, (01)00884838099210(21)379, (01)00884838099258(21)711, (01)00884838085268(21)703439, (01)00884838085268(21)63, (01)00884838085367(21)346, (01)00884838099241(21)103, (01)00884838099258(21)2700, (01)00884838099241(21)603, (01)00884838099241(21)27, (01)00884838099272(21)6, (01)00884838085268(21)397, (01)00884838085350(21)32, (01)00884838099241(21)1158, (01)00884838099258(21)1534, (01)00884838099227(21)143, (01)00884838099265(21)231, (01)00884838099272(21)1089, (01)00884838099272(21)913, (01)00884838099241(21)623, (01)00884838099241(21)968, (01)00884838099258(21)1736, (01)00884838099241(21)983, (01)00884838099258(21)97, (01)00884838099272(21)840, (01)00884838099272(21)24, (01)00884838085282(21)381, (01)00884838099258(21)3574, (01)00884838099258(21)203, (01)00884838099241(21)1009, (01)00884838099258(21)2344, (01)00884838099241(21)993, (01)00884838099241(21)972, (01)00884838085268(21)1384, (01)00884838099258(21)1685, (01)00884838099210(21)385, (01)00884838085251(21)432, (01)00884838085251(21)433, (01)00884838099258(21)3119, (01)00884838099258(21)1066, (01)00884838099272(21)364, (01)00884838085268(21)704035, (01)00884838099258(21)1932, (01)00884838085251(21)636, (01)00884838085268(21)444, (01)00884838085268(21)282, (01)00884838085268(21)971, (01)00884838099265(21)342, (01)00884838099227(21)86, (01)00884838085268(21)1215, (01)00884838099241(21)1199, (01)00884838110564(21)230, (01)00884838099272(21)307, (01)00884838099241(21)1230, (01)00884838110564(21)42, (01)00884838085268(21)704188, (01)00884838099265(21)396, (01)00884838099227(21)286, (01)00884838099272(21)1035, (01)00884838099203(21)81, (01)00884838099203(21)22, (01)00884838099227(21)96, (01)00884838099227(21)227, (01)00884838099234(21)5, (01)00884838099258(21)821, (01)00884838099203(21)28, (01)00884838099203(21)165, (01)00884838085251(21)954, (01)00884838099241(21)636, (01)00884838099203(21)196, (01)00884838099272(21)1055, (01)00884838085275(21)54, (01)00884838099234(21)43, (01)00884838099241(21)1257, (01)00884838099258(21)1801, (01)00884838085275(21)46, (01)00884838099258(21)162, (01)00884838099265(21)7, (01)00884838099241(21)267, (01)00884838099241(21)268, (01)00884838085251(21)447, (01)00884838085275(21)26, (01)00884838099258(21)3237, (01)00884838099265(21)159, (01)00884838085251(21)333, (01)00884838099258(21)1388, (01)00884838085251(21)358, (01)00884838099258(21)348, (01)00884838099241(21)1209, (01)00884838099258(21)838, (01)00884838099227(21)231, (01)00884838099272(21)532, (01)00884838099272(21)531, (01)00884838099241(21)293, (01)00884838085251(21)540, (01)00884838110571(21)124, (01)00884838099258(21)2756, (01)00884838099272(21)745, (01)00884838099258(21)1618, (01)00884838099265(21)480, (01)00884838099272(21)1033, (01)00884838085275(21)71, (01)00884838085251(21)619, (01)00884838099241(21)321, (01)00884838099258(21)1284, (01)00884838099258(21)2309, (01)00884838099234(21)477, (01)00884838110564(21)481, (01)00884838085268(21)703814, (01)00884838099234(21)330, (01)00884838099210(21)296, (01)00884838099227(21)329, (01)00884838099241(21)639, (01)00884838085350(21)224, (01)00884838099258(21)2218, (01)00884838099203(21)173, (01)00884838103276(21)264, (01)00884838085282(21)488, (01)00884838099258(21)3012, (01)00884838099234(21)386, (01)00884838110564(21)220, (01)00884838110564(21)213, (01)00884838103276(21)140, (01)00884838085268(21)703717, (01)00884838103276(21)68, (01)00884838110564(21)347, (01)00884838110564(21)346, (01)00884838103276(21)117, (01)00884838110571(21)23, (01)00884838099234(21)288, (01)00884838103276(21)49, (01)00884838085268(21)1136, (01)00884838103276(21)257, (01)00884838099258(21)3772, (01)00884838099241(21)1202, (01)00884838099258(21)3321, (01)00884838110571(21)52, (01)00884838110571(21)53, (01)00884838099234(21)167, (01)00884838099241(21)182, (01)00884838085268(21)703925, (01)00884838085268(21)703922, (01)00884838085268(21)703920, (01)00884838099234(21)549, (01)00884838110564(21)411, (01)00884838099258(21)2670, (01)00884838110564(21)362, (01)00884838099227(21)126, (01)00884838099227(21)162, (01)00884838099272(21)791, (01)00884838099241(21)1040, (01)00884838099265(21)274, (01)00884838099265(21)144, (01)00884838099241(21)158, (01)00884838099241(21)156, (01)00884838085268(21)1056, (01)00884838085251(21)492, (01)00884838099265(21)142, (01)00884838099234(21)784, (01)00884838099272(21)1105, (01)00884838110564(21)184, (01)00884838110564(21)185, (01)00884838099258(21)2984, (01)00884838085268(21)1428, (01)00884838099241(21)224, (01)00884838110564(21)30, (01)00884838110564(21)31, (01)00884838099258(21)1358, (01)00884838085268(21)796, (01)00884838099234(21)390, (01)00884838099258(21)2565, (01)00884838085275(21)106, (01)00884838110564(21)25, (01)00884838099203(21)46, (01)00884838099234(21)368, (01)00884838085268(21)704016, (01)00884838099234(21)261, (01)00884838085268(21)703729, (01)00884838085251(21)527, (01)00884838099203(21)53, (01)00884838099203(21)8, (01)00884838099258(21)393, (01)00884838099258(21)2252, (01)00884838099258(21)2104, (01)00884838099272(21)591, (01)00884838099258(21)2547, (01)00884838085350(21)207, (01)00884838085282(21)416, (01)00884838110564(21)483, (01)00884838085275(21)83, (01)00884838085251(21)600, (01)00884838099258(21)344, (01)00884838099241(21)163, (01)00884838099258(21)2782, (01)00884838085268(21)703479, (01)00884838099241(21)1157, (01)00884838085268(21)704243, (01)00884838099210(21)315, (01)00884838099210(21)314, (01)00884838099227(21)299, (01)00884838099258(21)903, (01)00884838099241(21)389, (01)00884838085268(21)703461, (01)00884838099258(21)2625, (01)00884838099258(21)491, (01)00884838099258(21)474, (01)00884838099241(21)1216, (01)00884838099258(21)3418, (01)00884838099258(21)3456, (01)00884838085251(21)816, (01)00884838085268(21)308, (01)00884838099227(21)35, (01)00884838099241(21)958, (01)00884838099258(21)2716, (01)00884838099241(21)510, (01)00884838085268(21)657, (01)00884838099227(21)11, (01)00884838085251(21)434, (01)00884838085268(21)695, (01)00884838085367(21)49, (01)00884838085275(21)37, (01)00884838085275(21)36, (01)00884838085282(21)80, (01)00884838085268(21)616, (01)00884838099258(21)2137, (01)00884838085251(21)892, (01)00884838099227(21)259, (01)00884838085251(21)990, (01)00884838085268(21)703748, (01)00884838099258(21)1181, (01)00884838099227(21)180, (01)00884838085367(21)326, (01)00884838085268(21)775, (01)00884838085268(21)1139, (01)00884838085367(21)286, (01)00884838099258(21)1953, (01)00884838099210(21)77, (01)00884838099258(21)3127, (01)00884838099258(21)601, (01)00884838099258(21)988, (01)00884838085268(21)492, (01)00884838085268(21)491, (01)00884838099258(21)3, (01)00884838085268(21)743, (01)00884838099258(21)2568, (01)00884838085268(21)1229, (01)00884838085251(21)124, (01)00884838099272(21)772, (01)00884838099258(21)2482, (01)00884838085350(21)76, (01)00884838099210(21)354, (01)00884838099241(21)1039, (01)00884838099258(21)3195, (01)00884838085251(21)122, (01)00884838085268(21)1426, (01)00884838099241(21)1081, (01)00884838099272(21)878, (01)00884838099241(21)444, (01)00884838085268(21)238, (01)00884838099241(21)1192, (01)00884838099241(21)1238, (01)00884838099241(21)1045, (01)00884838099241(21)83, (01)00884838085268(21)321, (01)00884838085268(21)311, (01)00884838085251(21)146, (01)00884838085251(21)201, (01)00884838099241(21)491, (01)00884838099272(21)86, (01)00884838099258(21)498, (01)00884838085251(21)509, (01)00884838099258(21)2531, (01)00884838085268(21)872, (01)00884838099258(21)936, (01)00884838099241(21)420, (01)00884838099258(21)559, (01)00884838099258(21)2604, (01)00884838085268(21)240, (01)00884838099241(21)1233, (01)00884838099241(21)969, (01)00884838099258(21)2944, (01)00884838099258(21)2400, (01)00884838099258(21)929, (01)00884838099258(21)930, (01)00884838099258(21)920, (01)00884838099258(21)919, (01)00884838099258(21)3382, (01)00884838099258(21)834, (01)00884838085268(21)703372, (01)00884838099258(21)1735, (01)00884838099272(21)1093, (01)00884838085268(21)703622, (01)00884838099258(21)126, (01)00884838085268(21)703537, (01)00884838085268(21)703538, (01)00884838085251(21)495, (01)00884838085268(21)703517, (01)00884838099258(21)345, (01)00884838099272(21)7, (01)00884838099258(21)2960, (01)00884838099241(21)967, (01)00884838099241(21)952, (01)00884838099258(21)117, (01)00884838099234(21)192, (01)00884838099258(21)1643, (01)00884838099258(21)2395, (01)00884838085268(21)1209, (01)00884838099241(21)1023, (01)00884838099258(21)934, (01)00884838099258(21)925, (01)00884838099241(21)512, (01)00884838099272(21)768, (01)00884838099241(21)610, (01)00884838099241(21)569, (01)00884838099258(21)241, (01)00884838085268(21)1418, (01)00884838099241(21)1030, (01)00884838085268(21)703440, (01)00884838085268(21)1364, (01)00884838099258(21)3122, (01)00884838099241(21)551, (01)00884838099258(21)2399, (01)00884838099258(21)1031, (01)00884838099258(21)931, (01)00884838099258(21)927, (01)00884838099241(21)1012, (01)00884838099258(21)2034, (01)00884838099258(21)924, (01)00884838085251(21)515, (01)00884838085268(21)703523, (01)00884838085268(21)703524, (01)00884838085268(21)703502, (01)00884838099241(21)410, (01)00884838099258(21)2526, (01)00884838099258(21)576, (01)00884838099258(21)92, (01)00884838085367(21)125, (01)00884838085251(21)353, (01)00884838099258(21)1248, (01)00884838085268(21)386, (01)00884838099234(21)66, (01)00884838085268(21)582, (01)00884838099241(21)1044, (01)00884838099241(21)330, (01)00884838085251(21)303, (01)00884838085268(21)494, (01)00884838085251(21)304, (01)00884838099234(21)115, (01)00884838099234(21)116, (01)00884838099241(21)896, (01)00884838099258(21)51, (01)00884838099258(21)1817, (01)00884838099241(21)1178, (01)00884838099258(21)3136, (01)00884838099258(21)923, (01)00884838099241(21)251, (01)00884838099258(21)570, (01)00884838099258(21)918, (01)00884838099258(21)1951, (01)00884838099258(21)535, (01)00884838099258(21)536, (01)00884838099258(21)1928, (01)00884838099258(21)457, (01)00884838099272(21)664, (01)00884838099241(21)406, (01)00884838085251(21)186, (01)00884838099258(21)2022, (01)00884838099241(21)404, (01)00884838099241(21)1004, (01)00884838099272(21)85, (01)00884838099241(21)86, (01)00884838085268(21)703377, (01)00884838085268(21)703376, (01)00884838099258(21)213, (01)00884838099258(21)3097, (01)00884838099241(21)397, (01)00884838099258(21)2366, (01)00884838085268(21)405, (01)00884838099241(21)33, (01)00884838099227(21)160, (01)00884838099258(21)2232, (01)00884838099241(21)748, (01)00884838099227(21)196, (01)00884838099258(21)2475, (01)00884838099241(21)547, (01)00884838085251(21)286, (01)00884838085251(21)548, (01)00884838099265(21)178, (01)00884838099258(21)1023, (01)00884838099203(21)56, (01)00884838099258(21)976, (01)00884838099272(21)52, (01)00884838099272(21)53, (01)008848380 99258(21)2214, (01)00884838099234(21)334, (01)00884838099227(21)322, (01)00884838099227(21)208, (01)00884838099258(21)3349, (01)00884838099258(21)1767, (01)00884838099241(21)1000, (01)00884838099241(21)400, (01)00884838099241(21)399, (01)00884838085251(21)518, (01)00884838085367(21)102, (01)00884838085268(21)562, (01)00884838099241(21)366, (01)00884838099258(21)2537, (01)00884838099272(21)890, (01)00884838099241(21)1018, (01)00884838099258(21)928, (01)00884838099241(21)513, (01)00884838085268(21)703503, (01)00884838099258(21)2054, (01)00884838099241(21)345, (01)00884838099258(21)1592, (01)00884838085268(21)703829, (01)00884838085367(21)113, (01)00884838085282(21)12, (01)00884838085251(21)486, (01)00884838085350(21)108, (01)00884838099272(21)63, (01)00884838099234(21)467, (01)00884838099272(21)1069, (01)00884838085268(21)1330, (01)00884838085251(21)475, (01)00884838085251(21)474, (01)00884838085268(21)24, (01)00884838099210(21)380, (01)00884838099241(21)380, (01)00884838085268(21)703952, (01)00884838085282(21)404, (01)00884838099258(21)2167, (01)00884838099258(21)2833, (01)00884838085251(21)853, (01)00884838099272(21)637, (01)00884838099258(21)2161, (01)00884838085350(21)226, (01)00884838085251(21)180, (01)00884838099258(21)749, (01)00884838085251(21)243, (01)00884838085268(21)495, (01)00884838085251(21)244, (01)00884838085268(21)704018, (01)00884838085268(21)703957, (01)00884838099258(21)2612, (01)00884838085251(21)335, (01)00884838085367(21)124, (01)00884838110571(21)15, (01)00884838099258(21)1751, (01)00884838103276(21)60, (01)00884838099234(21)307, (01)00884838110564(21)106, (01)00884838099258(21)3221, (01)00884838099227(21)282, (01)00884838099241(21)94, (01)00884838085268(21)932, (01)00884838099234(21)304, (01)00884838099272(21)815, (01)00884838110564(21)142, (01)00884838099258(21)1845, (01)00884838110564(21)128, (01)00884838110564(21)36, (01)00884838099234(21)350, (01)00884838099258(21)1528, (01)00884838085268(21)704033, (01)00884838085268(21)704032, (01)00884838085268(21)703695, (01)00884838099234(21)212, (01)00884838110571(21)100, (01)00884838099258(21)2455, (01)00884838099258(21)2553, (01)00884838085282(21)285, (01)00884838099258(21)1501, (01)00884838099241(21)621, (01)00884838099265(21)348, (01)00884838085367(21)138, (01)00884838099241(21)95, (01)00884838085251(21)549, (01)00884838099265(21)17, (01)00884838099258(21)984, (01)00884838099241(21)428, (01)00884838099241(21)13, (01)00884838099241(21)672, (01)00884838099227(21)13, (01)00884838085268(21)720, (01)00884838099258(21)2078, (01)00884838099227(21)178, (01)00884838099258(21)863, (01)00884838099227(21)4, (01)00884838099258(21)3746, (01)00884838099241(21)1200, (01)00884838099241(21)977, (01)00884838099258(21)868, (01)00884838099258(21)2698, (01)00884838099265(21)31, (01)00884838085350(21)48, (01)00884838085251(21)46, (01)00884838099241(21)927, (01)00884838099258(21)712, (01)00884838085251(21)616, (01)00884838085268(21)680, (01)00884838085251(21)972, (01)00884838099258(21)549, (01)00884838085275(21)15, (01)00884838099258(21)1765, (01)00884838099258(21)3327, (01)00884838099258(21)2323, (01)00884838085251(21)533, (01)00884838099241(21)1120, (01)00884838099241(21)1119, (01)00884838099258(21)70, (01)00884838085367(21)362, (01)00884838099258(21)2849, (01)00884838099241(21)1029, (01)00884838099241(21)190, (01)00884838099258(21)405, (01)00884838099241(21)1208, (01)00884838099258(21)3394, (01)00884838099272(21)953, (01)00884838099265(21)386, (01)00884838110564(21)473, (01)00884838085275(21)135, (01)00884838085268(21)703475, (01)00884838099258(21)2554, (01)00884838085268(21)703947, (01)00884838103276(21)103, (01)00884838085251(21)777, (01)00884838099272(21)943, (01)00884838099203(21)174, (01)00884838099272(21)729, (01)00884838099258(21)1290, (01)00884838099258(21)3368, (01)00884838099258(21)2422, (01)00884838099258(21)781, (01)00884838085251(21)182, (01)00884838110564(21)84, (01)00884838085268(21)704063, (01)00884838099258(21)1683, (01)00884838085268(21)703890, (01)00884838085282(21)515, (01)00884838099258(21)1741, (01)00884838103276(21)237, (01)00884838085268(21)703589, (01)00884838099234(21)340, (01)00884838099234(21)348, (01)00884838110564(21)507, (01)00884838099258(21)3139, (01)00884838099258(21)2935, (01)00884838103276(21)72, (01)00884838103276(21)160, (01)00884838110564(21)16, (01)00884838103276(21)81, (01)00884838103276(21)114, (01)00884838103276(21)137, (01)00884838085268(21)704204, (01)00884838099234(21)695, (01)00884838099241(21)901, (01)00884838110564(21)475, (01)00884838085268(21)703536, (01)00884838099258(21)2757, (01)00884838085268(21)703849, (01)00884838085282(21)449, (01)00884838103276(21)58, (01)00884838085268(21)704287, (01)00884838085268(21)703659, (01)00884838085268(21)704025, (01)00884838099258(21)1823, (01)00884838110564(21)257, (01)00884838099258(21)3730, (01)00884838099258(21)3222, (01)00884838103276(21)73, (01)00884838099203(21)43, (01)00884838099210(21)110, (01)00884838085350(21)264, (01)00884838099234(21)506, (01)00884838099234(21)516, (01)00884838099272(21)1079, (01)00884838110571(21)50, (01)00884838085268(21)703926, (01)00884838099258(21)389, (01)00884838099258(21)390, (01)00884838099272(21)478, (01)00884838099203(21)159, (01)00884838085268(21)703978, (01)00884838099272(21)1106, (01)00884838085275(21)123, (01)00884838099241(21)1224, (01)00884838085268(21)704260, (01)00884838085268(21)703991, (01)00884838085268(21)703987, (01)00884838110564(21)262, (01)00884838110564(21)241, (01)00884838110564(21)243, (01)00884838099258(21)1131, (01)00884838085268(21)704250, (01)00884838099258(21)2008, (01)00884838085251(21)1032, (01)00884838085282(21)477, (01)00884838099258(21)1852, (01)00884838099258(21)2207, (01)00884838085251(21)1051, (01)00884838099234(21)515, (01)00884838085282(21)338, (01)00884838099272(21)546, (01)00884838110564(21)165, (01)00884838103276(21)35, (01)00884838085268(21)704195, (01)00884838085268(21)703672, (01)00884838085268(21)703467, (01)00884838085268(21)703468, (01)00884838099234(21)292, (01)00884838110564(21)464, (01)00884838099234(21)33, (01)00884838099234(21)173, (01)00884838099241(21)296, (01)00884838099241(21)297, (01)00884838085251(21)850, (01)00884838099241(21)263, (01)00884838085268(21)703386, (01)00884838099258(21)382, (01)00884838085268(21)703652, (01)00884838099258(21)3739, (01)00884838099258(21)2112, (01)00884838099234(21)543, (01)00884838085268(21)703714, (01)00884838099258(21)1859, (01)00884838110564(21)147, (01)00884838099210(21)184, (01)00884838099210(21)180, (01)00884838110564(21)248, (01)00884838103276(21)22, (01)00884838099234(21)219, (01)00884838099258(21)2371, (01)00884838103276(21)94, (01)00884838099210(21)131, (01)00884838110564(21)288, (01)00884838099258(21)855, (01)00884838099210(21)284, (01)00884838085251(21)696, (01)00884838085268(21)1197, (01)00884838085268(21)307, (01)00884838099258(21)853, (01)00884838085275(21)107, (01)00884838099272(21)650, (01)00884838085275(21)87, (01)00884838085275(21)85, (01)00884838099272(21)26, (01)00884838085268(21)703481, (01)00884838099272(21)883, (01)00884838099241(21)342, (01)00884838099258(21)724, (01)00884838085275(21)89, (01)00884838099258(21)1722, (01)00884838099210(21)376, (01)00884838099272(21)171, (01)00884838085367(21)127, (01)00884838085251(21)345, (01)00884838085275(21)88, (01)00884838085367(21)145, (01)00884838085268(21)604, (01)00884838099258(21)810, (01)00884838085251(21)575, (01)00884838099203(21)1, (01)00884838085275(21)86, (01)00884838099272(21)453, (01)00884838099241(21)539, (01)00884838099258(21)2886, (01)00884838099272(21)889, (01)00884838099258(21)1431, (01)00884838085275(21)90, (01)00884838099241(21)726, (01)00884838085268(21)419, (01)00884838085275(21)91, (01)00884838085367(21)299, (01)00884838099258(21)1373, (01)00884838099258(21)1480, (01)00884838099234(21)661, (01)00884838085268(21)724, (01)00884838099272(21)753, (01)00884838085268(21)1057, (01)00884838085268(21)1115, (01)00884838099258(21)688, (01)00884838099234(21)160, (01)00884838099258(21)3826, (01)00884838085268(21)771, (01)00884838099234(21)634, (01)00884838085268(21)1339, (01)00884838099234(21)82, (01)00884838099258(21)1227, (01)00884838085282(21)419, (01)00884838099258(21)1022, (01)00884838099258(21)697, (01)00884838085268(21)1176, (01)00884838099234(21)135, (01)00884838085268(21)1160, (01)00884838099234(21)133, (01)00884838085268(21)703490, (01)00884838099258(21)2006, (01)00884838099258(21)699, (01)00884838085268(21)703631, (01)00884838085268(21)703549, (01)00884838085268(21)703443, (01)00884838085268(21)1025, (01)00884838085268(21)1336, (01)00884838099258(21)2440, (01)00884838099241(21)494, (01)00884838085268(21)703339, (01)00884838085282(21)215, (01)00884838085268(21)1161, (01)00884838085282(21)164, (01)00884838099234(21)733, (01)00884838085282(21)434, (01)00884838099234(21)208, (01)00884838085251(21)939, (01)00884838085268(21)703352, (01)00884838099234(21)478, (01)00884838099258(21)992, (01)00884838099234(21)354, (01)00884838085282(21)114, (01)00884838099241(21)1260, (01)00884838085268(21)1091, (01)00884838099234(21)162, (01)00884838085268(21)1390, (01)00884838085268(21)1178, (01)00884838085268(21)703649, (01)00884838099258(21)1428, (01)00884838099227(21)132, (01)00884838099203(21)176, (01)00884838099227(21)131, (01)00884838099234(21)165, (01)00884838099241(21)319, (01)00884838099241(21)320, (01)00884838085282(21)327, (01)00884838085268(21)703531, (01)00884838085282(21)144, (01)00884838085275(21)79, (01)00884838085268(21)703763, (01)00884838085268(21)703762, (01)00884838099258(21)700, (01)00884838085268(21)703534, (01)00884838099258(21)2100, (01)00884838099272(21)673, (01)00884838085275(21)133, (01)00884838085268(21)703706, (01)00884838085268(21)703704, (01)00884838085268(21)703705, (01)00884838099210(21)214, (01)00884838099234(21)635, (01)00884838085251(21)869, (01)00884838099234(21)597, (01)00884838085268(21)703554, (01)00884838085282(21)227, (01)00884838099265(21)226, (01)00884838099258(21)2209, (01)00884838099258(21)701, (01)00884838085268(21)703444, (01)00884838099258(21)1242, (01)00884838085268(21)703353, (01)00884838099234(21)378, (01)00884838085282(21)295, (01)00884838099241(21)586, (01)00884838085268(21)1058, (01)00884838099241(21)300, (01)00884838085282(21)294, (01)00884838099234(21)466, (01)00884838085251(21)874, (01)00884838085275(21)81, (01)00884838085268(21)703787, (01)00884838085251(21)871, (01)00884838099272(21)844, (01)00884838085282(21)319, (01)00884838099210(21)276, (01)00884838099258(21)2517, (01)00884838099210(21)116, (01)00884838085275(21)140, (01)00884838099234(21)636, (01)00884838099258(21)3491, (01)00884838085268(21)1075, (01)00884838085268(21)703431, (01)00884838085268(21)827, (01)00884838099258(21)2342, (01)00884838099234(21)163, (01)00884838085282(21)241, (01)00884838085282(21)126, (01)00884838085268(21)703785, (01)00884838085268(21)979, (01)00884838099210(21)73, (01)00884838085268(21)1024, (01)00884838099234(21)168, (01)00884838085282(21)92, (01)00884838085282(21)284, (01)00884838085268(21)703410, (01)00884838085268(21)844, (01)00884838099234(21)479, (01)00884838099234(21)164, (01)00884838099210(21)365, (01)00884838099234(21)81, (01)00884838085268(21)1094, (01)00884838099234(21)228, (01)00884838099258(21)743, (01)00884838085268(21)1069, (01)00884838099241(21)276, (01)00884838099234(21)401, (01)00884838085268(21)703522, (01)00884838085282(21)175, (01)00884838085251(21)1010, (01)00884838099234(21)161, (01)00884838085282(21)199, (01)00884838099241(21)493, (01)00884838085268(21)703354, (01)00884838085282(21)372, (01)00884838099258(21)751, (01)00884838085268(21)703590, (01)00884838085268(21)1180, (01)00884838099210(21)398, (01)00884838099241(21)624, (01)00884838085282(21)188, (01)00884838099234(21)97, (01)00884838099258(21)2389, (01)00884838099234(21)139, (01)00884838099234(21)540, (01)00884838099234(21)121, (01)00884838099234(21)147, (01)00884838085282(21)429, (01)00884838099210(21)157, (01)00884838099210(21)87, (01)00884838099210(21)114, (01)00884838085282(21)296, (01)00884838085282(21)172, (01)00884838099234(21)141, (01)00884838099258(21)689, (01)00884838085268(21)1067, (01)00884838099258(21)2909, (01)00884838099258(21)1207, (01)00884838085268(21)723, (01)00884838085268(21)1159, (01)00884838085268(21)703795, (01)00884838085268(21)703570, (01)00884838085268(21)703454, (01)00884838085268(21)703722, (01)00884838085268(21)703721, (01)00884838085268(21)1131, (01)00884838099234(21)734, (01)00884838085268(21)1173, (01)00884838099258(21)3396, (01)00884838099272(21)276, (01)00884838099258(21)705, (01)00884838099258(21)1021, (01)00884838085268(21)703422, (01)00884838085268(21)703518, (01)00884838085268(21)1404, (01)00884838085268(21)1437, (01)00884838099210(21)274, (01)00884838085282(21)283, (01)00884838085282(21)357, (01)00884838099234(21)285, (01)00884838099258(21)1176, (01)00884838099210(21)399, (01)00884838099234(21)65, (01)00884838085268(21)1092, (01)00884838099234(21)140, (01)00884838099234(21)598, (01)00884838085268(21)703484, (01)00884838085282(21)306, (01)00884838085367(21)317, (01)00884838085268(21)1085, (01)00884838099272(21)649, (01)00884838099234(21)294, (01)00884838085282(21)437, (01)00884838085282(21)210, (01)00884838085282(21)315, (01)00884838099210(21)366, (01)00884838085268(21)703325, (01)00884838085268(21)568, (01)00884838099258(21)1465, (01)00884838085275(21)52, (01)00884838085268(21)1207, (01)00884838085268(21)703865, (01)00884838085282(21)134, (01)00884838085282(21)255, (01)00884838099210(21)135, (01)00884838099234(21)146, (01)00884838085275(21)48, (01)00884838085268(21)703403, (01)00884838085268(21)1068, (01)00884838099258(21)759, (01)00884838085268(21)703519, (01)00884838085268(21)1099, (01)00884838085268(21)1126, (01)00884838099234(21)175, (01)00884838085282(21)207, (01)00884838099258(21)2119, (01)00884838085268(21)703437, (01)00884838085367(21)571, (01)00884838085268(21)703916, (01)00884838085251(21)1029, (01)00884838085268(21)703907, (01)00884838085251(21)1030, (01)00884838099234(21)170, (01)00884838099234(21)637, (01)00884838099234(21)295, (01)00884838099234(21)290, (01)00884838085282(21)192, (01)00884838099258(21)698, (01)00884838085282(21)116, (01)00884838099234(21)134, (01)00884838099258(21)760, (01)00884838099234(21)136, (01)00884838085282(21)221, (01)00884838099258(21)1087, (01)00884838099210(21)159, (01)00884838085268(21)703451, (01)00884838099234(21)658, (01)00884838085282(21)189, (01)00884838085282(21)117, (01)00884838085268(21)1071, (01)00884838085282(21)313, (01)00884838099234(21)316, (01)00884838099227(21)209, (01)00884838099258(21)703, (01)00884838099258(21)704, (01)00884838099210(21)158, (01)00884838085268(21)1307, (01)00884838099234(21)289, (01)00884838085268(21)612, (01)00884838099234(21)686, (01)00884838099234(21)207, (01)00884838085282(21)187, (01)00884838099258(21)1068, (01)00884838085282(21)166, (01)00884838099234(21)492, (01)00884838085268(21)703373, (01)00884838099258(21)1407, (01)00884838099258(21)3682, (01)00884838099272(21)69, (01)00884838099265(21)330, (01)00884838085268(21)277, (01)00884838085350(21)102, (01)00884838099241(21)1111, (01)00884838085251(21)18, (01)00884838099258(21)1552, (01)00884838099258(21)1496, (01)00884838099258(21)2622, (01)00884838085350(21)84, (01)00884838085367(21)187, (01)00884838085350(21)105, (01)00884838085275(21)148, (01)00884838085268(21)917, (01)00884838085251(21)551, (01)00884838099258(21)1792, (01)00884838085282(21)96, (01)00884838099210(21)78, (01)00884838085275(21)55, (01)00884838099210(21)125, (01)00884838099210(21)218, (01)00884838099265(21)114, (01)00884838099258(21)691, (01)00884838099210(21)124, (01)00884838085282(21)135, (01)00884838099210(21)126, (01)00884838085251(21)1013, (01)00884838099272(21)1058, (01)00884838085251(21)917, (01)00884838085268(21)225, (01)00884838085367(21)183, (01)00884838099272(21)556, (01)00884838099241(21)292, (01)00884838099258(21)3709, (01)00884838099265(21)266, (01)00884838085350(21)99, (01)00884838099272(21)528, (01)00884838099272(21)660, (01)00884838099234(21)416, (01)00884838099272(21)74, (01)00884838099272(21)358, (01)00884838099241(21)368, (01)00884838099272(21)855, (01)00884838099272(21)568, (01)00884838099265(21)219, (01)00884838099265(21)20, (01)00884838099241(21)777, (01)00884838099258(21)477, (01)00884838099265(21)32, (01)00884838099265(21)155, (01)00884838099258(21)2991, (01)00884838099272(21)353, (01)00884838085367(21)45, (01)00884838085350(21)21, (01)00884838085268(21)256, (01)00884838099265(21)121, (01)00884838099272(21)286, (01)00884838099241(21)84, (01)00884838099265(21)192, (01)00884838099272(21)317, (01)00884838099258(21)556, (01)00884838099258(21)721, (01)00884838099265(21)117, (01)00884838099210(21)192, (01)00884838099272(21)42, (01)00884838099272(21)115, (01)00884838085268(21)1079, (01)00884838085367(21)331, (01)00884838099265(21)164, (01)00884838099258(21)677, (01)00884838099258(21)2083, (01)00884838099241(21)789, (01)00884838099241(21)788, (01)00884838099265(21)82, (01)00884838099272(21)229, (01)00884838099265(21)85, (01)00884838099272(21)285, (01)00884838099258(21)2068, (01)00884838099272(21)440, (01)00884838099265(21)194, (01)00884838099258(21)1157, (01)00884838099272(21)606, (01)00884838085268(21)918, (01)00884838099258(21)769, (01)00884838099265(21)132, (01)00884838099265(21)131, (01)00884838099272(21)156, (01)00884838099272(21)908, (01)00884838099258(21)3242, (01)00884838085367(21)106, (01)00884838099241(21)184, (01)00884838085251(21)543, (01)00884838099265(21)355, (01)00884838099272(21)43, (01)00884838099234(21)298, (01)00884838099265(21)222, (01)00884838085268(21)703365, (01)00884838099272(21)288, (01)00884838085282(21)108, (01)00884838099272(21)933, (01)00884838099258(21)280, (01)00884838099265(21)40, (01)00884838099258(21)646, (01)00884838099272(21)161, (01)00884838099258(21)168, (01)00884838099241(21)338, (01)00884838099258(21)384, (01)00884838085350(21)159, (01)00884838085367(21)465, (01)00884838085268(21)983, (01)00884838099272(21)162, (01)00884838085367(21)281, (01)00884838099258(21)3185, (01)00884838099265(21)353, (01)00884838085367(21)112, (01)00884838085350(21)49, (01)00884838085367(21)501, (01)00884838085251(21)516, (01)00884838085350(21)185, (01)00884838099258(21)917, (01)00884838085251(21)12, (01)00884838085367(21)393, (01)00884838085367(21)116, (01)00884838085350(21)51, (01)00884838099272(21)828, (01)00884838099258(21)1505, (01)00884838099258(21)61, (01)00884838099265(21)212, (01)00884838085367(21)273, (01)00884838099272(21)421, (01)00884838085350(21)239, (01)00884838085350(21)40, (01)00884838099272(21)818, (01)00884838085350(21)94, (01)00884838099272(21)540, (01)00884838085268(21)636, (01)00884838085350(21)63, (01)00884838099234(21)30, (01)00884838099258(21)392, (01)00884838099258(21)654, (01)00884838085350(21)96, (01)00884838085367(21)188, (01)00884838085350(21)129, (01)00884838099265(21)445, (01)00884838099265(21)106, (01)00884838085268(21)614, (01)00884838099258(21)568, (01)00884838099265(21)370, (01)00884838099265(21)44, (01)00884838099258(21)676, (01)00884838099258(21)562, (01)00884838085367(21)264, (01)00884838099265(21)317, (01)00884838099258(21)487, (01)00884838099272(21)73, (01)00884838085268(21)640, (01)00884838085251(21)416, (01)00884838099272(21)452, (01)00884838099258(21)1973, (01)00884838099272(21)405, (01)00884838099265(21)13, (01)00884838099272(21)46, (01)00884838099258(21)96, (01)00884838099258(21)142, (01)00884838099241(21)467, (01)00884838099272(21)694, (01)00884838085367(21)139, (01)00884838099265(21)430, (01)00884838085268(21)864, (01)00884838099272(21)722, (01)00884838099258(21)397, (01)00884838099241(21)107, (01)00884838085282(21)37, (01)00884838085367(21)503, (01)00884838085350(21)245, (01)00884838099265(21)213, (01)00884838099265(21)54, (01)00884838099272(21)583, (01)00884838085251(21)155, (01)00884838099272(21)784, (01)00884838099272(21)525, (01)00884838099272(21)312, (01)00884838085350(21)186, (01)00884838085251(21)446, (01)00884838085367(21)122, (01)00884838099258(21)1718, (01)00884838099265(21)245, (01)00884838099272(21)570, (01)00884838099265(21)244, (01)00884838099234(21)59, (01)00884838099272(21)160, (01)00884838099234(21)419, (01)00884838099241(21)166, (01)00884838099265(21)69, (01)00884838085350(21)130, (01)00884838085367(21)439, (01)00884838085268(21)1407, (01)00884838099265(21)319, (01)00884838099272(21)658, (01)00884838099265(21)64, (01)00884838085350(21)252, (01)00884838099265(21)166, (01)00884838099258(21)914, (01)00884838099265(21)165, (01)00884838099272(21)804, (01)00884838099265(21)224, (01)00884838099272(21)216, (01)00884838085367(21)135, (01)00884838099272(21)95, (01)00884838085268(21)288, (01)00884838099258(21)3239, (01)00884838099272(21)709, (01)00884838099241(21)1193, (01)00884838099272(21)803, (01)00884838099265(21)334, (01)00884838099272(21)856, (01)00884838099272(21)39, (01)00884838099265(21)19, (01)00884838099265(21)18, (01)00884838099272(21)853, (01)00884838099272(21)653, (01)00884838099258(21)3719, (01)00884838099258(21)2642, (01)00884838099272(21)502, (01)00884838099265(21)77, (01)00884838099272(21)70, (01)00884838099272(21)128, (01)00884838085350(21)161, (01)00884838099272(21)287, (01)00884838099258(21)137, (01)00884838099272(21)40, (01)00884838099272(21)1032, (01)00884838099265(21)76, (01)00884838099265(21)280, (01)00884838099272(21)655, (01)00884838099265(21)236, (01)00884838099272(21)89, (01)00884838099272(21)279, (01)00884838085350(21)120, (01)00884838099272(21)851, (01)00884838085367(21)248, (01)00884838099272(21)807, (01)00884838085350(21)170, (01)00884838085251(21)685, (01)00884838099272(21)581, (01)00884838099272(21)265, (01)00884838099241(21)898, (01)00884838099272(21)237, (01)00884838099272(21)106, (01)00884838085367(21)404, (01)00884838085251(21)715, (01)00884838085268(21)1304, (01)00884838085268(21)1302, (01)00884838085350(21)197, (01)00884838085350(21)195, (01)00884838085350(21)196, (01)00884838085268(21)1303, (01)00884838099272(21)304, (01)00884838099258(21)3679, (01)00884838085251(21)225, (01)00884838099265(21)161, (01)00884838099272(21)137, (01)00884838099265(21)311, (01)00884838085367(21)390, (01)00884838085367(21)263, (01)00884838099241(21)622, (01)00884838099272(21)110, (01)00884838099272(21)336, (01)00884838099258(21)2075, (01)00884838085367(21)90, (01)00884838099265(21)316, (01)00884838099265(21)10, (01)00884838099241(21)1062, (01)00884838099210(21)212, (01)00884838099272(21)257, (01)00884838085350(21)171, (01)00884838085367(21)300, (01)00884838085367(21)224, (01)00884838099272(21)68, (01)00884838085282(21)245, (01)00884838085350(21)200, (01)00884838099265(21)150, (01)00884838085268(21)1130, (01)00884838099234(21)11, (01)00884838099234(21)119, (01)00884838085367(21)67, (01)00884838099272(21)841, (01)00884838099265(21)119, (01)00884838085367(21)61, (01)00884838085367(21)157, (01)00884838099272(21)234, (01)00884838099272(21)885, (01)00884838099258(21)388, (01)00884838099272(21)654, (01)00884838099265(21)88, (01)00884838085367(21)154, (01)00884838099272(21)64, (01)00884838085350(21)193, (01)00884838099258(21)969, (01)00884838099265(21)190, (01)00884838099265(21)125, (01)00884838085367(21)100, (01)00884838085350(21)80, (01)00884838099272(21)916, (01)00884838099265(21)151, (01)00884838099272(21)3, (01)00884838085251(21)1082, (01)00884838099272(21)819, (01)00884838099241(21)1082, (01)00884838099265(21)281, (01)00884838085367(21)217, (01)00884838099265(21)75, (01)00884838099272(21)192, (01)00884838099272(21)193, (01)00884838085350(21)28, (01)00884838085268(21)1313, (01)00884838099272(21)618, (01)00884838099272(21)32, (01)00884838099272(21)292, (01)00884838085367(21)253, (01)00884838099272(21)206, (01)00884838099272(21)904, (01)00884838099265(21)211, (01)00884838099258(21)1430, (01)00884838099265(21)39, (01)00884838085350(21)229, (01)00884838099258(21)2988, (01)00884838099265(21)127, (01)00884838099272(21)49, (01)00884838099272(21)541, (01)00884838085268(21)647, (01)00884838099210(21)149, (01)00884838099272(21)646, (01)00884838099272(21)456, (01)00884838099265(21)369, (01)00884838099272(21)870, (01)00884838099272(21)258, (01)00884838099265(21)60, (01)00884838099258(21)2834, (01)00884838099265(21)272, (01)00884838099272(21)172, (01)00884838099234(21)28, (01)00884838099272(21)529, (01)00884838099258(21)1134, (01)00884838099265(21)193, (01)00884838099265(21)118, (01)00884838085268(21)1383, (01)00884838099272(21)38, (01)00884838085350(21)145, (01)00884838085268(21)794, (01)00884838099258(21)3620, (01)00884838085367(21)391, (01)00884838099272(21)866, (01)00884838099265(21)341, (01)00884838099234(21)144, (01)00884838099265(21)210, (01)00884838099241(21)200, (01)00884838099241(21)509, (01)00884838099265(21)86, (01)00884838099258(21)473, (01)00884838085268(21)651, (01)00884838085350(21)79, (01)00884838099258(21)1471, (01)00884838099203(21)11, (01)00884838099272(21)382, (01)00884838099265(21)170, (01)00884838099265(21)171, (01)00884838085268(21)692, (01)00884838099258(21)794, (01)00884838085367(21)144, (01)00884838099272(21)524, (01)00884838085367(21)379, (01)00884838085268(21)703357, (01)00884838099258(21)197, (01)00884838085350(21)93, (01)00884838085350(21)137, (01)00884838099272(21)659, (01)00884838099258(21)143, (01)00884838099272(21)33, (01)00884838099272(21)656, (01)00884838099265(21)61, (01)00884838099258(21)2614, (01)00884838099265(21)332, (01)00884838099272(21)801, (01)00884838099272(21)802, (01)00884838085350(21)50, (01)00884838099272(21)290, (01)00884838099265(21)361, (01)00884838099258(21)2042, (01)00884838085367(21)202, (01)00884838099272(21)30, (01)00884838085282(21)72, (01)00884838085268(21)487, (01)00884838085367(21)105, (01)00884838085350(21)44, (01)00884838099258(21)1211, (01)00884838099241(21)532, (01)00884838099265(21)115, (01)00884838099258(21)692, (01)00884838099272(21)92, (01)00884838085268(21)703363, (01)00884838085367(21)226, (01)00884838099265(21)307, (01)00884838099241(21)653, (01)00884838099265(21)221, (01)00884838085251(21)913, (01)00884838099272(21)269, (01)00884838099258(21)2413, (01)00884838099265(21)198, (01)00884838085350(21)38, (01)00884838085367(21)315, (01)00884838099265(21)35, (01)00884838085268(21)500, (01)00884838099265(21)59, (01)00884838085367(21)464, (01)00884838099272(21)240, (01)00884838085367(21)197, (01)00884838099265(21)36, (01)00884838085268(21)980, (01)00884838085367(21)63, (01)00884838099265(21)363, (01)00884838085251(21)582, (01)00884838099258(21)1065, (01)00884838099265(21)154, (01)00884838085251(21)908, (01)00884838085367(21)252, (01)00884838099272(21)163, (01)00884838085367(21)175, (01)00884838099272(21)852, (01)00884838099258(21)1177, (01)00884838099265(21)62, (01)00884838099265(21)15, (01)00884838099272(21)62, (01)00884838085367(21)274, (01)00884838085268(21)703452, (01)00884838099272(21)1034, (01)00884838099265(21)57, (01)00884838085367(21)163, (01)00884838085350(21)175, (01)00884838085350(21)227, (01)00884838085367(21)186, (01)00884838099272(21)436, (01)00884838099258(21)1006, (01)00884838099265(21)335, (01)00884838099272(21)806, (01)00884838099258(21)2896, (01)00884838099272(21)210, (01)00884838085350(21)167, (01)00884838099265(21)279, (01)00884838085367(21)260, (01)00884838085367(21)502, (01)00884838099241(21)1096, (01)00884838099272(21)187, (01)00884838099265(21)191, (01)00884838085367(21)152, (01)00884838099272(21)264, (01)00884838099241(21)695, (01)00884838085251(21)19, (01)00884838099234(21)68, (01)00884838099258(21)1582, (01)00884838085268(21)463, (01)00884838099265(21)160, (01)00884838085350(21)78, (01)00884838099265(21)275, (01)00884838099258(21)2062, (01)00884838099272(21)578, (01)00884838099272(21)928, (01)00884838099241(21)592, (01)00884838099265(21)94, (01)00884838099265(21)95, (01)00884838085367(21)498, (01)00884838099258(21)150, (01)00884838099265(21)23, (01)00884838099265(21)139, (01)00884838099272(21)847, (01)00884838085350(21)86, (01)00884838085367(21)156, (01)00884838099272(21)90, (01)00884838099272(21)587, (01)00884838099241(21)884, (01)00884838085251(21)435, (01)00884838085251(21)16, (01)00884838099272(21)499, (01)00884838099265(21)197, (01)00884838099265(21)402, (01)00884838099272(21)154, (01)00884838099258(21)557, (01)00884838099258(21)2558, (01)00884838085350(21)91, (01)00884838099272(21)273, (01)00884838085350(21)189, (01)00884838085367(21)621, (01)00884838099272(21)186, (01)00884838085350(21)198, (01)00884838085367(21)296, (01)00884838099265(21)11, (01)00884838099265(21)391, (01)00884838099258(21)1723, (01)00884838099258(21)3451, (01)00884838099234(21)55, (01)00884838099272(21)34, (01)00884838099272(21)44, (01)00884838099272(21)65, (01)00884838085268(21)26, (01)00884838085251(21)11, (01)00884838099272(21)71, (01)00884838085268(21)541, (01)00884838085350(21)37, (01)00884838099241(21)633, (01)00884838099272(21)558, (01)00884838099265(21)120, (01)00884838099272(21)425, (01)00884838099265(21)58, (01)00884838099272(21)196, (01)00884838099265(21)34, (01)00884838099272(21)91, (01)00884838099272(21)657, (01)00884838099272(21)35, (01)00884838099272(21)356, (01)00884838085268(21)893, (01)00884838099272(21)500, (01)00884838099258(21)1400, (01)00884838085367(21)208, (01)00884838099265(21)21, (01)00884838085350(21)72, (01)00884838099265(21)329, (01)00884838099272(21)537, (01)00884838099258(21)764, (01)00884838085367(21)430, (01)00884838085367(21)161, (01)00884838099265(21)322, (01)00884838099258(21)2035, (01)00884838099258(21)2036, (01)00884838099258(21)910, (01)00884838085350(21)199, (01)00884838099272(21)375, (01)00884838099265(21)67, (01)00884838099241(21)327, (01)00884838099272(21)567, (01)00884838099272(21)668, (01)00884838099272(21)454, (01)00884838099265(21)278, (01)00884838085367(21)436, (01)00884838085268(21)1145, (01)00884838099258(21)2521, (01)00884838085367(21)409, (01)00884838085268(21)373, (01)00884838099272(21)479, (01)00884838085350(21)88, (01)00884838085367(21)155, (01)00884838085251(21)167, (01)00884838099272(21)31, (01)00884838085268(21)294, (01)00884838099272(21)450, (01)00884838099265(21)162, (01)00884838099272(21)496, (01)00884838099265(21)90, (01)00884838099265(21)89, (01)00884838099272(21)302, (01)00884838099258(21)1187, (01)00884838085268(21)635, (01)00884838099272(21)80, (01)00884838085367(21)298, (01)00884838099272(21)406, (01)00884838099258(21)2657, (01)00884838099234(21)27, (01)00884838085350(21)27, (01)00884838099272(21)857, (01)00884838099258(21)1421, (01)00884838085350(21)121, (01)00884838099265(21)92, (01)00884838099258(21)2626, (01)00884838099265(21)337, (01)00884838099258(21)2832, (01)00884838099265(21)91, (01)00884838099258(21)83, (01)00884838085275(21)113, (01)00884838085367(21)191, (01)00884838099258(21)157, (01)00884838099265(21)137, (01)00884838099258(21)170, (01)00884838099272(21)41, (01)00884838085367(21)278, (01)00884838099272(21)770, (01)00884838099265(21)93, (01)00884838099265(21)108, (01)00884838099265(21)73, (01)00884838085350(21)179, (01)00884838085350(21)183, (01)00884838085268(21)768, (01)00884838085367(21)368, (01)00884838099272(21)56, (01)00884838099265(21)371, (01)00884838099234(21)233, (01)00884838099258(21)1515, (01)00884838085350(21)117, (01)00884838085350(21)118, (01)00884838085350(21)85, (01)00884838099265(21)179, (01)00884838099258(21)718, (01)00884838085367(21)94, (01)00884838099272(21)155, (01)00884838099241(21)749, (01)00884838099272(21)619, (01)00884838099258(21)455, (01)00884838099272(21)177, (01)00884838085367(21)66, (01)00884838099265(21)360, (01)00884838099241(21)394, (01)00884838099265(21)143, (01)00884838099258(21)2973, (01)00884838099272(21)811, (01)00884838099265(21)359, (01)00884838099272(21)96, (01)00884838099272(21)28, (01)00884838099265(21)303, (01)00884838099272(21)560, (01)00884838099272(21)433, (01)00884838085367(21)108, (01)00884838099265(21)163, (01)00884838099258(21)2315, (01)00884838099265(21)286, (01)00884838099272(21)308, (01)00884838085367(21)201, (01)00884838085367(21)200, (01)00884838085350(21)90, (01)00884838085350(21)89, (01)00884838099272(21)998, (01)00884838099265(21)432, (01)00884838099265(21)379, (01)00884838099272(21)324, (01)00884838085350(21)119, (01)00884838099265(21)138, (01)00884838099258(21)1646, (01)00884838085367(21)98, (01)00884838085367(21)189, (01)00884838085350(21)87, (01)00884838099265(21)325, (01)00884838085251(21)664, (01)00884838099272(21)157, (01)00884838099258(21)731, (01)00884838099241(21)1215, (01)00884838085251(21)901, (01)00884838099265(21)100, (01)00884838099272(21)357, (01)00884838099258(21)879, (01)00884838099272(21)340, (01)00884838085350(21)31, (01)00884838099265(21)247, (01)00884838099272(21)588, (01)00884838099258(21)889, (01)00884838085268(21)929, (01)00884838085350(21)110, (01)00884838085350(21)178, (01)00884838085367(21)311, (01)00884838085350(21)73, (01)00884838085251(21)508, (01)00884838099265(21)98,  (01)00884838085367(21)367, (01)00884838085350(21)194, (01)00884838099258(21)3217, (01)00884838099272(21)934, (01)00884838085268(21)609, (01)00884838099265(21)16, (01)00884838099258(21)809, (01)00884838099272(21)79, (01)00884838099265(21)27, (01)00884838085350(21)39, (01)00884838099258(21)3589, (01)00884838099265(21)80, (01)00884838099272(21)77, (01)00884838099258(21)118, (01)00884838085251(21)530, (01)00884838099241(21)943, (01)00884838099272(21)476, (01)00884838099272(21)236, (01)00884838099258(21)2802, (01)00884838099265(21)282, (01)00884838099258(21)212, (01)00884838099265(21)42, (01)00884838085251(21)198, (01)00884838099272(21)146, (01)00884838099265(21)105, (01)00884838099272(21)863, (01)00884838085350(21)56, (01)00884838099265(21)377, (01)00884838099272(21)165, (01)00884838099272(21)402, (01)00884838099241(21)971, (01)00884838099272(21)652, (01)00884838099265(21)424, (01)00884838085367(21)297, (01)00884838085350(21)139, (01)00884838085350(21)61, (01)00884838085367(21)132, (01)00884838099265(21)74, (01)00884838099265(21)277, (01)00884838099265(21)63, (01)00884838099265(21)323, (01)00884838099272(21)241, (01)00884838099258(21)144, (01)00884838099265(21)240, (01)00884838099265(21)70, (01)00884838085367(21)507, (01)00884838099265(21)12, (01)00884838099265(21)9, (01)00884838085367(21)451, (01)00884838099258(21)538, (01)00884838099265(21)204, (01)00884838099265(21)238."
    }
  ]
}