      "status": "Ongoing",
      "city": "Pymble",
      "state": "",
      "country": "Australia",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99372",
      "recalling_firm": "Kico Knee Innovation Company",
      "address_1": "Suite 3, Building 1, 20 Bridge St",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Consignee Consignee Number  Region Device  Quantity RFK MEDICAL  CE102 RICHMOND VA  IN-29200, ARVIS SURGICAL NAVIGATION 1 VALOR MEDICAL  CE118 SAN ANTONIO TX  IN-29200, ARVIS SURGICAL NAVIGATION  2 JOINT VENTURES MEDICAL CE121 OSHKOSH WI  IN-29200, ARVIS SURGICAL NAVIGATION  1 CASTLE MEDICAL  CE122 WATERLOO IA  IN-29200, ARVIS SURGICAL NAVIGATION  6 GREAT LAKES MEDICAL GL108 NAPERVILLE IL   IN-29200, ARVIS SURGICAL NAVIGATION  2 CERTUS MEDICAL  GL112 INDIANAPOLIS IN   IN-29200, ARVIS SURGICAL NAVIGATION  5 I&S MEDICAL  GL113 LOUISVILLE KY   IN-29200, ARVIS SURGICAL NAVIGATION  2 PROCORE MEDICAL MA102 PIEDMONT SC  IN-29200, ARVIS SURGICAL NAVIGATION  1 FOX MEDTECH  MW001 ENGLEWOOD CO   IN-29200, ARVIS SURGICAL NAVIGATION  2 MEDSOURCE  NE131 TIVERTON RI   IN-29200, ARVIS SURGICAL NAVIGATION  4 IMPLANT MEDICAL SALES NE190 HICKSVILLE NY   IN-29200, ARVIS SURGICAL NAVIGATION  2 KINETIC MEDICINE NW105  OMAHA NE   IN-29200, ARVIS SURGICAL NAVIGATION  4 CALSURG NW115 SANTA BARBARA CA  IN-29200, ARVIS SURGICAL NAVIGATION  4 MIDWEST MEDICAL RESOURCES SC181 OVERLAND PARK KS  IN-29200, ARVIS SURGICAL NAVIGATION  3 ORTHO & SURGICAL SOLUTIONS SE175 GUAYNABO PR  IN-29200, ARVIS SURGICAL NAVIGATION  3 JOINT MEDICAL SOLUTIONS SE180 BIRMINGHAM AL  IN-29200, ARVIS SURGICAL NAVIGATION  4 FLT ORTHOPEDICS SE216 TAMPA FL   IN-29200, ARVIS SURGICAL NAVIGATION  6 MAC MEDICAL  W180 GILBERT AZ   IN-29200, ARVIS SURGICAL NAVIGATION  1 CP ORTHO  W214 HONOLULU HI   IN-29200, ARVIS SURGICAL NAVIGATION  1 UNIFIED ORTHOPEDICS W223 RIVERSIDE CA   IN-29200, ARVIS SURGICAL NAVIGATION  18 MOVEMENT MEDICAL W225 SEATTLE WA   IN-29200, ARVIS SURGICAL NAVIGATION  6 Total number of devices in the field 78",
      "recall_number": "Z-2957-2026",
      "product_description": "Brand Name: Kico Knee Innovation  Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2  Model/Catalog Number: IN-29200  Software Version: ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200)  ARVIS Shoulder Software v2.0.3 (IN-23630)  Product Description: The ARVIS system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components.          The ARVIS Surgical Navigation System is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements such as changes in leg length in Hip Arthroplasty. The system is compatible with straight acetabular impactors and with specific offset impactors, identified in the instructions for use, for which an adapter has been validated. Example orthopedic surgical procedures include but are not limited to:      \" Total Knee Arthroplasty      \" Unicompartmental Knee Arthroplasty: Tibial Transverse Resection      \" Hip Arthroplasty          The ARVIS\" head mounted display is for displaying augmented reality visualization and information to the user intraoperatively. The augmented/virtual displayed information should not be relied upon solely for absolute positional/alignment information and should always be used in conjunction with the displayed stereotaxic information.  Component: NA",
      "product_quantity": "78",
      "reason_for_recall": "Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing",
      "recall_initiation_date": "20260713",
      "center_classification_date": "20260814",
      "report_date": "20260826",
      "code_info": "Lot Code: GMDN - 45200 UDI-DI -  00810180351287  Software ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200) ARVIS Shoulder Software v2.0.3 (IN-23630)  Serial Number 6006A1118 6006A1036C3 6006A1117 6006A1045C2 6006A1106 6006A1126 6006A1149 6006A1034C2 6006A1046C2 6006A1086C2 6006A1093 6006A1079C3 6006A1096C2 6006A1052C2 6006A1058C3 6006A1030C2 6006A1088 6006A1100C 6006A1094 6006A1041C2 6006A1060C3 6006A1092C 6006A1105 6006A1140 6006A1040C2 6006A1090 6006A1116 6006A1109 6006A1037C2 6006A1053C3 6006A1102 6006A1108 6006A1120 6006A1155 6006A1127 6006A1139 6006A1095C2 6006A1131 6006A1132 6006A1142 6006A1152 6006A1193 6006A1194 6006A1208 6006A1209 6006A1107 6006A1064C 6006A1068C 6006A1081C 6006A1011C2  6006A1104 6006A1101 6006A1087C 6006A1075C 6006A1035C2 6006A1057C2 6006A1044C2 6006A1024C2 6006A1059C3 6006A1083C3 6006A1119 6006A1124 6006A1128 6006A1071C2 6006A1089 6006A1084 6006A1042C2 6006A1078C2 6006A1082C2 6006A1027C2 6006A1010C2 6006A1125 6006A1147 6006A1025C2 6006A1077C 6006A1076C",
      "more_code_info": ""
    }
  ]
}