  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-07",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Loudon",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99371",
      "recalling_firm": "Fort Defiance Industries, LLC",
      "address_1": "2411 Maremont Pkwy",
      "address_2": "",
      "postal_code": "37774-1066",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.",
      "recall_number": "Z-2983-2026",
      "product_description": "Fort Defiance Industries (FDI),  Automated Field Steam Sterilizer  REF: P2131",
      "product_quantity": "",
      "reason_for_recall": "The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.",
      "recall_initiation_date": "20260708",
      "center_classification_date": "20260818",
      "report_date": "20260826",
      "code_info": "Lot Code: All serial numbers/  GTIN: (01) 00862992000327",
      "more_code_info": ""
    }
  ]
}