  "results": [
    {
      "status": "Ongoing",
      "city": "Marietta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99341",
      "recalling_firm": "Erbe USA Inc",
      "address_1": "2225 Northwest Pkwy SE",
      "address_2": "",
      "postal_code": "30067-8764",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia,   Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.",
      "recall_number": "Z-2938-2026",
      "product_description": "ERBECRYO 2 Cryosurgical Unit  Model/Catalog Number: 10402-000",
      "product_quantity": "1126 units",
      "reason_for_recall": "A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector  resulting in CO2 leaking into  the return path, leading to excessive pressure causing  the cryoprobe to rupture.",
      "recall_initiation_date": "20260626",
      "center_classification_date": "20260812",
      "report_date": "20260819",
      "code_info": "UDI-DI 04050147013797/  Serial Numbers:  All serial numbers up to 11608268 Software Version: 1.0.3",
      "more_code_info": ""
    }
  ]
}