  "results": [
    {
      "status": "Ongoing",
      "city": "Garden City",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100222",
      "recalling_firm": "Beyond Air Inc.",
      "address_1": "900 Stewart Ave",
      "address_2": "Ste 301",
      "postal_code": "11530-4869",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of California, Tennessee, Arkansas, Nebraska, New York, North Carolina, Delaware, New Jersey, Texas, Indiana, Minnesota, Illinois, Oklahoma, Georgia, Florida, Utah and the countries of Canada, Hong Kong, Australia, Manila, Vietnam, Israel, Taiwan, Panama, Singapore.",
      "recall_number": "Z-0436-2027",
      "product_description": "NO Calibration Gas Cylinder, 45ppm;  Model/Catalog Number: 80006;  Software Version: N/A - No firmware/software;  Product Description: Nitric Oxide Calibration Gas, 45 ppm   Intended to be used with LungFit PH device (FDA PMA 200044) during pre verification checks and monthly maintenance to calibration internal NO sensors.  Component: No;",
      "product_quantity": "181",
      "reason_for_recall": "Some NO calibration tanks may have been contaminated with O2 (from the air).",
      "recall_initiation_date": "20261020",
      "center_classification_date": "20261112",
      "report_date": "20261118",
      "code_info": "The Beyond Air product code is 8006;   Mesa Gas part number is: Z166045; UDI-DI: N/A; Cylinder Numbers: C418368, C547918, C385448, C249243, C391758, C391775, C418404, C385315, C418603, C418303, C418750, C391573, C385388, C391749, C418707, C391651, C548078, C418298, C385354, C385373, C391773, C391805, C418606, C418313, C418346, C418720, C391819, C391910, C547921, C391636, C391820, C391907, C547912, C547896, C391928, C548077, C385314, C418675, C385425, C548059, C391911, C391800, C418676, C548069, C548034, C547980, C391745, C391801, C391772, C418627, C418427, C418406, C418376, C418663, C391752, C418677, C391703, C141887, C385130, C391762, C418698, C385372, C385440, C547891, C385337, C385353, C418684, C385441, C391803, C391767, C547881, C418390, C385385, C391717, C418494, C391757, C418349, C418751, C391886, C391701, C418296, C418291, C418685, C391724, C391864, C391927, C385142, C385331, C385131, C391768, C391779, C391614, C418514, C418403, C418353, C418604, C418742, C391885, C547934, C547894, C547933, C547908, C547885, C385358, C418607, C391750, C391780, C391753, C391751, C391740, C391746, C391571, C418686, C418697, C418708, C418719, C418314, C418355, C418743, C418741, C418678, C548057, C548075, C548068, C548061, C548056, C548063, C548074, C548066, C548064;",
      "more_code_info": ""
    }
  ]
}