  "results": [
    {
      "status": "Ongoing",
      "city": "New York",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100141",
      "recalling_firm": "VB SPINE LLC",
      "address_1": "505 Park Ave Fl 14",
      "address_2": "",
      "postal_code": "10022-1106",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: Nationwide Distribution;  International: Australia, Belgium, France, Ireland, Italy, Japan, Netherlands, Spain, United Kingdom;",
      "recall_number": "Z-0440-2027",
      "product_description": "NITINOL K-WIRE BLUNT;  Catalog/Part Number: 48280232; used in spinal surgery",
      "product_quantity": "12268",
      "reason_for_recall": "Potential for the K-wire to fracture during use resulting in procedural delay and possibility of a piece of fractured K-wire remaining in the patient's bone.",
      "recall_initiation_date": "20261002",
      "center_classification_date": "20261112",
      "report_date": "20261118",
      "code_info": "Catalog/Part Number: 48280232; UDI-DI: 07613327096200; Serial/Lot Number(s) Affected: 25A459, 25A460, 25A461, 25A462, 25A463, 25A464, 25A465, 25A466, 25A467, 25A468, 25A469, 25A470, 25A471, 25A472, 25A473, 25A474, 25A475, 25A476, 25A477, 25A478, 25A479, 25A480, 25A481, 25A482, 25A483, 25A484, 25A485, 25A486, 25A487, 25A488, 25A489, 25A490, 25A491, 25A492, 25A493, 25A494, 25A495, 25A496, 25A497, 25A498, 25A499, 25A500, 25A501, 25A502, 25A503, 25A504, 25A505, 25A506, 25A507, 25A508, 25A509, 25A513, 25A512, 25A514, 25A515, 25A511, 25A510, 25A516, 25A517, 25A518, 25A519, 25A522, 25A525, 25A520, 25A521, 25A523, 25A524, 25A526;",
      "more_code_info": ""
    }
  ]
}