      "status": "Ongoing",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100116",
      "recalling_firm": "Hologic, Inc",
      "address_1": "250 Campus Dr",
      "address_2": "",
      "postal_code": "01752-3020",
      "voluntary_mandated": "",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Dominican Republic, France, Germany, Greece, India, Iran, Islamic Republic of, Israel, Italy, Japan, Korea, Republic of, Lebanon, Malaysia, Mexico, Morocco, Nepal, New Zealand, Pakistan, Peru, Philippines, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-0423-2027",
      "product_description": "Horizon QDR X-Ray Bone Densitometer;  Models: 1. Horizon W, 2. Horizon-WI;",
      "product_quantity": "198",
      "reason_for_recall": "A nonconformance was identified for systems manufactured from December 2024 to January 2025 that were shipped without the validated VGA cables. Some installations were completed using VGA cables that may not meet Hologic s applicable performance and safety standards for this system.",
      "recall_initiation_date": "20261020",
      "center_classification_date": "20261112",
      "report_date": "20261118",
      "code_info": "1. Model Horizon W;  UDI:15420045505834; Serial Numbers: 308912M, 308911M, 308909M, 308908M, 308905M, 308899M, 308892M, 308890M, 308887M, 308925M, 308921M, 308918M, 308969M, 308968M, 308965M, 308964M, 308960M, 308956M, 308947M, 308945M, 308944M, 308942M, 308940M, 308939M, 308936M, 308934M, 308932M, 308928M, 309026M, 309017M, 309033M, 309029M, 309066M, 309045M, 309071M, 309048M, 309094M, 309109M, 309106M, 309104M, 309099M, 309146M, 309144M, 309136M, 309133M, 309126M, 309122M, 309117M, 309113M, 309184M, 309181M, 309177M, 309150M, 309704M, 309702M, 309700M, 309709M, 309707M, 309719M, 309715M, 309713M, 309741M, 309740M, 309738M, 309736M, 309735M, 309734M, 309733M, 309732M, 309731M, 309730M, 309729M, 309727M, 309726M, 309725M, 309724M, 309723M, 309722M, 309717M, 309711M, 308823M, 309743M, 309739M, 309751M, 309749M, 309747M, 309745M;  2. Model Horizon WI;  UDI: 15420045505865; Serial Numbers: 308848M, 308823M, 308818M, 308866M, 308862M, 308858M, 308855M, 308906M, 308902M, 308889M, 308884M, 309003M, 309002M, 309001M, 308946M, 308943M, 308941M, 308938M, 308937M, 308933M, 309008M, 309007M, 309006M, 309005M, 309004M, 309027M, 309023M, 309041M, 309037M, 309035M, 309032M, 309030M, 309069M, 309064M, 309062M, 309043M, 309039M, 309082M, 309080M, 309078M, 309076M, 309072M, 309070M, 309084M, 309187M, 309180M, 309176M, 309197M, 309196M, 309194M, 309193M, 309191M, 309189M, 309244M, 309243M, 309241M, 309240M, 309236M, 309265M, 309263M, 309251M, 309282M, 309255M, 309303M, 309298M, 309296M, 309293M, 309315M, 309312M, 309309M, 309306M, 309329M, 309326M, 309320M, 309346M, 309343M, 309340M, 309338M, 309336M, 309335M, 309362M, 309357M, 309356M, 309354M, 309374M, 309370M, 309359M, 309385M, 309379M, 309675M, 309703M, 309701M, 309699M, 309708M, 309705M, 309718M, 309716M, 309728M, 309742M, 309737M, 309721M, 309744M, 309714M, 309712M, 309710M, 309752M, 309750M, 309748M, 309746M, 309754M, 309753M;",
      "more_code_info": ""
    }
  ]
}