  "results": [
    {
      "status": "Ongoing",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100092",
      "recalling_firm": "TransMedics, Inc.",
      "address_1": "200 Minuteman Rd Ste 302",
      "address_2": "",
      "postal_code": "01810-1046",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the state of NE and the country of Italy.",
      "recall_number": "Z-0391-2027",
      "product_description": "OCS Liver System for organ transport   OCS Solution Delivery Cassettes   Model/Catalog Number: 1467     Product Description: OCS Solution Delivery Cassettes   The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes.  The BD syringe is a component of the SDS Cassette PN 100004459",
      "product_quantity": "55 OCS Solution Delivery Cassettes",
      "reason_for_recall": "Faulty Solution Delivery System (SDS) Cassettes  within the device set were assembled with syringes that  were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.",
      "recall_initiation_date": "20251120",
      "center_classification_date": "20261106",
      "report_date": "20261118",
      "code_info": "Lot Code: Lot: 4460-4-737189;  Expiration Date: 2028-10-08;  UDI: +B00944605/$$74460-4-737189/14D20281008/Q3L;   Lot: 4460-4-738008;  Expiration Date: 2028-11-07;  UDI: +B00944605/$$74460-4-738008/14D20281107/Q3C;   Lot: 4460-4-738009;  Expiration Date: 2028-11-07;  UDI: +B00944605/$$74460-4-738009/14D20281107/Q3D;   Lot: 4460-4-739728;  Expiration Date: 2029-01-14;  UDI: +B00944605/$$74460-4-739728/14D20290114/Q3K;",
      "more_code_info": ""
    }
  ]
}