  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100091",
      "recalling_firm": "Tandem Diabetes Care, Inc.",
      "address_1": "12400 High Bluff Dr",
      "address_2": "",
      "postal_code": "92130-3077",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide and the country of Czech Republic.",
      "recall_number": "Z-0437-2027",
      "product_description": "Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app)  Model/Catalog Number: 1015472",
      "product_quantity": "4086 users",
      "reason_for_recall": "A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.",
      "recall_initiation_date": "20261013",
      "center_classification_date": "20261112",
      "report_date": "20261118",
      "code_info": "Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1",
      "more_code_info": ""
    }
  ]
}