  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100070",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "",
      "distribution_pattern": "USA and OUS.",
      "recall_number": "Z-0327-2027",
      "product_description": "Senographe Pristina",
      "product_quantity": "",
      "reason_for_recall": "GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following:  the collimator blades might be visible in the image, or the collimator might be open more than expected, and  an incorrect collimator filter might be used with respect to the anode track selected  This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).",
      "recall_initiation_date": "20260828",
      "center_classification_date": "20261104",
      "report_date": "20261111",
      "code_info": "GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.",
      "more_code_info": ""
    }
  ]
}