  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100063",
      "recalling_firm": "TransMedics, Inc.",
      "address_1": "200 Minuteman Rd Ste 302",
      "address_2": "",
      "postal_code": "01810-1046",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.",
      "recall_number": "Z-0381-2027",
      "product_description": "OCS Liver Perfusion Set   Model/Catalog Number: 3200   Software Version: rev 3    OCS Liver Perfusion Module included in the Liver Perfusion Set used for organ transport.",
      "product_quantity": "18",
      "reason_for_recall": "The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.",
      "recall_initiation_date": "20260515",
      "center_classification_date": "20261106",
      "report_date": "20261118",
      "code_info": "1. Perfusion Set SN: 4470-4-24374; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24374/14D202711100*;  2. Perfusion Set SN: 4470-4-23845; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27;  UDI: *+B00944705/$$+74470-4-23845/14D202710235*;   3. Perfusion Set SN: 4470-4-24226; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27;  UDI: *+B00944705/$$+74470-4-24226/14D20271110$*;   4. Perfusion Set SN: 4470-4-24559; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24559/14D202711026*;  5. Perfusion Set SN: 4470-4-24636; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24636/14D202711022*;  6. Perfusion Set SN: 4470-4-24508; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24508/14D202711020*;  7. Perfusion Set SN: 4470-4-24384; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24384/14D202711022*;  8. Perfusion Set SN: 4470-4-23182; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23182/14D20271023%*;  9. Perfusion Set SN: 4470-4-23650; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23650/14D202710194*;  10. Perfusion Set SN: 4470-4-23593; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23593/14D20271019A*;  11. Perfusion Set SN: 4470-4-23471; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23471/14D202710195*;  12. Perfusion Set SN: 4470-4-22403; Description: OCS Liver Perfusion Set; EXP Date: 9/10/27; UDI: *+B00944705/$$+74470-4-22403/14D20270910+*;  13. Perfusion Set SN: 4470-4-22433; Description: OCS Liver Perfusion Set; EXP Date: 9/3/27; UDI: *+B00944705/$$+74470-4-22433/14D202709033*;  14. Perfusion Set SN: 4470-4-22303; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22303/14D20270820/*;  15. Perfusion Set SN: 4470-4-22280; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22280/14D202708201*;",
      "more_code_info": ""
    }
  ]
}