  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Durham",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "100047",
      "recalling_firm": "restor3d Inc.",
      "address_1": "4001 E Nc 54 Hwy",
      "address_2": "Suite 3160",
      "postal_code": "27709",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.",
      "recall_number": "Z-0348-2027",
      "product_description": "Veritas Reverse Total Shoulder Replacement,  Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1,  size 1 -2, 40mm Polymer Liner",
      "product_quantity": "240 units",
      "reason_for_recall": "(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.",
      "recall_initiation_date": "20261001",
      "center_classification_date": "20261103",
      "report_date": "20261111",
      "code_info": "All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409",
      "more_code_info": ""
    }
  ]
}