  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99279",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.",
      "recall_number": "Z-2726-2026",
      "product_description": "Optima XR200amx, Mobile Digital-Ready Radiographic System",
      "product_quantity": "4 units",
      "reason_for_recall": "GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.",
      "recall_initiation_date": "20260605",
      "center_classification_date": "20260713",
      "report_date": "20260722",
      "code_info": "UDI-DI: 00840682115278; System IDs: 419383DP3, 773792XR200, 856355OPT, FEB520100",
      "more_code_info": ""
    }
  ]
}