  "results": [
    {
      "status": "Ongoing",
      "city": "Northridge",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99228",
      "recalling_firm": "Medtronic MiniMed, Inc.",
      "address_1": "18000 Devonshire St",
      "address_2": "N/A",
      "postal_code": "91325-1219",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.",
      "recall_number": "Z-2778-2026",
      "product_description": "MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)",
      "product_quantity": "3,465",
      "reason_for_recall": "For MiniMed Go app version 1.0.0 and 1.1.0, when the  Set day and night  is turned OFF and  missed dose  or  Correct high glucose  alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.",
      "recall_initiation_date": "20260616",
      "center_classification_date": "20260723",
      "report_date": "20260729",
      "code_info": "Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0  Software Versions: 1.0.0, 1.1.0"
    }
  ]
}