  "results": [
    {
      "status": "Ongoing",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99219",
      "recalling_firm": "Think Surgical, Inc.",
      "address_1": "47201 Lakeview Blvd",
      "address_2": "N/A",
      "postal_code": "94538-6530",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, TX, UT, VA, and WA.",
      "recall_number": "Z-2716-2026",
      "product_description": "TMINI Femur Tracking Array REF 109003  UDI-DI code: M9461090030    TMINI Tibia Tracking Array REF 109004  UDI-DI code: M9461090040    TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components",
      "product_quantity": "360 devices",
      "reason_for_recall": "Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.",
      "recall_initiation_date": "20260508",
      "center_classification_date": "20260709",
      "report_date": "20260715",
      "code_info": "Model Number: 109003 UDI-DI code: M9461090030   Model Number: 109004 UDI-DI code: M9461090040"
    }
  ]
}