      "status": "Ongoing",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99141",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Eugeniavagen 18c",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2646-2026",
      "product_description": "RayCare 2024A SP1. Oncology Information Systems.",
      "product_quantity": "3 units",
      "reason_for_recall": "Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.",
      "recall_initiation_date": "20260520",
      "center_classification_date": "20260701",
      "report_date": "20260708",
      "code_info": "Software Version 8.0.1.60193, UDI 0735000201083920240906, Expiration Date: 2030-01-24"
    }
  ]
}