  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-09",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Palm Harbor",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99102",
      "recalling_firm": "Oscor Inc.",
      "address_1": "4875 Palm Harbor Blvd",
      "address_2": "",
      "postal_code": "34683-1217",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Distribution in the US state of MA",
      "recall_number": "Z-2678-2026",
      "product_description": "Abiomed REF: 0052-3025, Introducer Kit for Impella   Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm.  heart pump and accessories.",
      "product_quantity": "355 units",
      "reason_for_recall": "Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.",
      "recall_initiation_date": "20260610",
      "center_classification_date": "20260710",
      "report_date": "20260722",
      "code_info": "All Lots within Expiry/GTIN:00885672009793, 00813502010664",
      "more_code_info": ""
    }
  ]
}