{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Madison",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99075",
      "recalling_firm": "Accuray Incorporated",
      "address_1": "1240 Deming Way",
      "address_2": "N/A",
      "postal_code": "53717-1954",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.",
      "recall_number": "Z-2640-2026",
      "product_description": "Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator",
      "product_quantity": "48 units",
      "reason_for_recall": "In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.",
      "recall_initiation_date": "20260616",
      "center_classification_date": "20260701",
      "report_date": "20260708",
      "code_info": "All systems prior with software version 11X that utilize the Exchange Table"
    }
  ]
}