      "status": "Ongoing",
      "city": "Billings",
      "state": "MT",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98741",
      "recalling_firm": "Windstone Medical Packaging, Inc.",
      "address_1": "1602 4th Ave N",
      "address_2": "",
      "postal_code": "59101-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of Arizona.",
      "recall_number": "Z-2066-2026",
      "product_description": "Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes)    Angio Pack  Part numbers AMS6908E, AMS6908F",
      "product_quantity": "150 kits (US only)",
      "reason_for_recall": "Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.",
      "recall_initiation_date": "20260402",
      "center_classification_date": "20260508",
      "report_date": "20260520",
      "code_info": "Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875  Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088",
      "more_code_info": ""
    }
  ]
}