  "results": [
    {
      "status": "Ongoing",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97132",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: North Carolina  OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands",
      "recall_number": "Z-2152-2025",
      "product_description": "Estrone RIA, REF: DSL8700.  Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.",
      "product_quantity": "496 units",
      "reason_for_recall": "A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.",
      "recall_initiation_date": "20250612",
      "center_classification_date": "20250721",
      "report_date": "20250730",
      "code_info": "UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C"
    }
  ]
}