{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-25",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "third_party_flag": "N",
      "city": "St. Paul",
      "advisory_committee_description": "Anesthesiology",
      "address_1": "501 W. County Rd. E",
      "address_2": "",
      "statement_or_summary": "",
      "product_code": "BXW",
      "openfda": {
        "registration_number": [
          "9710567",
          "3010381606",
          "3020614",
          "3012302888",
          "3002808458",
          "3010082909",
          "3009443693",
          "3008317830",
          "3021284283",
          "1922553",
          "3033959233",
          "1924066",
          "3012309966",
          "1836161",
          "9615102",
          "3010187802",
          "3013298431",
          "2133542"
        ],
        "fei_number": [
          "3010381606",
          "3020614",
          "3012302888",
          "1000271197",
          "3010082909",
          "3009443693",
          "3008317830",
          "3002808458",
          "3021284283",
          "1922553",
          "3033959233",
          "1924066",
          "3012309966",
          "3002750084",
          "3010187802",
          "3013298431",
          "3003256355",
          "1000520092"
        ],
        "device_name": "Calibrator, Volume, Gas",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.1870",
        "device_class": "1"
      },
      "zip_code": "55112",
      "applicant": "Medical Graphics Corp.",
      "decision_date": "1985-12-05",
      "decision_code": "SESE",
      "country_code": "US",
      "device_name": "RESPIRATORY DIAGNOSTIC EQUIPMENT CALIBRATOR",
      "advisory_committee": "AN",
      "contact": "RONALD  BERGESON",
      "expedited_review_flag": "",
      "k_number": "K853848",
      "state": "MN",
      "date_received": "1985-09-16",
      "review_advisory_committee": "AN",
      "postal_code": "55112",
      "decision_description": "Substantially Equivalent",
      "clearance_type": "Traditional"
    }
  ]
}